Title: Photoneutron Distributions around 18 MV Xray Radiotherapy Accelerators using Nuclear Track Detectors
1Photoneutron Distributions around 18 MV X-ray
Radiotherapy Accelerators using Nuclear Track
Detectors
- Fazal-ur-Rehman,
- H. Al-Ghamdi,
- M. I. Al-Jarallah (Presenting Author)
- N. Maalej
King Fahd University of Petroleum and Minerals
2Outline
- Introduction
- Methods
- Results
- Conclusion
3Introduction
- Neutrons contaminate the high energy Linear
Accelerator (LINAC) photon beam in radiation
therapy and contribution dose to the patient - We studied the distribution of thermal and fast
neutrons in the LINAC room during irradiation
with an 18 MeV photon beam
4Methods
- Neutrons are generated by high energy photon
interaction with the high Z materials of the
LINAC head - Fast Neutron Measurement
- Bare NTDs are used
- Tracks are generated by protons from the (n,p)
recoil in the detector material (C12H18O7 ) - Thermal Neutron Measurement
- NTDs are covered with Li2B4O7
- Tracks are generated by ?-particles from
10B(n,?)7Li and 6Li(n,?)3H nuclear reactions
5Methods
Fig.1 Experimental arrangement showing the
location of CR-39 NTDs around a radiotherapy
linear accelerator.
6Methods
x
Fig. 2 Experimental arrangement showing the
location of CR-39 NTDs around a radiotherapy
linear accelerator.
7Methods
- NTDs advantages over active detectors
- There is no pulse pileup problem
- No photon interference with neutron measurement
- No electronics are required
- Less prone to noise and interference
8Results
Fig. 4 Fast neutron relative intensity as a
function of transversal distance from isocenter
of linear accelerators located in a small and
large treatment rooms.
9Results
Fig. 5 Fast neutron relative intensity as a
function of longitudinal distance from the
isocenter of linear accelerators located in a
small and large treatment rooms.
10Results
Fig. 6 Fast neutron relative intensity as a
function of longitudinal distance from the
isocenter of linear accelerators located in a
small and large treatment rooms.
11Results
Fig. 7 Thermal neutron relative intensity as a
function of longitudinal distance from isocenter
of linear accelerators located in a small and
large treatment rooms.
12Conclusion
- The fast neutron distribution along the
transverse and longitudinal directions is
symmetrical about the beam axis. - The fast neutron relative intensities fall in an
exponential-like fashion with distance from the
isocenter of the accelerators. - The thermal neutron distribution is uniform at
all the locations along transverse and
longitudinal directions. - The larger size LINAC room had lower fast and
thermal neutron relative intensities at all
locations.
13Acknowledgment
- Physics Department of King Fahd University of
Petroleum and Minerals (KFUPM). - The Department of Radiotherapy, King Fahd
Specialist Hospital. - Dhahran Health Center, Dhahran, for utilizing
their accelerator facilities in this study.