Title: THE MEASUREMENT OF pH
1THE MEASUREMENT OF pH
2General Principles
- Convenient way to express hydrogen ion
concentration, or acidity - pH - log H
- Where concentration is expressed in moles/liter
3Ph SCALE
- Is logarithmic
- Example, pH 5.0 solution has ten times more
hydrogen ions than pH 6.0 solution - As hydrogen ion concentration, or acidity,
increases, the pH value decreases
4Consider Ph OfPure Water
- The H of pure water is 1 X 10-7 mole/L What is
pH?
5Consider Ph OfPure Water Cont
- pH - log H
- The log of 1 X 10-7 is -7
- The negative log of 10-7 is -(-7) 7
- The pH of pure water is 7, which we call neutral
6Questions
- What is the pH of a solution with an H ion
concentration of 10-4 mole/L? - What is the pH of solution with an H ion
concentration of 5.0 X 10-6 mole/L?
7Answer
- pH - log H - log 10-4 -(-4) 4
- pH - log H - log 5.0 X 10-6
- -(-5.3) 5.3
8Question
- What is the concentration of H ions in a
solution with a pH of 9.0?
9Answer
- pH - log H
- 9.0 - log H
- - 9.0 log H
- antilog (-9.0) 1 X 10-9 mole/L
10Design OfpH Meter/ElectrodeMeasuring Systems
- pH meter systems measure hydrogen ion
concentration electrochemical
11(No Transcript)
12Animations
- http//www.umd.umich.edu/casl/natsci/slc/slconline
/PHM/theory.html
13Characteristics Types of Electrodes Ph Meters
- Many types of electrodes and meters
- Meters
- Conventional
- Portable
- Solid state
- Microprocessor controlled
14Electrodes
- Heart of the system
- Can buy separate measuring and reference
electrodes or - Combination both electrodes are in one housing
- Combination almost always found in biology labs
- But not necessarily right
15Separate Electrodes
- Use them for difficult samples that
- Require different type of glass electrode or
- Require a different type of junction
- Junction is small hole from which filling
solution slowly flows out of electrode - Different types of junctions have different flow
rates - Tendency to clog
16ReferenceElectrodes
- Two major types
- Ag/AgCl and calomel (mercury)
- General purpose usually silver/silver chloride
17ReferenceElectrodes Cont
- Calomel electrodes are said to be more accurate
for measuring pH of Tris buffers - But calomel electrodes cannot be used in
solutions above 80o C and contain mercury - Both types are available in combination
electrodes or separately
18ProperOperation Avoiding Error
- Consider electrodes
- Consider sample
- Consider meter operation
19Electrodes
- Maintain electrodes properly
- Filling Solution. Filling solution, often
saturated KCl or AgCl/KCl - Refillable electrodes are periodically filled to
nearly the top - Filling hole must be open to pH and closed for
storage - Gel-filled electrodes contain gelled filling
solution, never refilled
20(No Transcript)
21Electrodes
- Storage. Consult the manufacturer's instructions
- Dont store electrodes in distilled water
- New combination electrodes often stored dry
- Then are conditioned before use by soaking at
least 8 hours in pH 7 buffer
22Sample
- Important part of the system
- Must be homogenous
- May be chemical changes in sample
- May be temperature effects
23Difficult Samples
- Non-aqueous solvents
- High purity water. High purity water does not
readily conduct current and it absorbs CO2 from
the atmosphere - High salt samples. Sample ions compete with the
reference filling solution ions
24Difficult Samples Cont
- Sample-electrode compatibility. Tris buffer,
sulfides, proteins, Br-, and I- , can complex
with silver in Ag/AgCl electrodes leading to a
clogged junction. A calomel electrode is often
recommended - Slurries, sludges, viscous and colloidal samples.
Use a fast flowing junction. Keep measuring
electrode bulb clean
25OperationCalibration
- User calibrates frequently
- Two buffers of known pH forms calibration line.
Try this in lab - Older pH meter first calibration buffer should be
pH 7.00 - For acidic samples, second is pH 4.00
- For basic samples, second is pH 10.00
26OperationCalibration Cont
- Newer, microprocessor-controlled meters
- Any two standard buffers that bracket pH of
samples may be used - Some microprocessor-controlled meters allow use
of more than two standard buffers
27The Relationship Between pH andMillivolt Readings
The X axis is pH and the Y axis is millivolts
28Calibration Buffers
- Accuracy of pH meter depends on calibration
buffers - Some buffers react with CO2 from air
- Keep buffer containers closed
- Throw away buffer after use
- pH 10.0 buffer is particularly sensitive to CO2
29Buffers
- For best accuracy, do not use buffers after their
expiration date - Avoid contamination
- pH of a buffer will change as its temperature
changes
30Temperature And pH
- Temperature has two important effects
- Measuring electrode's response to pH is affected
by the temperature - pH of solution that is being measured change as
its temperature changes
31Effect of Temperature on pH Readings
32CompensationFor Temperature
- Meter needs to "know" temperature of solution
- Can measure solution temperature with thermometer
and "tell" pH meter - Alternatively, use ATC (automatic temperature
compensating) probes that is connected to the pH
meter - Compact devices may have a temperature probe
built into the electrode housing
33Temperature
- The pH of some solutions changes with temperature
- Usually measure pH of samples at the temperature
at which they will be used
34A ConventionalMethod or Measuring pH
- 1. Warm-up meter
- 2. Open filling hole check filling solution
level - 3. If meter has a "standby mode, use it when
the electrodes are not immersed in sample
35A ConventionalMethod or Measuring pH cont
- 4. Calibrate each day or before use
- A. Adjust temperature setting.
- B. Rinse electrodes and blot dry.
- C. Dont wipe.Immerse electrodes in pH 7.00
calibration buffer. Junction must be immersed
and level of sample must be below filling
solution level. Allow reading to stabilize.
36A ConventionalMethod or Measuring pH cont
- D. Adjust meter to read 7.00
- E. Remove electrodes, rinse, blot dry.
- F. Place electrodes in second standardization
buffer Adjust meter to pH. Remove, rinse, and
blot.
37A ConventionalMethod or Measuring pH cont
- G. Older meters Recheck pH 7.00 buffer and
readjust as necessary. Recheck second buffer and
readjust as necessary. - H. Readjust as needed up to three times. If
readings are not within 0.05 pH units of what
they should be, electrode probably needs
cleaning.
38Quality Control
- 5. Optional Quality control checks.Try this
in lab. - A. Linearity Check. Take reading of a third
calibration buffer. For example, if you
calibrated with pH 7.00 and 10.00 buffers, check
pH 4.00 buffer. - B. Allow the reading to stabilize and record
value. Do not readjust meter. If reading is
outside proper range, service electrodes.
39Quality Control
- C. Check pH of a control buffer whose pH is
known and that has a pH close to the pH of the
sample. - Common to set maximum allowable error of control
buffer to 0.10 pH units. - Do not adjust meter to pH of control buffer.
- If pH of control buffer is not within required
tolerance, service electrodes.
40Sample
- 6. Set meter to temperature of sample or use
ATC probe.
41Sample
- 7. Place electrodes in sample allow reading to
stabilize. - A. Wait too long, pH of some samples will
change due to air, chemical reactions, etc. - B. Difficult solutions may require longer to
stabilize. - C. Many new pH meters have an autoread
feature.
42Sample
- 9. Record all relevant information.
- 10. Remove electrodes from sample, rinse and
blot, store properly with filling hole closed
(refillable electrodes only).
43Iv Trouble-shooting
- First step is to know you have trouble.
- Symptoms of pH system problems include
- Reading drifts and won't stabilize.
- Reading fluctuates.
- Meter cannot be adjusted to both calibration
buffers. - pH value for a buffer or sample seems wrong.
44Trouble-shooting Tips
- Look for and correct simple (embarrassing)
mistakes - Electrode measuring bulb and junction are not
immersed in sample. - Meter is not turned on or plugged in, or the
electrode cables are not connected to meter.
45Trouble-shootingTips Cont.
- Reference electrode is not filled with
electrolyte. - Reference electrode filling hole is closed.
- Sample is not well-stirred.
- Calibration buffers are not good.
- Electrode is cracked or broken.
46Next
- What is most likely cause of problems?
- Problems can arise in
- The reference electrode
- The measuring electrode
- The calibration buffers
- The sample
- The meter
47Reference Electrode
- Reference electrode junction most common source
of problems. - Occluded junction causes long stabilization time
reading drifts slowly towards correct pH. - If junction is completely plugged, reading may
never stabilize. - This is also caused by broken electrodes and by
some problems within meter.
48Reference Electrode cont
- Slow equilibration also caused by changes in
sample temperature, reactions in sample, or
sample-electrode incompatibility.
49Reference Electrode cont
- Poor calibration buffers will cause inaccuracy.
This may be detected if solution of known pH is
checked. - If sample is not homogenous, or if its
temperature is unstable, then pH readings will
fluctuate or drift. - Difficult samples may be slow to equilibrate or
give incorrect readings.
50Reference Electrode cont
- Complete lack of response likely caused by
problems with meter, but meter is least likely
component to cause problems. - To check which component is at fault substitute
in a new reference or combination electrode.