Title: Trouble Shooting (Mechanical Ventilation)
1Trouble Shooting (Mechanical Ventilation)
-
-
-
- Arthur Sadhanandham
- Medical ICU, CMC.
-
2- WHEN NOAH BUILT THE ARK
- IT WAS NOT RAINING
3Precautions that would reduce troubles
- I. Power
- Plug into a grounded AC power with
correct voltage receptacle. - Secure the power cord properly.
- Battery Back up
- Check the battery level before connecting.
- Charging should be carried out regularly.
- Remember it is for short term use.
-
4II. Gas Source
- Preferable to have centralised supply.
- If cylinders used, should be full
- Spare cylinders should be available.
- Gas hoses should be in good condition.
- Hoses not contaminated with grease or oil
(combustible) - Availability of compressors should be ensured.
- Gases should remain dry and clean.
5III. Personnel
- Properly trained personnel should only use.
- Familiarising staff with operators manuel before
using on a patient. - (One manufacturers manual may not exactly match
with other brands). - Appropriate monitoring the functioning state of
the ventilator while in use.
6 Contd
- Familiarizing staff with alarm system.
- Do not place ventilators in a combustible or
explosive environment. - Do not use with flammable anaesthetic agents such
as nitrous oxide and ether.
7IV Servicing and Testing
- Qualified personnel should undertake servicing.
- Ventilator housing should not be opened while it
is still connected with power. - Follow the specifications mentioned in the
service manual. - Use replacement parts supplied by the
manufacturer only.
8Contd.
- General servicing at regular intervals should be
done. - Run the prescribed tests and calibrations before
using the ventilator on a patient. - Ensure that the ventilators pass all the tests
before putting them in to clinical use.
9ALARMS
- All ventilators are equipped with visual and
audible alarms which notify the user problems.
10Points to remember
- Never ignore an alarm.
- Never mute the alarm on regular basis.
- Find out for yourself what alarm is on.
- Check the patient.
- Silence the alarm.
11Act Swiftly
- Depending upon the patients status and nature of
the alarm, act appropriately. - This includes disconnecting the ventilator and
connecting another means of ventilation to
patient Bains/ Ambu.
12Do not forget
- The use of an alarm monitoring system does not
give absolute assurance of warning for every form
of trouble that may occur with the ventilator.
13Do not be like this !
But hear the alarm and respond See the problem
and Ask if you do not know what to do
14Common Troubles and Shooting
- Ensure Alarm knobs / switches are turned on and
functional. -
Alarm Cause Shooting
Apnoea No breath was delivered for the operator set apnoea time in spont, SIMV, AC, CMV NIV modes Because spontaneous Ventilation is too high or patient effort is too minimal Trigger level set improperly. Check the patient- Arouse if needed Activate back up facility if it was not done already. Consider switching over to any mandatory mode Or go up on rate Set trigger level appropriately
15 Low SpO2 Air / O2 Blender continuous alarm Delivery of O2 FiO2, PEEP High resistance due to various clinical reasons Supply pressures are inadequate. Disconnect patient from ventilator Manually bag with Bains and Ambu. Insert the gas hose fittings (air O2) correctly into the wall outlets. Ensure wall outlets has adequate pressure
16Low pressure or Low min.Vent Or Low exhaled volume or Disconnection The measured PIP is lesser than the set minimum level because of cuff leak. Leak in the circuit Connections may be loose ET tube displacement Disconnection Inadequate flow Evaluate cuff pressure at regular intervals. Reinflate if leak / ruptured is noticed change ET tube. Check circuits, junctions- tighten or replace. Check water traps Check ET tube placement. Position it properly. Reconnect ventilator. Patient may require higher flow.
17High Pressure Alarm The measured peak inspiratory pressure is great than set level because of ?Secretions in airway Partial block (ETt) Kinking of tube Biting the tube Water in the tube Cuff herniation Deep Rt. sided intubation Fighting the ventilator Suctioning, Irrigation Release tubings Bite block insertion Empty the tubings and water traps Deflate reinflate cuff 3-4 times Reposition the ET tube Reposition the patient Re assurance Sedation medication (pain)
18 High pressure alarm Cough Increased airway resistance or decreased compliance because of Bronchospasm Atelectasis Fluid overload Pneumothorax Medication Bronchodilators Adjust the settings VT ? Rate Adjust the settings VT ? Rate, ? PEEP (Peak pressure to be monitored) Immediate intervention
19Auto Cycling High Tidal Volume Leak Improper trigger setting Patient trying to take more volume of air Secure all tubings tight Set proper trigger level Increase flow rate or Increase tidal volume
20Trouble shooting!
THANK YOU FOR YOUR PATIENT HEARING. . . . .