Title: MiniFlex Training
1(No Transcript)
2MiniFlex Training
- Instructions on the care and operation of the
Rigaku MiniFlex X-ray Diffractometer
3Topics to be covered
- X-ray Safety
- System Maintenance
- Sample Preparation
- Data Collection
- Data Processing using MDI JADE
4X-ray Safety
5Hazards of X-ray Radiation
- X-rays are energetic electromagnetic radiation
that ionize matter by ejecting electrons from
atoms - The extent of ionization, absorption and
molecular change depends upon the quality
(spectral distribution) and quantity (flux and
intensity) of the radiation
6Hazards of X-rays (contd)
- Living organisms can be injured by exposure
- Type and extent of injury is a function not only
of quality and quantity of radiation, but also
duration of exposure and distance from source
7Hazards of X-rays (contd)
- X-rays are INVISIBLE, so its impossible to see
the path they take - However, the strongest intensity lies along the
path of the DIRECT BEAM - Intensity from a scattered beam is MUCH LESS than
the direct beam
8Inverse Square Law applies. Intensity decreases
with the square of the distance.
9MiniFlex Hardware
- How to identify various system components
10MiniFlex
11Water Chiller
12X-ray Tube Cross-section
Here is a picture of the glass insulated tube,
sometimes called the European Standard tube. It
comes in Long and Short Anode versions. The
MiniFlex uses the Long Anode version. The Anode
is a copper block onto which is welded the
target. If no material is added the tube is
copper. Other typical target materials are
chromium, cobalt, molybdenum, and tungsten.
13MiniFlex (with the hood up)!
14Tube Height Adjustment
15Divergence Slit andSample Stage
16Diffracted Beam Opticsand Detector
17Kß Filter Only
18Absorber on Kß Filter
19Correction Angle(adjust with every stage change)
- The Correction Angle is determined by scanning
over a major peak of a standard (like Si 111
which occurs at 28.4422? for Cu K? radiation)
and calculating the correction. - Si 111 for Chrome42.832 2? for Fe35.965
Co33.151Mo 12.989
20Correction Angle (contd)
- In MANUAL MEASUREMENT Under Control select
Measurement - Setup scan as above
- Press Execute
21Correction Angle (contd)
Press Execute under Peak Search to locate top of
peak
22Correction Angle (contd)
- Correction Peak Position - 28.44 0.02
- Enter 0.02 as Correction Angle in Setting
- Press Execute
- Re-measure
23Correction Angle (contd)
Peak is within 0.01 of reference value for
standard
24Detector HV/PHA
- The Pulse Height Analyzer (PHA) is hard wired
and cannot be adjusted. - Detector High Voltage (HV) is adjusted to give
the greatest intensity.
25Detector HV/PHA (contd)
- In Manual Measurement Under Control select
Independent mode. Select 28.44 as the target
angle (for Cu tube). Press Execute. - Press SCHV/PHA.
26Detector HV/PHA (contd)
- Setup HV Scan as indicated below.
27Detector HV/PHA (contd)
Press Execute under Peak Search to locate HV value
28Detector HV/PHA (contd)
Enter HV value in Control Setting
29MiniFlexSystem Maintenance
- Procedures to ensure safe operation and optimal
performance
30MiniFlex System Maintenance
- Cleaning
- Chiller
- MiniFlex
- System Checks
- Log Records
31Cleaning the Chiller
- Check the grill for dust buildup. Remove with a
paint brush if necessary. - Check water. If it looks cloudy due to green or
brown algae, flush and change with fresh
DISTILLED WATER.
32Cleaning the Chiller (contd)
- Never add TAP or DEIONIZED WATER. Tap water
contains minerals that may clog a filter inside
the tube cap. Deionized waters pH can actually
damage the x-ray tube.
33Cleaning the MiniFlex
- Check for dust and dirt buildup
- Card cage
- Floor of measuring chamber
34Cleaning the MiniFlex
- To clean dust from card cage
- Use a can of compressed air to blow out as much
as possible. If big particles remain you can
call Rigaku Service to remove them or carefully
remove the board (POWER OFF!) and do it yourself.
35Cleaning the MiniFlex
- Cleaning the floor of the measuring chamber
- Usually consists of sample debris
- Recommend using rubber glover and face mask
- Use a piece of paper and a small brush
- Carefully sweep the debris away using the paper
as a dustpan.
36System Checks
- System should be checked
- Weekly A sample (like Silicon) should be checked
for - Peak Position of 111 reflection (28.442?)
- Intensity of 111 reflection
- Peak Width of 111 reflection
- Log results along with operator and date
37 System Checks (contd)
- Monthly Same standard is thoroughly checked for
- Weekly check plus
- Scan from 25 to 902?
- Use JADEs Theta Calibration (Linear Fit) to
determine Calibration Curve
38Sample Preparation
- How to prepare various kinds of samples to obtain
optimal data
39Applications
X-ray Diffraction (XRD) analyzes a wide variety
of solid samples. Why solid? Because it needs
the interaction of x-rays with the crystalline
structure. Most solids have a crystalline
structure that is stable. Glasses are the
exception, but even they have structure. Some
examples where XRD is used Chemicals Geology
(mineralogy, oil exploration) Metallurgy Polymers
Catalysts
40XRD Information
- Each diffraction pattern contains sets of
information - Set of local diffraction maximum positions
- Set of intensities for those positions
- Set of intensity distribution as a function of
the diffraction angle (2?)
41Common Errors found in XRD Data
- Sample Displacement
- Sample Transparency
- Sample Flatness
- Sample Particle Size
- Axial Divergence
- Preferred Orientation
42Sample Displacement
- Probably the most common and largest source of
error in diffraction data. - Sample sits above or below Measuring Circle
- Shifts peak position 0.01 for every 60 ?m
43Sample Displacement (contd)
Measuring Circle
44Sample Displacement (contd)
45Sample Transparency
- Error caused by diffraction from below the
surface of a low absorbing sample - Asymmetrically broadens peaks
- Change in 2? is a function of sin2?
- Maximum effect at 902?
46Sample Transparency (contd)
Measuring Circle
47Sample Transparency (contd)
48Sample Flatness
- Sample surface should be as Flat as possible
- Measuring circle changes curvature with 2?
- Peaks shift towards lower 2?
- Broadens peaks asymmetrically
- Maximum effect at small 2? angles
49Sample Flatness (contd)
Measuring Circle
50Sample Flatness (contd)
51Particle Size
- Particles must be small enough to
- allow x-rays to penetrate to core
- pack to provide a flat surface
- Usually that means -325 mesh (lt45?m), but if the
sample has heavy elements -500 mesh may be
required - Spinning the sample helps
52Axial Divergence
- X-rays travel transverse as well a straight,
which means they travel a longer distance - This error can be reduced with finer soller slits
(hardware) (but not on the MiniFlex) - Maximum at low 2?. Minimum at 90 2?.
53Axial Divergence (contd)
- Soller Slits prevent x-rays the deviate more than
X from straight from being counted
54Preferred Orientation
- Samples crystals line-up or stack in such a way
that certain reflections appear more intense than
normal - Typical in clays and micas
- Spinning sometimes helps
- Special sample preparation may be required
55Sample Holders
56Zero Background Holder
57Top Loading Holder
58Back Filled Holder
59Steps for Preparing a Back Filled Holder
- Place empty holder on a flat surface, like a
large glass slide - Pour in just enough sample to fill cavity.
ESTIMATE! - Using a second glass slide pack the sample into
the cavity as tightly as possible
60Steps for Preparing a Back Filled Holder (contd)
- While holding onto both slides plus sample,
INVERT carefully. - Remove top slide exposing the surface that was on
the bottom. - If sample holds in place, remove bottom slide.
Otherwise, insert holder plus bottom slide into
system
61MiniFlex Data Collection
- How to collect XRD data using the Standard
Measurement program
62Standard Measurement Icon
- Locate this icon on the screens desktop
- Double-click to start program
63Standard Measurement
- Press Measurement condition to setup conditions
Main Screen
64Measurement Conditions
Measurement Method CONTINUOUS degrees
per Minute. Column header reads Scan Speed
Continuous Measurements
65Measurement Conditions
Measurement Method FIXED TIME Seconds dwell per
step. Column header reads pre-set time.
Fixed Time Measurements
66Measurement Conditions
- Click on scan to duplicate
- Check Dupli
- Press Execute
Duplicate Scan
67Measurement Conditions
- Click on scan to delete
- Check Delete
- Press Execute
Delete scan
68Setup Samples
- Make sure Use is set to YES
- Specify Directory name, if different
- File name must end with a .raw extension for
JADE to read the file.
Setting up First Sample
69Setup Samples
- Select sample to duplicate
- Check Dupli
- Press Execute
- Complete File name, Sample name, and Cond. No.
fields
Setting up Additional Samples
70Setup Samples
- Click on sample to delete
- Check Delete
- Press Execute
Deleting samples
71Setup with Sample Changer
Press here to setup Spinning
- Specify samples position on carousel (1-6).
- Samples do not have to be in order
72Start Data Collection
- Press Measurement Start to start data collection
73Data Collection Tips
- Pre-Scan sample to determine efficient data
collection conditions (2-80 2?, 0.05 step at
20 /minute). - To increase counting statistics, count 5 times
longer. - To reduce noise, do 5 quick repeats. Crystalline
signal is constant noise signal is variable.
Average scans together in JADE.
74Utilize The Extra Scan Conditions
- There are 50 measuring conditions and 20
individual conditions for each1000 scans. - If you have multiple users, consider giving each
person a separate condition number. - Set up frequently used conditions and turn them
On or Off as needed. - Keep Standards, such as the Weekly or Monthly Si
(111) on a separate condition.