Title: The Determination of 226Ra in Water Samples
1 The Determination of 226Ra in Water Samples
eichrom
TECHNOLOGIES INC
- Anil H. Thakkar Michael Fern
- Eichrom Technologies, Inc.
- David McCurdy
- Duke Engineering Services
2Objectives
eichrom
TECHNOLOGIES INC
- To design a procedure for radium in water which
would - be a simple, short and quick procedure
- allow to measure Ra-226 and Ra-224 via alpha
spectroscopy and Ra-228 via Ac-228 measurements - eliminate the tedious steps of precipitation and
metathesis
3Current Procedure Used
eichrom
TECHNOLOGIES INC
- Barium sulfate precipitation
- Conversion to BaCO3- tedious and time consuming
4Current Procedure Used
eichrom
TECHNOLOGIES INC
- Sample Ba carrier Ba-133 Conc.H2SO4_ Heat
BaSO4 - Precipitate allowed to settle for several hours
or overnight - Precipitate collected and converted to BaCO3
5New Approach
eichrom
TECHNOLOGIES INC
- Water samples
- Cation exchange resin column
- Eichrom Ln Resin column
6Preparation of Water Samples
eichrom
TECHNOLOGIES INC
- 0.5 L / 1.0 L water samples, acidified to
- pH 2
- Add 133Ba as a tracer
7Eichrom Cation Exchange Column
eichrom
TECHNOLOGIES INC
- Load solution 0.5 L/ 1.0 L of sample
- Rinse with 50 mL of 0.1M HNO3
- Strip Ra and Ba with 50 mL of 8M HNO3
8eichrom
TECHNOLOGIES INC
Preparation of Load Solution for Ln Resin Column
- Evaporate eluent from cation exchange column to
dryness - Dissolve residue in 10 ml of 0.095M HNO3 acid
- Load on Eichrom Ln Resin column
9eichrom
TECHNOLOGIES INC
Ln Resin Column Separation
(1) Load 10 mL of mL 0.095M HNO3,
Collect (1) and (2) for Ba 226Ra
(2) Rinse 15 mL of 0.095M HNO3
(3) Strip 10 mL of 0.35M HNO3
collect (3) for 228Ac
10226Ra Via Alpha Spectrometry
eichrom
TECHNOLOGIES INC
- Evaporate 226Ra/133Ba fraction to 10 mL volume
- Add 0.1 ml of 75 mg/ml barium carrier
- Add 3 mL of 40 sodium sulfate, 4 drops of 11
acetic acid. Swirl - Add 0.2 mL of seeding suspension. Mix
- Place tubes in cold water bath for 30 minutes
- Filter on 0.1 micron filters, count via alpha
spectrometry
11133Ba Via Gamma Counter
eichrom
TECHNOLOGIES INC
- Count 133Ba tracer yield via gamma spectrometry
12228Ac Via Gas-Flow Proportional Counter
eichrom
TECHNOLOGIES INC
- Add 50 micrograms of cerium carrier
- Add 1.0 mL of HF
- Filter on 0.1 micron filters
- Measure for beta radiation
13226Ra Recoveries-0.5L DI Water
eichrom
TECHNOLOGIES INC
14226Ra Alpha Spectrum
eichrom
TECHNOLOGIES INC
15133Ba Gamma Spectrum
eichrom
TECHNOLOGIES INC
16226Ra, 224Ra, and Daughters
eichrom
TECHNOLOGIES INC
224Ra
226 Ra
220Rn
218Po
222Rn
17226Ra Recoveries in Tap water
eichrom
TECHNOLOGIES INC
18226Ra in EPA SampleSept.18, 1998
eichrom
TECHNOLOGIES INC
19226Ra in NJ Sample
eichrom
TECHNOLOGIES INC
20226Ra in Georgia Inst. Of Tech. Sample S8933
eichrom
TECHNOLOGIES INC
21226Ra in Georgia Inst. Of Tech. Sample WS 14776
eichrom
TECHNOLOGIES INC
22Summary
eichrom
TECHNOLOGIES INC
- Cation exchange resin option
- provides a good replacement for the time
consuming BaSO4 precipitation steps - it eliminates inconsistent metathesis steps
- reduces usage of chemicals/reagents
- provides consistency and saves time
- Acceptable 226Ra recoveries
23Future Work
- To evaluate the effect of Ba/Ra uptake in high
TDS samples especially with high Ca and Mg
content - To measure Ra-228 concentrations in samples
24Acknowledgements
- Special thanks to
- Dr. B. Parsa, New Jersey department of health, NJ
- Dr. B. Kahn, Georgia Institute of Technology, GA