Title: FECAL OCCULT BLOOD TESTING USING SERACULT SLIDES
1FECAL OCCULT BLOOD TESTING USING SERACULT SLIDES
WellOne Primary Medical and Dental Care
Click here to move on
- For Medical Clinical Staff
2WARNING
- Conducting this test exposes the operator to
potentially infectious material. Standard
precautions, including glove use are required for
this procedure. - .
Click here to move on
Click here to go back
3Fecal Occult Blood Testing General Overview of
Testing
- As with any fecal occult blood test, results are
not to be considered conclusive evidence for the
presence or absence of any gastrointestinal
pathology. The test is not a substitute for
other diagnostic procedures such as endoscopy or
colonoscopy, barium enema or other studies. - The fecal occult blood test will be considered a
screening test at WellOne. Positive results must
be followed with diagnostic procedures to confirm
and when indicated, negative results should be
followed with recommended testing. - Since bleeding from gastrointestinal lesions may
be intermittent, fecal samples for testing
should be collected from three consecutive bowel
movements or three bowel movements closely spaced
in time. - To further increase the probability of detecting
occult blood, separate samples should be taken
from two different sections of each fecal
specimen. - Foods which contain peroxides may leave residues
in the feces that may interfere with the test. - The expected/normal result is a sample that is
negative for fecal occult blood.
Click here to move on
Click here to go back
4A positive fecal occult blood test
Click here to select this answer
- Should be repeated for validation
- Requires follow up testing such as colonoscopy
and CBC - Can be ignored if the patient has not followed
dietary restrictions - Is almost always related to hemorrhoids
Click here to select this answer
Click here to select this answer
Click here to select this answer
Click here to go back
5The correct answer is
- Requires follow up testing such as colonoscopy
and CBC - Since bleeding from the intestinal tract
- may be intermittent (come and go), even
- one slightly positive test should not be
- ignored
Click here to move on
Click here to go back
6Fecal Occult Blood Testing Storage and Stability
- Test slides
- Store slides at room temperature (15-30ºC),
protected from light, heat, and UV radiation - Slides are stable until expiration date unless
general integrity is interrupted. - Developer Solution
- Store developer at room temperature (15-30ºC)
- If properly stored, developer will remain stable
until expiration date on bottle.
Click here to move on
Click here to go back
7Fecal Occult Blood Testing Storage and Stability
- Prior to testing a slide, be sure to check that
the slide and developer solution have not expired
Click here to move on
Click here to go back
8Each stool card is imprinted with an expiration
date on the back of the card that can be
difficult to see without close inspection-
marking over the area with ink will help you to
see the date
Click here to go back
Click here to move on
9Fecal Occult Blood Testing Procedural Steps
- Open the front flap of the slide and apply a very
thin layer of stool smear to one window and apply
a very thin smear of stool to the other window
then close front flap of slide.
BACK
FRONT
Click here to move on
Click here to go back
10For accurate fecal occult blood testing, the
fecal smear should be
Click here to select this answer
- thinly spread on window
- applied thickly on window
- applied to both sides of the card
- applied with a cotton swab
Click here to select this answer
Click here to select this answer
Click here to select this answer
Click here to go back
11The correct answer is
Click here to select this answer
- thinly spread on window
- applied thickly on window (this will interfere
with reading results) - applied to both sides of the card (the front flap
is for the test specimen and the back flap is for
test and control interpretation) - applied with a cotton swab (the specimen should
be applied with a gloved finger or wooden
spatula)
Click here to select this answer
Click here to select this answer
Click here to select this answer
Click here to move on
Click here to go back
12Fecal Occult Blood Testing Procedural Steps
- Open perforated cover on back of slide.
- Apply 2 drops of developer solution to each smear
in specimen test area and read at 30-60 seconds. - A positive result is indicated by the appearance
of any blue color in the testing window. A lack
of blue color indicates a negative result.
Click here to go back
Click here to move on
13When performing fecal occult blood testing, ____
drops of solution are to be applied to each of
the test areas on the back of the card
Click here to select this answer
Click here to select this answer
Click here to select this answer
Click here to go back
14The correct answer is
2 drops for testing
Click here to go back
Click here to move on
15The fecal occult blood test result should be read
at __________
Click here to select this answer
- 2 minutes
- 5 minutes
- 5-10 seconds
- 30-60 seconds
Click here to select this answer
Click here to select this answer
Click here to select this answer
Click here to go back
16The correct answer is
30-60 seconds for test interpretation
Click here to move on
Click here to go back
17A positive test result is indicated by the
appearance of
Click here to select this answer
- any black color
- any yellow color
- any blue color
- any white color
Click here to select this answer
Click here to select this answer
Click here to select this answer
Click here to go back
18The correct answer is any blue color
Click here to move on
Click here to go back
19Fecal Occult Blood Testing Procedural Steps
- Develop the performance control for each slide
(only after specimen tests have been completed
and interpreted) by applying 1 drop of developer
solution to the performance Control line. - A blue color should appear within 30 seconds
(indicating a valid test slide). - Remove and dispose of gloves in regular trash
perform hand hygiene. - Document results in the patient record (verbally
alert provider of any positive results) and mark
on the encounter form.
Click here to move on
Click here to go back
20Fecal Occult Blood Testing More On Control
Testing
- The internal control must be performed with each
test done by applying 1 drop of developer
solution to the performance control line only
after the interpretation of the actual test
results. A blue color should appear within 30
seconds. A positive control test lets us know
that both the slide and solution are likely to
give accurate results. A negative control tells
us that there is a problem with either the test
slide or solution and the test results are not
valid. - If the control is negative, disregard the testing
results and perform control testing on another
slide from the same lot. If the control is
positive, repeat testing using another slide and
interpret test and control results accordingly. - If the control remains negative, immediately
notify the COO or designee. Action to taken
includes the determination of the source of the
problem (ie. developer verses test slides). Use
a different developer solution from same lot if
possible on same lot of test slides-if control is
now positive, discard the original developer
solution bottle. If control remains negative,
test the same lot of test slides using a
developer solution from a different lot. If
control is now positive, segregate and return all
developer solution bottles of this lot number.
If control remains negative, run the control on a
test slide from a different lot of stool cards.
If control is now positive, segregate and return
to manufacturer all test slides of the affected
lot number.
Click here to move on
Click here to go back
21Internal control testing must be performed
Click here to select this answer
- with the opening of each new box
- with the opening of a new solution
- with each test performed
- never there is no control
Click here to select this answer
Click here to select this answer
Click here to select this answer
Click here to go back
22The correct answer is..
- With each test performed.
Click here to move on
Click here to go back
23Control testing must be performed
Click here to select this answer
- Prior to interpreting the test results
- Only after interpreting the test results
Click here to select this answer
Click here to go back
24The correct answer is..
- Only after interpreting the test results
Click here to go back
Click here to move on
25To perform control testing, use ______ drop/s of
solution a blue color must appear in _______ to
be counted as a valid test
Click here to select this answer
- One, 30-60 seconds
- Two, 30-60 seconds
- One, 30 seconds
Click here to select this answer
Click here to select this answer
Click here to go back
26The correct answer is..
- One (drop of solution) with a blue coloring
appearing within 30 seconds
Click here to go back
Click here to move on
27If the control on the fecal occult blood card is
negative
Click here to select this answer
- the test is positive
- the test is negative
- the test is not valid
- the test is sensitive
Click here to select this answer
Click here to select this answer
Click here to select this answer
Click here to go back
28The correct answer is..
- The test is NOT VALID- The results MUST NOT BE
INTERPRETED
Click here to move on
Click here to go back
29You have reached the end of this program..
- Once you are confident that you are comfortable
with all the materials presented, proceed to the
Fecal Occult Blood Testing post-test at