Title: Plain Film


1
Resident Physics Lectures
  • Plain Film
  • Tomography

George David Associate Professor Department of
Radiology Medical College of Georgia
2
Body Section Tomography
  • Not CT
  • Body Section Imaging
  • Goal
  • keep plane of interest in focus
  • blur all other plans
  • enhances contrast
  • Popularity decreasing because of
  • CT
  • MRI

3
Tomography History
  • Predates CT by decades
  • Used to be popular for inner ear studies
  • Still used at MCG for IVPs

4
Tomography Blurring
  • Blurring accomplished by synchronous movement of
    tube film
  • tube bucky physically connected by rod
  • rod pivots around fulcrum
  • tube moves one direction
  • film moves in other direction

5
Conventional Tomography Blurring
  • Image produced on film
  • Objects above or below fulcrum plane change
    position on film thus blur

6
Tomography Blurring
  • the further from the fulcrum an object is, the
    more it blurs!

7
Tomography Blurring
  • objects shaped oriented in direction of motion
  • elongate
  • do not blur

8
Tomographic Blurring
  • Blurring improved by use of complex motions
  • circular
  • tri-spiral
  • elliptical

Hypocycloidal
9
Complex Blurring
  • The more complex the blurring motion
  • the sharper the tomographic image
  • better blurring
  • the more expensive the machine


10
Tomo Patient Doses
  • Tomo can be high dose procedure
  • several rads not unusual
  • one image per cut with film
  • cuts at many levels routinely employed to find
    cut of interest
  • each cut exposes entire field
  • Can do multiple images per cut with digital
    receptor

11
Thickness of Cut
  • Same as thickness of region in focus
  • Determined by angle tube moves
  • Larger angles yield thinner cuts
  • more motion more blurring
  • Smaller angles yield thicker cuts
  • 0o (stationary) yields conventional film

12
Thickness of Cut
Smaller Angle Thicker Cut
Larger Angle Thinner Cut
13
Cut Thickness
  • Approximate cut thickness for linear tomo

Tomo Angle Thickness of Cut (mm)
(degrees) ----------------------------------------
--------------- 0 Infinite (non-tomo)
2 31 4 16 6 11 10 6 20 3 35 2 50
1
14
Grids Tomo
  • Linear grid used
  • Grid lines parallel with table for linear tomo
  • Grid must change orientation (rotate) in complex
    motion tomo
  • eliminates grid cutoff