Title: MECHANICAL MEASUREMENTS
1MECHANICAL MEASUREMENTS
Prof. Dr. Ing. Andrei Szuder Tel.
40.2.1.4112604 Fax. 40.2.1.4112687 www.labsmn.pub.
ro szuder_at_labsmn.pub.ro
2INTRODUCTION TO MEASUREMENT
LECTURE 1
Prof. Dr. Ing. Andrei Szuder Tel.
40.2.1.4112604 Fax. 40.2.1.4112687 www.labsmn.pub.
ro szuder_at_labsmn.pub.ro
3INTRODUCTION TO MEASUREMENT
Aims What is measurement? Why measurement?
4INTRODUCTION TO MEASUREMENT
Measurement Systems Measurement
Standards Variables Tests Measurement
Instruments Accuracy Calibration
5AIM OF COURSE
To develop the following skills To understand
existing systems To design outline systems To
communicate with system designers and
manufacturers
6MEASUREMENT SYSTEM
Measures a physical quantity Displays and/or
records the result of the measurement in an
appropriate form
7MEASUREMENT IN MANUFACTURING
- To ensure compatibility of processes
- To ensure compatibility of products
- Essential for quality control
- Frequently required in automatic control
8MEASUREMENT STANDARDS
- All measurement is relative
- Measurement involves the comparison of a physical
quantity with a similar physical quantity - Agreed reference standards are required
- A combination of several international agencies
are responsible for maintaining the primary
standard measures of various quantities. The
standard kilogram and the standard second are
maintained by the French. Others are kept
elsewhere. It extremely important that these
standards do not change with time, even over
hundreds of years.
9HIERARCHY OF STANDARDS
These primary standards cant be passed around to
any that wants to take some measurements if we
expect them to maintain their values, so
secondary standards are kept which may be
somewhat less accurate, but much more accessible.
These are calibrated against the primary
standards. In this manner, a hierarchy of
standards exist. In addition, there are Test
Standards which specify test procedures,
terminology, methods of construction and data
reduction. These are often dictated from
professional societies such as the ASME.
10MEASURAND (definition)
- This is the quantity being measured
11PRIMARY STANDARDS (SI Units)
- Length Metre
- Defined as the distance travelled by light in
1/299 792 458 seconds - Reproducible to 1 in 108
- Time Second
- Defined by the rotation time of a specific atom
- Reproducible to 1 in 1011
- Mass Kilogram
- Defined as mass of cylinder held in France
- Reproducible to 1 in 109
12PRIMARY STANDARDS (SI Units)
- Temperature K
- International practical scale
- Defined by various physical properties
- Reproducibility varies with temperature
13DERIVED STANDARDS
- All other quantities are defined in terms of
primary standards. - Examples
- Force Mass x length/time
- Current - defined as the force between two
current carrying wires.
14STANDARDS LABS
- In all major industrial countries.
- ROMANIA
- National Physical Laboratory
- National Engineering Laboratory
15VARIABLES
CONTROL VARIABLE A variable that we are able to
hold constant at a known value. In math,
this is called an independent
variable. DISCRETE VARIABLE Takes on discrete
values, like theroll of the dice. More
important examples are things like different
pieces of equipment that may effect
measurements. CONTINUOUS VARIABLE Not
discrete, anything else. EXTRANEOUS
VARIABLE Cannot be controlled and can effect the
value (e.g. room temp).
16EXEMPLE
The book talks about measuring the gas mileage of
a car. In order to determine this quantity, we
need to measure the miles driven and the volume
of gas used. These are both continuous
variables. Several examples of extraneous
variables are given, including the weather and
the driver.
17MORE DEFINITIONS
PARAMETER A functional relationship between
several variables.
For example, in fluid mechanics, say we have
measured/determined the variables velocity, u,
kinematic viscosity, v, and the length scale D.
Then the Reynolds number is a parameter ReD
uD/v
Noise a random variation of the value due to
effects of extraneous variables. Interference
a deterministic variation in the value due to
extraneous variables.
18TEST MATRICES
If we have a phenomenon y that is a function of
one control variable x, then in order to
determine how y varies with x, we will set x to
several values, x1, x2, x3 and measure the
corresponding y values y1, y2, y3 until we have
enough data to determine the functional form of
y(x). Generally, phenomenon depend on more than
one control variable. Say z(x,y). In this case,
it is necessary to hold one of the control values
(say x x1) constant while varying the other
(y). Then, a second value of x is chosen, and in
best case, the same values of y are repeated.
19RANDOM TESTS
Random tests attempt to minimize the effect of
extraneous variables. If tests are made by
randomly varying the independent variables
(controls) rather than making the tests in order,
we should make interference effects look more
like noise. Noise can be averaged out of a
result by making more and more tests, while
interference can not.
REPETITION Decreases random errors.
20CONCOMITANT METHODS
- It is never smart to believe any single piece of
information. - Instruments let you down more often that you
would hope, as do other methods of getting
results (e.g. computer simulations). - Do not bank on any result until you have verified
it independently with a different method.
21MEASUREMENT INSTRUMENTS
- A measurement instrument gives an estimate of the
value of a physical quantity defined in terms of
the standard - It is essential that the accuracy of this
estimate is known
22ACCURACY
- Accuracy is an estimate of the likely difference
between the measured value and the true value.
- The true value is that which would be obtained
by direct comparison with a primary standard. - Error is basically the same thing as accuracy -
it is a better term but is used less often. - Accuracy is determined by calibration
23ACCURACY
Accuracy can be determined during a calibration.
It is the ability of the system to indicate the
true value exactly. We define the error as e
true value - indicated value. The accuracy is
the error relative to the true value expressed in
a percentage, or
24ACCURACY
- Deviation of the reading from a known input.
Instrument reading
45 deg
accuracy
Known Input
25CALIBRATION
- This is the process which relates the measurement
given by an instrument to the measurement
standard - It enables a set of parameters for the
instruments to be quantified.
26CALIBRATION DEFINITIONS
- Say we have a sensor that has the relationship
y(x) and that we calibrate it by varying x
between xmin and xmax. As a result, we measure
values of y between ymin and ymax. - (Note that ymin is not necessarily the value when
x xmin). Then our input range ri is xmax- xmin
and our output range ro is ymax- ymin. Any
measurements outside of this domain are not
valid.
27CALIBRATION
A sensor is something that changes its value as a
result of some phenomenon that we want to
measure. A good sensor is sensitive to only one
variable. The manner in which the value changes
with the control variable is not theoretically
known in most cases. Therefore, we compare the
output of our sensor/transducer to a known value
(standard) as we vary the control variable. A
thermometer is an example.
28CALIBRATION EXAMPLES
29CALIBRATION EXAMPLES
Thermal Anemometry uses the fact that many
materials change their resistance with
temperature. A hot-wire anemometer is a device
that heats a wire by pumping current through it
and keeps its resistance (and thus its
temperature) constant. When air blow on the wire
the current required to keep the wire hot goes
up. Therefore this instrument is sensitive to
velocity. It needs to be calibrated against a
known velocity, however.
A2, V2, P2
A1, V1, P1
A1 is 100 times larger than A2, so V1 is close to
zero. So, Bernoulli says that
30TRACEABILITY
- All measurement systems should be traceable to
the primary standard via a calibration chain.
31CALIBRATION
- Two kinds of calibration will be considered
- calibrating a new instrument
- calibrating an existing instrument
- We will also look at how the specification of a
commercial instrument is defined and assessed
32MAKING A NEW INSTRUMENT
- An instrument is a device which has an output
which varies in response to changes in the
measurand - The output may be a voltage, a current instrument
- , the position of a pointer, etc.
- The relationship between the output value and the
measurand must be established - This is done by calibration of the instrument
33HOW CALIBRATION IS DONE
- A "known" instrument which measures the same
measurand is required - this instrument must have
already been traceably calibrated - The quantity being measured must be variable over
the range of interest - All other parameters which might affect the
performance of the instrument must be kept
constant as far as possible
34CALIBRATION DATA
35CALIBRATION DATA
- Output may be in Volts, mm, degrees, mA etc.
- The operating range of the instrument can be
found from this data - The calibration data enables the instrument to be
used subsequently for measurement - Systematic and random errors can also be
determined - these define the accuracy of the
instrument
36OPERATING RANGE (definition)
- This defines the range of measurand values which
the instrument can measure - Examples
- Temperature 50 - 250C
- Length 5 - 200 mm
- Pressure 10 - 200 Kpa
- Also called Span, Full Scale Output (FSO), Full
Scale deflection (FSD)
37DETERMINING THE CALIBRATION PARAMETERS
- This is usually done using regression analysis
- If this is not possible, they will have to be
individually estimated.
38LINEAR REGRESSION ANALYSIS
- The value of the measurand, M, is varied in
suitable steps over the operating range,
increasing and decreasing several times. - The instrument output X (usually, but not always
volts) is noted. - A straight line is fitted to the data
X SMb - S is the slope of the line, i.e. the sensitivity
- b is the intercept with the axis
39LINEAR REGRESSION ANALYSIS
- The regression data can then be used to find the
measurand value when the output has a particular
value - output value X
- measurand (X-b)/S
- This is only an approximation since the
relationship is not truly linear - The errors incurred in using this approximation
enable the sources of error to be determined and
quantified. - This is done using the residual values
40(No Transcript)
41ACCURACY AND ERRORS
LECTURE 2
Prof. Dr. Ing. Andrei Szuder Tel.
40.2.1.4112604 Fax. 40.2.1.4112687 www.labsmn.pub.
ro szuder_at_labsmn.pub.ro
42SYSTEMATIC AND RANDOM ERRORS
- Systematic errors are constant for a given
instrument - The values are found by the calibration
- Their effects can be removed
- Random errors cannot be removed
- However, their effects can be reduced by taking
several measurements
43ERROR
- The deviation of a reading from a known input.
Systematic errors can be reduced by calibration.
Reading (cts)
EU value
error
44PARAMETERS DEFINING ACCURACY (ERROR)
- Parameters found from calibration which are used
to define the accuracy (error) - - Precision
- Resolution
- Sensitivity
- Linearity (or non-linearity)
- Hysteresis
- Repeatability/precision/reproducibility
- Environmental errors
45DEFINING ACCURACY
- For a given instrument, only a few of these
errors will be significant - The overall error (and accuracy) will be given by
the combination of the individual errors.
46ACCURACY
- Deviation of the reading from a known input.
Instrument reading
45 deg
accuracy
Known Input
47PRECISION
- Ability to reproduce a certain reading with a
given accuracy.
Standard deviation
samples
Readings
Mean reading
Input value
48LEAST COUNT
- Smallest difference that can be detected.
Counts
1 ct
Volts
Least Count
49RESOLUTION
- The smallest change in the measurand which can be
detected by the instrument - This is normally determined by how the output is
displayed - Straightforward with a digital display
- Not fully defined with an analogue system
- The resolution is sometimes limited by a
particular aspect of the instrument itself (e.g.
a potentiometer) - Sometimes called discrimination
50RESOLUTION OF AN ANALOGUE DISPLAY
51SENSITIVITY AND LINEARITY
- Sensitivity is the rate of change of output vs
input - May be in mV/mm, A/mbar, mm/C
- An ideal instrument has constant sensitivity
- Linearity is a measure of how the sensitivity
varies over the operating range of the instrument
52HYSTERESIS
- This is a measure of the variation in output for
a given measurand value when approached both
upwards and downwards. - It arises due to stiffness, backlash, friction,
magnetic effects
53REPEATABILITY (PRECISION,
REPRODUCIBILITY)
- Repeatability is a measure of the random error
- Strictly speaking, it should be called
repeatability error
54(No Transcript)
55ENVIRONMENTAL ERRORS
- Changes in instrument output due to changes in
environment, not measurand changes - Humidity
- Acceleration
- Vibration
- Temperature
- Pressure
- Mounting effects
56RESIDUAL VALUES AND ERRORS
- The residuals show the difference between the
real points and the fitted curve - For a perfect straight line, the residuals would
all be zero. - The bigger the residual values, the greater the
errors - The form of the residual curve is related to the
types of error
57ERROR
- The deviation of a reading from a known input.
Can be reduced by calibration.
Reading (cts)
EU value
error
58UNCERTAINITY
- The portion of the error that cannot or is not
corrected for by calibration.
Uncertainty 3 sd
Standard deviation (sd)
samples
Readings
Mean calibrated reading
Input EU value
59CALCULATING ERRORS FROM RESIDUALS
- The real value is not on the straight line
- The difference between the output value and the
value indicated by the straight line is ?X, the
residual value - Using the straight line gives an error in the
measured value of ?M - The error, ?M, is related to the residual value,
?X by ?M ?X/S where S is the slope or
sensitivity
60(No Transcript)
61(No Transcript)
62SOURCES OF ERROR
- These can be determined by looking at the form of
the residuals - Each type of error produces a specific form of
residual variation - If there is more than one type of error, these
will be combined
63FORM OF RESIDUALS
- Linearity error shows a systematic deviation from
the straight line - Hysteresis error shows as residuals with opposite
signs for increasing and decreasing measurand - Repeatability error has random variation in the
residuals
64QUANTIFYING THE ERRORS
- If there is one predominant error
- find the largest positive and negative residual
values - divide by S, the sensitivity, to give the
measurand errors - quote as ?m or mmax , -mmin
65QUANTIFYING THE ERRORS
- If there are several sources of errors
- estimate the amount of residual error due to each
source - divide each value by S to give the measurand
error due to that source - The total error is then found by combining these
errors - see Unit 3.
66 67CALIBRATION OF A POSITION MEASURING SYSTEM
- The sensitivity is found to be 3.35V/mm
- The residual errors are plotted on the next slide
- Assume that the vertical axis is in Volts
68Position measuring system - residuals
69SAQ 3
- The maximum error is
- a) ? 0.3 mm
- b) ? 0.03V
- c) ? 0.003mm
- d) ? 0.003mV
70Speed measurement system
- The output of the system is the position of a
needle on a circular display - Calibration gives the sensitivity as 0.5420 per
mph. - The residuals are shown on the next slide
- Assume the vertical scale is in 0 (degrees)
71Speed measurement calibration - residuals
72SAQ 4
- The main source of error here is
- a) non-linearity
- b) repeatibility
- c) speeding
- d) hysteresis
73SAQ 5
- The maximum error is
- a) ? 5mph
- b) ? 0.50
- c) ? 1mph
- d) ? 0.01mph
74Calibration of a mercury thermometer
- Output is the position of the mercury level on a
vertical scale - The sensitivity is found to be 0.132mm/0C
- The residuals are plotted on the next slide
- Assume the vertical scale is in mm
75Mercury thermometer - residuals
76SAQ 6
- The main source of error here is
- a) non-linearity
- b) draughts
- c) repeatability
- d) hysteresis
77SAQ 7
- The maximum error is
- a) ?.04mm
- b) ? 0.040 C
- c) ? 0.4 0 C
- d) ?.04V
78Position sensor
- The output is in the form of a voltage
- The sensitivity is found to be 2.43V/mm
- The residuals are plotted on the next slide
- Assume the vertical axis is in V
79Position sensor - residuals
80SAQ 8
- The main source(s) of error is/are
- a) non-linearity
- b) hysteresis
- c) repeatibility and hysteresis
- d) hysteresis and non-linearity
81SAQ 9
- The maximum error is
- a) .02V
- b) .008mm
- c) 0.5mm
- d) 0.02mm
82Pressure sensor - residuals
83SAQ 10
- The main source(s) of error is/are
- a) Repeatability
- b) Repeatability and hysteresis
- c) Hysteresis
- d) Non-linearity and hysteresis
84SAQ 11
- The maximum error is
- a) ?.012V
- b) ? 0.2kPa
- c) ? 0.07kPa
- d) ? 0.1mm
85MEASURING ERRORS DIRECTLY
- Applies to
- repeatability
- hysteresis
- May be necessary if it is not possible to perform
a complete calibration.
86REPEATABLITY
- Random departures from the ideal line indicate a
repeatability error
87REPEATABILITY AND RESIDUALS
- When the variation in the residuals is random,
repeatability is the only source of error - The repeatability error can be calculated
statistically - If ? is the rms residual value
- The random error (repeatability) is less than
2?/S with a confidence of 95
88MEASURING REPEATABILITY DIRECTLY
- The instrument output is noted for a fixed value
of the measurand several times - This is repeated for several different values of
the measurand - The repeatability error can be taken as the
largest difference which occurs for a given
measurand value - Usually quoted as ? this difference
89HYSTERESIS
- A difference in output for increasing and
decreasing measurand indicates hysteresis
90MEASURING HYSTERESIS DIRECTLY
- The measurand is set to the same value approached
from both directions - The instrument output is noted
- This is done for several values
- The hysteresis error can be taken as the largest
difference which occurs for a given measurand
value when it is approached in opposite
directions. - Usually quoted as ? this difference
91SENSITIVITY ERROR
- If the line is not at 450, then there is a
sensitivity error
92Sensitivity error
- The manufacturer may quote a sensitivity error
- e.g. sensitivity 12.3V/mm, error ??0.5
- When the manufacturers sensitvity value is used
in setting up the instrument, each value will
have an associated error - e.g. position 12.5 mm, sensitivity error
12.5mm x 0.5 0.06mm - Can be avoided by doing your own calibration
93MEASURING ENVIRONMENTAL ERRORS
- If the appropriate environmental parameter can be
varied, its effect on the instrument can be
measured - For example, the variation in output of the
instrument may be noted as the temperature
changes
94MEASURING ENVIRONMENTAL ERRORS
- It is also possible to correct for such errors,
but this makes the measurement much more complex - These errors may be negligible
- They are normally treated as part of the cause of
random errors (repeatability)
95CALIBRATION RESULTS
- An ideal instrument gives a straight line at 45o
through the origin - Deviations from this indicate the errors that are
present
96BIAS
- If the points lie on a line which is parallel but
displaced from the ideal line, then the
instrument is biased
97NON-LINEARITY
- Systematic deviation from the ideal line
indicates non-linearity error
98SPECIFICATION OF A COMMERCIAL INSTRUMENT
- This is found partly from the manufacturer's data
sheet, and partly from investigation in use
99MANUFACTURER'S DATA SHEET
- Commercial transducers and instruments are
supplied with a specification sheet containing
values for relevant parameters - It is not always easy to find the information you
are looking for! - May quote an overall accuracy which combines all
errors - May quote various sources of error, (linearity,
hysteresis, temperature etc) which will need to
be combined.
100HOW CALIBRATION PARAMETERS ARE QUOTED
- range or FSO (Full Scale Output)
- reading
- number of l.s.d. (least significant
digits) - units of measurand
- combinations of these
101HOW CALIBRATION PARAMETERS ARE QUOTED - examples
- 2lsd
- 0.02 Range
- 0.1 FSO
- 0.1 mm
- 20 kPa
- 0.1 FSO 0.2 reading
- 0.2 reading 3lsd
102Accuracy of the reference instrument
- The accuracy of the known instrument should be 10
times the accuracy you expect in the instrument
you are calibrating - The minimum acceptable is 3 times.
103CALIBRATING AN EXISTING INSTRUMENT
- The instrument gives a direct reading of the
quantity being measured (mm, oC, etc) - Calibration enables its accuracy to be assessed
- It may also indicate that the instrument should
be adjusted - Another known instrument is required
- It must be possible to vary the measurand in
appropriate steps
104Assessing a commercial instrument
- Check the range
- Check all the different sources of error
- non-linearity error
- repeatability error
- hysteresis error
- sensitivity error - see later
- temperature error - see later
- any other environmental errors
- any other sources of error mentioned
105Assessing a commercial instrument
- Other important parameters
- Dimensions
- Weight
- Drive voltage required
- Sensitivity
- Temperature operating range
- Cost!!!
106Temperature error
- This is a measure of how the sensitivity varies
with temperature - Such a variation in sensitivity give srise to an
error - If the sensitivity error is 0.01/oC, and the
temperature varies during use by 10oC, this will
give rise to an error of 0.1 of the reading
107CALIBRATION AND QUALITY ASSURANCE (ISO9000)
- An effective system for control and calibration
of measurement standards and equipment must be
maintained - All personnel shall have adequate training
- The calibration system should be reviewed to
maintain its effectiveness - All measurements should take into account errors
and uncertainties in the measurement
108CALIBRATION AND QUALITY ASSURANCE
- Calibration procedures shall be documented
- Objective evidence of the measurement system's
effectiveness should be available - Calibration must be done using equipment
traceable to national standards