Methods and Tehniques in Surface Science
Prof. Dumitru LUCA Alexandru Ion Cuza
University, Iasi, Romania

2
Ioni Spectroscopies
  • SIMS/SNMS (Secondary Ion/Neutral Mass
    Spectroscopy)
  • the most sensitive elemental.
  • Difficulties in interpretation of spectra
  • Sputtering incident ion ? emergent ion.
  • Requires HV/UHV conditions.
  • Detectioin mass spectrometry
  • RBS (Rutherford Backscattering Spectroscopy)
    scattering of high-energy (MeV) incident ions on
    sample NUCLEI.
  • Probing depth a copuple of ?m!
  • Scattering incident ion emergent ion
  • Sputtering effect is minor (sputtering
    cross-section sectiunea is almost nil at that
    energy values).
  • Detection solid-state scintillators.
  • Equipment particle accelerator.
  • LEISS (ISS) (Low-Energy Ion scattering
    Spectroscopy) scattering of incident ions
    imprastierea on the atoms in the topmost layer of
    the surface.
  • Requires UHV.
  • Equipment Dedicated LEIS spectrometer
  • Detection electrostatic analizer.

3
Low Energy Ion Scattering Spectroscopy (LEIS)
?ltlt dij ? Classical Mechanics Interaction
potential ??
Coulmb Potential Screening function a0 Bohr
radius of the scattering atom. Incident ions
experience to a lesser extent the presence of
nucleus due to electrostatic screening. (in RBS
the screening function 1)
E1 kinetic energy of scattered ions E0
kinetic energy of the incident ion (100 10 000
eV) M1 mass of the incident iont M2 mass
of the scattering atom ?L scattering angle.
Allows for qualitative and semi-quantitative
analysis of surface composition.
4
Low Energy Ion Scattering Spectroscopy (LEIS)
  • For a scattering angle values of ?L 900
    (forward scattering) and ?L 1800 (backward
    scattering), the previous equation becomes even
    simpler
  • The optimum performance in terms of mass
    discrimination involves
  • Due to very high neutralization probability of
    the incident ions by impact with the atoms in the
    surface, the LEIS technique provides information
    on the nature of the ions in the topmost layer of
    the sample surface, exclusively.

5
Low Energy Ion Scattering Spectroscopy
  • The intensity of the detected current, I, is a
    function of the number of the atoms of a species
    k, Nk, via the equuation

  • I K Ip Nk ?
    Pi W
  • where
  • ? - scattering cross-section ( probability
    that an incident ion be scattered towards the
    detector, after a collision with an atom of
    species k),
  • Ip incident beam current,
  • Pi the probability that an ion remains
    un-neutralized after a collision,
  • W entrance solid angle of the detector.
  • The above equation is seldom used for
    quantitative analysis, since the Pi parameter is
    hardly known.
  • Data processing involves to know the scattering
    cross-section and the probability for impact
    neutralization/re-ionization. Usually we rather
    want to calibrate the LEISS machine by using
    standard samples.
  • Most frequently, LEISS is associated
    with complementary techniques.

6
LEIS - Instrumentation
Schematics of the LEIS setup, using the TOF
spectroscopy to detect forward- and backward
scattered particles.
Nuclear Instruments and Methods, Vol. 162, 1979,
p 587.
7
TOF basics
In reality, correction factors should be taking
into accountm ostly in the case of reflectron
configuration
8
LEISS applications
Incident beam 3 keV 3He Detection angle 1350
A LEIS spectrum showing the evolution of a
topmost layer of the Ti during titanium
nitridation
The evolution of the LEIS Ti peak area with
pressure. A Ti surface is exposed to a nitrogen
atmosphere in UHV.
9
N/O substitution at Ti surface