Neonatal Intensive Care Monitoring - PowerPoint PPT Presentation

1 / 23
About This Presentation
Title:

Neonatal Intensive Care Monitoring

Description:

Neonatal Intensive Care Monitoring. Overview. Neonatal Blood ... Methemoglobin. Caboxyhemoglobin. Jaundice. Reads low. Medical dyes. Other causes of inaccuracy ... – PowerPoint PPT presentation

Number of Views:945
Avg rating:3.0/5.0
Slides: 24
Provided by: marily51
Learn more at: http://classes.kvcc.edu
Category:

less

Transcript and Presenter's Notes

Title: Neonatal Intensive Care Monitoring


1
Neonatal Intensive Care Monitoring
  • Overview
  • Neonatal Blood Gases
  • Pulse Oximeters
  • Neonatal Hemodynamic Equipment
  • Transcutaneous Monitors

2
Neonatal Blood Gases - Sampling Possibilities
  • Arterial Gases
  • Venous Gases
  • Capillary

3
Arterial Gases
  • Radial, Brachial, Temporal Punctures
  • Radial Artery Line
  • Umbilical Artery Gases
  • Umbilical Artery Catheter (UAC)
  • Preductal placement vs postductal placement

4
Venous Gases
  • Drawn from Umbilical Venous Catheter (UVC)
  • Not desirable but......

5
Capillary Gases
  • Drawn from heel
  • Procedure
  • heel warmed to arterialize blood
  • lancet puncture
  • blood flows, trapped in capillary tube

Preferred Sites
6
Variability in Cap Gases
  • Warming time
  • Amount of contact with air
  • Squeezing blood
  • As a result, not desired but .......

7
Comparative
pH pCO2 HCO3 PO2 Arterial 7.4 40 24 60-80 (term)
Arterial 7.4 40 24 50-70 (preterm) Capillary 7
.4 40 24 40-50 Venous 7.35 45 24 35-45
8
Pulse Oximeters
  • Sites of attachment
  • (foot and hand)
  • Preductal placement in first twelve hours
  • (right hand)

9
Pulse Oximeters
  • Reads high
  • Methemoglobin
  • Caboxyhemoglobin
  • Jaundice
  • Reads low
  • Medical dyes
  • Other causes of inaccuracy
  • Motion
  • Hypothermia/vasoconstriction
  • Hypotension
  • Excessive ambient light on sensor probe

10
Hemodynamic Monitoring
  • Umbilical Artery Catheter (UAC) preferred

11
UAC Insertion Procedure
  • Insertional position 1/3 length heel to crown
  • Procedure
  • sterile field and drape
  • purse string suture around umbilicus
  • cut cord and snug
  • tease umbilical artery open
  • insert catheter
  • fix position
  • follow with CXR

12
Monitoring UAC Post Insertion
  • Position of catheter tip
  • (aortic arch is preductal and not preferred)
    Normal position above diaphragm
  • (low position is L3-L4)
  • Monitor leg color of infant
  • (blanching indicates obstruction of flow)

13
Indwelling UAC Gases
  • Orange Medical Company
  • PO2 electrode at tip of catheter
  • Provides continuous reading

Cathode
Anode
14
Transcutaneous Gas Monitors
  • Useful as trend monitor
  • Can detect hypoxemia, hyperoxemia
  • Can detect hypocarbia, hypercarbia
  • Also responds to changes in blood flow

15
Types of Transcutaneous Monitors
  • Single Electrode Models
  • PO2 most common

16
Types of Transcutaneous Monitors
  • Dual element electrodes
  • PO2 and PCO2
  • Called TcPO2 and TcPCO2

17
Principle of Operation Tc Monitors
  • Heated electrode placed on skin
  • Temperature 43 to 45 C
  • Arterializes sample
  • Gas diffuses through skin

18
Calibration of Transcutaneous Monitors
  • Requires high and low calibration
  • TcPO2
  • Can be done with chemical zero and room air
  • Most commonly done with cylinders

Calibration value Concentration of gas in
cylinder x Pb
Using a cylinder that contains 10 O2, what would
be the calibration value of a TcPO2 device if the
barometric pressure was 760?
Calibration value .1 x 760 76 mm Hg
19
Calibration of TcPCO2 Devices
  • Similar to TcPO2 except.......
  • 1.6 is the factor that accounts for heating
    increasing CO2 production

Calibration value Concentration of CO2 x Pb
1 .6
20
Normal Transcutaneous Gases
  • TcPCO2 is 35 to 45 torr
  • TcPO2 is 50 to 70 torr

21
Advantages of Transcutaneous Monitors
  • Decreased number of sticks
  • cost reduction
  • lower infant risk (less invasive)
  • Trend tool
  • blood sample provides view at one moment
  • gases values wander ( 7 torr)
  • infant reaction to sample varies

22
Problems with Transcutaneous Monitors
  • Labor Intensive
  • Change site every 4 to 6 hours or more
  • Limited choices for attachment
  • (site must have perfusion)
  • Air leak around electrode
  • Burns
  • called hookies after Huch

23
Interpretation of Tc Results
  • Air leak under electrode
  • TcPCO2 reading near zero
  • TcPO2 reading near PbO2
  • Decreased perfusion under electrode
  • TcPCO2 will increase
  • TcPO2 will decrease
Write a Comment
User Comments (0)
About PowerShow.com