Title: Servoi%20ventilator
1Servoi ventilator
- Charles Gomersall
- Dept of Anaesthesia Intensive CareThe Chinese
University of Hong KongPrince of Wales Hospital
Version 1.1
2Disclaimer
- Although considerable effort has been taken to
ensure that the material in this tutorial
accurately reflects the information provided by
Maquet AB in its publications the author takes no
responsibility for the accuracy of the
information in this tutorial nor for any adverse
event that occurs as a result of using this
tutorial. Users are advised to consult the manual
that accompanies their ventilator. Please note
that this tutorial has not been endorsed by
Maquet AB.
3Navigation
Whats different from a 300A?
4User interface a quick tour
Whats different from a 300A?
Setting the mode
Modes
Changing parameters
Alarms
Advanced
Starting up
Click on Maquet to end the tutorial at any stage
5A quick tour
6Menu touch pad
7Using the menu touch pad
ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Additionalsettings
Additionalvalues
8ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Additionalsettings
Additionalvalues
9ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Set ventilation mode
Volume control ?
Automode
Ti 1.33 s (33)
Basic
IE
Trigger
.
Tidal volume 500
IE12.0
Trigger sensitivityV
Resp. Rate15
T. pause10
PEEP5
T. Insp. rise5
O2 conc.40
Additionalvalues
Cancel
Accept
10ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Set ventilation mode
Volume control ?
Automode
Ti 1.33 s (33)
Basic
IE
Trigger
.
Tidal volume 500
IE12.0
Trigger sensitivityV
Resp. Rate15
T. pause10
PEEP5
T. Insp. rise5
O2 conc.40
Additionalvalues
Cancel
Accept
11ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Set ventilation mode
Volume control ?
Automode
Ti 1.33 s (33)
Basic
IE
Trigger
.
Tidal volume 500
IE12.0
Trigger sensitivityV
Resp. Rate15
T. pause10
PEEP5
T. Insp. rise5
O2 conc.40
.
Additionalvalues
Cancel
Accept
12ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Set ventilation mode
Volume control ?
Automode
Ti 1.33 s (33)
Basic
IE
Trigger
.
Tidal volume 550
IE12.0
Trigger sensitivityV
Resp. Rate15
T. pause10
PEEP5
T. Insp. rise5
O2 conc.40
.
Additionalvalues
Cancel
Accept
13ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Set ventilation mode
Volume control ?
Automode
Ti 1.33 s (33)
Basic
IE
Trigger
.
Tidal volume 550
IE12.0
Trigger sensitivityV
Resp. Rate15
T. pause10
PEEP5
T. Insp. rise5
O2 conc.40
.
Additionalvalues
Cancel
Accept
14ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Set ventilation mode
Volume control ?
Automode
Ti 1.33 s (33)
Basic
IE
Trigger
.
Tidal volume 550
IE12.0
Trigger sensitivityV
Resp. Rate15
T. pause10
PEEP5
T. Insp. rise5
O2 conc.40
.
Additionalvalues
Cancel
Accept
15Text and alarm messages appear here
16Alarmpofile
Fixed keys
Additionalsettings
Additionalvalues
Startbreath
O2breaths
Exp.hold
Insp.hold
17Silence/pre-silence alarms
Alarmpofile
- Silences or pre-silences alarms for 2 minutes
Additionalsettings
Startbreath
O2breaths
Exp.hold
Insp.hold
18Silence/pre-silence alarms
Alarmpofile
158
- Silences or pre-silences alarms for 2 minutes
- After silencing an alarm the Alarm silence symbol
is shown with the remaining time
Startbreath
O2breaths
Exp.hold
Insp.hold
19Silence/pre-silence alarms
055
- When alarms are pre-silenced a pre-silence symbol
is shown as well as the remaining mute time
20Alarm profile
- Press this button to
- Display upper and lower alarm limits
- Adjust preset limits
- Adjust the alarm sound level
21Alarm profile
Alarm sound level
Upper
Lower
100
40
Pressure cm H2O
Minute Ventilation l/min
5.0
40.0
Respiratory Rate b/min
5
30
End Exp. Pressure cmH2O
3
Autoset
Additionalvalues
Cancel
Accept
22Save
- Saves
- Current curve values
- Current parameter values
- Setting values
- 10 secs before and 10 secs after pressing key
23ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Recording
T
50 cmH2O
70 l/min
-70
700 ml
Additionalsettings
Additionalvalues
Startbreath
O2breaths
Exp.hold
Insp.hold
24ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Additionalsettings
Additionalvalues
25ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Menu
Alarm
Review
Options
Compliancecomp.
Copy data
Biomed
Panel lock
Changepatient category
Close
Additionalsettings
Additionalvalues
26ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Menu
Alarm
Trends
Review
Recordedwaveforms
Options
Compliancecomp.
Copy data
Biomed
Panel lock
Changepatient category
Close
Additionalsettings
Additionalvalues
27ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Recorded waveforms 12/25/02 143210
50 cmH2O
40 cmH2O
70 l/min
0 l/min
-70
700 ml
475 ml
0.0 s
Additionalvalues
Settings
Cursor
Close
28(No Transcript)
29Trends graphs
1432 Recording saved
l/min 2
?
Vee
0
0
l/min 20
10.5 10.0
MVi MVe
0
ml 800
VTi VTe
300 296
?
0
1400
1500
Cursor
Hours 1
Close
30ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Recorded waveforms 12/25/02 143210
50 cmH2O
40 cmH2O
70 l/min
0 l/min
-70
700 ml
475 ml
0.0 s
Additionalvalues
Settings
Cursor
Close
31Trends graphs
1442
l/min 2
?
Vee
0
0
l/min 20
10.5 10.0
MVi MVe
0
ml 800
VTi VTe
300 296
?
0
1400
1500
Cursor
Hours 1
Close
32Trends graphs
1442
l/min 2
?
Vee
0
0
l/min 20
10.5 10.0
MVi MVe
0
ml 800
VTi VTe
300 296
?
0
1400
1500
Cursor
Hours 1
Close
33Trends graphs
1442
l/min 2
?
Vee
0
0
l/min 20
10.5 10.0
MVi MVe
0
ml 800
VTi VTe
300 296
?
0
1400
1500
Cursor
Hours 1
Close
34Trends graphs
1432
l/min 2
?
Vee
0
0
l/min 20
10.5 10.0
MVi MVe
0
ml 800
VTi VTe
300 296
?
0
1400
1500
Cursor
Hours 1
Close
35Trends graphs
1438
l/min 2
?
Vee
0
0
l/min 20
10.5 10.0
MVi MVe
0
ml 800
VTi VTe
300 296
?
0
1400
1500
Cursor
Hours 1
Close
36Trends graphs
1432
l/min 2
?
Vee
0
0
l/min 20
10.5 10.0
MVi MVe
0
ml 800
VTi VTe
300 296
?
0
1400
1500
Cursor
Hours 1
Close
37Trends graphs
1438
l/min 2
?
Vee
0
0
l/min 20
10.5 10.0
MVi MVe
0
ml 800
VTi VTe
300 296
?
0
1400
1500
Cursor
Hours 1
Close
38Trends graphs
1438
l/min 2
?
Vee
0
0
l/min 20
10.5 10.0
MVi MVe
0
ml 800
VTi VTe
300 296
?
0
1400
1500
Cursor
Hours 1
Close
39i
- This function is not currently available
40Main rotary dial
41Using the main rotary dial
ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
.
Additionalsettings
Additionalvalues
42ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
.
Additionalsettings
Additionalvalues
43ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
.
Additionalsettings
Additionalvalues
44ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Set ventilation mode
Volume control ?
Automode
Ti 1.33 s (33)
Basic
IE
Trigger
.
Tidal volume 500
IE12.0
Trigger sensitivityV
Resp. Rate15
T. pause10
PEEP5
T. Insp. rise5
O2 conc.40
.
Additionalvalues
Cancel
Accept
45ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Set ventilation mode
Volume control ?
Automode
Ti 1.33 s (33)
Basic
IE
Trigger
.
Tidal volume 500
IE12.0
Trigger sensitivityV
Resp. Rate15
T. pause10
PEEP5
T. Insp. rise5
O2 conc.40
.
Additionalvalues
Cancel
Accept
46ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Set ventilation mode
Volume control ?
Automode
Ti 1.33 s (33)
Basic
IE
Trigger
.
Tidal volume 500
IE12.0
Trigger sensitivityV
Resp. Rate15
T. pause10
PEEP5
T. Insp. rise5
O2 conc.40
.
Additionalvalues
Cancel
Accept
47ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Set ventilation mode
Volume control ?
Automode
Ti 1.33 s (33)
Basic
IE
Trigger
.
Tidal volume 500
IE12.0
Trigger sensitivityV
Resp. Rate15
T. pause10
PEEP5
T. Insp. rise5
O2 conc.40
.
Additionalvalues
Cancel
Accept
48ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Set ventilation mode
Volume control ?
Automode
Ti 1.33 s (33)
Basic
IE
Trigger
.
Tidal volume 550
IE12.0
Trigger sensitivityV
Resp. Rate15
T. pause10
PEEP5
T. Insp. rise5
O2 conc.40
Additionalvalues
Cancel
Accept
.
49ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Set ventilation mode
Volume control ?
Automode
Ti 1.33 s (33)
Basic
IE
Trigger
.
Tidal volume 550
IE12.0
Trigger sensitivityV
Resp. Rate15
T. pause10
PEEP5
T. Insp. rise5
O2 conc.40
Additionalvalues
Cancel
Accept
.
50ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Set ventilation mode
Volume control ?
Automode
Ti 1.33 s (33)
Basic
IE
Trigger
.
Tidal volume 550
IE12.0
Trigger sensitivityV
Resp. Rate15
T. pause10
PEEP5
T. Insp. rise5
O2 conc.40
Additionalvalues
Cancel
Accept
.
51ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Set ventilation mode
Volume control ?
Automode
Ti 1.33 s (33)
Basic
IE
Trigger
.
Tidal volume 550
IE12.0
Trigger sensitivityV
Resp. Rate15
T. pause10
PEEP5
T. Insp. rise5
O2 conc.40
.
Additionalvalues
Cancel
Accept
52ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Set ventilation mode
Volume control ?
Automode
Ti 1.33 s (33)
Basic
IE
Trigger
.
Tidal volume 550
IE12.0
Trigger sensitivityV
Resp. Rate15
T. pause10
PEEP5
T. Insp. rise5
O2 conc.40
.
Additionalvalues
Cancel
Accept
53Measured values appear here
54Pmean
12
PEEP
6
RR (b/min)
30
15
5
O2 ()
48
42
36
IE
12.0
MVe (l/min)
8.5
7.4
6.0
VTi
501
VTe
493
Additionalsettings
Additionalvalues
55Pmean
12
PEEP
6
RR (b/min)
30
15
5
O2 ()
48
42
36
IE
12.0
MVe (l/min)
8.5
7.4
6.0
VTi
501
VTe
493
Additionalsettings
Additionalvalues
56Expiratory MinuteVolume Low
Pmean
12
PEEP
6
RR (b/min)
30
15
5
O2 ()
48
42
36
IE
12.0
MVe (l/min)
8.5
5.2
6.0
VTi
356
VTe
346
Additionalsettings
Additionalvalues
57Respiratory Rate High
Pmean
12
PEEP
6
RR (b/min)
30
31
5
O2 ()
48
42
36
IE
12.0
MVe (l/min)
8.5
8.5
6.0
VTi
292
VTe
274
Additionalsettings
Additionalvalues
58Respiratory Rate High
Pmean
12
PEEP
6
RR (b/min)
30
31
5
O2 ()
48
42
36
IE
12.0
MVe (l/min)
8.5
8.5
6.0
VTi
292
VTe
274
Additionalsettings
Additionalvalues
59Pmean
12
PEEP
6
RR (b/min)
30
15
5
O2 ()
48
42
36
IE
12.0
MVe (l/min)
8.5
7.4
6.0
VTi
501
VTe
493
Additionalsettings
Additionalvalues
60Pmean
12
PEEP
6
RR (b/min)
30
15
5
O2 ()
48
42
36
IE
12.0
MVe (l/min)
8.5
7.4
6.0
VTi
501
VTe
493
Additionalsettings
Additionalvalues
61Ppeak
34
Pplat
24
Pmean
12
PEEP
5
RR
15
O2
21
Vee
0
IE
12.0
MVe
7.0
MVi
7.5
VTi
501
VTe
471
Additionalsettings
Nextpage
62Ppeak
34
Pplat
24
Pmean
12
PEEP
5
RR
15
O2
21
Vee
0
IE
12.0
MVe
7.0
MVi
7.5
VTi
501
VTe
471
Additionalsettings
Nextpage
63Additionalsettings
Nextpage
64Additionalsettings
Nextpage
65Additionalsettings
Nextpage
66Alarmprofile
Save
Trends
i
Additionalsettings
Additionalvalues
Special function keys
!
Startbreath
O2breaths
Exp.hold
Insp.hold
67Special function keys
- Start breath
- Activates a set breath
- O2 breaths
- Changes the FiO2 to 1.0 for one minute
- Can be interrupted by re-pressing within one
minute
68Special function keys
- Expiratory hold
- Produces an expiratory pause hold for measuring
total PEEP - Lasts as long as the key is pushed up to a
maximum of 30 secs - Inspiratory hold
- Produces an inspiratory pause hold for measuring
plateau pressure - Lasts as long as the key is pushed up to a
maximum of 30 secs
69Alarmprofile
Save
Trends
i
Quickstart
Menu
Mainscreen
Additionalsettings
Additionalvalues
Direct access knobs
!
Startbreath
O2breaths
Exp.hold
Insp.hold
70Direct access knobs
ModePressure Control
Automode
Admitpatient
Nebulizer
Status
- Parameter controlled by each direct access knob
depends on ventilation mode selected - Turn the knob to the desired value shown in the
box above the knob
Additionalsettings
Additionalvalues
O2 conc. 30
PEEP 5
Resp. rate 15
PC above PEEP 15
mmHg
b/min
cm H2O
71Direct access knobs
ModePressure Control
Automode
Admitpatient
Nebulizer
Status
- Parameter controlled by each direct access knob
depends on ventilation mode selected - Turn the knob to the desired value shown in the
box above the knob
Additionalsettings
Additionalvalues
PEEP 5
Resp. rate 15
PC above PEEP 15
mmHg
b/min
cm H2O
72Direct access knobs
ModePressureControl
Automode
Admitpatient
Nebulizer
Status
- Parameter controlled by each direct access knob
depends on ventilation mode selected - Turn the knob to the desired value shown in the
box above the knob
Additionalsettings
Additionalvalues
PEEP 5
Resp. rate 15
PC above PEEP 15
mmHg
b/min
cm H2O
73Direct access knobs
ModePressure Control
Automode
Admitpatient
Nebulizer
Status
- If the bar in the box is white the setting is
within what is generally considered safe limits
Additionalsettings
Additionalvalues
PEEP 5
Resp. rate 15
Tidal volume 500
ml
b/min
cm H2O
mmHg
74Direct access knobs
ModePressure Control
Automode
Admitpatient
Nebulizer
Status
- If the bar turns yellow the setting is outside
what are generally considered safe limits
Additionalsettings
Additionalvalues
PEEP 5
Resp. rate 15
Tidal volume 500
ml
b/min
cm H2O
mmHg
75Direct access knobs
ModePressure Control
Automode
Admitpatient
Nebulizer
Status
- If the bar turns red the setting is significantly
outside what are generally considered safe limits - An audible alarm and alarm message is given
Additionalsettings
Additionalvalues
PEEP 5
Resp. rate 15
Tidal volume 500
ml
b/min
cm H2O
mmHg
76Direct access knobs
ModePressure Control
Automode
Admitpatient
Nebulizer
Status
- When you come to a safety limit the knob is
inactive for 2 seconds in order to make you aware
that you have passed a safety limit
Additionalsettings
Additionalvalues
PEEP 5
Resp. rate 15
Tidal volume 500
ml
b/min
cm H2O
mmHg
77Direct access knobs
ModePressure Control
Automode
Admitpatient
Nebulizer
Status
- NB When you turn a direct access knob the changes
in ventilation take place from the next breath
without additional confirmation
Additionalsettings
Additionalvalues
O2 conc. 30
PEEP 5
Resp. rate 15
PC above PEEP 15
mmHg
b/min
cm H2O
78Alarmprofile
Save
Trends
i
Quickstart
Menu
Mainscreen
Additionalsettings
Additionalvalues
Mains indicator
!
Startbreath
O2breaths
Exp.hold
Insp.hold
79Alarmprofile
Save
Trends
i
Quickstart
Menu
Mainscreen
Additionalsettings
Additionalvalues
Start/stop ventilation (standby)
!
Startbreath
O2breaths
Exp.hold
Insp.hold
80Waveform area
Additionalsettings
Additionalvalues
Startbreath
O2breaths
Exp.hold
Insp.hold
81ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
T
50 cmH2O
70 l/min
-70
700 ml
Additionalsettings
Additionalvalues
Startbreath
O2breaths
Exp.hold
Insp.hold
82ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
T
50 cmH2O
70 l/min
-70
700 ml
Additionalsettings
Additionalvalues
Startbreath
O2breaths
Exp.hold
Insp.hold
83ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
T
50 cmH2O
70 l/min
-70
700 ml
Additionalsettings
Additionalvalues
Startbreath
O2breaths
Exp.hold
Insp.hold
84ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
T
50 cmH2O
70 l/min
-70
700 ml
Additionalsettings
Additionalvalues
Startbreath
O2breaths
Exp.hold
Insp.hold
85ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
T
50 cmH2O
70 l/min
-70
700 ml
Additionalsettings
Additionalvalues
Startbreath
O2breaths
Exp.hold
Insp.hold
86ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
T
50 cmH2O
70 l/min
-70
700 ml
Additionalsettings
Additionalvalues
Startbreath
O2breaths
Exp.hold
Insp.hold
87Alarmprofile
Save
Trends
Patient category
i
Quickstart
Menu
Mainscreen
Additionalsettings
Additionalvalues
!
Startbreath
O2breaths
Exp.hold
Insp.hold
88Whats different from a Servo 300/300A?
89Volume control
Volume support
Pressure support
Inspiratory rise time
User interface
Apnoea time
Automode
SIMV
90Volume control
91Volume control
- 300 and 300A
- If patients breathing effort reaches 2 cm H2O
below preset PEEP ventilator permits a higher
flow - pressure is then regulated to 2 cm H2O above PEEP
- Servoi
- If a pressure drop of ?3 cm H2O is detected
during inspiration the ventilator switches to
pressure support, resulting in increased flow - When flow ? to calculated target level flow
maintained until set tidal volume delivered
92Volume support
93Minute volume
- 300 or 300A
- If RR lt set RR (ie minute vol. lt set minute vol.)
- new target VT is calculated (based on preset
minute volume) used to regulate insp. pressure - Max. VT 150 of preset
- Servoi
- Neither RR nor minute volume are set
- Inspiratory pressure is regulated purely based on
preset VT
94Inspiratory rise time
- 300 and 300A
- Inspiratory rise time set as a of breath cycle
time, which is in turn dependent on set RR
- Servoi
- Inspiratory rise time set as
- in control and SIMV modes
- absolute time (s) in support modes. (NB. As a
result the proportion of inspiration taken up by
rise time will increase with RR)
95I-E cycling
- 300 and 300A
- I-E cycling occurs when inspiratory flow 5 of
maximum inspiratory flow
- Servoi
- I-E cycling occurs when inspiratory flow set
of maximum inspiratory flow - Set inspiratory cycle-off
96Apnoea back-up
- 300 and 300A
- Apnoea gt 20 secs (adults)? pressure regulated
volume control
- Servoi
- Apnoea gt set apnoea time? Back-up mode (volume
control)
97Pressure support
98Apnoea back-up
- Servoi
- Apnoea gt set apnoea time? Back-up mode (pressure
control)
99Inspiratory rise time
- 300 and 300A
- Inspiratory rise time set as a of breath cycle
time, which is in turn dependent on set RR
- Servoi
- Inspiratory rise time set as an absolute time (in
seconds). - NB. As a result the proportion of inspiration
taken up by rise time will increase with RR
100I-E cycling
- 300 and 300A
- I-E cycling occurs when inspiratory flow 5 of
maximum inspiratory flow
- Servoi
- I-E cycling occurs when inspiratory flow set
of maximum inspiratory flow - Set inspiratory cycle-off
101Inspiratory rise time
102Inspiratory rise time
- 300 and 300A
- Always set as a of cycle time
- Servoi
- Set as a of cycle time control modes and SIMV
- Set as an absolute time in support modes
103Apnoea time
- 300 and 300A
- Preset
- Adult 20 s
- Paediatric 15 s
- Neonate 10 s
- Servoi
- Adjustable in volume support, pressure
support/CPAP and SIMV modes - Press Alarm profile to set time
- Adult 15-45 s
- Infant 5-15 s
104SIMV
- 300 and 300A
- SIMV cycle
- SIMV period
- Servoi
- SIMV cycle
- SIMV period set breath cycle time
105SIMV
- 300 and 300A
- Mandatory (SIMV) breath given at end of SIMV
period if patient does not trigger ventilator
during SIMV period
- Servoi
- Mandatory (SIMV) breath given after 90 of SIMV
period if patient does not trigger ventilator in
initial 90 of SIMV period
106Automode
- 300A
- Ventilator shifts from control to spontaneous
mode if patient triggers 2 consecutive breaths
- Servoi
- Ventilator shifts from control to spontaneous
mode if patient triggers ventilator
107Automode
- 300A
- Ventilator shifts from spontaneous to control
mode after a set period of apnoea - Adult 12 s
- Paediatric 8 s
- Neonate 5 s
- Servoi
- Ventilator shifts from control to spontaneous
mode after trigger time-out period - Increases with each successive breath until the
set trigger timeout limit is reached
108Setting the mode
109ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Additionalsettings
Additionalvalues
110ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Set ventilation mode
Volume control ?
Automode
Ti 1.33 s (33)
Basic
IE
Trigger
.
Tidal volume 500
IE12.0
Trigger sensitivityV
Resp. Rate15
T. pause10
PEEP5
T. Insp. rise5
O2 conc.40
Additionalvalues
Cancel
Accept
111ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Set ventilation mode
Volume control ?
Automode
Ti 1.33 s (33)
Select ventilation mode
Volume control
SIMV (Vol. Control)Pressure Support
.
Pressure control
SIMV (Press. Control)Pressure Support
PRVC
SIMV (PRVC)Pressure Support
Pressure Support/CPAP
Volume Support
Additionalvalues
Cancel
Accept
112ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Set ventilation mode
Pressure control ?
Automode
Ti 1.33 s (33)
Basic
IE
Trigger
.
PC above PEEP 18
IE12.0
Trigger sensitivityV
Resp. Rate15
T. Insp. rise5
PEEP5
O2 conc.40
Additionalvalues
Cancel
Accept
113ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Set ventilation mode
Pressure control ?
Automode
Ti 1.33 s (33)
Basic
IE
Trigger
.
PC above PEEP 18
IE12.0
Trigger sensitivityV
Resp. Rate15
T. Insp. rise5
PEEP5
O2 conc.40
Additionalvalues
Cancel
Accept
114ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Set ventilation mode
Pressure control ?
Automode
Ti 1.33 s (33)
Basic
IE
Trigger
.
PC above PEEP 18
IE12.0
Trigger sensitivityV
Resp. Rate15
T. Insp. rise5
PEEP5
O2 conc.40
Additionalvalues
Cancel
Accept
115ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Set ventilation mode
Pressure control ?
Automode
Ti 1.33 s (33)
Basic
IE
Trigger
.
PC above PEEP 18
IE12.0
Trigger sensitivityV
Resp. Rate15
T. Insp. rise5
PEEP5
O2 conc.40
Additionalvalues
Cancel
Accept
116- If the bar in the box is white the setting is
within what is generally considered safe limits
117- If the bar turns yellow the setting is outside
what are generally considered safe limits
118- If the bar turns red the setting is significantly
outside what are generally considered safe limits
119Inspiratory rise time
- Can be altered to increase patient comfort
- Some patients find sudden bolus of gas
uncomfortable - Excessively long rise time can result in
receiving preset levels of ventilatory support
for relatively short periods
120Inspiratory rise time
- Control modes and SIMV
- Rise time set as a of cycle time
- SIMV
- Cycle time is set
- Control modes
- Cycle time 60/RR
121Inspiratory rise time
- Support modes
- Rise time set as an absolute time
- Proportion of inspiratory time that is taken up
by rise time increases as cycle time shortens (ie
increased RR)
122Triggering
123ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Set ventilation mode
Pressure control ?
Automode
Ti 1.33 s (33)
Basic
IE
Trigger
.
PC above PEEP 18
IE12.0
Trigger sensitivityV
Resp. Rate15
T. Insp. rise5
PEEP5
O2 conc.40
Additionalvalues
Cancel
Accept
124ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Set ventilation mode
Pressure control ?
Automode
Ti 1.33 s (33)
Trigger sensitivity -2 -20
0
Basic
IE
Trigger
.
PC above PEEP 18
IE12.0
Trigger sensitivityV
Resp. Rate15
T. Insp. rise5
PEEP5
O2 conc.40
Additionalvalues
Cancel
Accept
125ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Automode
Additionalsettings
Additionalvalues
126Automode
- When automode has been selected the soft key
turns white
127ModeVolume Control
Automode Volume support
Admitpatient
Nebulizer
Status
Additionalsettings
Additionalvalues
O2 conc. 30
PEEP 5
Resp. rate 15
Tidal volume 500
ml
b/min
cm H2O
128ModeVolume Control
Automode Volume support
Admitpatient
Nebulizer
Status
Additionalsettings
Additionalvalues
O2 conc. 30
PEEP 5
Resp. rate 15
Tidal volume 500
ml
b/min
cm H2O
129Automode Volume support
ModeVolume Control
Automode Volume support
Admitpatient
Nebulizer
Status
Additionalsettings
Additionalvalues
O2 conc. 30
PEEP 5
Resp. rate 15
Tidal volume 500
ml
b/min
cm H2O
130ModeVolume Control
Automode
Admitpatient
Nebulizer
Status
Set ventilation mode
Pressure control ?
Automode
Ti 1.33 s (33)
Basic
IE
Trigger
.
PC above PEEP 18
IE12.0
Trigger sensitivityV
PS above PEEP20
Resp. Rate15
T. Insp. rise5
Insp. cycle off5
PEEP5
Trigger timeout7
O2 conc.40
Additionalvalues
Cancel
Accept
131Automode
- Automode can only be set in association with the
following modes of ventilation - Volume control
- Pressure regulated volume control
- Pressure control
132Modes
Volume control
Volume support
Pressure control
Pressure support/CPAP
PRVC
SIMV
Automode
Back up ventilation
133Volume control
- volume preset assist/control ventilation
- ventilator delivers preset volume in preset
inspiratory time with preset pause time and
preset flow pattern - when triggered by the patients respiratory
efforts - as determined by the preset ventilatory rate
134Volume control
50 cmH2O
70 l/min
-70
700 ml
135Volume control
- Ventilator will deliver breaths to ensure preset
frequency is achieved - In general preset volume is delivered according
to preset waveform for all breaths - With the Servo i it possible for the patient to
obtain a higher inspiratory flow and tidal volume
from the ventilator than is pre-set
136Volume control
- Flow is usually dependent on preset volume and
the inspiratory time. - Inspiratory time is determined by the IE ratio
and the respiratory rate - If the flow does not meet patient demand the flow
is automatically increased.
137Volume control
- If a pressure drop of ?3 cm H2O is detected
during inspiration the ventilator switches to
pressure support, resulting in increased flow - When flow ? to calculated target level flow
maintained until set tidal volume delivered
138Volume control
Flow
Time
Insp.time
Pausetime
Exp time
139Volume control
Flow
Time
Insp.time
Pausetime
Exp time
140Pressure control
- Pressure preset assist/control ventilation
- Similar to volume control except pressure is
preset - Inadequate flow unlikely to be a problem
141Pressure control
PC above PEEP
Pressure
PEEP
Flow
142Pressure control
- Theoretical decrease in risk of barotrauma
- Improved oxygenation due to decelerating flow
pattern - Risk of hypoventilation as lung compliance falls
- monitor tidal and minute volumes
143Pressure regulated volume control
- Preset tidal volume delivered in preset
inspiratory time at preset frequency - Ventilator automatically adapts inspiratory
pressure control level to changes in mechanical
properties of chest - Inspiratory pressure constant
- Inspiratory flow decelerating
144PRVC
Pressure
Flow
Volume
145PRVC
Upper Pressure limit
5 cm H2O
Pressure
Flow
Volume
146PRVC
- Has the advantages of pressure control
ventilation without the disadvantage of potential
hypoventilation/over-ventilation as chest
compliance changes
147PRVC
- With automode off
- PRVC
- With automode on
- PRVC if patient is apnoeic
- VS if patient is triggering the ventilator
148PRVC
- In this mode the pressure is titrated to the
tidal volume - should not be used in a situation where the
inspiratory tidal volume leaving the ventilator
is not an accurate reflection of delivered tidal
volume (eg circuit leak, uncuffed tube)
149Volume support
- Every breath is patient triggered
- Pressure square wave with decelerating flow
- Ventilator automatically titrates pressure to
achieve preset tidal volume with the lowest
possible pressure
150Volume support
Upper Pressure limit
5 cm H2O
Pressure
10 cm H2O
Flow
Volume
151Volume support
Upper Pressure limit
5 cm H2O
Pressure
10 cm H2O
Flow
Volume
152Volume support
- If apnoea time is exceeded ventilator
automatically - Alarms and switches to Back up mode (volume
control with default settings) if Automode is off - Remains in Back up mode until alarm is reset
153Pressure support/CPAP
- With both PS and PEEP gt0
- pressure support ventilation with PEEP
- With PEEP 0
- pressure support ventilation
- With PS 0
- PS of 2 cmH2O plus PEEP
154Pressure support
- inspiratory assist
- constant preset airway pressure supplied during
inspiration
155Pressure support
PS above PEEP
Pressure
PEEP
Flow
Volume
156Pressure support
PS above PEEP
Pressure
PEEP
Flow
Volume
157Pressure support
- allows patient to control inspiratory and
expiratory times, inspiratory flow and tidal
volumes. This control can improve patient comfort - can be combined with SIMV or CPAP or used on its
own in patients whose gas exchange is improving
and in whom weaning is being attempted
158Pressure support
- pressure support of 3.5-14.5 cmH2O required to
overcome the additional work of breathing due to
breathing through ETT and demand valve. Patients
who require pressure support of lt 6 cmH2O can
probably be extubated - may unload respiratory muscles and reduce
diaphragmatic fatigue
159Pressure support
- disadvantages
- VT and minute ventilation variable, excessively
large VT may be delivered - high initial flow and termination algorithm may
be unsuitable for severe airflow obstruction
160Back-up ventilation
- Operates in association with volume support and
pressure support/CPAP - Does not operate when Automode is on
161Back-up ventilation
Pressure controlDefault pressure PS or 20
cmH2O Default IE ratio Default respiratory
rate Default inspiratory rise time
Pressuresupport/CPAP
Apnoeagtset apnoea time
162Back-up ventilation
Volume controlVT set VT Default IE
ratio Default respiratory rate Default
inspiratory rise time
Volumesupport
Apnoeagtset apnoea time
163SIMV (VC) PS
- patient can breathe spontaneously but also
receives a set number of volume controlled
mechanical breaths - volume controlled breaths are synchronized with
the patients breathing to prevent the
possibility of a mechanical breath on top of a
spontaneous breath - remaining breaths are pressure supported
164SIMV
Pressure support above PEEP
Pressure
PEEP
Time
Flow
Time
Trig
Trig
165SIMV (VC) PS
- If the patient is apnoeic the ventilator delivers
volume controlled breaths at the preset SIMV rate - If the patient triggers the ventilator the
response of the ventilator depends on the point
in the ventilator cycle at which triggering occurs
166SIMV (VC) PS
- Time is split up into a series of SIMV cycles
- duration of each SIMV cycle
167SIMV (VC) PS
- Each SIMV cycle is made up of 2 parts
- SIMV period
- breath cycle time
- spontaneous period
168SIMV cycle
SIMV period
Volume controlled breath
Pressure supported breath
T
T
169Volume controlled breath
90 of SIMV period
170SIMV (PC) PS
- Very similar to SIMV (VC) PS except that
controlled breaths are pressure rather than
volume controlled
171SIMV (PRVC) PS
- Very similar to SIMV (VC) PS except that
controlled breaths are PRVC breaths rather than
volume controlled breaths
172Automode
- Active in VC, PC and PRVC modes
- When automode is on ventilator switches between
control and spontaneous modes based on presence
or absence of triggering
173Automode
Pressurecontrol
Pressuresupport
Apnoea
Triggering
Volumecontrol
Volumesupport
Apnoea
Triggering
PRVC
Volumesupport
Apnoea
Triggering
174Automode
- Trigger timeout
- duration of apnoea before ventilator switches
from support mode to control mode - The limit increases with each successive breath
until the set trigger timeout limit is reached
175Automode
- When automode has been selected the soft key
turns white
176ModeVolume Control
Automode Volume support
Admitpatient
Nebulizer
Status
Additionalsettings
Additionalvalues
O2 conc. 30
PEEP 5
Resp. rate 15
Tidal volume 500
ml
b/min
cm H2O
177ModeVolume Control
Automode Volume support
Admitpatient
Nebulizer
Status
Additionalsettings
Additionalvalues
O2 conc. 30
PEEP 5
Resp. rate 15
Tidal volume 500
ml
b/min
cm H2O
178Automode Volume support
ModeVolume Control
Automode Volume support
Admitpatient
Nebulizer
Status
Additionalsettings
Additionalvalues
O2 conc. 30
PEEP 5
Resp. rate 15
Tidal volume 500
ml
b/min
cm H2O
179Changing the parameters
180Changing parameters
ModePressure Control
Automode
Admitpatient
Nebulizer
Status
- Direct access knobs
- Additional settings
Additionalsettings
Additionalvalues
O2 conc. 30
PEEP 5
Resp. rate 15
PC above PEEP 15
mmHg
b/min
cm H2O
181Changing parameters
ModePressure Control
Automode
Admitpatient
Nebulizer
Status
- Direct access knobs
- Additional settings
Additionalsettings
Additionalvalues
O2 conc. 30
PEEP 5
Resp. rate 15
PC above PEEP 15
mmHg
b/min
cm H2O
182AlarmsThis module will be available in later
versions of this tutorial
183Starting up
184(No Transcript)
185Pre-use check
186Standby
Do you want to start Pre-use check?
Yes
No
to start ventilation
Press
Additionalsettings
Additionalvalues
187Standby
Do you want to start Pre-use check?
Yes
No
to start ventilation
Press
Additionalsettings
Additionalvalues
188(No Transcript)
189Pre-use check
- Do not use the ventilator if it does not pass the
pre-use check
190Check batteries
191(No Transcript)
192Status
Additionalsettings
Additionalvalues
193Status
Status
General
Version Universal
Serial no. 2017
Pre-use check Passed 2002-02-06 1452
Operating time 167 hours
Next Preventative Maintenance 1423 hours
Battery backup time 312 min
Barometric pressure 1023 hPa
Air Pressure 394 kPa100
O2 pressure 401 kPa100
O2 cell
Exp. cassette
Modules
Options
Pre-use check
Additionalsettings
Additionalvalues
194Status
Status
General
Version Universal
Serial no. 2017
Pre-use check Passed 2002-02-06 1452
Operating time 167 hours
Next Preventative Maintenance 1423 hours
Battery backup time 312 min
Barometric pressure 1023 hPa
Air Pressure 394 kPa100
O2 pressure 401 kPa100
O2 cell
Exp. cassette
Modules
Options
Pre-use check
Additionalsettings
Additionalvalues
195Status
Status
General
Modules status
Battery 52 min, Slot position 1
Battery 51 min, Slot position 2
Battery 49 min, Slot position 3
Battery 52 min, Slot position 4
Battery 52 min, Slot position 5
Battery 52 min, Slot position 6
O2 cell
Exp. cassette
Modules
Options
Pre-use check
Additionalsettings
Additionalvalues
196AdvancedThis module will be available in later
versions of this tutorial
197Alarm messages
- This module will be available in a later version