THE MEASUREMENT OF pH - PowerPoint PPT Presentation

1 / 50
About This Presentation
Title:

THE MEASUREMENT OF pH

Description:

Ph SCALE. Is logarithmic ... What is the pH of a solution with an H ion concentration of 10-4 mole/L? ... to be more accurate for measuring pH of Tris buffers ... – PowerPoint PPT presentation

Number of Views:255
Avg rating:3.0/5.0
Slides: 51
Provided by: lisase7
Category:
Tags: measurement | the | ph | scale

less

Transcript and Presenter's Notes

Title: THE MEASUREMENT OF pH


1
THE MEASUREMENT OF pH
2
General Principles
  • Convenient way to express hydrogen ion
    concentration, or acidity
  • pH - log H
  • Where concentration is expressed in moles/liter

3
Ph SCALE
  • Is logarithmic
  • Example, pH 5.0 solution has ten times more
    hydrogen ions than pH 6.0 solution
  • As hydrogen ion concentration, or acidity,
    increases, the pH value decreases

4
Consider Ph OfPure Water
  • The H of pure water is 1 X 10-7 mole/L What is
    pH?

5
Consider Ph OfPure Water Cont
  • pH - log H
  • The log of 1 X 10-7 is -7
  • The negative log of 10-7 is -(-7) 7
  • The pH of pure water is 7, which we call neutral

6
Questions
  • What is the pH of a solution with an H ion
    concentration of 10-4 mole/L?
  • What is the pH of solution with an H ion
    concentration of 5.0 X 10-6 mole/L?

7
Answer
  • pH - log H - log 10-4 -(-4) 4
  • pH - log H - log 5.0 X 10-6
  • -(-5.3) 5.3

8
Question
  • What is the concentration of H ions in a
    solution with a pH of 9.0?

9
Answer
  • pH - log H
  • 9.0 - log H
  • - 9.0 log H
  • antilog (-9.0) 1 X 10-9 mole/L

10
Design OfpH Meter/ElectrodeMeasuring Systems
  • pH meter systems measure hydrogen ion
    concentration electrochemical

11
(No Transcript)
12
Animations
  • http//www.umd.umich.edu/casl/natsci/slc/slconline
    /PHM/theory.html

13
Characteristics Types of Electrodes Ph Meters
  • Many types of electrodes and meters
  • Meters
  • Conventional
  • Portable
  • Solid state
  • Microprocessor controlled

14
Electrodes
  • Heart of the system
  • Can buy separate measuring and reference
    electrodes or
  • Combination both electrodes are in one housing
  • Combination almost always found in biology labs
  • But not necessarily right

15
Separate Electrodes
  • Use them for difficult samples that
  • Require different type of glass electrode or
  • Require a different type of junction
  • Junction is small hole from which filling
    solution slowly flows out of electrode
  • Different types of junctions have different flow
    rates
  • Tendency to clog

16
ReferenceElectrodes
  • Two major types
  • Ag/AgCl and calomel (mercury)
  • General purpose usually silver/silver chloride

17
ReferenceElectrodes Cont
  • Calomel electrodes are said to be more accurate
    for measuring pH of Tris buffers
  • But calomel electrodes cannot be used in
    solutions above 80o C and contain mercury
  • Both types are available in combination
    electrodes or separately

18
ProperOperation Avoiding Error
  • Consider electrodes
  • Consider sample
  • Consider meter operation

19
Electrodes
  • Maintain electrodes properly
  • Filling Solution. Filling solution, often
    saturated KCl or AgCl/KCl
  • Refillable electrodes are periodically filled to
    nearly the top
  • Filling hole must be open to pH and closed for
    storage
  • Gel-filled electrodes contain gelled filling
    solution, never refilled

20
(No Transcript)
21
Electrodes
  • Storage. Consult the manufacturer's instructions
  • Dont store electrodes in distilled water
  • New combination electrodes often stored dry
  • Then are conditioned before use by soaking at
    least 8 hours in pH 7 buffer

22
Sample
  • Important part of the system
  • Must be homogenous
  • May be chemical changes in sample
  • May be temperature effects

23
Difficult Samples
  • Non-aqueous solvents
  • High purity water. High purity water does not
    readily conduct current and it absorbs CO2 from
    the atmosphere
  • High salt samples. Sample ions compete with the
    reference filling solution ions

24
Difficult Samples Cont
  • Sample-electrode compatibility. Tris buffer,
    sulfides, proteins, Br-, and I- , can complex
    with silver in Ag/AgCl electrodes leading to a
    clogged junction. A calomel electrode is often
    recommended
  • Slurries, sludges, viscous and colloidal samples.
    Use a fast flowing junction. Keep measuring
    electrode bulb clean

25
OperationCalibration
  • User calibrates frequently
  • Two buffers of known pH forms calibration line.
    Try this in lab
  • Older pH meter first calibration buffer should be
    pH 7.00
  • For acidic samples, second is pH 4.00
  • For basic samples, second is pH 10.00

26
OperationCalibration Cont
  • Newer, microprocessor-controlled meters
  • Any two standard buffers that bracket pH of
    samples may be used
  • Some microprocessor-controlled meters allow use
    of more than two standard buffers

27
The Relationship Between pH andMillivolt Readings
The X axis is pH and the Y axis is millivolts
28
Calibration Buffers
  • Accuracy of pH meter depends on calibration
    buffers
  • Some buffers react with CO2 from air
  • Keep buffer containers closed
  • Throw away buffer after use
  • pH 10.0 buffer is particularly sensitive to CO2

29
Buffers
  • For best accuracy, do not use buffers after their
    expiration date
  • Avoid contamination
  • pH of a buffer will change as its temperature
    changes

30
Temperature And pH
  • Temperature has two important effects
  • Measuring electrode's response to pH is affected
    by the temperature
  • pH of solution that is being measured change as
    its temperature changes

31
Effect of Temperature on pH Readings
32
CompensationFor Temperature
  • Meter needs to "know" temperature of solution
  • Can measure solution temperature with thermometer
    and "tell" pH meter
  • Alternatively, use ATC (automatic temperature
    compensating) probes that is connected to the pH
    meter
  • Compact devices may have a temperature probe
    built into the electrode housing

33
Temperature
  • The pH of some solutions changes with temperature
  • Usually measure pH of samples at the temperature
    at which they will be used

34
A ConventionalMethod or Measuring pH
  • 1. Warm-up meter
  • 2. Open filling hole check filling solution
    level
  • 3. If meter has a "standby mode, use it when
    the electrodes are not immersed in sample

35
A ConventionalMethod or Measuring pH cont
  • 4. Calibrate each day or before use
  • A. Adjust temperature setting.
  • B. Rinse electrodes and blot dry.
  • C. Dont wipe.Immerse electrodes in pH 7.00
    calibration buffer. Junction must be immersed
    and level of sample must be below filling
    solution level. Allow reading to stabilize.

36
A ConventionalMethod or Measuring pH cont
  • D. Adjust meter to read 7.00
  • E. Remove electrodes, rinse, blot dry.
  • F. Place electrodes in second standardization
    buffer Adjust meter to pH. Remove, rinse, and
    blot.

37
A ConventionalMethod or Measuring pH cont
  • G. Older meters Recheck pH 7.00 buffer and
    readjust as necessary. Recheck second buffer and
    readjust as necessary.
  • H. Readjust as needed up to three times. If
    readings are not within 0.05 pH units of what
    they should be, electrode probably needs
    cleaning.

38
Quality Control
  • 5. Optional Quality control checks.Try this
    in lab.
  • A. Linearity Check. Take reading of a third
    calibration buffer. For example, if you
    calibrated with pH 7.00 and 10.00 buffers, check
    pH 4.00 buffer.
  • B. Allow the reading to stabilize and record
    value. Do not readjust meter. If reading is
    outside proper range, service electrodes.

39
Quality Control
  • C. Check pH of a control buffer whose pH is
    known and that has a pH close to the pH of the
    sample.
  • Common to set maximum allowable error of control
    buffer to 0.10 pH units.
  • Do not adjust meter to pH of control buffer.
  • If pH of control buffer is not within required
    tolerance, service electrodes.

40
Sample
  • 6. Set meter to temperature of sample or use
    ATC probe.

41
Sample
  • 7. Place electrodes in sample allow reading to
    stabilize.
  • A. Wait too long, pH of some samples will
    change due to air, chemical reactions, etc.
  • B. Difficult solutions may require longer to
    stabilize.
  • C. Many new pH meters have an autoread
    feature.

42
Sample
  • 9. Record all relevant information.
  • 10. Remove electrodes from sample, rinse and
    blot, store properly with filling hole closed
    (refillable electrodes only).

43
Iv Trouble-shooting
  • First step is to know you have trouble.
  • Symptoms of pH system problems include
  • Reading drifts and won't stabilize.
  • Reading fluctuates.
  • Meter cannot be adjusted to both calibration
    buffers.
  • pH value for a buffer or sample seems wrong.

44
Trouble-shooting Tips
  • Look for and correct simple (embarrassing)
    mistakes
  • Electrode measuring bulb and junction are not
    immersed in sample.
  • Meter is not turned on or plugged in, or the
    electrode cables are not connected to meter.

45
Trouble-shootingTips Cont.
  • Reference electrode is not filled with
    electrolyte.
  • Reference electrode filling hole is closed.
  • Sample is not well-stirred.
  • Calibration buffers are not good.
  • Electrode is cracked or broken.

46
Next
  • What is most likely cause of problems?
  • Problems can arise in
  • The reference electrode
  • The measuring electrode
  • The calibration buffers
  • The sample
  • The meter

47
Reference Electrode
  • Reference electrode junction most common source
    of problems.
  • Occluded junction causes long stabilization time
    reading drifts slowly towards correct pH.
  • If junction is completely plugged, reading may
    never stabilize.
  • This is also caused by broken electrodes and by
    some problems within meter.

48
Reference Electrode cont
  • Slow equilibration also caused by changes in
    sample temperature, reactions in sample, or
    sample-electrode incompatibility.

49
Reference Electrode cont
  • Poor calibration buffers will cause inaccuracy.
    This may be detected if solution of known pH is
    checked.
  • If sample is not homogenous, or if its
    temperature is unstable, then pH readings will
    fluctuate or drift.
  • Difficult samples may be slow to equilibrate or
    give incorrect readings.

50
Reference Electrode cont
  • Complete lack of response likely caused by
    problems with meter, but meter is least likely
    component to cause problems.
  • To check which component is at fault substitute
    in a new reference or combination electrode.
Write a Comment
User Comments (0)
About PowerShow.com