Title: Reimaging Lens Polarization
1As-Built Performance of the FPP
Spectro-Polarimeter October, 2004 FPP Team
Bruce W. Lites 303 497 1517 lites_at_ucar.edu
2FPP Spectro-Polarimeter Performance
- AS-BUILT PERFORMANCE OF THE FPP-SP
- Slit Scanning mechanism
- Vignetting as the image is scanned across the
slit - Spectral response function
- Signal/Noise (polarimetric precision)
- Scattered Light
- System Polarization Calibration
3Performance Requirements
- Polarization Accuracy Matrix (for measured Stokes
parameters I,Q,U,V) - Spectrum Range 6301-6303 Å
- Spectral Purity FWHM ? 35 mÅ
- Spectral Sample ? 25 mÅ
- Undispersed light scatter lt 0.01
- Polarization signal-to-noise, continuum gt 10001
4Performance of the Slit Scanning Mechanism
- In order to produce maps of the vector magnetic
field, the slit scanner should provide regular
and reproducible samples of the image plane in
the direction perpendicular to the slit. - The slit scanning mechanism nominal range is 2000
steps of 0.16 arcsec on the Sun - Requirements
- Repeatability 0.33 step
- Linearity no two adjacent steps separated by gt
0.32 arcsec - Performance
- Step size on Sun 0.148 arcsec (average)
- Range to limit switches 3233 steps (478.9 arcsec
on Sun only 320 arcsec are needed)
5Theodolite Measurements of Slit Scanner
Measured deflection of the beam as a function of
slit scan step number. Also shown is the linear
fit to the points. (Scan mirror motion of 1
arcsec corrseponds to 1/13.3 arcsec motion of the
solar image on the slit)
Departures of the measured beam deflection are
shown as a function of scan mirror step position.
A periodic error in the deflection corresponds
to one rotation of the ball screw. Amplitude of
this error is of order 1.2 arcsec on the sun per
400 steps (60 arcsec of scanning)
6Theodolite Measurements of Slit Scanner
Repeatability of the scanner is measured for 4
short scans around the center of the slit scan
range. Measured positions agree within 1
arcsec, corresponding to 0.075 arcsec on the
Sun. Some of this error may arise from the
measurement process.
7Optical Verification of Slit Scanner In Completed
FPP
Map of reticle image test central range of slit
scanner
8Optical Verification of Slit Scanner In Completed
FPP
Deviation of measured reticle line from a linear
fit
black steep slope line red shallow slope line
9Slit Scanner Performance
- Large-scale sinusoidal variations in the slit
scan step size 1.2 arcsec deviation from linear
over 400 steps (60 arcsec) - Short scans positions are repeatable within
about 0.15 arcsec peak (0.025 arcsec rms) - No adjacent steps deviate by more than 1 step
10Slit Scan Vignetting
Scan Mirror Step Number
Scan Mirror Step Number
19 August 2004 NAOJ SP intensity vs. scan mirror
position before pre-slit repair. FPP on OBU with
solar feed.
26 May 2005 NAOJ SP intensity vs. scan mirror
position after pre-slit repair. FPP on optical
bench. Solar feed with telescope simulator.
11FPP-SP Spectral Resolution
- Spectral resolution was monitored by measurement
of the spectrum from the tunable laser - Rotating diffuser in beam to reduce laser speckle
- Short integrations to minimize the effects of
drift of the laser wavelength - Image at right shows sample laser measurement in
both polarization images - Tuning the laser wavelength allows measurement of
spectral response over the entire image plane
12FPP-SP Spectral Resolution
- Fitted Laser Line Width Along Slit Periodic
Width Variation - Spectrum is undersampled (instrumental width
is25mÅ, sampling is 21.3mÅ) - Spectrum curvature samples the profile along the
slit at differing shifts relative to the peak of
the emission line - Gaussian fitting procedure has some sensitivity
to phase of undersampled points relative to the
emission line center
13FPP-SP Spectral Resolution
- Measured Variation of Spectral Resolution along
Slit Length and Within Spectral Field of Each
CCDSIDE - Post-vibration data, 8 February 2006
- Average spectral width in 100-pixel bins along
slit - No significant variations either along the slit,
or as a function of wavelength - All widths considerably smaller than 35mÅ
requirement
14FPP-SP Spectral Resolution
Use the Spectrum Curvature to Advantage
- Assume small variation of the spectral resolution
profile along the length of the slit - Normalize each spectral profile
- Shift each profile to a common line center
position (shift from fitting procedure) - Arrive at an instrumental resolution profile
highly sampled in wavelength - Red wing asymmetry a feature of the spectrograph
design
15FPP-SP Polarization S/N
- FPP-SP flux levels measured during Sun tests
- Measurements of transmission of heliostat and
window - Solar radiance measurements simultaneously during
Sun tests - Radiance measurements calibrated to zero airmass
- These measurements and extrapolations allow one
to extrapolate to the on-orbit S/N in
polarimetric measurements
- Inputs
- Dark-corrected continuum intensity in raw
measurement units (DN) - Measured SP CCD scale factor 100 e-/DN
- Read Noise 110 e-
- Heliostat transmission at 630 nm 0.51
- Atmospheric transmission at time of measurements
0.58 - Sum two sides of CCD
- Polarization modulation efficiency 0.5
- Anticipated S/N, Typical 4.8 sec Integration
- Continuum, Quiet Sun 1100
- Line center, Quiet Sun 580
- Line center, Umbra 164
16Scattered/Stray Light in the FPP-SP
- Measurements made of scattering inside the
spectrograph by means of a mask placed over the
slit. - Intensity as a function of distance from the mask
edge shows - Maximum intensity adjacent to the mask edge is
0.017 of the illuminated intensity - Scattering falls below the required 10-2 level
at a distance of 60 pixels (10 arcsec) from the
edge
17FPP-SP Polarization Calibration
- Calibration of entire SOT-FPP-SP optical system
during Sun tests, June 2005 - Linear, right-, and left-circular polarizers over
entrance of OTA
- Observations
- Clear (no polarizer)
- Linear, right-, left-circular polarizers
- Four rotation stations of each 0º, 45º, 90º,
135º - Standard mode observations (4.8s integrations)
- Two spectral ROIs covering entire spectral range
- Nine such data sets covering entire slit scan
range
18FPP-SP Polarization Calibration
- Data Analysis
- Bin data spatially along slit (16x) and
spectrally (2x) to reduce data volume and
increase S/N - Normalize each Stokes vector observation to the
measured Stokes I - Each binned pixel for each CCDSIDE was subjected
to a non-linear least-squares fitting procedure
to determine - 15 polarization response matrix elements (0,0
element I?I crosstalk set to unity) - Mount offset error of angular orientation for
right- and for left-circular polarizers
19FPP-SP Polarization Calibration
Polarization Accuracy Requirement and Measured
Polarization Response Matrix
1.00000 0.21996 0.01587 0.00437
-0.00033 0.48083 0.07739 -0.00125
-0.00072 0.06002 -0.47494 -0.00561
-0.00028 0.00437 -0.00792 0.52939
Typical system response matrix for CCDSIDE0, for
the left ROI (without strong spectral lines).
Off-diagonal values highlighted in red exceed
those of the polarization accuracy matrix.
Polarization Accuracy Matrix Requirement
- It is almost the case that no polarization
calibration is necessary response matrix is
nearly diagonal, within the required accuracy! - Calibration should improve the polarization
precision by at least an order of magnitude
better than the requirement
20FPP-SP Polarization Calibration
Features of the polarization response matrix
1.00000 0.21996 0.01587 0.00437
-0.00033 0.48083 0.07739 -0.00125
-0.00072 0.06002 -0.47494 -0.00561
-0.00028 0.00437 -0.00792 0.52939
Typical system response matrix for CCDSIDE0, for
the left ROI (without strong spectral lines).
Off-diagonal values highlighted in red exceed
those of the polarization accuracy matrix.
Polarization Accuracy Matrix Requirement
- The off-diagonal Q?U and U?Q elements (in red)
represent a simple rotation of coordinate system,
in this case by only about 2.5º. - The second element of the first row shows the
large value of Q?I crosstalk typical of a
single-beam polarimeter. When the two CCDSIDEs
are combined, this response is largely cancelled. - Diagonal elements are close to their anticipated
values. - Intensity to polarization crosstalk (first column
of the matrix) is very small.
21FPP-SP Polarization Calibration
Spatial/Spectral Variation of the Measured
Polarization Response Matrix
ROI 0-112
ROI 112-224
- Variations of the matrix across the field are
smaller than the uncertainties, except for first
column (I?Q,U,V, easily calibrated on orbit) - Residual smoothed variations are on the order of
few x 10-4 of the continuum intensity
----- 0.00040 0.00053 0.00037 0.00432
0.00308 0.00115 0.00077 0.00173 0.00139
0.00289 0.00130 0.00203 0.00045 0.00035
0.00166
RMS fluctuations of the response matrix for ROI
0-112, CCDSIDE0
22FPP Spectro-Polarimeter Performance
- SUMMARY
- The FPP-SP meets (and exceeds) its performance
requirements - Polarization calibration will be carried out as a
function of wavelength, dimension along the slit,
and slit scan position to account for small
variations - The spectral response function, highly sampled in
wavelength, will be used in the inversion process - Signal/Noise is higher than anticipated because
of excess read noise in the cameras.
Nonetheless, the goal of 10001 in the continuum
per spectral/spatial pixel is achieved