Title: CsI Compton Veto Detector for A low Mass WIMP Experiment


1
CsI Compton Veto Detector for A low Mass WIMP
Experiment
  • He Dao
  • KPS_at_Jeju University
  • 23 Oct 2004

2
Content
  • Introduction of our low mass wimp experiment
  • Structure of CsI detector
  • Some test and conclusion
  • plan

3
Status of WIMP detection
  • The thresholds are
  • a few keV
  • Corresponding to
  • the 10GeV mass
  • limit measurable

4
CRESST lowest threshold
  • Cryogenic bolometer
  • 262g Sapphire detector
  • The lowest threshold
  • gotten until now
  • 500keV

5
Our Idea
  • Using low-threshold detector to find low-mass
    WIMP

6
Our detector
  • 5g Ge detector Shielding

7
Background Control
  • Cosmic Ray
  • Underground lab
  • Material background
  • purest material Ge
  • Other particles
  • Multi-layer shielding
  • Pb, Cu, CsI veto detector
  • PSD(Pulse Shape discrimination)

8
Energy Threshold of Detector
  • Threshold?
  • less than 100eV after PSD

PSD Parameter for every current shape
100eV
9
The Aim of Our experiment
  • Our estimated results
  • Target mass 5g
  • Counts rate 1
    count/keVkgday
  • Threshold 100 eV
  • Upgrade to 1kg Ge

10
CsI Compton veto Detector
11
CsI Crystal Structure
  • Top
  • ?150mm?50mm
  • 29 degree slope from height 30mm
  • ?76mm at top
  • Bottom
  • ?150mm?30mm

12
CsI Crystal Structure
  • Wall (6 parts)
  • R75 mm
  • r45 mm
  • 5 L220 mm
  • 1 L171 mm
  • Outlet hole
  • (see pic before)

13
Photon Collection
  • PMT performance
  • CR119, one, at the top of CsI crystal
  • low background, good stability
  • Minus HV -1.5KV
  • Anode resistance 10K
  • Current lt 1mA
  • Reflecting film covers all surface of crystal .

14
Compton Veto Function
  • Shielding for high energy gamma.

15
Primary Performance Test
  • Place Tsinghua, China
  • Time 16-18,Aug,2004
  • Source Cs-137, Co-60
  • Bias 12V
  • HV -1000V
  • Void inside

16
Co-60 test result
  • Resolution becomes better when the source moves
    form bottom to top.
  • Peak moves since amplifier threshold adjusted.

17
Cs-137 test result
  • The result seems bad while source near the
    bottom. That is because of the energy which is
    not high enough.

18
Comparison of two sources
19
Test Conclusion
  • Photon collection efficiency
  • Efficiency becomes worse while source moving
    from top to bottom because only one PMT is used,
    at top.
  • Ray Energy
  • Higher energy ray can be detected more easily
    than lower one, which is good for high energy
    gamma ray shielding in this experiment.

20
Plan
  • Now
  • CsI detector was sent to Korea
  • Nov
  • Installation in Y2L and advanced test
  • MC simulation
  • Dec
  • Built up with HPGe detector

21
  • Thank you!