Title: Integrating Dosimeters I
1Integrating Dosimeters I
- Thermoluminescence Dosimetry
2The Thermoluminescence Process Phosphors
- The sensitive volume of a TLD consists of a small
mass (1-100 mg) of crystalline dielectric
material containing suitable activators to make
it perform as a thermoluminescent phosphor - The activators provide two kinds of centers
- Traps for the electrons and holes, which can
capture and hold the charge carriers in an
electrical potential well for usefully long
periods of time - Luminescence centers, located at either the
electron traps or the hope traps, which emit
light when the electrons and holes are permitted
to recombine at such a center
3Phosphors (cont.)
- The following energy-level diagram illustrates
the TL process - At left it shows an ionization event elevating an
electron into the conduction band, where it
migrates to an electron trap - The hole left behind migrates to a hole trap
- At the temperature existing during irradiation,
these traps should be deep enough in terms of
potential energy to prevent the escape of the
electron or hole for extended periods of time,
until deliberate heating releases either or both
of them
4Energy-level diagram of the TL process (A)
ionization by radiation, and trapping of
electrons and holes (B) heating to release
electrons, allowing luminescence production.
5Phosphors (cont.)
- At right in the diagram the effect of such
heating is shown - We will assume that the electron is released
first, that is, that the electron trap in the
phosphor is shallower than the hole trap - The electron again enters the conduction band and
migrates to a hole trap, which may be assumed
either to act as a luminescence center or to be
closely coupled to one - In that case recombination is accompanied by the
release of a light photon
6TL Process Randall-Wilkins Theory
- The simple first-order kinetics for the escape of
such trapped charge carriers at a temperature
T(K) were first described by Randall and Wilkins
using the equation - where p is the probability of escape per unit
time (s-1), ? is the mean lifetime in the trap, ?
is called the frequency factor, E is the energy
depth of the trap (eV), and k is Boltzmans
constant
7Randall-Wilkins Theory (cont.)
- It is evident from this equation that, on the
assumption of constant values for k, E, and ?,
increasing T causes p to increase and ? to
decrease - Thus if the temperature T is scanned upward
linearly vs. time, starting at room temperature,
an increase in the rate of escape of trapped
electrons will occur, reaching a maximum at some
temperature Tm followed by a decrease as the
supply of trapped electrons is gradually exhausted
8Randall-Wilkins Theory (cont.)
- Assuming that the intensity of light emission is
proportional to the rate of electron escape, a
corresponding peak in TL brightness will also be
observed at Tm - This is called a glow peak, as shown in the
following diagram - The presence of more than one trap depth E gives
rise to plural glow peaks, which may be
unresolved or only partially resolved from one
another in the glow curve
9A TL glow curve vs. temperature that results from
the gradual heating of an irradiated
thermoluminescent phosphor that contains two trap
depths.
10Randall-Wilkins Theory (cont.)
- The value of Tm is related to the linear heating
rate q (K/s) by the following relation from R-W
theory - which simplifies to the following approximate
relationship on the assumption of ? 109/s and q
1 K/s - hence Tm 216 C for E 1 eV
11Randall-Wilkins Theory (cont.)
- Tm increases gradually with q, so that Tm 248
C at q 5 K/s, and 263C at 10 K/s for the same
values of ? and E - The light-emission efficiency may be found to
decrease with increasing temperature by a process
called thermal quenching - Thus at higher heating rates some loss of total
light output may be noticed - Aside from this effect, varying the heating rate
leaves total light output constant, preserving
the area of glow curves in terms of brightness
vs. time
12Glow curves vs. time obtained with a CaF2Mn TL
dosimeter at eight linear heating rates. The
dose to the phosphor was adjusted to be inversely
proportional to the heating rate in each case.
13Glow curves vs. temperature recorded
simultaneously with the previous curves
14Randall-Wilkins Theory (cont.)
- It can be seen that employing the light sum
rather than the height of the glow peak as a
measure of the absorbed dose is less subject to
errors caused by fluctuations in the heating rate - However, the peak height may be used if the
heating rate is very stable, and this may be
advantageous in measuring small doses for which
the upper limb of the glow curve rises due to IR
and spurious effects
15TL Process Trap Stability
- The usefulness of a given phosphor trap (and its
associated TL glow peak) for dosimetry
applications depends on its independence of time
and ambient conditions - If the traps are not stable at room temperature
before irradiation, but migrate through the
crystal and combine with other traps to form
different configurations, changes in radiation
sensitivity and glow-curve shape will be observed
16Trap Stability (cont.)
- LiF (TLD-100) is such a phosphor, requiring
special annealing (e.g., 400 C for 1 h, quick
cooling, then 80 C for 24h) to minimize
sensitivity drift - In general TL phosphors give best performance as
dosimeters if they receive uniform, reproducible,
and optimal (depending on the phosphor) heat
treatment before and after use
17Trap Stability (cont.)
- The inability of traps to hold charge carriers at
ambient temperature after irradiation is called
trap leakage, and of course it becomes greater if
the ambient temperature is increased - As a rule of thumb, in typical TLD phosphors a
glow peak at ?200-225 C is ordinarily found to
have small enough leakage for practical
room-temperature dosimetry, having a half-life of
trapped charge carriers measured in months or
years
18Trap Stability (cont.)
- A glow peak at ?150 ?C usually has a half-life
only of the order of a few days, while a 100 ?C
peak decays in a matter of hours - Although short-term dosimetry may still be
possible with rapidly leaking traps, careful
timing control is required
19Trap Stability (cont.)
- Higher-temperature traps than 200-225 C are
usually even more stable, and would be
advantageous for dosimetry except for the
existence of two competing effects - Heat (Infrared) Signal. As the phosphor and its
heating tray rise in temperature, the
short-wavelength tail of the blackbody radiation
begins to extend into the visible region and
produce a non-dose-related response in the PMT
used for measuring the TL light output - Spurious TL Signal. The combined effects of
adsorbed gases, humidity, dirt, and mechanical
abrasion of the phosphor surface tend to produce
a spurious TL emission
20Trap Stability (cont.)
- Flowing an oxygen-free inert gas such as N2 or Ar
through the space above the heater pan, thus
surrounding the phosphor during the TL readout
process, allows the stored energy due to these
surface effects to be released without light
emission - Thus N2 flow is often used to reduce spurious
background TL readings, especially when small
doses are to be measured
21TL Process Intrinsic Efficiency of TLD Phosphors
- Only a small part of the energy deposited as
absorbed dose in a TLD phosphor is emitted as
light when the substance is heated - The ratio (TL light energy emitted per unit
mass)/(absorbed dose) is called the intrinsic TL
efficiency - This has been measured as 0.039 in LiF
(TLD-100), 0.44 in CaF2Mn, and 1.2 in CaSO4Mn
22Intrinsic Efficiency (cont.)
- The energy budget in LiF (TLD-100) has been
estimated to account for the loss of the missing
99.96 of the energy deposited by ionizing
radiation that ultimately goes into heat
production - It should not be surprising that TLDs must be
used under reproducible conditions to obtain
consistent results, considering that such a small
fraction of the absorbed dose energy is relied
upon as a measure of the entire dose
23TLD Readers
- The instrument used to heat a TLD phosphor, and
to measure the resulting TL light emitted, is
simply called a TLD reader - Its design principle is shown schematically in
the following diagram - The TLD phosphor to be measured is placed in the
heater pan at room temperature, and heated while
the emitted light is measured with a
photomultiplier
24Schematic diagram of a typical TLD reader
25TLD Readers (cont.)
- Heating of the sample may be done by means of an
resistively heated pan as shown, or by preheated
N2 gas, or by an intense light spot from a
projection lamp or laser, or other suitable means - Often the heating program may be more complicated
than simply linear vs. time
26TLD Readers (cont.)
- One typical scheme is to heat the phosphor
rapidly through the unstable-trap region,
ignoring light emission until some preset
temperature is reached - Then the phosphor is either heated linearly or
abruptly raised to a temperature sufficient to
exhaust the glow peak of dosimetric interest,
while measuring the emitted-light sum, which is
displayed as a charge or dose reading - Finally, the phosphor may be heated further to
(say) 400 C to release any remaining charge from
deeper traps, while ignoring any additional light
emission
27Typical programmed readout cycle in a modern TLD
reader, consisting of a preheat period without
light integration to discriminate against
unstable low-T traps, a read period spanning
the emission of the part of the glow curve to be
used as a measure of the dose, an anneal period
during which the remainder of the stored energy
is dumped without light integration, and the
cooling-down period after the heater-pan power is
turned off.
28TLD Readers (cont.)
- Heating-program reproducibility is vital in
achieving reproducible TL dosimetry - In addition, one must provide constant light
sensitivity so that a given TLD light output
always gives the same reading - This requires constant PM-tube sensitivity, and a
clean optical system - A constant light source with an appropriate
spectrum may be built into the reader to
substitute for a TLD as a check on the constancy
of light sensitivity
29TLD Phosphors
- TLD phosphors consist of a host crystalline
material containing one or more activators that
may be associated with the traps, luminescence
centers, or both - Amounts of activators range from a few parts per
million up to several percent in different
phosphors - The host crystal almost entirely determines the
radiation interactions, since the activators are
usually present in such small amounts
30TLD Phosphors (cont.)
- Many different TLD phosphors have been studied
and reported in the literature - A few representative ones are listed in the
following table - The following curves show their glow curves at a
heating rate of 40 C/min and their approximate
light output vs. 60Co ?-ray exposure
31Characteristics of TL phosphors
32Glow curves vs. temperature (upper scale) and
time (lower scale) for four TL dosimetry
phosphors. Heating rate 40 C/min. The
amplitudes are arbitrary.
33Glow-peak-area response vs. 60Co ?-ray exposure
for several TL phosphors
34TLD Phosphors (cont.)
- All the phosphors show some degree of
supralinearity of response, this effect being
most pronounced in lithium borate - In CaF2Mn the rise is only ?4 in the
neighborhood of 104 R, which is too small to be
seen on this figure - This supralinearity may be due to the increased
availability of luminescence centers when the
charged-particle tracks become closer together,
or to radiation-induced trap formation, or to
other causes - At large enough doses all TL phosphors either
saturate in their output as all available traps
become filled, or maximize and then decrease due
to radiation damage of the phosphor
35TLD Forms
- The most common forms of TLD are
- Bulk granulated, sieved to 75-150- ?m grain size
- Compressed pellets or chips, usually 3.2 mm
square by 0.9 mm thick - A Teflon matrix containing 5 or 50 by weight of
lt40-?m grain-size TLD powder - A TLD pellet fastened on an ohmic heating element
in an inert-gas-filled glass bulb - Single-crystal plates, cleaved from a larger
grown crystal boule - Powder enclosed in plastic tubing that can be
heated
36Calibration of TLDs Form
- Solid TLD chips or Teflon-TLD discs are the
preferred forms of the phosphor for most
applications - They can be individually identified and
calibrated, they do not require containment, and
they are flat, so that they can be oriented
perpendicular to a monodirectional radiation
beam, thus presenting a known cross-sectional area
37Calibration of TLDs Basis for Calibration
- Most TL phosphors have some threshold dose level
below which the TL light output per unit mass is
proportional to the absorbed dose to the
phosphor, provided that - the LET of the radiation remains low or
practically constant, and - the phosphor sensitivity is kept constant by
using reproducible annealing procedures
38Basis for Calibration (cont.)
- Assuming TL-reader constancy, and negligible
attenuation of light in escaping from the
phosphor during heating, one can then say that
the same TL reading will result from a given
average absorbed phosphor dose in a TL dosimeter,
regardless of the spatial distribution of
absorbed dose within it, so long as the dose
throughout remains in the linear range
39Basis for Calibration (cont.)
- The practical consequence of this is that a 60Co
?-ray calibration in terms of average phosphor
dose in the TL dosimeter can then be used as an
approximate calibration for all low-LET
radiations, including x-rays, ?-rays, and
electron beams of all energies above 10 keV,
even if they deposit dose nonuniformly in the
dosimeter
40Basis for Calibration (cont.)
- Relating the phosphor dose so measured to the
dose in a similar mass of tissue hypothetically
substituted for the TLD requires a separate step
based on cavity theory - For the simplest (B-G) case of a thin TLD and
very penetrating electrons, the dose ratio
Dtiss/DTLD is proportional to the mass collision
stopping-power ratio, (dT/?dx)c,tiss/(dT/?dx)c,TLD
evaluated at the mean electron kinetic energy
41Basis for Calibration (cont.)
- If the incident radiation beam is completely
stopped by the TLD, then the incident energy
fluence can be derived - The 60Co calibration (under TCPE conditions)
gives the TL reading per unit of average phosphor
dose - Multiplying that dose by the mass of the TLD chip
allows relating the TL reading to a given
integral dose, or energy spent in the chip
42Basis of Calibration (cont.)
- If the chip area presented to the beam is A (m2),
its mass is m (kg), and the 60Co ?-ray
calibration factor is kCo (DTLD/r)Co
Gy/(scale division), where r is the TLD
reading, then the energy fluence of a stopped
beam is given by
43Calibration of TLDs 60Co ?-ray Calibration
- For a free-space 60Co ?-ray exposure X (C/kg) at
the point to be occupied by the center of the TLD
in its capsule, the average absorbed dose in the
TLD, in grays, under TCPE conditions is given by - where a is a correction for broad-beam ?-ray
attenuation in the capsule wall plus the half
thickness of the TLD
4460Co ?-ray Calibration (cont.)
- For a LiF TLD chip in a Teflon capsule 2.8 mm in
thickness (for TCPE) the average dose calculated
from this equation is approximately
4560Co ?-ray Calibration (cont.)
- If the resulting TLD reading is r scale
divisions, then the calibration factor is kCo
(DTLD/r)Co, which applies at the dose value used
in calibration and throughout the linear
response-vs.-dose range - For all low-LET radiations, the average absorbed
dose in the TLD can then be obtained from the
observed TLD reading r by
4660Co ?-ray Calibration (cont.)
- For higher-LET radiations than 60Co ?-rays, TLDs
typically show some variation in efficiency, and
consequently a reciprocal change in the low-LET
calibration factor kCo - The following diagram gives the results of
measurements for lithium fluoride, lithium
borate, and beryllium oxide - Their LET dependence is seen to be quite
different, with lithium borate coming closest to
constancy
47LET response of BeO, Li2B4O7Mn, and LiF. The
curves give values of kCo/kLET as a function of
LET in water, in keV/?m.
4860Co ?-ray Calibration (cont.)
- The following diagrams show the photon energy
dependence of lithium fluoride and lithium borate - Curves A were obtained from
- showing the CPE dose in the phosphors per
unit of exposure, normalized to 1.25 MeV (60Co)
49Thermoluminescent response of LiF per roentgen
and per rad for photon energies from 6 to 2800 keV
5060Co ?-ray Calibration (cont.)
- Curves B show the TL response per unit exposure,
and curves C the TL response per unit of absorbed
dose in the phosphors - Thus curves C represent the LET-dependence of the
TL efficiency relative to 60Co, or kCo/kLET
51Thermoluminescent response of Li2B4O7Mn per
roentgen and per rad for photon energies from 6
to 2800 keV
52TLD Advantages
- Specific characteristics vary from phosphor to
phosphor, and are available from the manufacturer - We will describe the most widely used, LiF
(TLD-100), where specifics are referred to - Wide useful dose range, from a few millirads to
103 rad linearly, plus another decade (103-104)
of supralinear response vs. dose. - Dose-rate independence, 0-1011 rad/s.
53TLD Advantages (cont.)
- Small size passive energy storage. Small TLDs
can be used as dose probes with little
disturbance of the radiation field in the medium.
They can be made thin enough to approach B-G
conditions at high energies, but TCPE is easier
to achieve because of their condensed state. - Commercial availability. TLDs and readers are
available from a number of suppliers.
54TLD Advantages (cont.)
- Reusability. By employing appropriate annealing
procedures to release all the prior stored
energy, and checking for possible alteration in
radiation sensitivity, TLD phosphors can normally
be reused many times until they become
permanently damaged by radiation, heat or
environment. Thus it is feasible to calibrate
individual dosimeters. - Readout convenience. TLD readout is fairly rapid
(lt 30 s) and requires no wet chemistry.
55TLD Advantages (cont.)
- Economy. Reusability usually reduces cost per
reading. - Availability of different types with different
sensitivities to thermal neutrons. - Automation compatibility. For large
personnel-monitoring operations automatic readers
are available. - Accuracy and precision. Reading reproducibility
of 1-2 can be achieved with care. Comparable
accuracy may be obtained through individual
calibration and averaging of several dosimeters
in a cluster, since their volume is small.
56TLD Disadvantages
- Lack of uniformity. Different dosimeters made
from a given batch of phosphors still show a
distribution of sensitivities, and different
batches of phosphor generally have different
average sensitivities. Thus individual dosimeter
calibration, or at least batch calibration, is
necessary for acceptable accuracy and precision.
57TLD Disadvantages (cont.)
- Storage instability. TLD sensitivity can vary
with time before irradiation in some phosphors,
as a result, for example, of gradual
room-temperature migration of trapping centers in
the crystals. Controlled annealing of the TLDs
can usually restore them to some reference
condition again.
58TLD Disadvantages (cont.)
- Fading. Irradiated dosimeters do not permanently
retain 100 of their trapped charge carriers.
This results in a gradual loss of the latent TLD
signal. This must be corrected for, especially
in applications (e.g., personnel monitoring) that
involve long time delays.
59TLD Disadvantages (cont.)
- Light sensitivity. TLDs all show some
sensitivity to light especially UV, sunlight,
or fluorescent light. This can cause accelerated
fading, or leakage of filled traps. Or it can
produce ionization and the filling of traps, thus
giving rise to spurious TL readings. - Spurious TL. Scraping or chipping of TLD
crystals or surface contamination by dirt or
humidity also can cause spurious TL readings.
However, the presence of an oxygen-free inert gas
during readout suppresses these signals.
60TLD Disadvantages (cont.)
- Memory of radiation and thermal history. The
sensitivity can be either increased or decreased
after receiving a large dose of radiation and
undergoing readout. Additional annealing
procedures are necessary to restore the original
sensitivity, if possible. It may be more
economical to throw away the phosphor after a
single use, especially for large doses.
61TLD Disadvantages (cont.)
- Reader instability. TLD readings depend on the
light sensitivity of the reader as well as on the
heating rate of the phosphor. Thus reader
constancy is difficult to maintain over long time
periods. - Loss of a reading. The measurement of light out
of a TLD erases the stored information. Unless
special provision is made (e.g., a spare TLD),
there is no second chance at getting a reading.
Reader malfunction can lose a reading.