Title: Principles of Biomedical Systems
1Principles of Biomedical Systems Devices
Safety in Clinical Environment
2Safety in Clinical Environment
- Definitions
- Safety freedom from unacceptable risk of harm.
- Basic Safety Protection against direct physical
hazards when medical electrical equipment is used
under normal or other reasonably foreseeable
conditions. - Hazard A situation of potential harm to people
or property. - Risk The probable rate of occurrence of a
hazard causing harm and the degree of severity of
the harm. - Effectiveness the ability of an item to meet a
service demand of given quantitative
characteristics. - Safety Integrity Level A qualitative measure of
the level of assurance that a component or system
will function as intended or, if it malfunctions,
will do so in a safe manner
3Types of Hazards
- Electrical hazards
- Electrical shocks (micro and macro) due to
equipment failure, failure of power delivery
systems, ground failures, burns, fire, etc. - Mechanical hazards
- mobility aids, transfer devices, prosthetic
devices, mechanical assist devices, patient
support devices - Environmental hazards
- Solid wastes, noise, utilities (natural gas),
building structures, etc. - Biological hazards
- Infection control, viral outbreak, isolation,
decontamination, sterilization, waste disposal
issues - Radiation hazards
- Use of radioactive materials, radiation devices
(MRI, CT, PET), exposure control
4Types of Hazards
- Many sources of energy, potentially hazardous
substances, instruments and procedures - Use of fire, compressed air, water, chemicals,
drugs, microorganisms, waste, sound, electricity,
radiation, natural and unnatural disaster,
negligence, sources of radiation, etc. - Medical procedures expose patients to increased
risks of hazards due to skin and membranes being
penetrated / altered - 10,000 device related injuries in the US every
year! Typically due to - Improper use
- Inadequate training
- Lack of experience
- Improper (lack of) use of manuals
- Device failure
5Electrical Safety
- Safety in the clinical environment Electrical
safety - Physiological effects of electricity
- Susceptibility parameters
- Distribution of electrical power
- Isolated power systems
- Macroshock hazards
- Microshock hazards
- Electrical safety codes and standards
- Protection
- Power distribution
- Ground fault circuit interrupters (GFCI)
- Equipment design
- Electrical safety analyzers / Testing electrical
systems
6Physiological Effects of Electricity
- For electricity to have an effect on the human
body - An electrical potential difference must be
present - The individual must be part of the electrical
circuit, that is, a current must enter the body
at one point and leave it at another. - However, what causes the physiological effect is
NOT voltage, but rather CURRENT. - A high voltage (K.103V) applied over a large
impedance (rough skin) may not cause much (any)
damage - A low voltage applied over very small impedances
(heart tissue) may cause grave consequences
(ventricular fibrillation) - The magnitude of the current is simply the
applied voltage divided by the total effective
impedance the current faces skin largest. - Electricity can have one of three effects
- Electrical stimulation of excitable tissue
(muscles, nerve) - Resistive heating of tissue
- Electrical burns / tissue damage for direct
current and high voltages
7Physiological Effects of Electricity
The real physiological effect depends on the
actual path of the current
Dry skin impedance93 kO / cm2 Electrode gel on
skin 10.8 kO / cm2 Penetrated skin 200 O / cm2
Physiological effects of electricity. Threshold
or estimated mean values are given for each
effect in a 70 kg human for a 1 to 3 s exposure
to 60 Hz current applied via copper wires grasped
by the hands.
8Physiological Effects of Electricity
- Threshold of perception The minimal current that
an individual can detect. For AC (with wet hands)
can be as small as 0.5 mA at 60 Hz. For DC, 2 10
mA - Let-go current The maximal current at which the
subject can voluntarily withdraw. 6 100 mA, at
which involuntary muscle contractions, reflex
withdrawals, secondary physical effects (falling,
hitting head) may also occur - Respiratory Paralysis / Pain / Fatigue At as low
as 20 mA, involuntary contractions of respiratory
muscles can cause asphyxiation / respiratory
arrest, if the current is not interrupted.
Strong involuntary contraction of other muscles
can cause pain and fatigue - Ventricular fibrillation 75 400 mA can cause
heart muscles to contract uncontrollably,
altering the normal propagation of the
electrical activity of the heart. HR can raise up
to 300 bpm, rapid, disorganized and too high to
pump any meaningful amount of blood ? ventricular
fibrillation. Normal rhythm can only return using
a defibrillator - Sustained myocardial contraction / Burns and
physical injury At 1 6 A, the entire heart
muscle contracts and heart stops beating. This
will not cause irreversible tissue damage,
however, as normal rhythm will return once the
current is removed. At or after 10A, however,
burns can occur, particularly at points of entry
and exit.
9Important Susceptibility Parameters
- Threshold and let-go current variability
Distributions of perception thresholds and let-go
currents These data depend on surface area of
contact, moistened hand grasping AWG No. 8 copper
wire, 70 kg human, 60Hz, 13 s. of exposure
10Important Susceptibility Parameters
- Frequency
- Note that the minimal let-go current happens at
the precise frequency of commercial power-line,
50-60Hz. - Let-go current rises below 10 Hz and above
several hundred Hz.
Let-go current versus frequency Percentile
values indicate variability of let-go current
among individuals. Let-go currents for women are
about two-thirds the values for men.
11Important Susceptibility Parameters
- Duration
- The longer the duration, the smaller the current
at which ventricular fibrillation occurs - Shock must occur long enough to coincide with the
most vulnerable period occurring during the T
wave. - Weight
- Fibrillation threshold increases with body weight
(from 50mA for 6kg dogs to 130 mA for 24 kg dogs.
Fibrillation current versus shock duration.
Thresholds for ventricular fibrillation in
animals for 60 Hz AC current. Duration of current
(0.2 to 5 s) and weight of animal body were
varied.
12Important Susceptibility Parameters
- Points of entry
- The magnitude of the current required to
fibrillate the heart is far greater if the
current is not applied directly to heart
externally applied current loses much of its
amplitude due to current distributions. Large,
externally applied currents cause macroshock. - If catheters are used, the natural protection
provided by the skin (15 kO 2 MO) is bypassed,
greatly reducing the amount of current reqd to
cause fibrillation. Even smallest currents (80
600 µA), causing microshock, may result in
fibrillation. Safety limit for microshocks is 10
µA. - The precise point of entry, even externally is
very important If both points of entry and exit
are on the same extremity, the risk of
fibrillation is greatly reduced even at high
currents (e.g. the current reqd for fibrillation
through Lead I (LA-RA) electrodes is higher than
for Leads II (LL-RA) and III (LL-LA).
13Important Susceptibility Parameters
Effect of entry points on current distribution
(a) Macroshock, externally applied current
spreads through-out the body. (b) Microshock, all
the current applied through an intracardiac
catheter flows through the heart.
14Distribution of Electrical Power
(230 V)
Simplified electric-power distribution for 115 V
circuits. Power frequency is 60 Hz
15Distribution of Power
- If electrical devices were perfect, only two
wires would be adequate (hot and return), with
all power confined to these two wires. However,
there are two major departures from this ideal
case - A fault may occur, through miswiring, component
failure, etc., causing an electrical potential
between an exposed surface (metal casing of the
device) and a grounded surface (wet floor, metal
case of another device etc.) Any person who
bridges these two surfaces is subject to
macroshock.
- Even if a fault does not occur, imperfect
insulation or electromagnetic coupling
(capacitive or inductive) may produce an
electrical potential relative to the ground. A
susceptible patient providing a path for this
leakage current to flow to the ground is subject
to microshock. - The additional ground line provides a good line
of defense! (how / why?)
16GROUND!
- The additional line that is connected directly to
the earth-ground provides the following - In case of a fault (short circuit between hot
conductor and metal casing), a large current will
use the path through the ground wire (instead of
the patient) and not only protect the patient,
but also cause the circuit breaker to open. The
ability of the grounding system to conduct high
currents to ground is crucial for this to work! - If there is no fault, the ground wire serves to
conduct the leakage current safely back to the
electrical power source again, as long as the
grounding system provides a low-resistance
pathway to the ground - Leakage current recommended by ECRI are
established to prevent injury in case the
grounding system fails and a patient touches an
electrically active surface (10 100 µA). -
17Isolated Power Distribution
Not grounded !
- In fact, in such an isolated system, if a single
ground-fault occurs, the system simply reverts
back to the normal ground-referenced system. - A line isolation monitor is used with such
system that continuously monitors for the first
ground fault, during which case it simply informs
the operators to fix the problem. The single
ground fault does NOT constitute a hazard!
- Normally, when there is a ground-fault from hot
wire to ground, a large current is drawn causing
a potential hazard, as the device will stop
functioning when the circuit breakers open ! - This can be prevented by using the isolated
system, which separates ground from neutral,
making neutral and hot electrically identical. A
single ground-fault will not cause large
currents, as long as both hot conductors are
initially isolated from ground!
18Macroshock
- Most electrical devices have a metal cabinet,
which constitutes a hazard, in case of an
insulation failure or shortened component between
the hot power lead and the chassis. There is then
115 230 V between the chassis and any other
grounded object. - The first line of defense available to patients
is their skin. - The outer layer provides 15 kO to 1 MO depending
on the part of the body, moisture and sweat
present, 1 of that of dry skin if skin is
broken, - Internal resistance of the body is 200O for each
limb, and 100O for the trunk, thus internal body
resistance between any two limbs is about 500O
(somewhat higher for obese people due to high
resistivity of the adipose tissue)! - Any procedure that reduces or eliminates the skin
resistance increases the risk of electrical
shock, including biopotential electrode gel,
electronic thermometers placed in ears, mouth,
rectum, intravenous catheters, etc. - A third wire, grounded to earth, can greatly
reduce the effect of macroshock, as the
resistance of that path would be much smaller
then even that of internal body resistance!
19Macroshock Hazards
- Direct faults between the hot conductor and the
ground is not common, and technically
speaking, ground connection is not necessary
during normal operation. - In fact, a ground fault will not be detected
during normal operation of the device, only
when someone touches it, the hazard becomes
known. Therefore, ground wire in devices and
receptacles must be periodically tested.
20Macroshock Hazards
21Microshock Hazards
Small currents inevitably flow between adjacent
insulated conductors at different potentials ?
leakage currents which flow through stray
capacitances, insulation, dust and
moisture Leakage current flowing to the chassis
flows safely to the ground, if a low-resistance
ground wire is available.
22Microshock Hazards
- If ground wire is broken, the chassis potential
rises above the ground a patient who has a
grounded connection to the heart (e.g. through a
catheter) receives a microshock if s/he touches
the chassis. - If there is a connection from the chassis to the
patients heart, and a connection to the ground
anywhere in the body, this also causes
microshock. - Note that the hazard for microshock only exists
if there is a direct connection to the heart.
Otherwise, even the internal resistance of the
body is high enough top prevent the microshocks.
23Microshock via Ground Potentials
Microshocks can also occur if different devices
are not at the exact same ground potential. In
fact, the microshock can occur even when a device
that does not connected to the patient has a
ground fault! A fairly common ground wire
resistance of 0.1O can easily cause a a
500mVpotential difference if initiated due to
a, say 5A of ground fault. If the patient
resistance is lessthen 50kO, this would cause an
above safe current of 10µA
24Microshock Via Ground Potentials
25Safety Codes Standards
- Limits on leakage current are instituted and
regulated by the safety codes instituted in part
by the National Fire Protection Association
(NFPA), American National Standards Institute
(ANSI), Association for the Advancement of
Medical Instrumentation (AAMI), and Emergency
Care Research Institute (ECRI).
26Basic Approaches to Shock Protection
- There are two major ways to protect patients from
shocks - Completely isolate and insulate patient from all
sources of electric current - Keep all conductive surfaces within reach of the
patient at the same voltage - Neither can be fully achieved ? some combination
of these two - Grounding system
- Isolated power-distribution system
- Ground-fault circuit interrupters (GFCI)
27Grounding Systems
- Low resistance (0.15 O) ground that can carry
currents up to the circuit-breaker ratings
protects patients by keeping all conductive
surfaces and receptacle grounds at the same
potential. - Protects patients from
- Macroshocks
- Microshocks
- Ground faults elsewhere (!)
- The difference between the receptacle grounds and
other surface should be no more then 40 mV)
All the receptacle grounds and conductive
surfaces in the vicinity of the patient are
connected to the patient-equipment grounding
point. Each patient-equipment grounding point is
connected to the reference grounding point that
makes a single connection to the building ground.
28Isolated Power Systems
- A good equipotential grounding system cannot
eliminate large current that may result from
major ground-faults (which are rather rare). - Isolated power systems can protect against such
major (single) ground faults - Provide considerable protection against
macroshocks, particularly around wet conditions - However, they are expensive !
- Used only at locations where flammable
anesthetics are used. Additional minor protection
against microshocks does not justify the high
cost of these systems to be used everywhere in
the clinical environment
29Ground Fault Circuit Interrupters (GFCI)
- Disconnects source of electric current when a
ground fault greater than about 6 mA occurs!
When there is no fault, IhotIneutral. The GFCI
detects the difference between these two
currents. If the difference is above a threshold,
that means the rest of the current must be
flowing through elsewhere, either the chassis or
the patient !!!. The detection is done through
the monitoring the voltage induced by the two
coils (hot and neutral) in the differential
transformer!
30 GFCI
The National Electric Code (NEC - 1996) requires
that all circuits serving bathrooms, garages,
outdoor receptacles, swimming pools and
construction sites be fitted with GFCI. Note that
GFCI protect against major ground faults only,
not against microshocks. Patient care areas are
typically not fitted with GFCI, since the loss of
power to life support equipment can also be
equally deadly!
31Protection through Equipment Design
- Strain-relief devices for cords, where cord
enters the equipment and between the cord and
plug - Reduction of leakage current through proper
layout and insulation to minimize the capacitance
between all hot conductors and the chassis - Double insulation to prevent the contact of the
patient with the chassis or any other conducting
surface (outer case being insulating material,
plastic knobs, etc.) - Operation at low voltages solid state devices
operating at lt10V are far less likely to cause
macroshocks - Electrical isolation in circuit design
32Electrical Isolation
- Main features of an isolation amplifier
- High ohmic isolation between input and output
(gt10MO) - High isolation mode voltage (gt1000V)
- High common mode rejection ration (gt100 dB)
33Transformer Isolation Amplifiers
34Optical Isolation Amplifier
35Electrical Safety AnalyzersWiring / Receptacle
Testing
- Three LED receptacle tester
- Simple device used to test common wiring problems
(can detect only 8 of possible 64 states) - Will not detect ground/neutral reversal, or when
ground/neutral are hot and hot is grounded (GFCI
would detect the latter)
36Electrical Safety AnalyzersTesting Electrical
Appliances
- Ground-pin-to-chassis resistance Should be
lt0.15O during the life of the appliance
Ground-pin-to-chassis resistance test
37Electrical Safety AnalyzersTesting Electrical
Appliances
- Chassis leakage current The leakage current
should not exceed 500µA with single fault for
devices not intended for patient contact, and not
exceed 300 µA for those that are intended for
patient contact.
Appliance power switch (use both OFF and ON
positions)
Open switch for appliances not intended
to contact a patient
Grounding-contact switch (use in OPEN position)
Polarity- reversing switch (use both positions)
Appliance
H (black)
To exposed conductive
H
surface or if none, then 10 by
Internal Circuitry
N
120 V
20 cm metal foil in contact
N (white)
with the exposed surface
G
Insulating surface
G (green)
I
H hot
Building ground
Current meter
N neutral (grounded)
G grounding conductor
Test circuit
This connection
is at service
I
lt 500 µA for facility
Ð
owned housekeeping and maintenance appliances
entrance or on
I
gt 300 µA for appliances intended for use in
the patient vicinity
supply side of
separately derived
system
38Electrical Safety AnalyzersTesting Electrical
Appliances
- Leakage current in patient leads
- Potentially most damaging leakage is the one with
patient leads, since they typically have low
impedance patient contacts - Current should be restricted to 50µA for
non-isolated leads and to 10 µA for isolated
leads (used with catheters / electrodes that make
connection to the heart) - Leakage current between any pair of leads, or
between a single lead and other patient
connections should also be controlled - Leakage in case of line voltage appearing on the
patient should also be restricted.
39Leakage current Testers
Test for leakage current from patient leads to
ground
40Leakage Current testers
Test for leakage current between patient leads
41Leakage Current Testers
Test for ac isolation current
Isolation current is the current that passes
through patient leads to ground if and when line
voltage appears on the patient. This should also
be limited to 50µA