Title: AFM Fundamental System Components
1AFM Fundamental System Components
- Outline
- Sample preparation
- Instrument setting
- Data acquisition
- Imaging software
2Elements of a Basic Atomic force Microscope
3(No Transcript)
4Potential Diagram
Repulsion
Distance
Potential
Attraction
5Piezoelectric Material
6Sample Preparation
- AFM Does require minimum of sample preparation
- No clean room handling
- No thin film metal coating
- Works in liquids, gases, and vacuum
- Works at elevated or sub ambient temperatures
- Dimensions are not critical
7Instrument Setting
- Sample Center it in the middle of the sample
plate and immobilize it using dual side sticky
pads - Laser Laser beam has to bounce on the tip of the
cantilever. - Photodiode Reflected beam signal has to be
shared equally between the 4 cells - System adjustment Servo Gain (PI values), Force,
Raster speed
8Two progressively greater magnifications
(Lowest magnification, over a 10µm grating)
(Highest magnification, over a 10µm grating)
9Camera and Lens Assembly
10Video System Overview
- The NAVITAR zoom lens system provides an optical
magnification range of 2.1x-13.5x to the camera. - The degree of magnification at the monitor
depends on the ratio of the monitor size to the
CCD chip size. The camera uses a 1/3" CCD (6mm
diagonal). Using a 12" monitor (305mm diagonal)
with the 1/3" CCD chip, the total magnification
of the system would then be (13.5) x (1.8) x
(305/6) 1230 (1 micron would be seen as 1.2 mm
on the screen)
11How the sample is scanned
- Positioning the probe
- The main challenge is to move the probe with
increments as small as 0.05 nm and keep it at the
right position - Resolution in the X-Y range is limited by the
radius of the probe tens nm - Resolution in the Z-range is limited by the noise
of the system 0.05 nm - Introduction to Piezoelectric materials
- Ceramic tube
- Pendulum design
12How to move and to maintain the probe at the
right position?
- For a full scale of 1 micron assuming an image
area of 1000 x 1000 pixel the x-y resolution is 1
nm - No mechanical positioning can meet this
specification - Piezoelectric ceramic actuators can meet these
requirements
13Introduction to Piezo-Electric Properties
- Electric dipoles in domains
- unpoled ferroelectric ceramic
- During and (3) after poling (piezoelectic ceramic)
PZT (Lead zirconium titanate) ceramics must be
poled at an elevated temperature. The ceramic
now exhibits piezoelectric properties and will
change dimensions when an electric potential is
applied.
14Piezoelectric materials and scanners
- The extension or contraction of a piezoelectric
element is small - For example, for a 5 cm long piezoelectric
element, a voltage of100 V will result in an
extension of 1 micron - Since voltages can be controlled on the level of
at least 10 mV, this gives a resolution of 0.1 nm
or 1 Angstrom
15Ceramic Tubular Actuator
The tube is deformed in a controlled way by
applying a voltage on the X electrodes
There are four electrically isolated parts on the
outside of the tube X, -X, Y, -Y and one
electrical electrode inside of the tube Z
16Errors Introduced by the PZT Scanner
Hysteresis
Voltage
Creep
Voltage
Top PZT materials have hysteresis. When a
voltage ramp is placed on the ceramic, the motion
is nonlinear. Bottom Creep occurs when a voltage
pulse on a PZT causes initial motion followed by
drift.
17Linearity Error
A test pattern with squares, A, will appear
severely distorted if the piezoelectric scanner
in the AFM is not linear as in B. A common
method for correcting the problems of X-Y
non-linearity and calibration is to add
calibration sensors to the X-Y piezoelectric
scanners (Close loop scanner).
18Error due to the Bow
The motion of the probe is nonlinear in the
Z axis as it is scanned across a surface. The
motion can be spherical
19Bow and Tilt
20Balanced Pendulum How Does It Work
- Laser tracking spot remains fixed
- relative to Z-piezo AFM cantilever
- Z-piezo does not bend
Pendulum Design
Tube Design
21Why Balance the Pendulum Moving weight
distribution for scanning accuracy and speed
Simple pendulum scans slower, less accurate
during turn around, more noise
- Balanced pendulum
- Scans faster, less noisy
- More accurate control in XYZ
- Low inertia
- Maintains rigidity
- Minimizes X-Y coupling
22Nose Assemblies
The nose assembly retains the cantilever and
enables its motion. A spring clip on the nose
assembly secures the probe in place. One-piece
nose assemblies are available for different modes
and may include additional electronics and/or
components.
Clockwise from upper left Top MAC, CSAFM,
Contact Mode, AC Mode, STM
23Mounting the Nose Assembly on the Scanner
Push evenly and straight down when inserting the
nose assembly. Small off-axis forces will create
LARGE torques about the anchor point for the
piezoes, where most breakage occurs. Do NOT push
as this will damage the spring clip and/or glass
down on the top of the nose assembly window.
24Controlling and Imaging Software
- PicoView provides control and the first line of
visual interpretation and has to be understood
before getting any further. It gives limited
information about results and requires the use of
a more sophisticated software to interpret the
experiments. - Gwyddion and Imaging Metrology provide sample
measurements and statistical data. They have to
be used to prepare professional reports.
25PicoView Powerful SPM Control Software
- Benefits
- Simultaneous real-time display of up to eight
channels (in all resolutions) - Simultaneously display real-time image and
post-processed data - Unlimited data points in spectroscopy
- 16x16 to 4096x4096 pixels in images
- Parametric data structure in Spectroscopy
- Allow flexible data presentation
- Temporal display of all channels
- Select any channel as x-axis and plot all the
rest against it.
26PicoView Powerful SPM Control Software
- PicoScript scripting interface for PicoView
- SPM I/O and control function library
- DLL (dynamically linked library) for VB,
LabView, and more. - Labview VI
- Allow interface with external acquisition cards
- Benefits
- Empower user to customize their own application
needs - No need to understand the source code structures
- Allows the popular LabView program to interface
with PicoView
27Data Acquisition PicoView Software
Data Types Topography Z height (quantitative
information) Amplitude (AC AFM) rms value of the
cantilevers oscillation at the set frequency
(qualitative info only) Phase (AC- AFM) Phase
difference between driving signal and the
waveform of the tips interaction with the
sample Other types Deflection, Current, Friction
etc
28Gwyddion Imaging Software
Free powerful Imaging Software!
http//gwyddion.net/ recommended by Agilent
29Image Metrology SPIP Imaging Software
Expensive but very versatile!. Free trial.
http//www.imagemet.com\ We have a license for
one station at the time