Title: Atomic%20Emission%20Spectroscopy%20Lecture%2018
1Atomic Emission SpectroscopyLecture 18
2Atomic Emission Spectroscopy )AES)
- - (AES), in contrast to AAS, uses the very high
temperatures of atomization sources to excite
atoms. - Thus excluding the need for lamp sources.
- Emission sources, which are routinely used in
AES - plasma, 2) arcs and 3) sparks, 4) flames.
- We will study the different types of emission
sources, their operational principles, features,
and operational characteristics. - Finally, instrumental designs and applications of
emission methods will be discussed.
3Plasma Sources
- The term plasma is defined as a homogeneous
mixture of (gaseous atoms, ions and electrons) at
very high temperatures. - Two types of plasma atomic emission sources are
frequently used - Inductively coupled plasma (ICP).
- Direct current plasma (DCP).
4(No Transcript)
5Inductively Coupled Plasma (ICP)
- A typical ICP consists
- Three concentric quartz tubes (???????? ????
??????) through which streams of argon gas flow
at a rate in the range from 5-20 L/min. - The outer tube is about 2.5 cm (1 inch) in
diameter and the top of this tube is surrounded
by a radiofrequency powered induction coil (RF)
producing a power of about 2 kW at a frequency in
the range from 27-41 MHz. - This coil produces a strong magnetic field as
well.
6- - Ionization of flowing argon is achieved by a
spark. - - The ionized argon interacts with the strong
magnetic field and is thus forced to move within
the vicinity of the induction coil at a very high
speed. - A very high temperature is obtained as a result
of the very high resistance experienced by
circulating argon (collisions of e- and cations
with ambient gas) (ohmic heating). - Why Ar used?? 1- inert 2- few emission lines
7- - The top of the quartz tube will experience very
high temperatures and should, therefore, be
isolated and cooled. - Cooling
- This can be accomplished by passing argon
tangentially around the walls of the tube. - A schematic of an ICP (usually called a torch
plasma) is shown below
8Approach spark or arc to ionize Ar and so Ar
ions circulate fatly and suffers from resistance
so temperature increases
Current passes in the coil and form MF ? to the
current in the coil
9Inductively Coupled Plasma Torch
Radio frequency induction coil 27-41 MHz 0.5-2KW
Plasma torch F 2.5 cm
Tangential flow isolates the torch from plasma
Plasma argon
Sample argon
Argon flow rate 5-20 L/min
Tangential argon plasma support flow
10- The torch is formed as a result of the argon
emission at the very high temperature of the
plasma. - The temperature gradients in the ICP torch can be
pictured in the following graphics
11(No Transcript)
12- Problems
- Needs very large amount of Ar
- Needs large amount of Energy
- Main Advantage
- Pure Atomization
Cross flow nebulizer
13Viewing region for working
14- The viewing region
- Used in elemental analysis is usually about 6000
oC, which is about 1.5-2.5 cm above the top of
the tube. - High cost of the ICP torch
- Because argon consumption is relatively high
which makes the running - Argon is a unique inert gas for plasma torches
- Since it has few emission lines.
- This decreases possibility of interferences with
other analyte lines.
15Sample Introduction
- Several methods for sample introduction
- The most widely used is
- The nebulization of an analyte solution into the
plasma. - However, other methods, as described earlier, are
fine where vapors of analyte molecules or atom
from electrothermal or ablation devices can be
driven into the torch for complete atomization
and excitation. - For your convenience, sample introduction methods
are summarized here again
16Samples in Solution
- Pneumatic Nebulizers
- Samples in solution are usually easily introduced
into the atomizer by a simple nebulization,
aspiration, process. - Nebulization converts the solution into an
aerosol of very fine droplets using a jet of
compressed gas. - The flow of gas carries the aerosol droplets to
the atomization chamber or region.
17(No Transcript)
18Ultrasonic Nebulizers
- In this case samples are pumped onto the surface
of a piezoelectric crystal that vibrates in the
kHz to MHz range. - Such vibrations convert samples into homogeneous
aerosols that can be driven into atomizers. - Ultrasonic nebulization is preferred over
pneumatic nebulization since finer droplets and
more homogeneous aerosols are usually achieved. - However, most instruments use pneumatic
nebulization for convenience.
19- Electrothermal Vaporization
- (Only for sample introd. Not for atomization)
- An accurately measured quantity of sample (few
mL) is introduced into an electrically heated
cylindrical chamber through which an inert gas
flows. - Usually, the cylinder is made of pyrolytic
carbon but tungsten cylinders are now available. - The vapors of molecules and atoms are swept into
the plasma source for complete atomization and
excitation.
20Electrothermal Vaporization
To ICP
Sample
Graphite rod
Heater power
Water coolant
Argon inlet
21- Hydride Generation Techniques
- Samples that contain arsenic, antimony, tin,
selenium, bismuth, and lead can be vaporized by
converting them to volatile hydrides by addition
of sodium borohydride. - Volatile hydrides are then swept into the plasma
by a stream of an inert gas.
22(No Transcript)
23Introduction of Solid Samples
- A variety of techniques were used to introduce
solid samples into atomizers. These include - 1. Conductive Samples
- If the sample is conductive and is of a shape
that can be directly used as an electrode (like a
piece of metal or coin), that would be the choice
for sample introduction in arc and spark
techniques. - Otherwise, powdered solid samples are mixed with
fine graphite and made into a paste. - Upon drying, this solid composite can be used as
an electrode. - The discharge caused by arcs and sparks interacts
with the surface of the solid sample creating a
plume of very fine particulates and atoms that
are swept into the plasma by argon flow.
24- Laser Ablation
- Sufficient energy from a focused intense laser
will interact with the surface of samples (in a
similar manner like arcs and sparks) resulting in
ablation. - The vapors of molecules and atoms are swept into
the plasma source for complete atomization and
excitation. - Laser ablation is becoming increasingly used
since it is applicable to conductive and
nonconductive samples.
25The Glow Discharge Technique
- The technique is used for sample introduction and
atomization as well. - The electrodes are kept at a 250 to 1000 V DC.
- This high potential is sufficient to cause
ionization of argon, which will be accelerated to
the cathode where the sample is introduced. - Collision of the fast moving energetic argon ions
with the sample (cathode) causes atomization by a
process called sputtering. - Samples should thus be conductive to use the
technique of glow discharge. - The vapors of molecules and atoms are swept into
the plasma source for complete atomization and
excitation by flowing argon. - However, nonconductive samples were reported to
be atomized by this technique where they were
mixed with a conductor material like graphite or
powdered copper.
26Plasma Appearance and Spectra
- A plasma torch looks very much like a flame but
with a very intense nontransparent brilliant ????
white color at the core (less than 1 cm above
the top of the tube). - In the region from 1-3 cm above the top of the
tube, the plasma becomes transparent. - The temperatures used are at least two to three
orders of magnitude higher than that achieved by
flames which may suggest efficient atomization
and fewer chemical interferences.
27- Ionization in plasma is not a problem
- It is may be thought to be a problem due to the
very high temperatures - But fortunately the large electron flux from the
ionization of argon will suppress ionization of
all species.
282) The Direct Current Plasma (DCP)
- The DCP is composed of three electrodes arranged
in an inverted Y configuration. - A tungsten cathode resides at the top arm of the
inverted Y. - The lower two arms are occupied by two graphite
anodes. - Argon flows from the two anode blocks and plasma
is obtained by momentarily bringing the cathode
in contact with the anodes. - Argon ionizes and a high current passes through
the cathode and anodes.
29- It is this current which ionizes more argon and
sustains the current indefinitely. - Samples are aspirated into the vicinity of the
electrodes (at the center of the inverted Y)
where the temperature is about 5000 oC. - DCP sources usually have fewer lines than ICP
sources, require less argon/hour, and have lower
sensitivities than ICP sources. - In addition, the graphite electrodes tend to
decay with continuous use and should thus be
frequently exchanged. - A schematic of a DCP source is shown below
30(No Transcript)
31(No Transcript)
32- DCP advantages
- Less argon consumption about 1/3 the ICP and
less E of power supply. - Simpler instrumental requirements.
- less spectral line interference (lower
atomization temp. about 5000 0C). - However
- ICP sources are more convenient to work with
- - ICP is free from frequent consumables (like
the anodes in DCPs which need to be frequently
changed) - - More sensitive than DCP sources.
33Advantages of Plasma Sources
- No oxide formation as a result of two factors
including - Very high temperature
- Inert environment inside the plasma (no oxygen)
- 2. Minimum chemical interferences (no or few
ionization) es from the ionized Ar suppress
ioniztion. - 3. Minimum spectral interferences except for
higher possibility of spectral line interference
due to exceedingly large number of emission lines
(because of high temperature)
34- 4. Uniform temperature which results in precise
determinations - 5. No self-absorption is observed which extends
the linear dynamic range to higher concentrations - 6. No need for a separate lamp for each element
- 7. Easily adaptable to multichannel analysis
(simultaneous measurements of many elements).
35Plasma Emission Instruments
- Three classes of plasma emission instruments can
be presented including - 1. Sequential instruments ???? ??????
- In this class of instruments a single channel
detector is used - The signal for each element is read using the
specific wavelength for each element
sequentially. - each element is measured after the another.
- Two types of sequential instruments are
available
36- Linear sequential scan instruments
- ???? ??????
- The wavelength is linearly changed with time.
- Therefore, the grating is driven by a single
speed during an analysis of interest. - b. Slew scan instruments
- The monochromator is preset to provide specific
wavelengths - moving very fast in between wavelengths.
- while moving slowly at the specific wavelengths.
- Therefore, a two-speed motor driving the grating
is thus used.
37Radial vs. and axial Viewing
Radial traditional side view, better for
concentrated samples. Axial direct view into
plasma, lower sensitivity, shifts detection range
lower.
38Axial view
- Much more light available
- This gives you the opportunity to achieve Lower
Detection Limits than Radial Plasma - Disadvantages
- More Matrix Interferences
- Slightly Reduced Dynamic Range
39Radial view
40Filter wheel, to remove orders of radiation
41Sequential vs. multichannel
- Sequential instrument
- PMT moved behind aperture plate,(slits found for
elements at their ? at the focal plane). - or grating prism moved to focus new l on exit
slit - Pre-configured exit slits to detect up to 20
lines, slew scan - characteristics
- Cheaper
- Slower
- Multichannel instrument
- Polychromators (not monochromator) (multiple
PMT's) - Array-based system - FT Instruments
- charge-injection device/charge coupled device
- characteristics
- Expensive ( gt 80,000-150,000)
- Faster
42Slew scan spectrometer
- Two slew-scan gratings
- Two PMTs for VIS and UV
- Most use holographic grating
43Slew Scan Spectrometer
Exit slit
Composite grating
Mirror
Filter wheel
Photomultiplier tubes
Motorized observation height
Plasma torch
Mirrors
Hg lamp
Entrance slit
442. Multichannel Instruments
- This class of instruments is also referred to as
simultaneous instruments in which all signals are
reported at the same time using two types of
configurations
45a. Polychromators (do not confused with
monochromator)
- Multiple detectors each measure 1 ?
- Usually photomultiplier tubes are used.
- Beams of radiation emerging from the grating are
guided to exit slits (each representing the
wavelength of a specific element) are focused at
several PMTs for detection. - Detection, thus, takes place simultaneously
46Many slits at each a detector present
simultaneous measring
47Grating
Many PMT for the elements to be analyzed
Measuring electronics
Dedicated computer
Instrument control electronics
Schematic of an ICP polychromator
48b. Array-based systems
- This multichannel type instrument uses a
multichannel detector like a charge injection
device or a charge-coupled device. - Diffracted beams from a grating pass through a
prism where further resolution of diffracted
beams takes place by a prism. - The prism will disperse the orders of each
diffracted beam. - The multichannel detector can also be a linear
photodiode array as in the figure below
49PDA
Problems Spectral line interferences should be
eliminated the detector need to be very small to
prevent more than one line to hit any detector
50To overcome the previous problem
Prism to further resolve unresolved lines and the
orders of any line
Echele grating (high Res, use the order)
Could be used for qualitative analysis
51(No Transcript)
52Long distance for G and CM give large angle and
good dispersion
53(No Transcript)
543. Fourier transform instruments (FT)
- Instruments in which the signal is coded will
need a decoding mechanism in order to see the
signal. - FT is a very common technique for decoding time
domain spectra. - In such instruments, the detector records the
change of signal with time, which is practically
not useful. - However, Fourier transformation of the time
domain signal yield a frequency domain spectrum,
which is the usual signal, obtained by
conventional methods. - Instruments that rely on decoding a coded signal
is also said to have a multiplex design.
55Atomic Emission Spectroscopy
56Applications of Plasma Sources
- 1. Since plasma sources result in a very large
number of emission lines, these sources can be
used for both qualitative and quantitative
analysis. - 2. The signal obtained from plasma sources is
stable, has a low noise and background, as well
as freedom from interferences. - 3. Requires sample preparation similar to AAS
57- 4. Plasma sources are usually best suited for
operation in the ultraviolet region - Therefore
- Elements having emission lines below 180 nm (like
B, P, S, N, and C) can only be analyzed under
vacuum since air components absorb under 180 nm. - Alkali metals are difficult to analyze since
their best lines under plasma conditions occur in
the visible or near infrared. - 5. An analytical emission line can easily be
located but will depend on the other elements
present since spectral line interferences are
encountered in plasma sources due to the very
high temperatures used.
58- 6. Linear calibration plots are usually obtained.
- Problem
- departure from linearity is observed at high
concentrations - due to self absorption as well as other
instrumental reasons. - Overcome
- An internal ( ?????? ???? ?????? ???????? ???? ??
???? ??? ??? ????)standard is often used in
emission methods to correct for fluctuations in
temperature as well as other factors. - The calibration plot in this case is a plot
between the concentration of analyte and the
ratio of the analyte to internal standard signal.
- The internal standard
- Is a substance that is added in a constant
amount to all samples, blanks, and standards
therefore it must be absent from initial sample
matrix. - The internal standard should have very close
characteristics (both chemically and physically)
to analyte.
59Elements by ICP-AES
Different elements have different emission
intensities. Alkalis (Na, K, Rb, Cs) are weakly
emitting. Alkaline Earths (Be, Mg, Ca, Sr, Ba )
are strongly emitting (sensitive to low con)
60background
61Typical Calibration Curves
Tl 334.94 nm
10
Photocurrent ( na)
1
V 437.92 nm
Nb 371.30 nm
Ce 456.24 nm
0.001
0.0001
0.1
0.01
Impurity concentration in iron ( wt)
62INDUCTIVELY COUPLED PLASMA-MASS
SPECTROMETRY (ICP-MS)
The detector is the MS (used to measure AW for
different atoms) - Very sensitive and good for
trace analysis. - Plasma produces analyte
ions. - Ions are directed to a mass
spectrometer. - Ions are separated on the basis
of their mass-to-charge ratio. - A very
sensitive detector measures ions. - Very low
detection limits.
63Emission Spectroscopy Based on Arcs and Sparks
- Samples are excited in the gap between a pair of
electrodes connected to a high potential power
supply (200 VDC or 2200- 4400 VAC). - The high potential applied forces a discharge
between the two electrodes to occur where current
passes between the two separated electrodes
(temperature rises due to very high resistance).
64One electrode is graphite and the other is the
sample the two electrodes approach to each others
Arc
65- The very high temperature (4000-5000 oC) realized
in the vicinity between the two electrodes
provide enough energy for atomization and
excitation of the samples in this region or when
the sample is, or a part of, one of the
electrodes. - Arc and spark methods are mainly used as
qualitative techniques and can also be used as
semiquantitative techniques.
66Sample Handling and Preparation
- If the sample is conductive and is of a shape
that can be directly used as an electrode (like a
piece of metal or coin), that would be the choice
for sample introduction in arc and spark
techniques. - Otherwise, powdered solid samples are mixed with
fine graphite and made into a paste. - Upon drying, this solid composite can be used as
an electrode. - The discharge caused by arcs and sparks interacts
with the surface of the solid sample creating a
plume of very fine particulates and atoms that
are excited and emission is collected. - The figure below shows some common shapes of
graphite electrodes used in arc and spark
sources.
67Carbon electrodes
- Sample pressed into electrode or mixed with Cu
powder and pressed - Briquetting (pelleting) - Cyanogen bands (CN) 350-420 nm occur with C
electrodes in air -He, Ar atmosphere - Arc/spark unstable
- each line measured gt20 s
- needs multichannel detection
68Instruments for Arcs and Sparks
- In most cases, emission from atoms in an arc or
spark is directed to a monochromator with a long
focal length and the diffracted beams are allowed
to hit a photographic film. - This typical instrument is called a spectrograph
since it uses a photographic film as the detector.
69Spectrograph
- Beginning 1930s
- (in some old universities)
- photographic film detector
- Cheap
- Long integration times (20-30 s to obtain stable
signal). - Difficult to develop/analyze
- Non-linearity of line "darkness on the
photograph film.
70Very long monochromator 2-3 m For good separation
71(No Transcript)
72- The blackness of the lines on the photographic
film is an indication of the intensity of the
atomic line and thus the concentration of the
analyte. - The location of emission lines as compared to
standard lines on a film serves to identify the
wavelengths of emission lines of analyte and thus
its identity. - The use of spectrographs is not very convenient
since a lot of time and precautions must be spent
on processing and calibrating the photographic
film.
73- Qualitative analysis
- Is accomplished by comparison of the wavelengths
of some emission lines to standards while the
line blackness serves as the tool for
semiquantitative analysis. - Polychromators are also available as multichannel
arc and spark instruments. - However, these have fixed slits at certain
wavelengths in order to do certain elements and
thus they are not versatile.
74(No Transcript)
75- Recently
- Arc and spark instruments based on charge
injection and charge coupled devices became
available. - These have extraordinarily high efficiency and
performance in terms of - - Easier calibration
- Short analysis time
- Superior quantitative results.
76For Qual. and Quant. Analysis
77(No Transcript)
78Characteristics of Arc Sources
- Typical temperatures between 4000-5000 oC are
high enough to cause atomization and excitation
of sample and electrode materials. - 2. Usually, cyanogens compounds are formed due to
reaction of graphite electrodes with atmospheric
nitrogen. - Emission bands from cyanogens compounds occur in
the region from 350-420 nm. - Disadvantage
- Several elements have their most sensitive lines
in this same region which limits the technique. - Overcome
- Use of controlled atmosphere around the arc
(using CO2, Helium, or argon) very much decreases
the effect of cyanogens emission.
79- 3. The emission signal should be integrated over
a minute or so since volatilization and
excitation of atoms of different species differ
widely. - While some species give maximum signal, others
may still be in the molecular state. - 4. Arc sources are very good for qualitative
analysis of elements while only semiquantitative
analysis is possible. - It is mandatory ?????? to compare the emission
spectrum of a sample with the emission spectrum
of a standard. - In some cases, a few milligrams of a standard is
added to the sample in order to locate the
emission lines of the standard and thus identify
the emission wavelengths of the different
elements in the sample. - A comparator?????? densitometer????? ??? ?????
can be used to exactly locate the wavelengths of
the standard and the sample components.
80The lines from the standard are projected on the
lines of the combined sample/standard emission
spectra in order to identify sample components.
Only few lines are shown in the figure.
81(No Transcript)
82(No Transcript)
83(No Transcript)
84Why use Carbon in Atomic Spectroscopy?
- We have previously seen the use of graphite in
electrothermal AAS as well as arc and spark AES,
even though molecular spectra are real problems
in both techniques due to cyanogens compounds
absorption and emission. - The reasons after graphite common use in atomic
spectroscopy can be summarized below
85- It is conductive.
- It can be obtained in a very pure state.
- Easily available and cheap.
- Thermally stable and inert.
- Carbon has few emission lines.
- Easily shaped.
86Spark Sources
- Most of the instruments in this category are arc
based instruments. - Spark based instruments are of the same idea
except for a spark source substituting an arc
source. - The spark source construction
- An AC potential in the order of 10-50 KV is
discharged through a capacitor which is charged
and discharged through the graphite electrodes
about 120 times/s resulting in a discharge
current of about 1000 A.
87This very high current will suffer a great deal
of resistance, which increase the temperature to
an estimated 40000 oC. Therefore, ionic spectra
are more pronounced.
Two electrodes
Capacitor Charged and discharge periodically in
the electrodes rapidly giving spark (1000 A)