Multileaf Collimator (MLC) Leaf Intrusion an Optimum Value? Variable MLC Leaf Width? - PowerPoint PPT Presentation

About This Presentation
Title:

Multileaf Collimator (MLC) Leaf Intrusion an Optimum Value? Variable MLC Leaf Width?

Description:

Multileaf Collimator MLC Leaf Intrusion an Optimum Value Variable MLC Leaf Width – PowerPoint PPT presentation

Number of Views:1096
Avg rating:3.0/5.0
Slides: 13
Provided by: mewe
Category:

less

Transcript and Presenter's Notes

Title: Multileaf Collimator (MLC) Leaf Intrusion an Optimum Value? Variable MLC Leaf Width?


1
Multileaf Collimator (MLC) Leaf Intrusion an
Optimum Value?Variable MLC Leaf Width?
  • Martin E Welch
  • Presented at the AAPM meeting 2001

2
Quality Assurance (QA)
  • Conformal fields created by the use of an (MLC)
    are only as accurate as the calibration of each
    Leaf. As treatments are performed at all Gantry
    angles the QA should be performed at the four
    cardinal points. Thus taking into account any
    Leaf droop or sag, and must be performed over the
    range of Leaf movement.
  • The Light Field Test Shape shown here, is one of
    three that allows this test to be performed and a
    hard copy (film) to be taken and checked in the
    same time as a standard 10x10 field. A Light
    Field Template of these fields is used to check
    the film.

3
Effect of Leaf Calibration Errors at the Central
Axis
  • The effect of Leaf calibration errors is easily
    seen on these two circular fields.
  • The errors seen are only due to Leaf calibration
    and not rotational displacement.
  • Although the calibration error will be constant
    over the Leaf range of travel. The effective
    error will be greater as the treatment field gets
    smaller.

4
Leaf Tip to Contour Angle
  • The angle that the Leaf tip is presented to the
    required contour defines the fit of the produced
    isodose.
  • As the angle increases the effective control
    decreases.
  • With a curved contour the problem is greater due
    to the error changing per Leaf.
  • For most shapes the control of the required
    isodose is established by the use of the Leaf tip
    and edges. It is the ratio of these that
    determines the effective control

5
The Effect of Depth on the Scalloped edge.
  • The Scallop effect or the dip and peak in the
    isodose produced by the width of the Leaf, does
    change with depth.
  • Figures a/b/c 6MV show results of a triangular
    shape measured in water at depths of Dmax/
    D10/D15. The effect is clearly visible.
  • Figures a2/b2 show the same field for 15Mv. To
    minimise variables no correction was made for
    penumbrae and the Leaves were not moved. The 80
    isodose represents 95 had the Leaves been moved.
    The effect may still be seen at D10 a typical
    patient separation.

6
The Effect of Depth on the Scalloped edge.
  • The figure on the right shows the scallop effect
    reducing as the field size increases.
  • Control of small fields such as head and neck
    work are significantly affected by the Leaf width
    and the scallop effect.

7
The Phased Field Technique Royal Free Hospital.
  • This technique has been in clinical use at this
    hospital since 1993. It is the basis of the
    Siemens HD270 MLC.
  • It is a method of artificially reducing the Leaf
    width at any point in the field.
  • Two fields are normally used. One field centre
    being shifted by 5.0mm.
  • There is no increase in treatment time as each
    field is 50 of applied dose.

8
Effect of Leaf Tip to Edge Ratio Applied to a
Clinical Field
  • The collimator angle is selected for optimum
    control.
  • Four Leaf Tip to Edge ratios are demonstrated.
  • Effect on penumbrae is easily seen on this
    clinical shape.

9
Leaf Intrusion Values of 0 and 50 applied to a
shape in three field sizes
  • The 11.0cm field demonstrates that 50 intrusion
    does give the best control.
  • The 5.0cm field produces a poor fit for both
    values.
  • The 3.0cm field suggests that the lesser evil is
    50 intrusion

10
A Demonstration of three Leaf Intrusion Values
applied to the same field
11
This Easily Recognisible Shape contains many of
the problems encountered with its clinical
counter parts.
12
Conclusions
  • The ratio of Leaf Tip to Edge controlling the
    isodose greatly affects the result.
  • If this ratio approaches 50 the field at this
    point is larger than required.
  • An intrusion value of 25 gives best all round
    control.
  • The Phased Field Technique applied with this
    value of intrusion enhances the effective
    control.
  • This same technique reduces the interleaf leakage
    by a factor equal to the number of fields
    employed ie two fields equals a 50 reduction.
Write a Comment
User Comments (0)
About PowerShow.com