Title: Spin-Warp Imaging
1Spin-Warp Imaging
- For each RF pulse
- Frequency encoding is performed in one direction
- A single phase encoding value is obtained
- With each additional RF pulse
- The phase encoding value is incremented
- The phase encoding steps still has the appearance
of stop-action motion
2Spin-Warp Pulse Sequence
3Spin-Warp Data Acquisition
- In 1D, the Fourier transform produced a 1D
image. - In 2D, the Fourier transform is applied in both
the frequency and phase encoding directions. - This is called the 2D Fourier transform.
- Commonly we structure the samples in a 2D grid
that we call k-space. - One line of k-space is acquired at a time.
4Spin-Warp Data Acquisition
2D FourierTransform
5Echo-Planar Imaging
- As with spin-warp imaging, echo-planar imaging
(EPI) is just the combination of two 1D
localization methods - EPI is also a combination of
- Frequency encoding in one direction (e.g.
Left-Right) - Phase encoding in the other direction (e.g.
Anterior-Posterior) - EPI uses a different phase encoding method.
6Echo-Planar Imaging
Frequency Encoding(in x direction)
Phase EncodingMethod 1(in y direction)
7Echo-Planar Imaging
- For each RF pulse
- Frequency encoding is performed many times
- All phase encoding steps are obtained
- The entire image is acquired
- With each additional frequency encoding (each
additional line in the k-space grid) - The phase encoding value is incremented
- The phase encoding steps still has the appearance
of stop-action motion
8EPI Pulse Sequence
9EPI Data Acquisition
- As with Spin-Warp imaging, we put the acquired
data for the frequency and phase encoding into
the 2D grid called k-space. - Also, the 2D Fourier transform is used to create
the image. - In EPI, the data is filled into k-space in a
rectangular zig-zag-like pattern.
10EPI Data Acquisition
11EPI Imaging
- In summary, EPI data is in many ways like
Spin-Warp imaging - They are combinations of two kinds of 1D
localization. - They have both frequency and phase encoding.
- Data are acquired on a 2D grid called k-space.
- Images are reconstructed by a 2D Fourier
transform.
12EPI Imaging
- It is also different from Spin-Warp Imaging
- The image can be acquired with a single RF pulse.
- The phase encoding steps all happen in rapid
succession. - The frequency direction alternates in sign.
- The time needed to acquire data after each RF
pulse is very long. - Special hardware is required.
- These differences are the focus of the rest of
this presentation.
13Variants on EPI
- There are many variations on EPI.
- One technique that is useful for Spin-Warp
imaging that also works for EPI is Partial
k-space or Half k-space acquisitions. - Like Spin-Warp imaging, this can be used to
- Reduce echo-time. (phase)
- Improve spatial resolution. (frequency)
14Partial k-space EPI
Fullk-space
PartialPhase Data
PartialFrequency Data
15Multi-shot EPI
- While possible to acquire an entire image with a
single RF pulse (single-shot), it is sometimes
necessary to use multiple shots. - There are two common ways of doing this
- Interleaving
- Mosaic
- Multi-shot EPI is useful to
- Improve spatial resolution
- Reduce artifacts
16Multi-shot EPI
InterleavedEPI
MosaicEPI
17Methods Similar to EPI
- One method that has very similar properties to
EPI is Spiral Imaging. - Like EPI
- All image data can be acquired in a single-shot.
- Multi-shot variants also exist.
- Many of the artifacts are similar.
- But
- Image reconstruction is complicated.
- Some artifacts are different.
18Spiral Imaging
k-SpaceData
Pulse Sequence
19EPI Parameters
- Many parameters are the same as in spin-warp
imaging - SE vs. GRE or IR
- TR, TE, Flip Angle, TI
- FOV, matrix size, spatial resolution
- Some parameters require extra thought, however
- If only a single image is acquired using
single-shot EPI, the TR might be meaningless. (TR
is infinite)
20Scan Time in EPI
- The scan time is most closely related to the
number of shots and not matrix size. - Scan Time (number of shots)(TR)
- Not (number of phase encodes)(TR)
- Consider 64x64 single-shot EPI and 128x128
single-shot EPI - both are single-shot and take a
single RF pulse to acquire an image. - If 128x128 has artifacts that are too severe,
however, multi-shot EPI may be required.
21Echo Time in EPI
- In EPI, it is often hard to achieve a short echo
time. - The TE is defined as the time between the RF
pulse and the acquisition of the center of
k-space. - In single-shot EPI, this could be a long time
(often a minimum TE of 15-20 ms). - This can be addressed by doing a partial k-space
acquisition in the phase encoding direction. - This will allow much shorter TEs (5-10 ms).
22Echo Time in EPI
Partialk-space
Full k-space
23Pulse Sequence Options in EPI
- Flow Compensation (Gradient Moment Nulling)
- Flow Comp (GNM) is often not as effective with
EPI due to the long echo times. - Partial k-space (phase) acquisitions reduce echo
time and make this technique more effective. - Spatial and chemical presaturation can also be
used (fat saturation is nearly always used). - There are also a 3D (volume) versions of EPI.
24T1 Weighting in EPI
- In EPI, short TEs are difficult to obtain and
the TR is often very long. - EPI is not well suited to T1-weighted imaging
with the usual short TR pulse sequences. - On the other hand, one shot (or a small number of
shots) is required for an image. - EPI is well-suited to inversion recovery
T1-weighted imaging.
25Artifacts in EPI
- The ability to acquire images very rapidly is the
strength of the EPI method. - As a result, artifacts resulting from subject
motion are nearly non-existent when imaging with
single-shot EPI. - Ghosting artifacts resulting from pulsatile blood
flow are also extremely rare with single-shot EPI.
26Artifacts in EPI
- There are however, several kinds of image
artifacts that are very different from those seen
in spin-warp imaging - N/2 or Nyquist ghosting
- Distortions from magnetic field inhomogeneity
- Chemical shift and susceptibility artifacts
27N/2 Ghosting
- N/2 (N over 2) or Nyquist ghosting artifacts
are unique to EPI. - Caused by imperfections in the image
acquisition. - There are two distinct kinds
- Even and Odd Ghosts
- Ghost tuning procedures can reduce or eliminate
these ghosts. - Tuning can be done for each day, subject, or
scan. - Might also be done automatically (with prescan).
28N/2 Ghosting
Even Ghost
Odd Ghost
Original Image
29Distortions from Inhomogeneity
- EPI is very sensitive to center frequency
adjustments and inhomogeneities. - For a misadjusted center frequency, the image is
shifted in the phase direction. - Careful prescan tuning is necessary.
- For misadjusted shims, the image can be twisted,
stretched or squeezed. - Shimming is often necessary (esp. at high
fields).
30Distortions from Inhomogeneity
Center FrequencyMisadjustment
OriginalImage
X shim(L/R shim)
Y shim(A/P shim)
31Fat and Susceptibility Artifacts
- In EPI, unsuppressed fat is often shifted 2 cm or
more. - Fat suppression (Fat Sat) is always required.
- At areas of high magnetic susceptibility, a
piling-up artifacts is often seen. - Prevalent near frontal sinuses, above ears, etc.
- Pulse sequence parameters can affect this
- Interleaving usually reduces this artifact.
- Increasing resolution in the frequency direction
often worsens the artifact.
32Fat and Susceptibility Artifacts
Susceptibility (piling-up) Artifact
Fat Artifact
Original Image
33EPI Hardware
- EPI is an extremely rapid and useful imaging
method. - It does, however, require some special, high
performance hardware. Why? - In spin-warp, we acquire a small piece of data
for an image with each RF pulse. - However in EPI, we try to acquire all of the data
for an image with a single RF pulse.
34Spin-Warp vs. EPI Pulse Sequences
EPI
Spin-Warp
Many acquisitionsto make a one image.
One acquisitionto make one image.
35T2 Decay and Acquisition Time
- In spin-warp imaging, only a single phase encode
need to be acquired. - Only takes a short time.
- In EPI, all phase encode lines need to be
acquired. - Takes longer.
- Without special hardware - 100 ms to 1 second.
- T2 decay reduces signal throughout data
acquisition.
36T2 Decay and Acquisition Time
DataAcq. takes longer.
Signal decays away during acquisition.