Thermogravimetric AnalysisTheory, Operation,
Calibration and Data Interpretation
  • Prepared by Kadine Mohomed, Ph.D
  • Thermal Applications Chemist
  • TA Instruments

2
Agenda TGA Theory, Operation and Calibration
  • Definitions and review of instrument
  • Balance, furnace and heat exchanger review
  • Mass and temperature calibration
  • Purge gas considerations
  • Baseline considerations
  • Sample preparation and pan selection
  • Method development

3
TGA The Technique
  • Thermogravimetric Analysis (TGA) measures the
    amount and rate of change in the weight of a
    material as a function of temperature or time in
    a controlled atmosphere.
  • Measurements are used primarily to determine the
    composition of materials and to predict their
    thermal stability at temperatures up to 1000C.
  • The technique can characterize materials that
    exhibit weight loss or gain due to decomposition,
    oxidation, or dehydration.

4
What TGA Can Tell You
  • Thermal Stability of Materials
  • Oxidative Stability of Materials
  • Composition of Multi-component Systems
  • Estimated Lifetime of a Product
  • Decomposition Kinetics of Materials
  • The Effect of Reactive or Corrosive Atmospheres
    on Materials
  • Moisture and Volatiles Content of Materials

5
Calcium Oxalate Example

6
Mechanisms of Weight Change in TGA
  • Weight Loss
  • Decomposition The breaking apart of chemical
    bonds.
  • Evaporation The loss of volatiles with elevated
    temperature.
  • Reduction Interaction of sample to a reducing
    atmosphere (hydrogen, ammonia, etc).
  • Desorption.
  • Weight Gain
  • Oxidation Interaction of the sample with an
    oxidizing atmosphere.
  • Absorption.
  • All of these are kinetic processes (i.e. there is
    a rate at which they occur).

7
Features of the Q500/ Q50 TGA
  • The Q500 is a research grade thermogravimetric
    analyzer, whose leading performance arises from a
    responsive low-mass furnace sensitive
    thermobalance, and efficient horizontal purge gas
    system (with mass flow control). Its convenience,
    expandability and powerful, results-oriented
    software make the Q500 perfect for the multi-user
    laboratory where a wide variety of TGA
    applications are conducted and where future
    expansion of analytical work is anticipated.

8
Features of the Q500 TGA
  • 1. Q Series Two Point Mass Adjustment
  • 200mg range
  • 1000mg. range
  • No need to do a mass recalibration
  • when switching from regular Pt pans to
  • Pt pans with Al hermetic pans.
  • Mass Loss Reference Materials
  • Materials with nominal 2, 50 and 98
  • mass loss are available for verification of TGA
  • weight calibration.
  • 2. Curie Point Transition Temperature
  • Calibration
  • ASTM 1582
  • Curie Temperature Reference Materials
  • TA Instruments is the exclusive worldwide
  • distributor for a set of six certified and
    traceable
  • Curie temperature materials developed by ICTAC

9
Q50/Q500 Features and Options
Feature Q500 Q50 Furnace low
mass Standard Standard Furnace EGA
Option Option Temperature Range RT-1000C
RT-1000C MFC / GSA Standard
Option Autosampler Option NA Hi-Res
TGA Option NA Modulated TGA Option
NA Touch-screen display Standard NA TGA
/ MS operation Option Option TGA / FTIR
operation 3rd Party 3rd Party
NA Not Available
10
TGA Furnaces
  • Standard Furnace
  • Low mass
  • Used for Hi-Res Runs
  • Cools down in lt20min
  • EGA Furnace
  • Higher Mass
  • Used for EGA runs due to quartz liner
  • Cools down in 40min

11
TGA Purge Gas Flow
10ml/min
40ml/min
90ml/min
60ml/min
EGA Furnace
Standard Furnace
12
Standard Furnace
13
EGA Furnace Schematic

Low internal Volume 15ml
14
TGA How the balance works
  • The balance operates on a null-balance principle.
    At the zero, or null position equal amounts of
    light shine on the 2 photodiodes.
  • If the balance moves out of the null position an
    unequal amount of light shines on the 2
    photodiodes. Current is then applied to the
    meter movement to return the balance to the null
    position.
  • The amount of current applied is proportional to
    the weight loss or gain.

15
TGA Q Series MFC and GSA
MFC and GSA standard on Q500 and optional on Q50
16
TGA Q-Series Purge Gas Plumbing
  • Instruments w/o MFC
  • The gas 1 port purges the sample area only.
  • The gas 2 port purges the balance area only.
  • Instruments w/ MFC
  • The gas 1 port purges both sample and balance
    areas.
  • The gas 2 port is used when a different purge gas
    is required or gas switching is used.
  • Selection of gas on NOTES page is critical for
    proper use of MFC calibration tables.

17
Heat Exchanger Cleaning
  • Check cleanliness (no algae growth) once every
    3-6 months.
  • To clean dump old water, fill with new and add
    conditioner (algae growth suppressor) if
    available.
  • For Q series, after filling, in software choose
    Control \ Prime Exchanger.
  • For 2xxx, after filling, continue starting a
    dummy run until error 119 (heat exchanger no
    flow) goes away.

18
TGA Performance Criteria
  • Baseline
  • Drift
  • Affected by TGA construction, balance quality,
    and buoyancy effect (minimized through proper
    construction techniques and purge gas control)
  • Sensitivity
  • Affected by TGA balance quality
  • Reproducibility
  • Affected by balance quality, temperature control,
    and construction quality
  • Temperature Accuracy
  • Affected by thermocouple placement, calibration
    stability, purge gas interaction

19
TGA Performance
  • TGA Performance is primarily a function of
    balance sensitivity and baseline stability
  • Balance sensitivity is optimized through design
    and construction techniques
  • Baseline stability is a function of instrument
    design, as well as purge gas control
  • TGA resolution is primarily a function of heating
    rate, but can be optimized using Hi-Res TGA

20
Quantifying TGA Baseline Performance
Drift
Unnormailzed Sample Mass
Temperature or Time
21
Measuring Q500 TGA Baseline Performance
Drift 19 mg
Q500, 20C/min Ramp
22
TGA Calibrations
  • Mass (Verify monthly)
  • Temperature (Verify monthly)
  • Platform (Perform if there is a problem picking
    up pans.)
  • Q series instruments w/ MFC will also have
    options to calibrate the sample and balance
    MFCs. These have been calibrated by TA
    Instruments and should not require further
    calibration. Contact TAI if a problem arises.

23
TGA Mass Calibration
  • Two point mass adjustment 2050, 2950, Q50, Q500
  • 100mg. (2XXX modules) or 200mg (Q series) range
    (use 100mg. weight)
  • 1000mg. range (use 1000mg. weight)
  • Q5000IR 100mg
  • Run TGA weight calibration routine
  • Follow screen instructions to tare and
    masscalibrate using two calibration weights (if
    known,enter exact mass of calibration weights)

24
Mass Loss And Residue Validation
P/N 952540.901 TGA / SDT Mass Loss Reference
Materials Kit 1,760
Mass Loss Reference Materials Materials with
nominal 2, 50 and 98 mass loss are available
for verification of TGA weight calibration.
25
Temperature Calibration Curie Point Transition
  • Paramagnetic - a material that is susceptible to
    attraction by a magnet
  • Curie Point Temperature - that temperature where
    the material loses its magnetic susceptibility
    (defined as offset point)
  • Requires a magnet and well characterized
    transition materials
  • ASTM 1582 - Standard Practice for Calibration of
    Temperature Scale for Thermogravimetry

26
TGA Temperature Calibration
Vertical Balance Configuration - TGA
2050/2950/Q50/Q500
Tare

Sample

Offset
Furnace
temp
Attraction of Sample to Magnet Results in Initial
Weight Gain
Magnet
27
TGA Temperature Calibration Important Points
  • Clear the Temperature Table before performing
    the calibration runs (TGA only).
  • Choose method end condition of Furnace Closed.
    This prevents the potential of the furnace
    opening onto the magnet at the end of the run and
    damaging the TGA.
  • Start run and then put magnet under furnace.
    This allows capture of the weight increase
    (decrease) at the beginning.
  • Use of a small labjack is recommended for holding
    the magnet in place under the furnace.

28
Standards Can Be Run Simultaneously
Alumel 157.00?C
Nickel 368.80?C
29
Calcium Oxalate Standard Analysis
  • Although Calcium Oxalate is not generally
    accepted as a Standard Material, it does have
    practical utility for INTRA-laboratory use
  • Carefully control the experimental conditions
    i.e. pan type, purge gases/flowrates, heating
    rate
  • Particularly control the amount (5mg) and the
    particle size of the sample and how you position
    it in the pan
  • Perform multiple runs, enough to do a statistical
    analysis
  • Analyze the weight changes and peak temperatures
    and establish the performance of YOU and YOUR
    instrument
  • When performance issues come up, repeat the
    Calcium Oxalate analysis

30
Calcium Oxalate Decomposition
  • 1st Step CaC2O4H2O (s) CaC2O4 (s) H2O (g)
  • Calcium Oxalate Monohydrate Calcium Oxalate
  • 2nd Step CaC2O4 (s) CaCO3 (s) CO (g)
  • Calcium Oxalate Calcium Carbonate
  • 3rd Step CaCO3 (s) CaO (s) CO2 (g)
  • Calcium Carbonate Calcium Oxide

31
Calcium Oxalate Repeatability
Overlay of 8 runs, same conditions
32
Calcium Oxalate Repeatability
33
General Considerations(Experimental Effects)
34
TGA Curves are not Fingerprint Curves
Because most events that occur in a TGA are
kinetic in nature (meaning they are dependent on
absolute temperature and time spent at that
temperature), any experimental parameter that can
effect the reaction rate will change the shape /
transition temperatures of the curve. These
things include
  • Pan material type, shape and size.
  • Ramp rate.
  • Purge gas.
  • Sample mass, volume/form and morphology.

35
Effect of Sample Size on Decomposition Temperature
36
Effect of Heating Rate on Decomposition
Temperature
37
Mass Effect Semi-crystalline PE
38
Shift in Onset with Ramp Rate
39
Typical Applications
  • Thermal Stability
  • Compositional Analysis
  • Oxidative Stability

40
Thermal Stability of Polymers
41
TGA of an Adhesive
25.18mg of an adhesive _at_ 10C/min
42
Inset View Shows Strange Result
Is this real?
43
Use time based derivative of temperature to plot
the heating rate
44
Aberration in Heating Rate
Usually means that the sample touched the
thermocouple
45
Typical Applications
  • Thermal Stability
  • Compositional Analysis
  • Oxidative Stability

46
PET w/ Carbon Black Filler
How much Carbon Black was in this sample?
47
PET
48
Comparison of Filled Un-Filled PET
49
Filled Polymer Analysis

Inert filler
Inert filler
Inert filler
50
Kinetic Analysis
  • The rate at which a kinetic process proceeds
    depends not only on the temperature the specimen
    is at, but also the time it has spent at that
    temperature.
  • Typically kinetic analysis is concerned with
    obtaining parameters such as activation energy
    (Ea), reaction order (k), etc. and/or with
    generating predictive curves.

51
Kinetic Analysis, cont.
Activation energy (Ea) can be defined as the
minimum amount of energy needed to initiate a
chemical process.
Ea
State 1
State 2
With Modulated TGA, Ea can be measured directly.
52
TGA Kinetics
  • 1st Order Kinetics based on Flynn and Wall method
  • Lifetime Estimation based on Toops and Toops
    method
  • PTFE tested at 1, 5, 10 and 20 deg/min
  • Sample sizes constant
  • Nitrogen purge
  • Conversion levels selected at 1, 2.5, 5, 10 and
    20

53
Common Thermogram with TGA Scans
54
Log Heating Rate versus 1/T
Check for linearity
55
Activation Energy by MTGA
56
Sample of TGA Application Briefs
  • H-16781 Thermogravimetry-Mass Spectrometry Using
    a Simple Capillary Interface
  • TA023 Thermal Analysis Review High Resolution
    TGA - Theory and Applications
  • TA075 High Resolution TGA Kinetics
  • TA 122 Determination of Carbon Black Pigment in
    Nylon 66 by TGA
  • TA 125 Estimation of Polymer Lifetime by TGA
    Decomposition Kinetics
  • TA231 TGA Evaluation of Zeolite Catalysts
  • TN6 Consideration of Subtle Experimental Effects
    (Simultaneous TGA-DTA)
  • TN24 TGA Temperature Calibration Using Curie
    Point Standards
  • TN40 Optimizing Stepwise Isothermal Experiments
    in Hi-Res? TGA
  • TS13 Clarification of Inorganic Decomposition by
    TG-MS
  • TS39 Characterization of Polyurethane by TGA and
    Hi-Res ? TGA

57
Common TGA Parts Accessories
58
Common TGA Parts Accessories
59
Need Help?
  • Check the online manuals and error help.
  • Contact the TA Instruments Hotline
  • Phone 302-427-4070 M-F 8-430 EST
  • Select Thermal or Rheology Support
  • Email thermalsupport_at_tainstruments.com or
  • rheologysupport_at_tainstruments.com
  • Call the TA Instruments Service Hotline
  • Phone 302-427-4050 M-F 8-430 EST
  • Call your local Technical or Service
    Representative
  • Check out our Website www.tainstruments.com