Title: Module 4
1Module 4
- Analytic Phase of Laboratory Testing
2Analytical Phase
- It may be surprising, but current technology has
resulted in the analytical phase now having the
least error of the 3 phases of laboratory testing - Current methods and equipment used to generate
test results have a high degree of - Accuracy
- Correctness how closely the measurement
approaches the true value of the substance being
analyzed (analyte) - Precision
- Reproducibility how closely together are
results from measurements of the same substance
in the same sample
3Test is neither precise nor accurate
Test is precise but not accurate
Test is precise and accurate
Which test is precise, which test is accurate?
4Precision, Accuracy or Both?
- Years ago, it was often said that precision was
more important than accuracy in lab testing - Why?
- A lab test which is precise but not accurate may
still be clinically useful if the data is used to
follow trends or if the lab has properly set its
own reference or normal range - Presently, most common laboratory tests are both
accurate and precise
5Automation
- Many laboratory tests are performed on automated
instruments. - Automation and information technology are widely
used to minimize the degree to which manual
techniques need to be employed in many testing
situations - There are a number reasons for this trend
including - Advances in technology
- The continuing shortage of allied healthcare
personnel entering the field of laboratory
medicine
6Automated Medical Laboratory Photo by GÜliz A.
Barkan, MD
7Case Scenario 1
- A 32-year old man presents with a 3 week history
of fatigue. On physical exam he appears pale and
has scattered petechiae. - His physician begins a diagnostic evaluation he
orders, among other tests, a CBC with
differential
8Case Scenario 1CBC with diff results
- WBC 1.0 L 4.0-10.0 k/ul
- RBC 2.04 L 3.60-5.50 m/ul
- Hgb 7.4 L 12.0-16.0 gm/dl
- Hct 22.2 L 34.0-51.0
- MCV 85 85-95 fl
- MCH 28.3 28.0-32.0 pg
- MCHC 33.3 32.0-36.0 gm/dl
- RDW 16.6 H 11.0-15.0
- Plt Count 7 LL 150-400 k/ul
- Diff Type Manual
- Blasts 20
- Blasts 0.2
- Gran 20 L 45-70
- Gran 0.2 L 2.0-7.0 k/mm3
- Lymph 45 20-45
- Lymph 0.45 1.0-4.0 k/mm3
- Mono 15 0-10
- Mono 0.15 0.0-1.0 k/mm3
- Eo 0.0 0-7
9Case Scenario 1
- Define manual differential.
- Why was it done in lieu of an automated
differential?
10Case Scenario 1Discussion
- Initial specimen analysis was performed by the
automated instrument
Recall the color of the collection tube caps
for a CBC?
Photo by Theresa Kristopaitis, MD
11Case Scenario 1Discussion
- Data generated by the instrument that are not
acceptable based on defined criteria are
flagged to alert the technologist that further
investigation is required - In this case the marked leukopenia,
thrombocytopenia and anemia resulted in the
flag
12Case Scenario 1flagged result
Automated CBC report
Photos by Theresa Kristopaitis, MD
13Case Scenario 1Discussion
- The results from the automated count are reviewed
by the laboratory technologist - A manual slide review is performed
14Stained Peripheral Blood Smear
Manual Cell Counter
Photos by Theresa Kristopaitis, MD
15CBC with manual differential reported
Photo by Theresa Kristopaitis,MD
16Despite Advances in Automated Laboratory
Techniques
- Some laboratory tests are still performed
manually or have manual components - Hematology (as the case illustrated),
microbiology and fluid labs employ medical
laboratory technologists who are well-trained in
microscopic analyses - In addition, staff techniques in pipetting,
vortexing or staining may affect lab results
(remember your days in chemistry lab!)
17 Gram stain reagents used by Microbiology Lab
Technologists to prepare smears
Gram stain of a positive blood culture What is
your interpretation?
Gram Positive Cocci in Clusters
Photos by Theresa Kristopaitis, MD
18Lab Testing is Highly Regulated
- It is critical that clinical laboratory staff
adhere to - A multitude of rigorous accreditation and
regulatory requirements such as - Use of standard operating procedures
- Appropriate performance of quality control
measurements - Instrument calibration schedules
- Instrument preventative maintenance schedules
- Participation in external reviews of analyte
testing (proficiency testing) - These measures help ensure the quality of the
process during the analytical phase of patient
testing
19- Proceed to Module 5 Post Analytic Phase of
Laboratory Testing