All QIBA PHANTOMS - PowerPoint PPT Presentation

About This Presentation
Title:

All QIBA PHANTOMS

Description:

All QIBA PHANTOMS. C52 Transducer held in ringstand verasonics system. Phantom on turntable 8 rotation angles. 7 transducer locations . 3 focal depths (40, 60 ... – PowerPoint PPT presentation

Number of Views:66
Avg rating:3.0/5.0
Slides: 4
Provided by: rsn4
Category:

less

Transcript and Presenter's Notes

Title: All QIBA PHANTOMS


1
All QIBA PHANTOMS
  • C52 Transducer held in ringstand verasonics
    system
  • Phantom on turntable 8 rotation angles
  • 7 transducer locations
  • 3 focal depths (40, 60, 80 mm)
  • SWS estimates
  • Bandpass filter 100-1000 Hz
  • Ransac algorithm fov 2 cmx2cm
  • Errorbars mean /- 1 std. dev from 56 measures

2
Results
1786-X
1787-X
3
Sim Results Lateral Regression Window Location
Window positions Blue 0-12 mm Red 1-12 Green
2-12 Black 4-12
  • Conclusion you must be outside the ROE (near
    field of shear wave) to get consistent speeds
  • Also- depth dependence here was within numerical
    error???