Gas phase infrared spectroscopy - PowerPoint PPT Presentation

1 / 42
About This Presentation
Title:

Gas phase infrared spectroscopy

Description:

Introduction to FTIR Gas phase infrared spectroscopy Molecules in gas phase vibrate and rotate at frequencies characteristic to each molecule. – PowerPoint PPT presentation

Number of Views:516
Avg rating:3.0/5.0
Slides: 43
Provided by: gasmetusa
Category:

less

Transcript and Presenter's Notes

Title: Gas phase infrared spectroscopy


1
(No Transcript)
2
Introduction to FTIR
3
Gas phase infrared spectroscopy
  • Molecules in gas phase vibrate and rotate at
    frequencies characteristic to each molecule. Each
    frequency is associated with an energy state of a
    molecule
  • Infrared radiation moves the molecules to higher
    energy states characteristic frequencies are
    absorbed by the molecule in the process
  • Each molecule absorbs infrared radiation at
    several characteristic frequencies (wavelengths)
  • The result is an IR absorption spectrum a
    fingerprint unique to each molecule

4
Gas phase infrared spectroscopy
  • All molecules can be identified on the basis of
    their characteristic absorption spectrum (except
    diatomic elements such as O2 and noble gases)
  • Each molecule absorbs infrared radiation at its
    characteristic frequencies
  • IR absorption spectrum is a fingerprint unique to
    each molecule
  • Beers law Absorption strength i.e absorbance is
    directly proportional to concentration

HCl molecule stretching vibration at 2880 cm-1

IR spectrum of HCl
Absorbance
Wavenumbers
All gases except O2, N2, H2, Cl2, F2, H2S, and
noble gases can be measured
5
IR technologies
  • Gas Filter Correlation IR (GFC)
  • Measures only separate wavelength bands with
    gas-filled filters
  • Only one component can be measured with each
    filter
  • Multiple gases can be measured and spectral
    interference resolved only with additional
    filters (typically maximum 6 gases)
  • Multiple gas filled filters means multiple
    calibration checks
  • Fourier Transform Infrared (FTIR)
  • Spectrometer measures all the IR wavelengths
    simultaneously and produces a full spectrum.
  • Any number of components (up to 50) can be
    analysed from single measurement and
    interferences are automatically resolved
  • Same optical elements used for each measurement,
    multiple calibration checks are not necessary

Optical filters
Broad band light source
Sample cell
B
A
C
B
Interferometer
6
FTIR spectroscopy
Interferometer
  • Based on the use of an optical modulator
    interferometer
  • Interferometer modulates radiation emitted by an
    IR-source, producing an interferogram that has
    all infrared frequencies encoded into it
  • Interferometer performs an optical Fourier
    Transform on the IR radiation emitted by the IR
    source
  • The whole infrared spectrum is measured at high
    speed
  • Spectral range is continuously calibrated with
    HeNe laser
  • Fast, extremely accurate measurements

Modulated IR Beam
Interferogram
Fourier Transformation
IR Spectrum
7
Michelson interferometer
  • Simplest interferometer design
  • Beamsplitter for dividing the incoming IR beam
    into two parts
  • Two plane mirrors for reflecting the two beams
    back to the beamsplitter where they interfere
    either constructively or destructively depending
    on the position of the moving mirror
  • Position of moving mirror is expressed as Optical
    Path Difference (OPD)

OPD Distance travelled by red beam minus
distance travelled by yellow beam
8
Interference
  • Electromagnetic (EM) radiation can be described
    as sine waves having definite amplitude,
    frequency and phase
  • When EM-waves interact, interference is observed
  • Depending on the relative phase of the waves,
    interference is either destructive or constructive

constructive interference
destructive interference
A
A
A
A
Interference signal
Interference signal
EM waves with same amplitude and frequency, out
of phase
EM waves with same amplitude and frequency, in
phase (OPD 0)
9
Mirror movement and interference of single
wavelength beam
When moving mirror is in the original position,
the two paths are identical and interference is
constructive When the moving mirror moves ¼ of
wavelength, the path difference is ½ wavelength
and interference is destructive Mirror moves
back and forth at constant velocity the
intensity of the interference signal varies as a
sine wave
OPD Distance travelled by red beam minus
distance travelled by yellow beam
10
Fourier transformation
  • The interferogram signal is recorded as a
    function of optical path difference
  • The interferogram is comparable to a time domain
    signal (eg. a recorded sound) and the spectrum
    represents the same information in frequency
    domain (eg. the frequency of the same sound)
  • Fourier transformation is the mathematical
    relation between the interferogram and the
    spectrum (in general, between time domain signal
    and frequency signal)
  • A pure cosine wave in the interferogram
    transforms to a perfectly sharp narrow spike in
    the spectrum

Intensity
Intensity
Fourier transformation pair
OPD / cm
Wave number / cm-1
11
Interferogram and spectrum
Spectrum consisting of three discrete frequencies
E(? )
Spectrometer IR source Continuous emission
?
FT
Each frequency contributes a cosine wave to the
interferogram
Observed interferogram of wide band of frequencies
FT
OPD
0
Observed interferogram with centerburst
OPD
OPD
0
Fourier transform analysis converts the recorded
interferogram back into a frequency spectrum by
reversing the process shown at left
12
IR and laser interferograms
  • IR interferogram is recorded after the IR beam
    passes through the interferometer and sample cell
  • IR interferogram contains the absorption of
    sample gas
  • Laser interferogram is produced by a helium-neon
    laser beam travelling through the interferometer
    into a special detector
  • Laser interferogram is a nearly ideal cosine wave
  • Laser interferogram tells the position of moving
    mirror with excellent accuracy

A
IR-interferogram
Laser-interferogram
OPD
?x 632.8 nm
13
Recording an interferogram
  • Laser interferogram signal is used to digitize
    the IR interferogram
  • Single mode HeNe-laser provides a constant
    wavelength output at 632.8 nm
  • Accurate and precise digitization interval
    provides high wavelength accuracy in the spectrum
  • The data points for IR interferogram are recorded
    every time the mirror has moved forward by one
    HeNe laser wavelength

Infrared source
Helium-neon laser
14
Recording an interferogram
  • The digitized IR interferogram (an XY table) is
    transmitted to computer where the Fast Fourier
    Transform (FFT) algorithm computes the spectrum

Infrared source
Infrared source
Helium-Neon laser
Helium-neon laser
0
-L
15
Measurement sequence
  • Transmittance spectrum is a single beam sample
    divided by background
  • Absorbance spectrum negative logarithm of
    transmittance
  • Calcmet automatically converts and displays
    spectra as absorbance spectra

16
Background and absorbance spectra
  • Absorbance spectrum is calculated from the
    background and a single beam sample spectrum
  • The absorbance peak height depends also on the
    concentration c of the sample, absorptivity
    epsilon (this is a physical constant specific to
    each gas and wavelength) and cell lenght l

Zero absorbance means that the amount of light
arriving at the detector is the same in both
sample and background. This is why the background
measurement is often called a zero calibration.
High absorbance means less light arriving at the
detector (-1 in the formula). If the baseline
(region of spectrum without peaks) is above zero,
transmission of light is less than in the
background.
17
Spectral resolution and signal-to-noise ratio
18
Resolution and interferogram
Absorption line shape due to truncation
Truncation points L, L (cm)
Resolution is limited by interferogram truncation
i.e. length of mirror movement
Resolution is also limited by aperture size
  • High resolution comes at a cost
  • long mirror movement (slow)
  • small aperture (little signal)
  • Gasmet design matches interferogram truncation
    with aperture size optimized for high signal to
    noise ratio

Source
Absorption line shape due to finite aperture
2? angle at which the source is seen from the
collimating lens
19
Signal-to-noise ratio
The most important property of the spectrum in
quantitative analysis
spectra are on same absorbance scale
1 cm-1
  • Due to low (8 cm-1) spectral resolution,
    Gasmet has an excellent signal-to-noise ratio
    (SNR).
  • SNR affects the uncertainty (error limits) of
    the analysis
  • - Precise and accurate
  • measurements
  • - Low detection limits
  • and reliable analysis

Absorbance (a.u.)
8 cm-1
3000
2800
2600
2400
2200
2000
1800
1600
1400
1200
1000
800
Wave number (cm-1)
Figure adapted from Instrumental Resolution
Considerations for Fourier Transform Gas-Phase
Spectroscopy. Applied Spectroscopy. Volume 51,
Number 8, 1997.
20
Resolution and dynamic Range
21
Spectral resolution
  • In high resolution spectrum, the band
    intensities are high and can get
  • saturated at relatively low concentrations
    (concentration ? path length)
  • Quantitative analysis precision deteriorates
    when band absorbance is
  • higher than approximately 0.434 A.U.
  • In low resolution spectrum, band intensities are
    low and get saturated
  • only at very high concentrations
  • Low resolution
  • High signal-to-noise ratio (SNR)
  • Wide dynamic range
  • Non-linear calibration
  • Strong spectral overlap
  • Short measurement times
  • High resolution
  • Low signal-to-noise ratio (SNR)
  • Low dynamic range
  • Linear calibration
  • Weak spectral overlap
  • Long measurement times

22
Interferogram and resolution
  • Only an envelope without any fine structure of a
    vibrational absorption band can be detected when
    only a short portion of the interferogram is
    recorded (low resolution)
  • When a longer interferogram is recorded
    containing the centerburst and the signatures,
    the rotational fine structure beneath the
    envelope becomes detectable (high resolution).
  • No information of the molecule is contained in
    the interferogram data points between the
    centerburst and the signatures

noise, no information
23
Advantages of FTIR spectroscopy
  • Speed (Felgett advantage) All the frequencies
    are recorded simultaneously a complete spectrum
    is measured in less than a second.
  • Sensitivity (Jacquinot or Throughput advantage)
    In the interferometer, the radiation power
    transmitted on to the detector is very high which
    results in high sensitivity.
  • Internally Calibrated (Connes advantage) FTIR
    spectrometers employ a HeNe laser as an internal
    wavelength calibration standard, no need to be
    calibrated by the user.
  • Multicomponent capability Since the whole
    infrared spectrum is measured continuously, all
    infrared active components can be identified and
    their concentrations determined.

24
Analysis of FTIR spectra
25
Analysis of FTIR spectra
  • Modified Classical Least Squares (CLS)
  • Use of single component library spectra
  • Use of both line shape and line intensity
  • Cross-correlation effects compensated
  • Residual spectrum and confidence intervals for
    QA/QC
  • Identification of unknowns

26
Cross-Interference compensation
  • Spectral analysis by a line-shape fitting
    modified CLS routine
  • The source of cross-interference is spectral
    overlap
  • Spectra of the interfering (overlapping) species
    used in the CLS routine as interfering
    components

27
Calcmet analysis
Sample spectrum
Reference Spectra (not to same scale)
Calcmet analysis 0.881 Water 10 vol-1.112
CO2 10 vol- 0.995 CO 1000 mg/Nm3 0.910 NO
300 mg/Nm3 0.810 SO2 300 mg/Nm3 0.660 NH3
100 mg/Nm3 0.082 HCl 50 mg/Nm3 0.210 Methane
50 mg/Nm3
Calculated spectrum
Concentrations Water 8.81 vol-
CO2 11.12 vol-
CO 955 mg/Nm3
NO 274 mg/Nm3
SO2 243 mg/Nm3 NH3
66.0 mg/Nm3 HCl 4.1 mg/Nm3 Methane
10.5 mg/Nm3
28
CLS analysis
Example mixture of 80 ppm propane and 150 ppm
ethane
Spectra are completely overlapping.
How to analyse?
29
CLS analysis
  • CLS analysis is an iterative process
  • At each step every individual reference spectrum
    is given a coefficient (k)
  • Model spectrum is calculated as a sum of
    reference spectra weighted by coefficient (k)
  • The difference between measured spectrum and
    model spectrum is called residual spectrum
    (residual)
  • The residual is calculated in every data point of
    the selected analysis area
  • The CLS algorithm searches for smallest possible
    residual by changing the k values
  • When the minimum residual is found, the
    concentrations in the sample spectrum are k times
    concentration of the reference spectra

30
Initial quess Propane 100 ppm, ethane 100 ppm
Residual Sample spectrum Calculated spectrum
Residual not in minimum -gt optimisation continued
31
  • The optimisation stops when
  • k for ethane is 1.5
  • k for propane is 0.8

Residual is only noise Succesful analysis!
Concentrations Ethane 1.5 X 100 ppm 150
ppm Propane 0.8 X 100 ppm 80 ppm
32
Cross interference correction
Methane not included
Successful analysis
  • Cross interference occurs when one or more gases
    are missing from the library
  • Incomplete library leads to large difference
    between measured and calculated spectrum ?
    analysis error
  • Cross interference may be avoided by selecting
    suitable analysis areas avoiding the interfering
    absorption if the library cannot be expanded.

33
Analysis Areas and typical sample spectrum
CO2, NH3, C2H4
CO, N2O
NO2 CH4 C3H8
H2O
HF
NO
HCl HCHO
12 September 2006
Gasmet Technologies Oy 2006
34
Extended CEM settings
35
Gasmet spectrum file (.spe)
Analyzer parameters
Analysis results
  • Stored ample spectrum is an absorbance spectrum,
    no need to ratio it againts background again
    (different from lab FTIRs)
  • Sample spectrum includes analyser hardware status
    information
  • Without sample spectrum, verification of results
    and re-analysis is impossible

36
Quality assurance and control
Toluene added into analysis. Positive reading and
no errors.
Residual error for methane. Invalid results.
Pure noise
Residual spectrum indicating missing gas
(toluene)
37
Gasmet FTIR structure
38
Gasmet structure and outline
FTIR spectrometer
IR source
SAMPLE CELL
Broad band infrared radiation
IR SOURCE
Detector
Signal and data processing
39
Corrosion resistant sample cells
  • Nickel-rhodium-gold plated
  • Fixed mirrors
  • Absorption lengths vary from 1 cm to 9.8 m
    according to application
  • Single pass and multipass (White cell)
  • Can be heated up to 180 oC

40
Sample cells and optical path length
Different path lengths for different measurement
ranges
High Sensitivity (Multipass) Sample Cell V
0.4 l L 60 980 cm T90lt 10 sec (4 lpm)
L 9.8 meter c 10 ppm A 0.0047 a.u
L 2.5 meter c 39 ppm A 0.0047 a.u
L 10 centimeters c 980 ppm A 0.0047 a.u
Single pass cellV 0.013 0.031 l L 1, 4, or
10 cm T90 lt 1 sec (4 lpm)
L 4 centimeters c 2450 ppm A 0.0047 a.u
41
Hot extractive sampling
Sample gas in
Hot 3-way solenoid valve
Flow restricted to 0.5 lpm
Pump
Zero / test gas in
to O2 analyser
Particle filters (two-stage filtration)
to FTIR
Hot zone maintained at 180 oC
  • Wet and hot sample gas is transferred through
    heated lines and pump to the analyzer
  • Sample gas must always be free of particles and
    in gaseous form
  • Sample gas should always be measured against
    ambient air

42
(No Transcript)
Write a Comment
User Comments (0)
About PowerShow.com