Title: Computed Tomography
1Computed Tomography
- FRCR Examination
- Craig Moore
- Clinical Scientist
- Radiation Physics Service
2Brief History of CT Scanning
3First CT Scanner - 1972
- 80 x 80 resolution
- 4 min. per rotation
- 8 grey levels
- overnight reconstruction
- dual slice.
4Here and Now
- 512 x 512 or 1024 x 1024 resolution
- 0.4 sec. per rotation
- 4096 grey levels
- Reconstruction16 secs for 30 slices
- 64 slices.
5Basic Principles of CT ScanningWhy CT?
- Conventional radiography suffers from the
collapsing of 3D structures onto a 2D image - Although lower resolution than general
radiography, it has extremely good low contrast
resolution, enabling the detection of very small
changes in tissue type - CT gives accurate diagnostic information about
the distribution of structures inside the body
6Collapse of 3D Data into 2D Plane
- Planar imaging
- 2D representation of 3D Distribution of Tissue
- No depth information
- Structures at different depths are superimposed
- Loss of contrast
Object Contrast 41
X rays
7(No Transcript)
8(No Transcript)
9(No Transcript)
10X-ray Tube
- X-rays are produced by firing electrons at a
metal target - Beam filtered (0.5 mm Cu and/or 1.3 mm Al
filters) to optimise spectrum - Stops attenuation coefficients varying with depth
via beam hardening
11- Compton scatter predominates over the kV range
used - CT is very demanding of X-ray tubes
- Up to 500 mA instantaneously and must provide
X-rays for 30 secs - Large heat capacity and fast cooling rates
- Up to 1.4 MHU/min
- Tube rotation causes mechanical stresses (up to
13 G)
12(No Transcript)
13CT Numbers
- Ideally the linear attenuation coefficient ?,
calculated by the reconstruction algorithm would
be absolutely correct - Coefficients can be regarded as correct in a
relative fashion only - Relative accuracy of modern scanners lt 0.1
- A CT number is a relative value of the
attenuation coefficient
14(No Transcript)
15Window Width and Level
- The appearance of the image on the screen can be
changed by altering the window width and level - Window width refers to the range of CT numbers
selected for display - This range of CT numbers is centred at a
particular level called the window level - e.g. if imaging bone window level should be 1000
- Can spread a small range of CT numbers over a
large range of grayscale values
16Window Level 593 Window Width 529 Good contrast
in lungs
Window Level 12 Window Width 400 Good soft
tissue contrast
17(No Transcript)
18How do we get the picture?
- Back Projection
- Reverse the process of measurement of projection
data to reconstruct image - Each projection if smeared back across the
reconstructed image
19(No Transcript)
20Back Projection
- Back Project each planar image onto three
dimensional image matrix
3
3
6
3
3
6
6
3
3
3
3
6
21Back Projection
- Back Project each planar image onto three
dimensional image matrix
3
3
6
1
1
2
1
1
2
1
1
2
22Back Projection
- Back Project each planar image onto three
dimensional image matrix
3
3
6
1
1
2
3
2
2
3
2
1
1
2
6
3
4
3
1
1
2
3
2
3
2
23Back Projection
- Back Project each planar image onto three
dimensional image matrix
3
3
6
3
3
4
4
6
6
6
6
6
8
6
3
3
4
4
3
3
6
24Back Projection
- Back Project each planar image onto three
dimensional image matrix
3
3
6
3
3
4
4
6
6
6
6
6
8
6
3
3
4
4
3
3
6
25Filtered Back Projection
- Back projection produces blurred transaxial
images - Projection data needs to be filtered before
reconstruction - Different filters can be applied for different
diagnostic procedures - Smoother filters for viewing soft tissue
- Sharp filters for high resolution imaging
- Back projection same as before
26(No Transcript)
27(No Transcript)
28Back Projection
- More views better reconstruction
- 1/r blurring, even with infinite number of views
29Filtered Back Projection
- Filter planar views prior to back projection
- Correction of 1/r blurring requires Ramp Filter
- Gives increasing weight to higher spatial
frequencies - Amplifies Noise
30Filtered Back Projection
- In Practice
- Use modifications of Ramp Filter
- Compromise between Noise and Spatial Resolution
31Problems with Filtered Back Projection
- Back projection is mathematically correct, but
real life images require Filtered Back Projection - Back Projection can introduce noise and streaking
artefacts - Not good with attenuation correction
- Filtered Back Projection can reduce noise and
artefacts, but may degrade resolution
32(No Transcript)
33(No Transcript)
34(No Transcript)
35(No Transcript)
36(No Transcript)
37(No Transcript)
38(No Transcript)
39(No Transcript)
40Xenon v Solid State
- Xenon Single detector chamber sub-divided
by electrodes - Solid state Detector array made up of
individual elements
41Number of detectors and projections
- Typically, for a 3rd generation scanner
- 650 900 detectors
- 1000 to 2000 projections per rotation
42Gantry Rotation
- Data transmission and power cables in old
scanners meant start stop rotation - Change direction at end of rotation to rewind
- Series of images acquired by rotating clockwise
and anti-clockwise - Scan times as fast as 1 s
43(No Transcript)
44(No Transcript)
45Helical and Multi-Slice Scanning
46Requirements for Helical Scanning
- Slip rings
- Continuous table feed through gantry
47(No Transcript)
48(No Transcript)
49(No Transcript)
50Advantages of Helical Scanning
- Speed
- No need to pause between scans for table movement
- Pitches greater than 1 allowed
- Longer scan lengths within breath hold
- Reduced patient movement artefacts
51(No Transcript)
52Disadvantages of Helical Scanning
- Broadening of Slice profile
- Effective slice thickness increases poorer z
axis resolution - Higher noise
- Helical artefacts not seen in axial scanning
53Multi Slice CT
54(No Transcript)
55(No Transcript)
56(No Transcript)
57Each element is 1.25 mm wide
58Beam width is varied
59Physical acquisition
Computer reconstruction
60(No Transcript)
61(No Transcript)
62Single slice
Multi slice
Single slice
Multi slice
3 rotations
One rotation
63(No Transcript)
64Dose and Multi-Slice Scanners
- Considerations similar to those of single slice
scanners - Dose utilisation on z axis usually poorer than
with single slice scanners - X ray beam width is generally broader than the
total imaged width - Geometric efficiency down to 50 for very small
slice thicknesses (sub mm)
65Geometric Efficiency
Broader beam to negate the effect of penumbra
Extra beam/x geo eff
66Dose and Helical CT
- All helical scanning requires extra irradiation
at the end of each run to obtain sufficient
interpolation data to reconstruct the required
volume - On multi-slice scanners this extra length can be
quite long
67(No Transcript)
68(No Transcript)
69Helical multi-slice pitch
- Pitch can be defined with respect to total beam
collimation (x), or individual detector width (d) - In practice pitchx is useful when looking at
patient dose, pitchd for image quality - Pitchx table travel/xray beam width
- Pitchd table travel/detector width
70(No Transcript)
71Helical Multi-Slice in the UK
- Helical Multi-Slice scanning represent
significant steps forward in CT - Better scanning of previous scans
- Expansion of workload
- Nearly all scanners sold in UK are multi-slice
- Technology is still advancing
- 32/40/64/256 slice scanners now available
- More slices in the future?