Title: Angiography Equipment
1Angiography Equipment
2Educational objectives
- What are equipment standards for cath equipment
(FDA, IEC), particular needs for pediatric
patients. - What to look for while establishing a cath lab.
- Importance of testing equipment performance.
3Equipment standards for Cath Lab
4X-Ray Equipment Standards And Regulations
- Standards are consensus guides from the
manufacturing community, not regulatory - Several groups set standards regarding equipment,
e.g., International Electrotechnical Equipment
(IEC) - Apply to electrical, mechanical, and radiation
safety - Apply to equipment at time of manufacture and
installation
5What to look for while establishing a cath. lab.
- If the relevant Standards are fulfilled.
- If a medical physicist is available.
- If radiation protection tools are available.
- If patient dose measuring and recording system is
available. - If acceptance tests, commissioning and quality
assurance programme have been foreseen.
6What to look for while establishing a cath. lab.
- If the X rays system selected is appropriate for
the procedures to be carried out in the
catheterization laboratory. - If some other relevant information described in
ACC/AHA Guidelines and AAPM-70 (described in this
lecture) have been taken into account.
7Outline
- FDA, IEC and ACR recommendations concerning X-ray
equipment for cardiology. - AAPM report and specific pediatric equipment
recommendations. - Key topics for cardiac X-ray equipment.
- IAEA survey and importance of testing equipment.
- Patient dose reports and DICOM header information.
8Limitation in entrance exposure rate
Federal Register May 19, 1994. 21 CFR Part
1020. Federal Performance Standard for Diagnostic
X-Ray Systems and Their Major Components Final
Rule. DEPARTMENT OF HEALTH AND HUMAN
SERVICES Food and Drug Administration
9Limitation in entrance exposure rate
- The Standard for Diagnostic X Ray Systems (May
19, 1994), limits the entrance exposure rate of
fluoroscopic x ray systems during normal
fluoroscopy to 10 R/min unless an optional
high-level control (HLC) is activated. - If HLC is activated, the entrance exposure rate
must be limited to 20 R/min. - The entrance exposure rate limits do not apply
during the recording of images.
10Measuring entrance dose and image quality
Test object to measure image quality, at the
isocenter
Flat ionisation chamber to measure phantom
entrance dose
11Typically we can find the following fluoroscopy
setting
12Typically we can find the following cine setting
13Proposed Rule December 10, 2002
14mGy (total) mGy/min (at 15 cm from the isocenter
towards the x-ray source) Fluoroscopy time
15(No Transcript)
16- Fluoroscopic equipment manufactured on or after
May 19, 1995 - Shall not be operable if AKR is higher than 88
mGy/min (10 R/min). - Exceptions
-
- When a mode a high-level control is activated
180 mGy/min (20 R/min). A continuous signal
audible to the fluoroscopist shall indicate that
the high-level control is being employed. - During the recording of images (archiving of
fluoroscopic or radiographic images in analog
format with a video-tape or video-disc recorder
does not qualify as an exception).
Limits 88 mGy/min 180 mGy/min
17IEC Standard 2000
18IEC standard on Interventional Radiology
- Radioscopically guided invasive (and
interventional) procedures. - Interventional reference point.
- Isokerma maps shall be provided.
- The anti-scatter grid should be removable without
the use of tools. - Dosimetric indications reference air kerma rate,
cumulative reference air kerma. cumulative area
kerma product, (shall be accurate to within ?
50 ). - Supplementary indications cumulative time of
radioscopy, cumulative number of radiographic
irradiations, integrated reference air kerma.
19Transmission chamber and collimators (Siemens
system)
20(No Transcript)
21- Collimation Dual-shape collimators incorporating
both circular and elliptical shutters may be used
to modify the field for cardiac contour
collimation. Partially absorbent contoured
filters are also available to control the bright
spots produced by the lung tissue bordering the
heart.
22Philips systems
23Example of the influence of wedge filter in the
skin dose (Vano)
24- Image intensifiers. Because of the necessity of
imaging large fields (e.g., for ventriculography,
aortography) as well as small fields (coronary
arteries), multimode (double or triple) cesium
iodide image intensifiers are recommended.
Formats available vary with the manufacturer but
are typically 9 in/ 6 in/4.5 in (9/6/4.5), 9/6,
10/4, and 9/5.
25- Patient and Equipment Support The ability to
obtain very steep sagittal plane angulation (in
excess of 45 ) is desirable. - An image intensifier with a diameter of more than
9 in is not recommended for cardiac
catheterization laboratories because its size
interferes with the ability to obtain steep
sagittal angulation.
26- The operator should be made aware of the
cumulative amount of exposure time during the
procedure. - In training programs there should be a limit to
the amount of fluoroscopic time granted to a
trainee to complete a specific task, based on a
number of considerations such as the progress
being made and the complexity of the procedure.
27- A freely movable lead glass or acrylic shield
suspended from the ceiling should be used. Its
sterility may be maintained by using disposable
plastic covers. - Each procedure room should have a detailed
determination of exposure levels performed by a
qualified radiation physicist. - There is a tendency in the busy laboratory to
assign a low priority to preventive maintenance
and quality assurance inspections.
28AAPM-70 (2001)
- The generator should be capable of generating 80
to 100 kilowatts (kW) of power. - The generator design should result in square
wave kVp pulses to achieve optimum patient dose
savings.
29AAPM-70 (2001)
- Several manufacturers are using relatively thick
copper filtration and reduced kVp during
fluoroscopy to generate an energy spectrum better
matched to the K-edge of iodine contrast media. - This technique requires high fluoroscopic tube
currents with the benefit of reducing patient
exposure to radiation while improving image
contrast.
30AAPM-70 (2001)
- For adult studies, a 9 to 11 inch (23 to 27 cm)
size is used. - Pediatric cardiac studies use smaller FoVs due to
the small size of the pediatric heart. - The 4.5 inch (11 cm) FoV would be commonly
employed for most pediatric imaging studies.
31AAPM-70 (2001). Pediatrics.
- If the cath lab will serve the pediatric
population, the generator design should allow
high quality imaging on patients which range in
size from 3 to 140 kilograms (kg). - This wide range of patient size places additional
demands on the design of the generator.
32AAPM-70 (2001). Pediatrics.
- The generator design should allow the mAs loading
of the tube per cine pulse to be varied from as
little as 0.1 mAs (100 mA and 1 msec) up to 6 mAs
(e.g., 800 mA and 7 msec) as a function of
patient size in order to maintain a kVp operating
range of 65 to 75 kVp.
33AAPM-70 (2001). Pediatrics.
- Cine frame rate capability should extend up to at
least 60 fps for small children. - The generator should support an x-ray tube with a
minimum of three focal spots. Patients up to 3 to
4 years old can be imaged with an 0.3 mm focal
spot size, and patients up to - 8 to 9 years old can be imaged with cine using an
0.6 mm focal spot.
34AAPM-70 (2001). Pediatrics.
- The 0.3 mm focal spot can also be used on small
children by removing the anti-scatter grid and
employing a geometric magnification factor up to
2. - The geometrical magnification method for small
children can also reduce patient dose because the
electronic magnification modes of the image
intensifier are avoided and the Bucky factor due
to the grid is eliminated.
35Cardiology equipmentkey topics
36Key topics
- Spatial beam modulation collimation (and virtual
collimation), wedge filters, etc. - Temporal beam modulation pulsed fluoroscopy
(grid controlled, temporal integration, etc). - Beam quality modulation extra filtering (Cu, Ta,
etc). - Last image hold.
- Patient dose measurement, display and archive.
- New detectors (dynamic flat panel), connectivity
and DICOM compliance.
37Control panel indications (Siemens system)
38X-ray room indications (Siemens system)
39Key topics
- Ergonomy in the room and system geometry.
- Accidental stop of the system.
- Dosimetric indications in the system and inside
the cath lab. - Protective tools in the system.
- Operational modes and how they are settled.
- DICOM header information.
- On line audit possibilities.
40High filtration
- The introduction of additional filtration in the
X ray beam (commonly copper filters) reduce the
number of low energy photons and as consequence,
saves skin dose for the patients.
41Reduction of Radiation Exposure with extra
filtration
- Additional Cu filters can reduce the skin dose by
more than 70. - Some systems affer variable extra filtration (0.2
mm - 0.9 mm) that is automatically set according
to patient weight and angulation of the C-arm. - Automatic filter insertion try to keep the dose
as low as possible without degrading image
quality.
42Pulsed fluoroscopy
- Pulsed fluoroscopy can be used as a method of
reducing radiation dose, particularly when the
pulse rate is reduced. - But pulsed fluoroscopy does not mean that dose
rate is lower in comparison with continuous
fluoroscopy!!. - Dose rate depends of the dose per pulse and the
number of pulses per second.
43Reduction of Radiation Exposure with virtual
collimation
- Radiation-free Collimation.
- Manipulation of diaphragms in Last Image Hold.
- No fluoroscopy required.
44 Example of X-ray system setting
45APR Philips Integris H5000 Room 3, December 2001
Fluo low
Fluo med
Fluo high
46IAEA survey 2001-2003
- X-ray systems evaluated
- 9-15 from 5 countries
47Very different dose/frame values have been found
48Also different dose increase with patient
thickness
49Also different dose increase with fluoroscopy
modes
50Conclusions from the IAEA survey
- Patient dose and image quality depend largely on
the settings made at the commissioning of the
radiological equipment. - For different systems and different operation
modes, entrance air kerma can increase by a
factor of 20 (including electronic magnification)
for the same patient thickness.
51Conclusions from the IAEA survey
- Increasing phantom thickness increases dose by an
additional factor of up to 12. - Differences in radiation doses from the evaluated
systems show a potential for dose reduction
whilst maintaining image quality.
52Importance of testing X ray equipment
- Characterization of the X- ray system, that
should be part of the acceptance and status
tests, should inform cardiologists about the dose
rates and dose/frame for the different operation
modes and for the different patient thicknesses.
Image quality shall also be evaluated. - Regular constancy checks should verify if
important changes could been occurred.
53Examples of patient dose reports
54Example of the data included in the study report
(Siemens)
55Example of the data included in the dosimetric
report (Philips) Philips Integris
5000 Coronary angiography 65 cine 35
fluoroscopy 13 series, 728 frames 1,54
Gy.cm2/min 0,368 Gy.cm2/10 fr 1 min fluoroscopy
39 fr 3 s cine
56Examples of information contained at the DICOM
header
57(No Transcript)
58(No Transcript)
59(No Transcript)
60Summary
- What to look for while establishing a cath lab
- FDA, IEC, ACR recommendations
- Specific aspects in paediatrics
- Examples of patient dose reports
- Dose variation in cine fluoro- IAEA survey
- DICOM header information