Mammography II - PowerPoint PPT Presentation

1 / 33
About This Presentation
Title:

Mammography II

Description:

Firm compression reduces overlapping anatomy and decreases tissue thickness of the breast. ... Wax insert composition and 'undegraded' radiograph. Phantom QA ... – PowerPoint PPT presentation

Number of Views:131
Avg rating:3.0/5.0
Slides: 34
Provided by: michae354
Category:

less

Transcript and Presenter's Notes

Title: Mammography II


1
Mammography II
  • Compression, scattered radiation, magnification
  • Screen-film cassettes processing
  • Quality assurance

2
(No Transcript)
3
Compression
  • Firm compression reduces overlapping anatomy and
    decreases tissue thickness of the breast. This
    results in
  • Fewer scattered x-rays
  • Less geometric blurring of anatomic structures
  • Lower radiation dose to the breast tissues
  • Uniform breast thickness lessens exposure dynamic
    range and allows the use of higher contrast film

4
(No Transcript)
5
Compression (cont.)
  • Compression paddle should match the size of the
    image receptor, be flat and parallel to the
    breast support table, and not deflect more than 1
    cm at any location when compression is applied
  • A right-angle edge at the chest wall produces a
    flat, uniform breast thickness when compressed
    with a force of 10 to 20 newtons (22 to 44 pounds)

6
(No Transcript)
7
Spot compression
  • Smaller compression paddle area ( 5 cm diameter)
    reduces even further the breast thickness in a
    specific area and redistributes the tissue for
    improved contrast and anatomic rendition
  • Valuable in delineating anatomy and achieving
    minimum thickness in an area of the breast
    presenting with suspicious findings on previous
    images

8
Compression of a suspicious are shows no visible
mass
9
Contrast
  • Maximum contrast with scatter

10
Scatter
  • The amount of scatter in mammography increases
    with increasing breast thickness and breast area,
    and is relatively constant with kVp
  • The fraction of scattered radiation in the image
    can be reduced by use of antiscatter grids or air
    gaps, as well as vigorous compression

11
Scatter depends on breast thickness and field area
12
Antiscatter grid
  • Parallel linear grids with a ratio of 41 or 51
    are commonly used
  • Aluminum and carbon fiber are common interspace
    materials
  • Grid frequencies range from 30 to 50 lines/cm for
    moving grids, and up to 80 lines/cm for
    stationary grids

13
Linear grid (typically 51), cellular grid for 2D
scatter rejection, and the air gap (intrinsic to
magnification procedures)
14
2D copper scatter grid with air interspaces
transmits 87 of primary photons (Argonne
National Labs, 2006)
15
Magnification
  • Magnification is achieved by
  • Placing a breast support platform to a fixed
    position above the detector
  • Selecting the small focal spot
  • Replacing the antiscatter grid with a cassette
    holder, and
  • Using an appropriate compression paddle

16
Geometric magnification
17
Magnification (cont.)
  • Advantages of magnification include
  • Increased effective resolution of the image
    receptor by the magnification factor
  • Reduction of effective image noise
  • Reduction of scattered radiation
  • Limitations of magnification include
  • Geometric blurring caused by the finite focal
    spot size
  • Small focal spot limits tube current, and extends
    exposure times. Even slight breast motion will
    cause blurring during the long exposure times

18
Illustration of the need for the 0.1 mm focal
spot in magnification studies
19
Screen-film cassettes
  • Most cassettes are made of low-attenuation carbon
    fiber and have a single high-definition phosphor
    screen used with a single emulsion film
  • The screen is positioned in the back of the
    cassette
  • Light spread varies with the depth of x-ray
    absorption
  • Most x-rays interact in the layers of the screen
    closest to the film, preserving spatial resolution

20
Film-screen system for mammography
21
(No Transcript)
22
Film processor QA
  • Must be done daily, prior to first patient use
  • Solution levels and temperatures checked
  • Base fog must be within 0.03 of baseline
  • Mid-density must be within 0.15 of baseline
  • Density difference (contrast) must be within
    0.15 of baseline

23
Film processor testing
24
(No Transcript)
25
Radiation dosimetry
  • Mean glandular dose preferred index since
    glandular tissue is most always the site of
    carcinogenesis
  • Speed of screen-film system and film OD preferred
    by radiologist are major factors
  • Should not exceed 3.0 mGy (300 mrad) per film

26
(No Transcript)
27
Mammography phantom
  • Used to determine adequacy of overall imaging
    system (including film processing) in terms of
    detection of subtle radiographic findings, and to
    assess reproducibility of image characteristics
    (e.g., contrast and optical density) over time

28
(No Transcript)
29
Phantom (cont.)
  • Intended to mimic attenuation characteristics of
    a standard breast of 4.2 cm compressed
    thickness composed of 50 adipose tissue and 50
    glandular tissue
  • Wax insert contains
  • 6 cylindrical nylon fibers of decreasing
    diameter
  • 5 simulated calcification groups (Al2O3 specks)
    of decreasing size
  • 5 low contrast disks, of decreasing diameter and
    thickness, that simulate masses

30
Wax insert composition and undegraded radiograph
31
(No Transcript)
32
Phantom QA
  • Should be done weekly must be done monthly
  • Minimum of 4 largest fibers, 3 largest speck
    groups, and 3 largest masses must be visible
  • Number of test objects of each group type visible
    in image should not decrease by more than one
    half
  • Background OD should be at least 1.4 and not vary
    more than 0.20 from operating level
  • Density difference due to a 4.0 mm acrylic disc
    should be at least 0.40 and not vary more than
    0.05 from established operating level

33
Mammographic image of composite phantom
Write a Comment
User Comments (0)
About PowerShow.com