Spiral CT - PowerPoint PPT Presentation

About This Presentation
Title:

Spiral CT

Description:

Spiral CT Bushong Chapter 5 Spiral CT Continuous source rotation with the patient translation through x-ray beam Patient couch moves as x-ray tube rotates Spiral CT ... – PowerPoint PPT presentation

Number of Views:333
Avg rating:3.0/5.0
Slides: 35
Provided by: aste151
Category:
Tags: scan | spiral

less

Transcript and Presenter's Notes

Title: Spiral CT


1
Spiral CT
  • Bushong Chapter 5

2
Spiral CT
  • Continuous source rotation with the patient
    translation through x-ray beam
  • Patient couch moves as x-ray tube rotates

3
Spiral CT
  • High voltage supplied by slip rings or on board
    generator
  • Slip ring replaces cables
  • No interscan delay
  • Interscan delay is a small delay between slices
    or volumes that is needed during standard axial
    scanning for the x-ray tube to stop and reverse
    direction. Interscan delay can also be used to
    allow extra time during a scan for tube cooling
  • Volumetric imaging within one breath-hold, at
    least 25s

4
Spiral CT
  • Hyperventilation may help patients extend
    breath-hold
  • Tell the patient to breath in and out several
    times before final breath hold
  • Contrast enhanced examination requires less
    contrast media
  • Contrast amounts between 100-150ml are still
    common
  • Some multislice/multidetector scanners are so
    fast that technologists need to be careful about
    not scanning faster than the contrast bolus moves
    through the vascular system
  • Detector data transferred to computer by slip
    rings

5
Spiral CT
  • Examination time is greatly reduced
  • Patient comfort is much improved
  • Because the patient is moved through the gantry
    while the x-ray tube rotates, a spiral pattern
    results

6
Spiral CT
  • Z-axis resolution is slightly reduced with spiral
    CT
  • Effective slice thickness increases with pitch

7
Pitch
  • Pitch is the patient couch movement per rotation
    divided by slice thickness
  • Contiguous spiral
  • Pitch 1, that is 10mm/10mm
  • Extended spiral
  • Pitch 2, that is 20mm/10mm
  • Overlapping spiral
  • ½, that is 5mm/10mm
  • Page 78 Bushong shows diagram

8
Pitch
  • Low pitch results in better z-axis resolution
  • Narrow collimation/high pitch results in better
    z-axis resolution than wide collimation/low pitch
  • Narrow collimation/high pitch is recommended for
    high contrast, thin slice examination, for
    example, lung nodules

9
Pitch
  • Data is collected continuously but not from a
    transverse plane
  • Instead a helical volume is created
  • As spiral CT pitch increases, patient dose is
    reduced
  • Patient dose is approximately proportional to
    1/pitch
  • Patient dose is proportional to slice thickness
    divided by couch movement

10
Pitch
  • Pitch in excess of 2 is not recommended for any
    clinical examination
  • Too much anatomy would be skipped
  • Couch incrementation is usually set to equal
    collimation, pitch 1
  • Couch speed (mm/s) should not exceed slice
    thickness (mm) in order to obtain best compromise
    between image quality and image volume

11
Pitch
  • Couch speed will not normally exceed 10mm/s
  • When pitch exceeds 1, 180 interpolation must be
    used to limit loss of z-axis resolution
  • Longitudinal (z-axis) image coverage is the
    product of couch velocity (mm/s) and image time
    (s)
  • If couch moves 10mm/s and the imaging time is 5s
    then 50mm (5cm) of axial coverage will result
  • The larger the pitch, the more anatomy is covered
    per examination

12
Interpolation
  • Reconstruction of spiral CT images is the same as
    that for conventional CT except for interpolation
  • A transverse planar image can be reconstructed at
    any position along the axis of the patient
    (z-axis)
  • The transverse image is reconstructed from spiral
    data first by interpolation, then by filtered
    back projection

13
Interpolation
  • Either 360 degrees or 180 degrees interpolation
    may be employed
  • Usually 180 degree interpolation is preferred
  • Contiguous reconstruction can result in partial
    volume effect when object is contained in
    adjacent slices

14
Interpolation
  • Overlapping reconstruction may be necessary to
    ensure that object is fully contained within a
    slice
  • Data acquisition is continuous along the z-axis
    therefore by interpolation, image reconstruction
    is at any z-axis position
  • Regardless of z-axis position, slice thickness is
    determined by collimation

15
Interpolation
  • Volume averaging increases with increasing pitch
  • By increasing the pitch the helix becomes more
    spread out and small structures can be missed
  • This is similar to stretching a slinky toy

16
Interpolation
  • Image noise is higher with spiral CT versus
    conventional CT regardless of pitch
  • Interpolation is the computation of an unknown
    value using known values on either side
  • Z-axis resolution is improved with 180 degree
    interpolation compared to 360 degree
    interpolation

17
Interpolation
  • Extrapolation is the computation of an unknown
    value using known values on one side
  • 180 degree interpolation results in a thinner
    slice than 360 degree interpolation
  • 180 degree interpolation results in a noisier
    image than 360 degree interpolation

18
Interpolation
  • 180 degree interpolation results in approximately
    20 higher noise than conventional CT
  • 360 degree interpolation results in approximately
    20 less noise than conventional CT
  • 180 degree interpolation results in better z-axis
    resolution on reformatted longitudinal images
    than 360 interpolation

19
Interpolation
  • 180 degree interpolation allows scanning at a
    higher pitch than 360 degree
  • 360 interpolation broadens sensitivity profile
    more than 180 degree interpolation
  • In general, image noise is less for 360 degree
    interpolation, spiral CT than for conventional CT
  • In general, image noise is much higher for 180
    degree interpolation, spiral CT than for
    conventional CT

20
Interpolation
  • Whether 180 degree or 360 degree interpolation,
    there are linear and higher order reconstruction
    algorithms
  • Two characteristic spiral CT artifacts have been
    identified as breakup and stair step
  • Both the breakup artifact and the stair step
    artifact occur as a consequence of reformatting
    interpolation transverse images to the
    longitudinal plane coronal or sagittal

21
Sensitivity Profile
  • Generally when covering a given length of
    anatomy, thinner collimation and higher pitch are
    preferred because the result is better spatial
    resolution
  • Pitch greater than 21 is not clinically useful
    because of a broadened sensitivity profile and
    reduced z-axis resolution
  • Generally, higher pitch results in thinner slice
    thickness and less partial volume artifact

22
Sensitivity Profile
  • During spiral CT with pitch gt1, the sensitivity
    profile (z-axis resolution) is wider than that of
    conventional CT
  • Spiral CT sensitivity is described by the full
    width at tenth maximum (FWTM) rather than the
    conventional full width at half maximum (FWHM)
  • The higher the pitch, the wider will be the
    sensitivity profile

23
Design Features Slip Ring Technology
  • Slip ring technology made spiral CT possible
  • Normal spiral CT gantry rotation is 1 revolution
    per second
  • Although .5s revolution is possible, the
    engineering required by the stress of centrifugal
    force is formidable

24
Slip Ring Technology
  • There may be multiple slip rings, both high
    voltage and low voltage
  • The design of spiral CT imagers is based on both
    third and fourth generation with no clear
    advantage to either
  • The slip ring contacts or brushes wear and are
    designed to be replaced during preventive
    maintenance

25
X-Ray Tube
  • Spiral CT requires less than 1s 360 degree
    rotation time and at least 5 MHU x-ray tubes
  • For very long scan times, mA must be reduced so
    that x-ray tube loading will not be exceeded
  • Regardless of heat capacity (MHU) and anode
    cooling (kHU/min), spiral CT is usually limited
    by the heat capacity of the focal track
  • High anode heat capacity (6-8 MHU) and rapid
    cooling (1 MHU/min) are required

26
X-Ray Tube
  • In spite of the high heat load, tube life is
    comparable to conventional CT at about 50,000
    exposures

27
Technique SelectionReconstruction
  • Index is the interval at which images are
    reconstructed
  • Index is reconstruction distance divided by
    collimation
  • An index of less than one indicates image overlap
  • An index of greater than one indicates a gap

28
Reconstruction
  • An index of less than one should be employed to
    visualize suspected lung nodules
  • Spiral CT significantly improves coronal and
    sagittal slice reconstruction
  • High quality two-dimensional and
    three-dimensional image reformation are made from
    overlapping transverse images

29
Reconstruction
  • Spiral CT operation requires the following unique
    technique selections
  • Scan time
  • Beam collimation
  • Couch feed velocity
  • Z-axis spacing for image reconstruction
  • Spiral images cannot be reconstructed as rapidly
    as they are acquired. Hence computer memory must
    be excessive

30
Reconstruction
  • Scan time cannot exceed the patients breath-hold
    capacity, usually about 25s
  • Collimation and couch velocity can be selected as
    pitch

31
Z-Axis Resolution
  • Z-axis resolution is compromised in spiral CT but
    not significantly
  • The ability to reconstruct images at any z-axis
    location improves small lesion detection by
    reducing partial volume effects

32
Advantages and Limitations
  • Image noise is usually less with spiral CT
  • More data is acquired in spiral CT therefore,
    image reconstruction takes a little longer
  • Spiral CT replaces single scan techniques with
    volume acquisition techniques
  • Spiral CT misses no anatomy in the scanned volume

33
Advantages and Limitations
  • Spiral CT images can be reconstructed at any
    z-axis position
  • Multiple overlapping transverse images are
    possible in a single breath-hold with no
    additional patient dose
  • Overall scan time is less with spiral CT
    resulting in improved patient throughput
  • Spiral CT takes a bit longer for image processing
    because of the required interpolation before
    planar image reconstruction

34
Advantages
  • Faster image acquisition
  • Contrast can be followed quicker
  • Reduced patient dose at pitch gt 1
  • Physiologic imaging
  • Improved 3d imaging
  • Angiographic imaging
  • Fewer partial volume artifacts
  • Freeze breathing
  • Fewer motion artifacts
  • No misregistration
  • Increased throughput
  • Improved patient comfort
  • Unlimited z-axis resolution
  • Real time CT biopsy
Write a Comment
User Comments (0)
About PowerShow.com