Title: Respironics BiPAP-Vision


1
Respironics BiPAP-Vision
2
Classification
  • Electrically powered internal battery powers
    Vent Inoperative audible alarms if AC power
    lost. Error code 205 if internal low battery
    condition.
  • 2 modes
  • CPAP
  • S/T
  • Set Rate
  • IT in seconds
  • IPAP
  • EPAP

3
Circuit
  • BiPAP circuit (with exhalation port or Whisper
    Swivel)
  • Humidifier or HME
  • With heated humidifier need circuit with water
    traps. Cannot cut into regular BiPAP circuit to
    add traps.

4
Start Up
  • Plug into AC mains Connect to 50 psi gas
    source
  • Connect
  • BiPAP circuit
  • - dont cap
  • Power switch
  • RU corner
  • on back
  • panel
  • ON

Dispos-able inlet filter
O2 module
Fuses
5
(No Transcript)
6
Power-on-self-test (POST) initiates
7
After POST, Start Test Exhalation Port
8
Follow steps on screen, block circuit, Start
Test.
Exhalation (castle) port
9
Exhalation port test in progress
10
Exhalation Port Test
  • Analyzes leak rate of exhalation port
  • If successful the value of Pt. Leak will be
    displayed
  • If unsuccessful Tot. Leak will be displayed
  • EPT must be performed when using the Plateau
    Exhalation Valve because the leak is
    significantly different than the port or swivel.

11
Test Complete, press Monitoring to exit.
Monitoring
12
Trigger Criteria in S/T
  • Time Trigger (Rate)
  • Flow Trigger
  • 6 ml inspired volume detected
  • Patient flow exceeds the shape signal (a shadow
    of the flow signal delayed by 300 ms)

13
In S/T mode
  • Only patient (flow) triggered breaths are flow
    cycled (PS).
  • Time triggered breaths are time cycled according
    to the set Inspiratory Time.

14
Cycle Criteria in S/T
  • Time cycled
  • Set Inspiratory Time
  • A 3 sec maximum IPAP time has occurred
  • Flow cycled
  • Flow reversal detected
  • Patient flow is less than signal flow
  • Spontaneous Expiratory Threshold achieved
    another way to flow cycle

SET
15
Alarm Silence
21
IPAP 8
ALARMS
Mode/Messages
EPAP 3
Alarm Reset
Rate 12
SB
Pt Data Display Area
options
Monitoring
Fz/ UnFz
Scale
Parameters
Alarms
Mode
16
To change settings, press Parameters then soft
key next to parameter and adjust with knob.
Soft keys
Knob
17
Can set Parameters while in Stand-by
18
Stand-by Mode
  • Can set parameters and alarms while in SB
  • Bumping the circuit may simulate pt inspiration,
    in which case Vision exits SB automatically and
    goes into ventilation mode.

19
To change Mode, press Mode, then parameter soft
key to adjust, and then Activate New Mode soft
key
20
Alarms displayed in Mode/Message Area
ALARM
Lo MinVen
21
Alarms
  • Lo Min
  • Vent
  • Hi Rate
  • Lo Rate
  • Hi Press
  • Lo Press
  • Lo Press
  • Delay
  • Apnea
  • Learn
  • Base
  • Flow

22
Cannot adjust Parameters when Monitoring displayed
23
IPAP Rise Time
Timed Inspiration
24
Can Change Bar Graphs to Waveforms with Options
soft key
25
Mini-heart Nebulizer Tx