Title: Instructions
1Instructions
2UV Infrared and Raman Spectroscopy
3Molecular energy
- Electronic
- Vibrational
- Rotational
-
- All molecular energies are quantised and give
rise to discrete energy levels in the molecule
4Molecular orbital bonding theory
Electronic energy
Atomic orbitals combined to form molecular
orbitals Allowed orbital combinations are
determined by symmetry, spatial distribution and
energy. Only valance orbitals are involved in
bonding.
Example O2
?
?
5Example H2O
6Example CH4
7Example C6H6
energy
LUMO
HOMO
8Possible electron transitions in organic compounds
- ? to ? Large ?E. Generally beyond the UV
visible range, 150 nm - n to ? Large ?E (less than above) but rarely
seen in UV - n to ? and ? to ? ?E in the 200-700 nm
range - CT (charge transfer) strong molar
absorbtivities, often near UV and may result in
strong colours - Allowed transitions are excited by incident
light with energy equal to the energy gap between
ground state and excited state. Measurement of
the light energies absorbed from a continuous
source of light provides a UV visible spectrum.
9Molecular vibrations
A molecule with N atoms will have 3N-6 (or 3N-5
linear) normal vibrations. Example H2O
N3, 3 vibrations
?3
?2
?1
10Molecular vibrations
- A molecule will have 3N-6(5) sets of vibrational
energy levels each characterised by a vibrational
quantum number vi
11Absorption of infrared radiation
If ?rad ?vib IR radiation will be absorbed
12IR selection rule
There must be a change in dipole moment
(?) during a vibration (qv) if it is to absorbe
IR radiation and thus be IR active.
13Vibrational transitions
Evib h?vib
To excite a vibrational transition
h?rad h?vib
The radiation frequency must equal the vibration
frequency to excite a transition.
Vibrational frequencies are 1013 Hz The
infrared region of the spectrum.
If we measure the infrared frequencies (normally
expressed in cm-1) absorbed by a molecule then we
will measure its vibrational frequencies
and obtain its Infrared spectrum.
14Molecular energy levels
e4
e3
energy
e2
e1
15Spectroscopy
- Spectroscopy measurement of wavelength
(frequency) and Intensity of light - Light (electromagnetic radiation) radiates from a
source with a speed of 3x108 m s-1 - Light travels as a waveform but may behave as
quantised energy packets (photons) depending upon
detection method - Energy Planks constant x frequency E h?
16Wavelength ?
Frequency ? number of wavelengths propagating
per second
Speed C ? ? 1/? ?/C
1/? ? ?
When ? is in cm, 1/? wavenumber ? (cm-1)
17(No Transcript)
18Interference
19Diffraction grating
2
1
N
N
?
?
?
?
a
For constructive interference between rays 1 and
2
K order of diffraction (normally 1 is used)
20Spectrometer principle
slit
detector
grating
21UV visible spectrometer
22Absorbance (A) and Molar Absorbtivity (?)
A log I0/ I
? A / c I
c sample concentration in moles/liter l
length of light path through the sample in cm.)
23Absorbance (A) and Molar Absorbtivity (?)
24(No Transcript)
25Measurement of infrared spectra
- Dispersion instrument spectrometer similar to
UV visible but using an IR source, IR detector.
Chopper alternates IR through reference and
sample paths, an attenuator is adjusted to
balance the beams and its movement is a measure
of absorbance. - FTIR instrument (Fourier Transform IR)
26Dispersive IR spectrometer
attenuator
27Simplified FTIR instrument
sample
detector
28Interference due to path difference between fixed
and moving mirrors
29 Individual wavelengths
Mirror
position (time)
intensity
Summed
wavelengths interferogram
Path difference 0
30Fourier transform (FT)
FT will transform time dependent data to
frequency dependent data
31Apodization function
- In theory an interferogram spans - ? to ?
- A limiting (apodization function) is used to deal
with this problem which in turn imparts shape to
the spectral bands and removes sidebands which
arise if a simple boxcar truncation function is
used .
32sample
detector
33FT
time
Continuous (all wavelengths) IR interferogram
34Background subtraction
35Characteristic IR Band Positions and Group
Frequency Ranges (cm-1)
OH stretching vibrations Free OH 3610-3645
(sharp) Intramolecular H bonds 3450-3600
(sharp) Intermolecular H Bonds 3200-3550
(broad) Chelate Compounds 2500-3200 (very
broad) NH Stretching vibrations Free NH
3300-3500 H bonded NH 3070-3350 CH
Stretching vibrations -C-H 3280-3340
C-H 3000-3100 C-CH3 2862-2882, 2652-2972
O-CH3 2815-2832 N-CH3 (aromatic)
2810-2820 N-CH3 (aliphatic) 2780-2805
CH2 2843-2863,2916-2936 CH 2880-2900
36SH Stretching Vibrations Free SH 2550-2600
C-N Stretching Vibrations Nonconjugated
2240-2260 Conjugated 2215-2240 C-C
Stretching Vibrations C-CH (terminal)
2100-2140 C-C-C-C 2190-2260
C-C-C-C-CH 2040-2200 CO Stretching
Vibrations Nonconjugated 1700-1900
Conjugated 1590-1750 Amides 1650 CC
Sretching Vibrations Nonconjugated
1620-1680 Conjugated 1585-1625
37 CH Bending Vibrations CH2 1405-1465
CH3 1355-1395, 1430-1470 C-O-C Vibrations in
Esters Formates 1175 Acetates 1240,
1010-1040 Benzoates 1275 C-OH Stretching
Vibrations Secondary Cyclic Alcohols
990-1060 CH out-of-plane bending vibrations
in substituted ethylenic systems -CHCH2
905-915, 985-995 -CHCH-(cis) 650-750
-CHCH-(trans) 960-970 CCH2 885-895
38Mercury cadmium telluride detector
39(No Transcript)
40(No Transcript)
41(No Transcript)
42(No Transcript)
43(No Transcript)
44(No Transcript)
45Raman effect
Classical Theory
When a molecule is placed in an electric field,
its electrons are displaced relative to the
nuclei thus developing an electric dipole moment.
For small fields, the induced moment (?i)
is proportional to the field strength (F)
(1)
?i ?F
46? polarisability (the ease with which the
electron cloud of a molecule can be distorted)
A fluctuating field will produce a fluctuating ?i
of the same frequency.
Electromagnetic radiation will induce a
fluctuating dipole (?i) in the molecule of
frequency ?0. This in turn will emit radiation of
frequency ?0.
47RAYLEIGH SCATTERING
48Consider a diatomic molecule vibrating with
a frequency ?v. If it performs simple
harmonic vibrations, then a coordinate along the
axis of vibration (bond length) at time t is
given by qv q0Cos2??vt
(3)
If the polarisability changes during the
vibration, its value for a small vibrational
amplitude will be
49qv q0Cos2??vt (3)
(4)
Substitute (3) into (4)
Consider incident radiation of frequency ?0 from
(1) and (2) ?i ?F ?F0Cos2??0t (6)
Substitute (5) into (6)
50Cos A Cos B ½(Cos (AB) Cos (A-B))
Therefore
The induced dipole moment will scatter light at
different frequencies
51Raman selection rule
There must be a change in polarisability
(?) during a vibration if it is to scatter Raman
shifted radiation and thus be Raman active.
52Raman effect
Quantum mechanical theory
The quantum mechanical theory of Raman
scattering takes account of the fact that the
energies of molecular vibrations are quantised.
Perturbation theory is used to generate the
new vibrational wave function which exists after
the scattering event.
Perturbation theory adds parts of the existing
electronic wave functions of the molecule to the
initial vibrational wave function until the new
vibrational wave function is obtained.
53The perturbing wave function can be considered
as giving rise to an enabling energy level -
Virtual Level
Rayleigh scatter
Stokes Raman scatter
Anti-Stokes Raman scatter
54Raman shift
By convention the exciting frequency (?0) is
regarded as zero and a shift to lower frequency
is assigned a positive value. RAMAN SHIFT
Stokes Raman shifts (0 - ?v) are thus positive
Anti-Stokes Raman shifts (0 ?v) are thus
negative
Raman shift Vibrational frequency (cm-1)
Raman shift corresponds to frequencies in the
infrared spectrum.
55Raman spectrum of carbon tetrachloride
Rayleigh
Stokes
Anti-Stokes
X 100
56Raman spectrum of carbon tetrachloride
intensity
0 1000
Raman shift cm-1
57Why is anti-Stokes scatter less intense than
Stokes scatter ?
The anti-Stokes transition starts from the less
populated v1 state
58Raman effect summary
Sample vibrations ?v
Incident laser light visible region h?o
59The fluorescence problem
- If the laser wavelength falls within the
fluorescence absorption band of the sample or an
impurity then a strong fluorescence emission
background will result which may obscure the
relatively weaker Raman spectrum.
60Spectrometer design
- Raman scatter is lt10-5 intensity of Rayleigh
scatter. - Need a spectrometer which will separate the
relatively small number of Raman photons from the
large number of Rayleigh and incident photons. - Spectrometer needs to be able to filter out the
Rayleigh and incident light and to spectrally
disperse the Raman light to obtain the spectrum. - A sensitive low noise photon counting detector is
essential. - High power monochromatic light source
61Conventional 90 Raman Spectrometry
double monochromator
Lens collects scattered light and focuses it onto
the entrance slit 1
The light is collimated by mirror 1 and
spectrally dispersed by grating 1
Mirror 2 focuses the dispersed light at slit 2
which removes the Rayleigh scattered light.
The second half of the double monochromator
further disperses the light and brings it to a
focus at the exit slit 4.
The gratings rotate to scan the dispersed light
onto the detector. The output (intensity) is
plotted against Raman shift (grating rotation) to
produce the Raman spectrum.
62Infrared and Raman mutual exclusion principal
For molecules with a centre of symmetry,
vibrations which are active in the IR spectrum
will be inactive in the Raman spectrum and vice
versa.
Example CO2
Raman active
Infrared active
?2
?1
?OCO?
Raman spectrum 1 band IR spectrum 2 bands
?3
63cyclohexane C6H12
IR
Raman
64ethanoic acid CH3-(CO)-OH
IR
Raman
652-butanone CH3CH2-(CO)-CH3
IR
Raman
66phenol C6H5OH
IR
Raman
67methylethanoate CH3-(CO)-OCH3
IR
Raman
68cyanoethene CH2CH-C?N
IR
Raman
69cyanopropane CH3CH2CH2-C?N
IR
Raman
70Application Examples
- Aqueous solutions
- High and low temperatures
- Liquids at high pressures
- Solid state properties
71(No Transcript)
72Aqueous KNO3
Aqueous Ca(NO3)2
73Aqueous KNO3
Aqueous Ca(NO3)2
74High temperatures (liquids lt250)
Heating elements
sample
Metal block
laser
Raman scatter
75Low temperatures gt77K
coolant
Copper block
vacuum
sample
Quartz windows
Raman scatter
laser
76Low temperature cryostat cell
Liquid N2 77K Liquid H2 20K Liquid He 4K
Raman scatter
Sample inlet
laser
77Vacuum furnace to study molten salts lt1500C
78Chlorocyclohexane C6H11Cl
C
l
H
H
C
l
equatorial
axial
110 C
22 C
-40 C
79High pressure Raman cell for liquids lt7kbar
80Phtosensitive samples
- Materials which absorbe the laser light will
often be raised in temperature and may decompose. - To counteract this
- Provide a heat sink (good conductor, under water)
- Minimise laser exposure.
81Rotating cell for light sensitive solutions
82Hydraulic press
powder
Spinning cell for light sensitive solids
Compacted crystal powder
Raman scatter
Focussed laser beam
83Raman microscope
84Raman microscope
2 view the magnified sample and position the
focused laser spot
spectrometer
monitor
TV camera
prism
ND filter
beam splitter
monitor
laser
beam splitter
Microscope objective lens
White light
sample
85Raman microscope
3 record the Raman spectrum
spectrometer
monitor
TV camera
prism
monitor
laser
Microscope objective lens
sample
A Raman microscope can record spectra from
samples as small as 0.5?m
86(No Transcript)
87(No Transcript)
88Hot stage (for use with Raman microscope to study
solids)
89Hot stage
90Hot stage
91(No Transcript)
92Temperature / electrical power plot for
polysilicon microheater
93Diamond anvil cell
Pressures up to 400 kbar easily achieved
94Infrared and Raman comparisons
95Infrared and Raman comparisons
96useful websites
- http//www.chemistry.oregonstate.edu/courses/ch361
-464/ch362/irinstrs.htm - http//www.forumsci.co.il/HPLC/FTIR_page.html
- http//mmrc.caltech.edu/mmrc_html/FTIR/FTIRintro.p
df - http//www.entek.chalmers.se/leam/ftir/ftir_pdf/o
wn_pub/theory_ftir.pdf - http//infrared.als.lbl.gov/FTIRinfo.html
- http//www.cem.msu.edu/reusch/VirtualText/Spectrp
y/UV-Vis/spectrum.htm