Title: TWM%20Basic%20Steps%20to%20%20%20%20%20%20Perform%20Valve%20Check%20Load%20Test.
1TWM Basic Steps to Perform Valve Check
Load Test.
Echometer How To
Reference Well Analyzer QuickRefs,
http//www.echometer.com/support/quickrefs/index.h
tml
2Valve Check Load Tests Answers the Following
Questions
- What rate does the Traveling Valve Leak?
- What is the condition of the traveling valve, the
pump barrel and plunger and the tubing string? - Is the fluid being held in the tubing?
- Does the Standing Valve Leak?
- Are the rod string lengths correct?
3Stop on the Upstroke to Monitor for Leakage from
the Tubing into the Pump
TV Check Load Test weighs the rod string buoyed
in tubing fluid plus the fluid load acting on the
traveling valve (across plunger).
4Stop on the Down Stroke to Monitor for Leakage
from the Pump into the Casing
SV Check Load Test weighs the rods in air reduced
by the buoyancy of the tubing fluid on the rod
string.
5Use Any of these Dynamometer Transducers to
Determine a Valve Check Load Test
6Accuracy of Acquired Load Data Depends on Type
of Load Cell Used
- Portable load cells are calibrated and accurately
determine polished rod loads. - The PRT acquired loads can drift and make Valve
Check Load Tests difficult to interpret. - Error in the measured loads are introduced if the
carrier bar and clamp do not uniformly contact
the load cell. - Donut load cell should have a spacer and
spherical washers to ensure the polished rod load
is centralized on the load cell.
7Connect the cables to Well Analyzer.
Far end on the coiled cable is attached to the
25-foot Y-cable or Straight cable, that is
connected into the MAIN INPUT on the Well
Analyzer.
8Turn on Well Analyzer and wait for GREEN LED.
9Turn on the computer.
10Start TWM in Acquire Mode
- 1. Select the Equipment Check tab.
- 2. The display of internal battery voltage and
battery temperature indicate that the computer is
communicating with the well analyzers internal
electronics and sensor measurements may be
acquired.
11TWM Software F3 to Select Well
- 1. Open Base Well File for the well where data
is to be acquired. Use New to create a Base
Well File if one does not exist. -
- 2. For acoustic surveys be sure to enter at
least avg. joint length, pump and formation
depths. -
12Select Wellbore Tab to Verify Data
- 1. From the Base Well File select the Wellbore
Tab and verify the well data is representative. - 2. For Dyno surveys be sure to enter Rod Type,
Rod Length, Rod Diameter, Pump Plunger Dia., Pump
Intake, Polished Rod Diameter, fluid gravities,
and production rates.
13Save Changes to Base Well File
- 1. Click Save button to be sure to save changes
to the base well file. -
14Select Type of Data Acquisition Test
- 1. Tap F4 to Select Test type to be acquired.
- 2. Click the Dynamometer Tab to indicate that
Dynamometer data will be acquired. - 3. Click Alt-2 to select the Valve Check Load
Test - 4. Select F5 to Acquire Data
15Read Instructions for Performing the Traveling
and Standing Valve Test
- 1. After choosing the Select Test option F4
the operator is presented with the instructions
performing the traveling and standing valve test. - 2. Click the Load tab to begin displaying
data.
16Tap Alt-S to Start Data Acquisition
17Traveling Valve Check Procedure
The traveling valve check load test is taken
during the upstroke by gently using the brake to
bring the pumping unit to a stop about ¼ from the
top of the stroke.
18Perform Two Traveling Valve Checks
19Upstroke TV Check Load Test
- 1. Pumping cycle was interrupted on the
upstroke when SV is open and TV is closed for a
traveling valve check load test. - 2. Static pressure above and below the
standing valve are equal. - 3. Acquired TV load the weight of the rods
buoyed in fluid plus the differential fluid load
on the plunger.
2500 psi
200 psi
200 psi
20Standing Valve Check Procedure
Standing valve check load is taken during the
down stroke by gently using the brake to bring
the pumping unit to a stop about ¼ from the
bottom of the stroke.
21Perform Two Standing Valve Checks
22Down Stroke SV Check Load Test
- 1. Pumping cycle was interrupted on the down
stroke when the TV was open and SV was closed for
a standing valve check load test. - 2. Static pressures across the plunger are
equal for SV Check Load (shown). - 3. Acquired SV load is the weight of the rod
string buoyed in the tubing fluid.
2500 psi
2500 psi
200 psi
23Alt-Q to Stop Acquisition, Then Save
24Press F6 to Analyze Valve Checks
Examine Display and Verify Correct SV and TV
Check Loads are Selected
25If PRT Used for Valve Checks
- 1. Verify Correct SV Check Load is Selected
- 2. Click Apply button to Adjust SV to Buoyant Rod
Wgt.
26If PRT Used for Valve Checks
- 1. After Clicking Apply button to Adjust SV to
Buoyant Rod Wgt. - 2. The Selected SV Check Load is adjusted to
exactly match the Calc. Buoyant Rod Wgt.
27If TV Leaks
Possible problems worn plunger or barrel,
pitted ball, cut seat, tubing leak, or well
flowing off.
- 1. Leakage rate indicated by changes in the
polished rod load due to changes in the
difference in pressure across the plunger - 2. Liquid leaking past traveling valve and/or
plunger enters pump barrel and increasing
pressure. - 3. Differential pressure decreases across
plunger. - 4. Pump load on the rods decreases, resulting
in a decreasing polished rod load.
2500 psi
500 psi
200 psi
28If TV Leaks
29If SV Leak
Possible problems pitted ball, cut seat, hole in
pump barrel, or damaged pump parts.
- 1. Liquid leaking past standing valve leaves
pump barrel and decreasing pressure. - 2. The leak causes increasing differential
pressure across plunger, and results in the
plunger applying fluid load to rods. - 3. Both pump rod load and polished rod load
increases.
2500 psi
2000 psi
200 psi
30If SV Leaks
31Problem TV Load OK, but NO SV Load
- Possible Problems
- 1. Severe SV leak.
- 2. Gas Locked Pump, P3gtP2gtP1.
- 3. SV is stuck open
- 4. Fluid level below pump intake.
- 5. Specialty Pump Gas Compression Chamber
2500 psi
200 psi
200 psi
32TV Load OK, but NO SV Load
33Problem SV Load OK, but NO TV Load
- Possible Problems
- 1. Completely worn out plunger/barrel or TV
- 2. TV is stuck open
- 3. Deep Rod Part or Pump Parted.
- 4. Well flowing through pump and both valves
open.
2500 psi
2500 psi
200 psi
34SV Load OK, but NO TV Load
35The END