Fourier Transform Infrared (FTIR) Spectrometer - PowerPoint PPT Presentation

About This Presentation
Title:

Fourier Transform Infrared (FTIR) Spectrometer

Description:

Fourier Transform Infrared (FTIR ... APPLICATIONS Identification of inorganic compounds and organic compounds Identification of components of an unknown ... – PowerPoint PPT presentation

Number of Views:1437
Avg rating:3.0/5.0
Slides: 31
Provided by: Subhashr
Category:

less

Transcript and Presenter's Notes

Title: Fourier Transform Infrared (FTIR) Spectrometer


1
Fourier Transform Infrared (FTIR) Spectrometer
  • Subhashree Mishra
  • ATMS Grad Student, UNR
  • W. P. Arnott
  • Physics, UNR

Introduction to Atmospheric Instrumentation (ATMS
360) University of Nevada Reno
2
Energy Levels Basic Ideas
Basic Global Warming The C02 dance
About 15 micron radiation
3
Wavelength and Wavenumber
  • Wavelength 1 / Wavenumber
  • For the IR, wavelength is in microns.
  • Wavenumber is typically in 1/cm, or cm-1.
  • 5 microns corresponds to 2000 cm-1.
  • 20 microns corresponds to 500 cm-1.
  • 15 microns corresponds to 667 cm-1. Much
    terrestrial IR energy at the wavenumber.

4
Carbon Dioxide Concentration
5
Example Problem Instantly Double CO2
Concentration.What is the effect on the infrared
spectrum at the surface?
Consequence The Earths surface warms because
of the additional IR comingto the surface from
the Atmosphere.
6
Example Problem Instantly Double CO2
Concentration.What is the effect on the infrared
spectrum from space?
Consequence The less IR radiation escapes to
space when the atmosphere has 800 ppmCO2 because
the atmosphere is less transparent to IR emitted
by the Earths surface. TheEarths surface
temperature must increase to again balance the
outgoing IR with the incoming solar radiation.
7
LEDs As Detectors
Each photon with enough energy will normally free
exactly one electron, and result in a free hole
as well. If this happens close enough to the
electric field, or if free electron and free hole
happen to wander into its range of influence, the
field will send the electron to the N side and
the hole to the P side. This causes further
disruption of electrical neutrality, and if we
provide an external current path, electrons will
flow through the path to their original side (the
P side) to unite with holes that the electric
field sent there, doing work for us along the
way. The electron flow provides the current, and
the cell's electric field causes a voltage. With
both current and voltage, we have power, which is
the product of the two.
From http//science.howstuffworks.com/solar-cell3.
htm
8
LEDs As Detectors Thermal Noise
9
FTIRs Often Use MCT Detectors Mercury Cadmium
Telluride
HgCdTe or Mercury cadmium telluride (also Cadmium
Mercury Telluride, MCT or CMT) is an alloy of
CdTe and HgTe and is sometimes claimed to be the
third semiconductor of technological importance
after Silicon and Gallium(III) arsenide. The
amount of cadmium (Cd) in the alloy (the alloy
composition) can be chosen so as to tune the
optical absorption of the material to the desired
infrared wavelength. (from http//en.wikipedia.org
/wiki/Mercury_cadmium_telluride)
10
Outline
  • Introduction
  • Theory
  • Design
  • Applications
  • Measurements
  • Discussions

11
What is a FTIR Spectrometer?
  • A spectrometer is an optical instrument used to
    measure properties of light over a specific
    portion of the electromagnetic spectrum, 5
    microns to 20 microns.
  • FTIR (Fourier Transform InfraRed) spectrometer is
    a obtains an infrared spectra by first collecting
    an interferogram of a sample signal using an
    interferometer, then performs a Fourier Transform
    on the interferogram to obtain the spectrum.
  • An interferometer is an instrument that uses the
    technique of superimposing (interfering) two or
    more waves, to detect differences between them.
    The FTIR spectrometer uses a Michelson
    interferometer.

12
FOURIER TRANSFORMS
  • Fourier transform defines a relationship between
    a signal in time domain and its representation in
    frequency domain.
  • Being a transform, no information is created or
    lost in the process, so the original signal can
    be recovered from the Fourier transform and vice
    versa.
  • The Fourier transform of a signal is a continuous
    complex valued signal capable of representing
    real valued or complex valued continuous time
    signals.

13
Fourier Transforms cont.
  • The Continuous Fourier Transform, for use on
    continuous signals, is defined as follows
  • And the Inverse Continuous Fourier Transform,
    which allows you to go from the spectrum back to
    the signal, is defined as
  • F(w) is the spectrum, where w represents the
    frequency, and f(x) is the signal in the time
    where x represents the time. i is sqrt(-1), see
    complex number theory.

14
Fourier Transforms cont.
  • A computer can only work with finite discrete
    signals, not with continuous signals. Thus, we
    need to define the Discrete Fourier Transform
    (DFT).
  • In DFT, the infinite borders of the integrals can
    be replaced by finite ones, and the integral
    symbol can be replaced by a sum. So the DFT is
    defined as
  • And the inverse DFT is defined as

15
FTIR Theory
16
  • The spectrometer described here is a modified
    Bomem MB-100 FTIR.
  • The heart of the FTIR is a Michelson
    interferometer (figure 2).
  • The mirror moves at a fixed rate. Its position is
    determined accurately by counting the
    interference fringes of a collocated Helium-Neon
    laser.
  • The Michelson interferometer splits a beam of
    radiation into two paths having different
    lengths, and then recombines them.
  • A detector measures the intensity variations of
    the exit beam as a function of path difference.
  • A monochromatic source would show a simple sine
    wave of intensity at the detector due to
    constructive and destructive interference as the
    path length changes (refer figure 3).

17
  • In the general case, a superposition of
    wavelengths enter spectrometer, and the detector
    indicates the sum of the sine waves added
    together.
  • Figure 3 shows some idealized light sources, and
    the interferograms that they would theoretically
    produce.
  • The difference in path length for the radiation
    is known as the retardation d (OM OF d) in
    figure 1 and 2.
  • When the retardation is zero, the detector sees
    a maximum because all wavenumbers of radiation
    add constructively.
  • When the retardation is l/2, the detector sees a
    minimum for the wavelength l. An interferogram is
    the sum of all of the wavenumber intensities.

18
Figure 1.
19
Schematic of Michelson Interferometer
Figure 2.
Source MS thesis submitted by Carl George
Schmitt, UNR , 1998.
20
Wave Interference
21
Sample interferograms and their theoretical
source intensity Source MS thesis submitted by
Carl George Schmitt, UNR , 1998.
Figure 3.
22
(No Transcript)
23
Calibration of the FTIR spectrometer
Source MS thesis submitted by Carl George
Schmitt, UNR , 1998.
24
  • The spectrometer produces a complex voltage at
    each wavenumber. A linear model for the
    spectrometer response is assumed, where A is an
    instrument offset, and C is a scaling factor,
  • V ACI (1)
  • If the spectrometer views a perfect blackbody,
    Eq. (1) gives
  • V A CBT (2)
  • where BT is the Planck emission curve for a
    blackbody of temperature T.
  • The two unknowns (A and C) can be determined from
    blackbody measurements at two different
    temperatures,
  • V1 A CBT1
  • V2 A CBT2

25
  • Solving for the unknowns yields
  • C (V1-V2)/(BT1-BT2)
  • and A V1(BT1-BT2)-BT1 (V1-V2)/(BT1-BT2)
  • Returning to Eq (1), The FTIR voltage of another
    target (Vtarget) is related to the target
    radiance (Itarget) by
  • Itarget(BT1-BT2)VtargetBT1V2BT2V1/(V1-V2)
  • Thus, with measurements of blackbodies at two
    temperatures, the calibrated radiance from a
    target (cloud) can be determined.

26
APPLICATIONS
  • Identification of inorganic compounds and organic
    compounds
  • Identification of components of an unknown
    mixture
  • Analysis of solids, liquids, and gasses
  • In remote sensing
  • In measurement and analysis of Atmospheric
    Spectra
  • - Solar irradiance at any point on earth
  • - Longwave/terrestrial radiation spectra
  • Can also be used on satellites to probe the space

27
Source UV thoughts from http//uvb.nrel.colostat
e.edu/UVB/publications/uvb_primer.pdf
28
MODIS Solar Irradiance Source
http//en.wikipedia.org/wiki/ImageMODIS_ATM_solar
_irradiance.jpg
29
Theoretical IR cross sections for absorption in
the wavenumber range most relevant to longwave
(terrestrial) radiation. Source
http//www.patarnott.com/atms749/TheoreticalSpectr
aIR.htm
30
Measurement Example from Reno
Write a Comment
User Comments (0)
About PowerShow.com