Title: Analytical method validation for FDA compliance
1????? ??
2Method Validation
- Validation of analytical procedures is the
process of determining the suitability of a given
methodology for providing useful - analytical data.
- J. Guerra, Pharm. Tech. March 1986
- Validation is the formal and systematic proof
that a method compiles with the requirements for
testing a product when - observing a defined procedures.
- G. Maldener, Chromatographia, July 1989
3- Method validation is the process of demonstrating
that analytical procedures are suitable for their
intended use and that they support the identity,
strength, quality, purity and potency of the - drug substances and drug products
- Method validation is primarily concerned with
- identification of the sources of potential errors
- quantification of the potential errors in the
method - An method validation describes in mathematical
and quantifiable terms the performance
characteristics of an assay
4Examples of Methods That Require Validation
Documentation
- Chromatographic Methods - HPLC, GC, TLC, GC/MS,
etc. - Pharmaceutical Analysis - In support of CMC.
- Bioanalytical Analysis - In support of
PK/PD/Clinical Studies. - Spectrophotometric Methods UV/VIS, IR, NIR, AA,
NMR, XRD,MS - Capillary Electrophoresis Methods - Zone,
Isoelectric Focusing - Particle Size Analysis Methods - Laser,
Microscopic, Sieving, SEC, etc. - Automated Analytical Methods - Robots, Automated
Analysis.
5Considerations Prior to Method Validation
- Suitability of Instrument
- Status of Qualification and Calibration
- Suitability of Materials
- Status of Reference Standards, Reagents, Placebo
Lots - Suitability of Analyst
- Status of Training and Qualification Records
- Suitability of Documentation
- Written analytical procedure and proper approved
protocol with pre-established acceptance criteria
6Validation Step
- Define the application, purpose and scope of the
method. - Analytes? Concentration? Sample matrices?
- Develop a analytical method.
- Develop a validation protocol.
- Qualification of instrument.
- Qualify/train operator
- Qualification of material.
- Perform pre-validation experiments.
- Adjust method parameters and/or acceptance
criteria if necessary. - Perform full validation experiments.
- Develop SOP for executing the method in routine
analysis. - Document validation experiments and results in
the validation report.
7Purpose of Method Validation
- Identification of Sources and Quantitation of
Potential errors - Determination if Method is Acceptable for
Intended Use - Establish Proof that a Method Can be Used for
Decision Making - Satisfy FDA Requirements
8What is not Analytical Method Validation?
- Calibration
- The Process of Performing Tests on Individual
System Components to Ensure Proper function - For example) HPLC Detector calibration
- Wavelength Accuracy/ Linear Range/ Noise Level/
Drift
9- System Suitability
- Test to verify the proper functioning of the
operating system, i.e., the electronics, the
equipment, the specimens and the analytical
operations. - Minimum Resolution of 3.0 between the analyte
peak and internal standard peaks - Relative Standard Deviation of replicate standard
injections of not more than 2.0
10System Suitability
Validation
Calibration
Pump
Injector
Detector
Data System
Method
Analyst
Sample
11Method Life Cycle
12Verification vs. Validation
- Compendial vs. Non-compendial Methods
- Compendial methods-Verification
- Non-compendial methods-Validation requirement
13Compendial Analytical Procedures
- The Analytical procedures in the USP 25/NF 20 are
legally recognized under section 501(b) of the
Federal Food, Drug and Cosmetic Act as the
regulatory analytical procedures for the
compendial items. The suitability of these
procedures must be verified under actual
conditions of use. When using USP 25/NF 20
analytical procedures, the guidance recommends
that information be provided for the following - characteristics
- Specificity of the procedure
- Stability of the sample solution
- Intermediate precision
14Published Validation Guidelines
- 1978 Current Good Manufacturing Practices
(cGMPs) - 1987 FDA Validation Guideline
- 1989 Supplement 9 to USP XXI
- 1994 CDER Reviewer Guidance
- Validation of Chromatographic
Method - 1995 ICH Validation Definitions
- Q2A, Text on Validation of
Analytical procedures - 1997 ICH Validation Methodology
- Q2B, Validation of Analytical
Procedures Methodology - 1999 Supplement 10 to USP 23 lt1225gt Validation
of Compendial Methods - 1999 CDER Bioanalytical Method Validation for
Human Studies - 2000 CDER Draft Analytical Procedures and
Method Validation
15Regulatory and Compliance Requirements Review
- Validation of an analytical method is the process
by which it is established, by laboratory
studies, that the performance characteristics of
the method meet the requirements for the intended
analytical - applications
USP 23 General Information lt1225gt
16- The accuracy, sensitivity, specificity, and
reproducibility of test methods employed by the
firm shall be established and documented. Such
validation and documentation may be accomplished
in accordance with 211.194(a)(2).
21 CFR PART 211 - CURRENT GOOD MANUFACTURING
PRACTICE FOR FINISHED PHARMACEUTICALS Subpart
I-Laboratory Controls 211.165 Testing and
release for distribution (e)
17- The objective of validation of an analytical
procedure is to demonstrate that it is suitable - for its intended purpose
ICH Guideline for Industry Q2A, Text on
Validation of Analytical Procedures March 1995
18- In practice, it is usually possible to design the
experimental work such that the appropriate
validation characteristics can be considered
simultaneously to provide a sound, overall
knowledge of the capabilities of the analytical
procedure, for instance Specificity, Linearity,
Range, Accuracy, and - Precision.
ICH Guideline for Industry Q2B, Validation of
Analytical Procedures Methodology
19Todays Validation Requirements
20ICH/USP Validation Requirements Parameters
ICH
USP
- Specificity
- Linearity
- Range
- Accuracy
- Precision
- Repeatability
- Intermediate Precision
- Reproducibility
- Limit of Detection
- Limit of Quantitation
- Specificity
- Linearity and Range
- Accuracy
- Precision
- Limit of Detection
- Limit of Quantitation
- Ruggedness
- Robustness
21USP Data Elements Required For Assay Validation
May be required, depending on the nature of the
specific test.
22USP Categories
- Category 1 Quantitation of major components or
- active ingredients
- Category 2 Determination of impurities or
- degradation products
- Category 3 Determination of performance
- characteristics
23ICH Validation Characteristics vs. Type of
Analytical Procedure
24Specificity/Selectivity
- Ability of an analytical method to measure the
analyte free from interference due to other
components. - Selectivity describes the ability of an
analytical method to differentiate various
substances in a sample - Original term used in USP
- Also Preferred by IUPAC and AOAC
- Also used to characterize chromatographic columns
- Degree of Bias (Used in USP)
- The difference in assay results between the two
groups - the sample containing added impurities,
degradation products, related chemical compounds,
placebo ingredients - the sample without added substances
25Specificity Impurities Assay
- Chromatographic Methods
- Demonstrate Resolution
- Impurities/Degradants Available
- Spike with impurities/degradants
- Show resolution and a lack of interference
- Impurities/Degradants Not Available
- Stress Samples
- For assay, Stressed and Unstressed Samples should
be compared. - For impurity test, impurity profiles should be
compared.
26Forced Degradation Studies
- Temperature (50-60?)
- Humidity (70-80)
- Acid Hydrolysis (0.1 N HCl)
- Base Hydrolysis (0.1 N NaOH)
- Oxidation (3-30)
- Light (UV/Vis/Fl)
- Intent is to create 10 to 30 Degradation
27Linearity
- Ability of an assay to elicit a direct and
proportional response to changes in analyte
concentration.
28Linearity Should be Evaluated
- By Visual Inspection of plot of signals vs.
analyte concentration - By Appropriate statistical methods
- Linear Regression (y mx b)
- Correlation Coefficient, y-intercept (b), slope
(m) - Acceptance criteria Linear regression r2 gt 0.95
- Requires a minimum of 5 concentration levels
29Range
- Acceptable range having linearity, accuracy,
precision. - For Drug Substance Drug product Assay
- 80 to 120 of test Concentration
- For Content Uniformity Assay
- 70 to 130 of test Concentration
- For Dissolution Test Method
- /- 20 over entire Specification Range
- For Impurity Assays
- From Reporting Level to 120 of Impurity
Specification for Impurity Assays - From Reporting Level to 120 of Assay
Specification for Impurity/Assay Methods
30Accuracy
- Closeness of the test results obtained by the
method to the true value.
31Accuracy
- Should be established across specified range of
analytical procedure. - Should be assessed using a minimum of 3
concentration levels, each in triplicate (total
of 9 determinations) - Should be reported as
- Percent recovery of known amount added or
- The difference between the mean assay result and
the accepted value
32Accuracy Data Set (1 of 3)
33Precision
- The closeness of agreement (degree of scatter)
between a series of measurements obtained from
multiple samplings of the same homogeneous
sample. - Should be investigated using homogeneous,
authentic samples.
34Precision Considered at 3 Levels
- Repeatability
- Intermediate Precision
- Reproducibility
35 Repeatability
- Express the precision under the same operating
conditions over a short interval of time. - Also referred to as Intra-assay precision
- Should be assessed using minimum of 9
determinations - (3 concentrations/ 3 replicates) or
- Minimum of 6 determinations at the 100 level.
36 Intermediate Precision
- Express within-laboratory variations.
- Expressed in terms of standard deviation,
relative standard deviation (coefficient of
variation) and confidence interval.
- Depends on the circumstances under which the
procedure is intended to be used. - Studies should include varying days, analysts,
equipment, etc.
37Repeatability Intermediate Precision
Mean 100.5 RSD 0.24
Mean 99.5 RSD 0.36
Grand Mean 100.0 RSD 0.59
38 Reproducibility
- Definition Ability reproduce data within the
predefined precision - Determination SD, RSD and confidence interval
- Repeatability test at two different labs.
- Note Data not required for BLA/NDA
39Detection Limit (LOD)/ Quantitation Limit (LOQ)
- LOD
- Lowest amount of analyte in a sample that can be
detected but not necessarily quantitated. - Estimated by Signal to Noise Ratio of 31.
- LOQ
- Lowest amount of analyte in a sample that can be
quantified with suitable accuracy and precision. - Estimated by Signal to Noise Ratio of 101.
40LOD and LOQ Estimated by
- Based in Visual Evaluations
- - Used for non-instrumental methods
- Based on Signal-to Noise-Ratio
- - 31 for Detection Limit
- - 101 for Quantitation Limit
- Based on Standard Deviation of the Response and
the Slope
41LOD and LOQ Estimated by
- S slope of calibration curve
- s standard deviation of blank readings or
- standard deviation of regression line
- Validated by assaying samples at DL or QL
42(No Transcript)
43 Robustness
- Definition Capacity to remain unaffected by
small but deliberate variations in method
parameters - Determination Comparison results under differing
conditions with precision under normal conditions - Examples of typical variations in LC
- Influence of variations of pH in a mobile phase
- Influence of variations in mobile phase
composition - Different columns (different lots and/or
suppliers) - Temperature
- Flow rate
44Ruggedness
- Degree of reproducibility of test results under a
variety of conditions - Different Laboratories
- Different Analysts
- Different Instruments
- Different Reagents
- Different Days
- Etc.
- Expressed as RSD
45ICH/USP System Suitability
- ICH
- Definition evaluation of equipment, electronic,
analytical operations and samples as a whole - Determination repeatability, tailing factor (T),
capacity factor (k), resolution (R), and
theoretical Plates (N)
46- USP 23 lt621gt
- System Suitability Requirements
47??? Re-validation
- When
- Method parameters have been changed
- The scope of the method has been changed
- Synthetic methods have been changed
- Impurity profile has been changed
- What
- Preferably everything. Exceptions should be
- scientifically justified
48How do we Know the expectations of the FDA?
- FDA Form 483
- FDA Warning Letters
- Personal Experiences
49483 Observations
- There was inadequate method validation
specificity data to demonstrate that each method
was capable of distinguishing the active
ingredient from its impurities and degradation
products. - Specificity studies did not include the minimum
stress conditions of acid and base hydrolysis,
oxidation, thermal degradation and photolysis,
degradation schematic for the active ingredient
that identifies the major degradation products - was not included for each product.
50FDA Waning Letter
- On addition to the example of modifying both
compendial methods and customer supplied methods,
we also observed the use of unvalidated in-house
methods as well as unvalidated - modifications to in-house methods.
- A statement indicating that the method has not
been validated in the particular formulation was
included in the certificate of analysis foruse
of this statement does not absolvefrom using
valid, accurate, and - reproducible methods. (June 2000)
51FDA Systems Based InspectionLaboratory System
Feb July 2002 212 Inspections (US)
Reference Albinus D Sa, FDA, CDER Office of
Compliance, from AAPS, Nov. 2002 presentation.
52Related Site
- www.fda.gov
- www.fda.gov/cder/
- www.waters.com
- www.usp.org
- www.ich.org
- www.aoac.org
- www.pharmweb.net