Title: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
1RADIATION PROTECTION INDIAGNOSTIC
ANDINTERVENTIONAL RADIOLOGY
IAEA Training Material on Radiation Protection in
Diagnostic and Interventional Radiology
2Introduction
- A review is made of
- The main elements of the of X Rays tube cathode
and anode structure - The technology constraints of the anode and
cathode material - The rating charts and X Ray tube heat loading
capacities
3Topics
- Basic elements of an X Ray source assembly
- Cathode structure
- Anode structure
- Rating chart
- X Ray generator
- Automatic exposure control
4Overview
- To become familiar with the technological
principles of the X Ray production
5Part 6 X Ray production
IAEA Training Material on Radiation Protection in
Diagnostic and Interventional Radiology
- Topic 1 Basic elements of an X Ray source
assembly
6Basic elements of the X Ray source assembly
- Generator power circuit supplying the required
potential to the X Ray tube - X Ray tube and collimator device producing the
X Ray beam
7X Ray tubes
8X Ray tube components
- Cathode heated filament which is the source of
the electron beam directed towards the anode - tungsten filament
- Anode (stationary or rotating) impacted by
electrons, emits X Rays - Metal tube housing surrounding glass (or metal) X
Ray tube (electrons are traveling in vacuum) - Shielding material (protection against scattered
radiation)
9X Ray tube components
housing
cathode
1 long tungsten filament 2 short tungsten
filament 3 real size cathode
1 mark of focal spot
10Part 6 X Ray production
IAEA Training Material on Radiation Protection in
Diagnostic and Interventional Radiology
- Topic 2 Cathode structure
11Cathode structure (I)
- Cathode includes filament(s) and associated
circuitry - tungsten material preferred because of its high
melting point (3370C) - slow filament evaporation
- no arcing
- minimum deposit of W on glass envelope
- To reduce evaporation the emission temperature of
the cathode is reached just before the exposure - in stand-by, temperature is kept at 1500C so
that 2700C emission temperature can be reached
within a second
12Example of a cathode
13Cathode structure (I)
- Modern tubes have two filaments
- a long one higher current/lower resolution
- a short one lower current/higher resolution
- Coulomb interaction makes the electron beam
divergent on the travel to the anode - lack of electrons producing X Rays
- larger area of target used
- focal spot increased ? lower image resolution
- Focalisation of electrons is crucial !
14Part 6 X Ray production
IAEA Training Material on Radiation Protection in
Diagnostic and Interventional Radiology
15X Ray tube characteristics
- Anode mechanical constraints
- Material tungsten, rhenium, molybdenum,
graphite - Focal spot surface of anode impacted by
electrons - Anode angle
- Disk and annular track diameter (rotation
frequency from 3,000 to 10,000 revolutions/minute)
- Thickness ? mass and material (volume) ? heat
capacity - Anode thermal constraints
- Instantaneous power load (heat unit)
- Heat loading time curve
- Cooling time curve
16Anode angle (I)
- The Line-Focus principle
- Anode target plate has a shape that is more
rectangular or ellipsoidal than circular - the shape depends on
- filament size and shape
- focusing cups and potential
- distance between cathode and anode
- Image resolution requires a small focal spot
- Heat dissipation requires a large spot
- This conflict is solved by slanting the target
face
17Anode characteristic
1 anode track 2 anode track
18Anode angle (II)
Angle
?
Â
Angle
?
Actual focal spot size
Actual focal spot size
Incident electron beam width
Incident electron beam width
Increased apparent focal spot size
Apparent focal spot size
Film
Film
THE SMALLER THE ANGLE THE BETTER THE RESOLUTION
19Anode heel effect (I)
- Anode angle (from 7 to 20) induces a variation
of the X Ray output in the plane comprising the
anode-cathode axis - Absorption by anode of X photons with low
emission angle - The magnitude of influence of the heel effect on
the image depends on factors such as - anode angle
- size of film
- focus to film distance
- Anode aging increases heel effect
20Anode heel effect (II)
- The heel effect is not always a negative factor
- It can be used to compensate for different
attenuation through parts of the body - For example
- thoracic spine (thicker part of the patient
towards the cathode side of the tube) - mammography
21Focal spot size and imaging geometry
- Focal spot finite size ? image unsharpened
- Improving sharpness ? small focal spot size
- For mammography focal spot size ? 0.4 mm nominal
- Small focal spot size ? reduced tube output
(longer exposure time) - Large focal spot allows high output (shorter
exposure time) - Balance depends on organ movement (fast moving
organs may require larger focus)
22Part 6 X Ray production
IAEA Training Material on Radiation Protection in
Diagnostic and Interventional Radiology
23Heat loading capacities
- A procedure generates an amount of heat depending
on - kV used, tube current (mA), length of exposure
- type of voltage waveform
- number of exposures taken in rapid sequence
- Heat Unit (HU) joule
- unit of potential x unit of tube current x unit
of time - The heat generated by various types of X Ray
circuits are - 1 phase units HU kV x mA x s
- 3 phase units, 6 pulse HU 1.35 kV x mA x s
- 3 phase units, 12 pulse HU 1.41 kV x mA x s
24X Ray tube rating chart (I)
- Tube cooling characteristics and focal spot size
- ? mA - time relationship at constant kV
- intensity decreases with increasing exposure time
- intensity increases with decreasing kV
- Note higher power ? reduced exposure time ?
reduced motion unsharpness
25X Ray tube rating chart (II)
- Manufacturers combine heat loading
characteristics and information about the limits
of their X Ray tubes in graphical representations
called Tube Rating Charts - Example
- Tube A a 300 mA, 0.5 s, 90 kV procedure would
damage the system operated from a 1-phase half
wave rectified generator (unacceptable) - Tube B a 200 mA, 0.1 s, 120 kV procedure comply
with the technical characteristics of the system
operated from a 3-phase fully rectified
generator (acceptable)
26X Ray tube rating chart (III)
X Ray tube A 1 f half-wave rectified 3000 rpm 90
kV 1.0 mm effective focal spot
700 600 500 400 300 200 100
70 kVp
50 kVp
Tube current (mA)
Unacceptable
90 kVp
120 kVp
0.01
0.05
0.1
0.5
1.0
5.0
10.0
Exposure time (s)
27X Ray tube rating chart (IV)
700 600 500 400 300 200 100
X Ray tube B 3f full-wave rectified 10.000 rpm
125 kV 1.0 mm effective focal spot
70 kVp
50 kVp
Tube current (mA)
90 kVp
Unacceptable
125 kVp
Acceptable
0.01
0.05
0.1
0.5
1.0
5.0
10.0
Exposure time (s)
28Anode cooling chart (I)
- Heat generated is stored in the anode, and
dissipated through the cooling circuit - A typical cooling chart has
- input curves (heat units stored as a function of
time) - anode cooling curve
- The following graph shows that
- a procedure delivering 500 HU/s can go on
indefinitely - if it is delivering 1000 HU/s it has to stop
after 10 min - if the anode has stored 120.000 HU, it will take
? 5 min to cool down completely
29Anode cooling chart (II)
Maximum Heat Storage Capacity of Anode
240 220 200 180 160 140 120 100 80 60 40
20
1000 HU/sec
Imput curve
500 HU/sec
350 HU/sec
Heat units stored (x 1000)
250 HU/sec
Cooling curve
1 2 3 4 5 6 7 8 9 10
11 12 13 14
Elapsed time (min)
30Part 6 X Ray production
IAEA Training Material on Radiation Protection in
Diagnostic and Interventional Radiology
31X-ray generator (I)
It supplies the X-ray tube with ? Current to
heat the cathode filament ? Potential to
accelerate electrons ? Automatic control of
exposure (power application time) ? Energy
supply ? 1000 ? X-ray beam energy (of which
99.9 is dissipated as thermal energy)
32X-ray generator (II)
- Generator characteristics have a strong influence
on the contrast and sharpness of the radiographic
image - The motion unsharpness can be greatly reduced by
a generator allowing an exposure time as short as
achievable - Since the dose at the image plane can be
expressed as - D k0 . Un . I . T
- U peak voltage (kV)
- I mean current (mA)
- T exposure time (ms)
- n ranging from about 1.5 to 3
33X-ray generator (III)
- Peak voltage value has an influence on the beam
hardness - It has to be related to medical question
- What is the anatomical structure to investigate ?
- What is the contrast level needed ?
- For a thorax examination 140 - 150 kV is
suitable to visualize the lung structure - While only 65 kV is necessary to see bone
structure - The ripple r of a generator has to be as low as
possible - r (U - Umin)/U x 100
34Tube potential wave form (I)
- Conventional generators
- single ? 1-pulse (dental and some mobile systems)
- single ? 2-pulse (double rectification)
- three ? 6-pulse
- three ? 12-pulse
- Constant potential generators (CP)
- HF generators (use of DC choppers to convert 50Hz
mains into voltages with frequencies in the kHz
range) ? Inverter technology
35Tube potential wave form (II)
Single phase single pulse
kV ripple ()
100
Single phase 2-pulse
13
Three phase 6-pulse
4
Three phase 12-pulse
Line voltage
0.01 s
0.02 s
36The choice of the number of pulses (I)
- Single pulse low power (lt2 kW)
- 2-pulse low and medium power
- 6-pulse uses 3-phase mains, medium and high
power (manual or automatic compensation for
voltage drop) - 12-pulse uses two shifted 3-phase system, high
power up to 150 kW
37The choice of the number of pulses (II)
- CP eliminates any changes of voltage or tube
current - high voltage regulators can control the voltage
AND switch on and off the exposure - voltage can be switched on at any moment
(temporal resolution) - kV ripple lt2 thus providing low patient exposure
- HF combines the advantages of constant
potential and conventional generator - reproducibility and consistency of tube voltage
- high frame rate possible
38Part 6 X-ray production
IAEA Training Material on Radiation Protection in
Diagnostic and Interventional Radiology
- Topic 6 Automatic Exposure Control (AEC)
39Automatic exposure control
- Optimal choice of technical parameters in order
to avoid repeated exposures (kV, mA) - Radiation detector behind (or in front of) the
film cassette (with due correction) - Exposure is terminated when the required dose has
been integrated - Compensation for kVp at a given thickness
- Compensation for thickness at a given kVp
40Automatic exposure control
41Automatic exposure control
- Optimal choice of technical parameters in order
to avoid repeated exposures (kV, mA) - Radiation detector behind (or in front of) the
film cassette (with due correction) - Exposure is terminated when the required dose has
been integrated - Compensation for kVp at a given thickness
- Compensation for thickness at a given kVp
42 Part 6 X-ray production
IAEA Training Material on Radiation Protection in
Diagnostic and Interventional Radiology
- Topic 7 X-ray equipment operation and mode
43X-ray equipment operation mode and application
(II)
- Radiography and Tomography
- Single and 3 ? generators (inverter technology)
- output 30 kW at 0.3 focus spot size
- output 50 - 70 kW at 1.0 focus spot size
- selection of kV and mAs , AEC
- Radiography and Fluoroscopy
- Under couch equipment, three ? generator
(inverter technology) - continuous output of 300
- 500 W - output 50 kW at 1.0 focus size for spot film
- output 30 kW at 0.6 for fluoroscopy (high
resolution) - priority given to contrast
- automatic settings of kV
44X-Ray equipment operation mode and application
(III)
- Radiography and Fluoroscopy
- Over couch equipment, three phase generator
(inverter technology) - continuous output of at
least 500 W - output 40 kW _at_ 0.6 focus size for spot film
- output 70 kW _at_ 1.0 for fluoroscopy (high
resolution) - priority given to contrast
- automatic settings of kV
- Cardiac angiography
- Three phase generator - continuous output ? 1kW
- output 30 kW _at_ 0.4 focus size
- output 80 kW _at_ 0.8 focus size
- frame rate up to 120 fr/s
45Summary
- The main parts of the system contributing to the
desired X Ray production - provide the required source of power
- deliver an appropriate X Ray spectrum
- ensure the optimum adjustment of exposure to
warrant the image quality
46Where to Get More Information
- Equipment for diagnostic radiology, E. Forster,
MTP Press, 1993 - IPSM Report 32, part 1, X-ray tubes and
generators - The Essential Physics of Medical Imaging,
Williams and Wilkins. Baltimore1994 - Manufacturers data sets for different X Ray tubes