Title: Chemicals, Apparatus, and Unit Operations
1- Chapter 2
- Chemicals, Apparatus, and Unit Operations
- of Analytical Chemistry
2Introduction
- At the heart of analytical chemistry is a core
set of operations and equipment. - The technology of analytical chemistry has
improved with the advent of electronic analytical
balances, automated titrators, and other
computer-controlled instruments. - The speed, convenience, accuracy, and precision
of analytical methods have improved as well.
3- 2A Selecting and handling reagents and other
chemicals - The purity of reagents influences the accuracy of
analysis. - Classifying Chemicals
- Regent grade conform to the minimum standards
set forth by the Reagent Chemical Committee of
the American Chemical Society (ACS). - Primary-standard carefully analyzed by the
supplier. The National Institute of Standards and
Technology (NIST) is an excellent source. - Special-purpose reagent chemicals are prepared
for a specific application such as solvents for
spectrophotometry and high-performance liquid
chromatography.
4- Rules for Handling Reagents and Solutions
- Select the best grade of chemical available. Pick
the smallest bottle that is sufficient to do the
job. - Replace the top of every container immediately
after removing reagent. - Hold the stoppers of reagent bottles between your
fingers. Never set a stopper on a desk top. - Unless specifically directed otherwise, never
return any excess reagent to a bottle.
5- Rules for Handling Reagents and Solutions (contd)
- 5. Never insert spatulas, spoons, or knives into
a bottle that contains a solid chemical. Instead,
shake the capped bottle vigorously or tap it
gently against a wooden table to break up an
encrustation. Then pour out the desired quantity.
- 6. Keep the reagent shelf and the laboratory
balance clean and neat. Clean up any spills
immediately. - 7. Follow local regulations concerning the
disposal of surplus reagents and solutions.
6- 2B Cleaning and marking of laboratory ware
- Each vessel that holds a sample must be marked.
Special marking inks are available for porcelain
surfaces. A saturated solution of iron(III)
chloride can also be used for marking. - Every apparatus must be thoroughly washed with a
hot detergent solution and then rinsed, initially
with large amounts of tap water and finally with
several small portions of deionized water. - Properly cleaned glassware will be coated with a
uniform and unbroken film of water. Do not dry
the interior surfaces of glassware. - An organic solvent, such as methyl ethyl ketone
or acetone, may be effective in removing grease
films.
7- 2C Evaporating liquids
- Evaporation is difficult to control because of
the - tendency of some solutions to overheat locally.
- Bumping can cause partial loss of the solution.
- Careful and gentle heating or use of glass beads
- will minimize such loss.
- Some unwanted substances can be eliminated
- during evaporation.
- Wet ashing is the oxidation of the organic
constituents of a sample with oxidizing reagents
such as nitric acid, sulfuric acid, hydrogen
peroxide, aqueous bromine, or a combination of
these reagents.
8- 2D Measuring mass
- An analytical balance must be used to measure
masses with high accuracy. - An analytical balance is used for determining
mass with a maximum capacity that ranges from 1 g
to a few kgs with a precision of at least 1 part
in 105 - A macrobalance has a maximum load of 160-200 g
and a precision of 0.1 mg. - A semimicroanalytical balance has a maximum load
of 10-30 g and a precision of 0.01 mg. - A microanalytical balance has a maximum load of
1-3 g and a precision of 0.001 mg, or 1 µg. - The traditional analytical balance had two pans
and is considered an equal-arm balance. - The single-pan analytical balance was far
superior and replaced the traditional balance. - The electronic analytical balance is the current
balance that is widely used.
9- Fig. 2-2 Electronic analytical balance. (a) Block
diagram. (b) Photo of electronic - balance.
- Placing an object on the pan causes the pan and
indicator arm to move downward, thus increasing
the amount of light striking the photocell of the
null detector. - The increased current from the photocell is
amplified and fed into the coil, creating a
larger magnetic field, which returns the pan to
its original null position.
10- Figure 2-3 Electronic analytical balances. (a)
Classical configuration with pan beneath the
cell. (b) A top-loading design.
11- In each electronic balance, the pan is tethered
to a system of constraints known collectively as
a cell. - The cell has several flexures that permit limited
movement of the pan. - At null, the beam is parallel to the
gravitational horizon. - Electronic balances feature an automatic taring
control that causes the display to read zero with
a container (such as a boat or weighing bottle)
on the pan. - Some electronic balances have dual capacities and
dual precisions. - These features permit the capacity to be
decreased from that of a macrobalance to that of
a semimicrobalance (30 g) with a corresponding
gain in precision to 0.01 mg.
12- Figure 2-4 Single-pan mechanical analytical
balance. - Components
- A light-weight beam is supported on
- a planar surface by a prism-shaped
- knife edge (A).
- A second knife edge (B) is located
- near the left end of the beam and
- support as a second planar surface.
- The two knife edges are prism-shaped
- agate or sapphire devices that form
- low-friction bearings with two planar surfaces
- contained in stirrups also of agate or sapphire.
13- Weighing with a Single-Pan Balance
- The beam of an adjusted balance is in horizontal
position. - Placing an object on the pan causes the left end
of the beam to move downward. - Masses are then removed systematically until the
imbalance is less than 100 mg. - The angle of deflection of the beam is directly
proportional to the additional mass that must be
removed to restore the beam to its horizontal
position. - The optical system measures this angle of
deflection. - A reticle is scribed with a scale that reads 0 to
100 mg. - A vernier makes it possible to read this scale to
the nearest 0.1 mg.
14- Precautions in using an analytical balance
- 1. Center the load on the pan as well as
possible. - 2. Protect the balance from corrosion.
- 3. Observe special precautions for the weighing
of liquids. - 4. Consult your instructor if the balance appears
to need adjustment. - 5. Keep the balance and its case scrupulously
clean. A camels hair brush is useful for
removing spilled material or dust. - 6. Always allow an object that has been heated to
return to room temperature before weighing it. - 7. Use tongs, finger pads, or a glassine paper
strip to handle dried objects to prevent
transferring moisture to them.
15- Sources of error in weighing
- Correction for Buoyancy
- A buoyancy error is an error that develops when
the object being weighed has a significantly
different density than the masses. - Buoyancy corrections may be accomplished with
the equation - W1 is the corrected mass of the object,
- W2 is the mass of the standard masses,
- dobj is the density of the object,
- dwts is the density of the masses, and
- dair is the density of the air displaced by
masses and object. - The value of dair is 0.0012 g/cm3.
16- Figure 2-5. Effect of buoyancy on weighing data
(density of weights 8 g/cm3). Plot of relative
error as a function of the density of the object
weighed.
17(No Transcript)
18- Temperature Effects
- Allow heated objects to return to room
temperature before you attempt to weigh them. - Convection currents within the balance case exert
a buoyant effect. - Warm air trapped in a closed container weighs
less than the same volume at a lower temperature.
Both effects cause the apparent mass of the
object to be low. - Figure 2-6. Absolute error as a function of time
after an - object was removed from a 110C drying
- oven.
- A porcelain filtering crucible.
- B weighing bottle containing about
- 7.5 g of KCl.
19- Other Sources of Error
- A porcelain or glass object will occasionally
acquire a static charge causing a balance to
perform erratically, especially when the relative
humidity is low. - Spontaneous discharge frequently occurs after a
short period. - A low level source of radioactivity in the
balance case will ionize enough ions to
neutralize the charge. - The optical scale of a single-pan mechanical
balance should be checked regularly for accuracy,
particularly under loading conditions that
require the full-scale range. - A standard 100-mg mass is used for this check.
20- Auxiliary Balances
- Less precise than analytical balances.
- Offer the advantages of speed, ruggedness, large
capacity, and convenience. - A sensitive top-loading balance will accommodate
150-200 g with a precision of about 1 mg. - Most are equipped with a taring device.
- Some are fully automatic, require no manual
dialing or mass handling, and provide a digital
readout of the mass. - Modern top-loading balances are electronic.
- A triple-beam balance that is less sensitive than
a typical top-loading auxiliary balance is also
useful. - This is a single-pan balance with three decades
of masses that slide along individual calibrated
scales. - The precision may be one or two orders of
magnitude less.
21- 2E Equipment and manipulations associated with
weighing - Drying or ignition to constant mass is a process
in which a solid is cycled through heating,
cooling, and weighing steps until its mass
becomes constant to within 0.2 to 0.3 mg. - Constant mass ensures that the chemical or
physical processes that occur during the heating
(or ignition) are complete. - Figure 2-7. Weighing Bottles are convenient for
- drying and storing solids.
- Plastic bottles are rugged but abrade easily
- and not easily cleaned
- as compared to glass.
22- Desiccators and Desiccants
- Oven drying is the most common way of removing
moisture from solids. - Dried materials are stored in desiccators while
they cool. - Figure 2-8. A typical dessciator. The base
section of a dessicator contains a chemical
drying agent, such as anhydrous calcium chloride
or calcium sulfate (Drierite). - Very hygroscopic materials should be
- stored in containers equipped with
- snug covers, such as weighing bottles.
- The bottles remain covered while in
- the desiccator.
23- Manipulating Weighing Bottles
- Heating at 105C to 110C for 1 hour is
sufficient to - remove the moisture.
- Figure 2-9. The recommended way to dry a sample.
- Figure 2-10. Avoid touching dried objects with
your - fingers. Instead, use tongs, chamois finger cots,
- clean cotton gloves, or strips of paper to handle
dried - objects for weighing.
24- Weighing by Difference
- The bottle and its contents are weighed.
- One sample is then transferred from the bottle to
a container. - The bottle and its residual contents are then
weighed. - The mass of the sample is the difference between
the two masses. - Weighing Hygroscopic Solids
- Place the approximate amount of sample needed in
the individual bottle and heat for an appropriate
time. - Then, quickly cap the bottles and cool in a
desiccator. - Weigh one of the bottles after opening it
momentarily to relieve any vacuum. - Quickly empty the contents of the bottle into its
receiving vessel, cap immediately, and weigh the
bottle again along with any solid that did not
get transferred. - Repeat for each sample and determine the sample
masses by difference.
25- Weighing Liquids
- The mass of a liquid is always obtained by
difference. - Liquids that are noncorrosive and relatively
nonvolatile can be transferred to previously
weighed containers. - The mass of the container is subtracted from the
total mass. - A volatile or corrosive liquid should be sealed
in a weighed glass ampoule. The ampoule is
heated, and the neck is then immersed in the
sample. - As cooling occurs, the liquid is drawn into the
bulb. - The ampoule is then inverted and the neck sealed
off with a small flame. The ampoule and its
contents, along with any glass removed during
sealing, are cooled to room temperature and
weighed. - The ampoule is then transferred to an appropriate
container and broken. - A volume correction for the glass of the ampoule
may be needed if the receiving vessel is a
volumetric flask.
26- 2F Filtration and ignition of solids
- Apparatus
- Simple crucibles serve only as containers.
Porcelain, aluminum oxide, silica, and platinum
crucibles maintain constant mass. - The solid is first collected on a filter paper.
The filter and contents are then transferred to a
weighed crucible, and the paper is ignited. - Filtering crucibles serve not only as containers
- but also as filters. A vacuum is used to hasten
the - filtration.
- Figure 2-11 Adaptors for filtering crucibles.
27- Sintered-glass crucibles are manufactured in
fine, medium, and coarse porosities. - The upper temperature limit is usually 200C.
- Made of quartz and can tolerate substantially
higher temperatures without damage. - A Gooch crucible has a perforated bottom that
supports a fibrous mat. - Small circles of glass matting are used in pairs
to protect against breaking during the
filtration. - Glass mats can tolerate temperatures in excess of
500C and are substantially less hygroscopic.
28- Filter paper
- Paper is an important filtering medium.
- Ashless paper is manufactured from cellulose
fibers that have been treated with hydrochloric
and hydrofluoric acids. - 9- or 11-cm circles of ashless paper leave a
residue that weighs less than 0.1 mg, which is
negligible under most circumstances. - Ashless paper can be obtained in several
porosities. - A coarse-porosity ashless paper is most effective
for filtering solids, but even with such paper,
clogging occurs. This problem can be minimized by
mixing a dispersion of ashless filter paper with
the precipitate prior to filtration.
29(No Transcript)
30- Heating Equipment
- Many precipitates can be weighed directly after
being brought to constant mass in a
low-temperature drying oven. - The maximum attainable temperature ranges from
140C to 260C. - Microwave laboratory ovens greatly shorten drying
cycles. - An ordinary heat lamp can be used to dry a
precipitate that has been collected on ashless
paper and to char the paper as well. - Burners are convenient sources of intense heat.
The maximum attainable temperature depends on the
design of the burner and the fuel combustion
properties. - The Meker burner provides the highest
temperatures, followed by the Tirrill and Bunsen
types. - A heavy-duty electric furnace (muffle furnace) is
capable of maintaining controlled temperatures of
1100C or higher.
31- Filtering and Igniting Precipitates
- Preparation of Crucibles
- A crucible used to convert a precipitate to a
form suitable for weighing must maintain a
constant mass throughout drying or ignition. - Backwashing a filtering crucible is done by
turning the crucible upside down in the adaptor
and sucking water through the inverted crucible. - Figure 2-12
- (a) Washing by decantation.
- (b) Transferring the precipitate.
- The steps in filtering an analytical
- precipitate are decantation, washing,
- and transfer.
32- The last traces of precipitate on the inside of
the beaker are dislodged with a rubber policeman. - Many precipitates possess the property of
creeping, or spreading over a wetted surface
against the force of gravity. - Filters are never filled to more than three
quarters of capacity to prevent the possible loss
of precipitate through creeping. - The addition of a small amount of nonionic
detergent, such as Triton X-100, to the
supernatant liquid or wash liquid can help
minimize creeping. - A gelatinous precipitate must be completely
washed before it is allowed to dry.
33- Figure 2-13. Preparation of a Filter Paper
34- Figure 2-14 Transferring Paper and Precipitate
to a Crucible
35- Ashing Filter Papers
- If a heat lamp is used, the crucible is placed on
a clean, nonreactive surface, such as a wire
screen covered with aluminum foil. - The lamp is positioned above the rim of the
crucible. - Charring is accelerated if the paper is moistened
with a drop of concentrated ammonium nitrate
solution. - A burner produces much higher temperatures than a
heat lamp. - Partial reduction of some precipitates can occur
- through reaction with the hot carbon
- of the charring paper.
- Figure 2-15 Ignition of a precipitate.
36- Using Filtering Crucibles
- A vacuum filtration train is used when a
filtering crucible can be used - instead of paper.
- Figure 2-16 Train for vacuum filtration. The trap
isolates the filter flask from the source of
vacuum.
37- Rules for Manipulating Heated Objects
- Practice unfamiliar manipulations before putting
them to use. - 2. Never place a heated object on the benchtop.
Instead, place it on a wire gauze or a
heat-resistant ceramic plate. - 3. Allow a crucible that has been subjected to
the full flame of a burner or to a muffle furnace
to cool momentarily. - 4. Keep the tongs and forceps used to handle
heated objects scrupulously clean. In particular,
do not allow the tips to touch the benchtop.
38- 2G Measuring volume
- The precise measurement of volume is as important
to many analytical methods as the precise
measurement of mass. - Units of Volume
- The unit of volume is the liter (L), defined as
one cubic decimeter. The milliliter(mL) is one
one-thousandth of a liter (0.001 L) and the
microliter (µL) is 1026 L or 10-3 mL. - The Effect of Temperature on Volume Measurements
- Most volumetric measuring devices are made of
glass, which has a small coefficient of
expansion.
39- The coefficient of expansion for dilute aqueous
solutions (approximately 0.025/C) is such that
a 5C change has a measurable effect on the
reliability of ordinary volumetric measurements.
40- Apparatus for Precisely Measuring Volume
- Volume may be measured with a pipet, a buret, or
a volumetric flask. - Volumetric equipment is marked TD for to
deliver or TC for to contain and also marked
for the temperature at which the calibration
strictly applies. - Pipets and burets are usually calibrated to
deliver specified volumes. Volumetric flasks are
calibrated to contain a specific volume. - Pipets
- Pipets permit the transfer of accurately known
volumes. - A volumetric pipet delivers a single fixed volume
between 0.5 and 200 mL. - Measuring pipets are calibrated in convenient
units to permit delivery of any volume up to a
maximum capacity ranging from 0.1 to 25 mL.
41Figure 2-17 Typical pipets (a) volumetric pipet,
(b) Mohr pipet, (c) serological pipet, (d)
Eppendorf micropipet, (e) OstwaldFolin pipet,
(f ) lambda pipet.
42(No Transcript)
43- All volumetric and measuring pipets are first
filled to a calibration mark. - A small amount of liquid tends to remain in the
tip after the pipet is emptied. - This residual liquid is never blown out of a
volumetric pipet or from some measuring pipets. - Handheld Eppendorf micropipets deliver adjustable
microliter volumes of liquid.
44Figure 2-18 Variable-volume automatic pipet,
1001000 µL.
45- Burets
- The precision attainable with a buret is
substantially greater than the precision with a
pipet. - A buret consists of a calibrated tube to hold
titrant plus a valve arrangement by which the
flow of titrant is controlled. - Figure 2-19 Burets
- (a) glass-bead valve,
- (b) Teflon valve.
46- Volumetric Flasks
- Figure 2-20 Typical volumetric
- flasks are manufactured with
- capacities ranging from 5 mL to 5 L.
- They are used for the preparation of standard
- solutions and for the dilution of samples to a
- fixed volume prior to taking aliquots with a pi-
- pet.
- Some are also calibrated on a to-deliver
- (TD) basis, and they are distinguished by
- two reference lines on the neck.
47- Using Volumetric Equipment
- Volume markings are blazed on clean volumetric
equipment by the manufacturer. - Only clean glass surfaces support a uniform film
of liquid. - Dirt or oil causes breaks in this film.
- Cleaning
- Brief soaking in a warm detergent solution is
usually sufficient. - The apparatus must be thoroughly rinsed with tap
water and then with three or four portions of
distilled water. - It is seldom necessary to dry volumetric ware.
- Prolonged soaking will cause the formation of a
ring at a detergent/air interface. - This ring cannot be removed and causes a film
break.
48- Avoiding Parallax
- It is common practice to use the bottom of the
meniscus as the point of reference in calibrating
and using volumetric equipment. - Parallax is the apparent displacement of a liquid
level or of a pointer as an observer changes
position. - Parallax is a condition that causes the volume to
appear smaller than its actual value if the
meniscus is viewed from above and larger if the
meniscus is viewed from below. - Figure 2-21 Reading a buret.
49- Directions for Using a Pipet
- Liquid is drawn into a pipet through the
application of a slight vacuum. Never pipet by
mouth because there is risk of accidentally
ingesting the liquid being pipetted instead, use
a rubber suction bulb.
Propipette consists of a rubber bulb (B)
attached to three short sections of tubing. Each
section of tubing contains a small chemically
inert ball (A, C, and D) that functions as a
valve to permit air to flow normally in the
directions indicated by the arrows. The valves
are opened by pinching with thumb and forefinger.
50- Cleaning
- Draw detergent solution to a level 2 to 3 cm
above the calibration mark of the pipet. - Drain this solution and then rinse the pipet with
several portions of tap water. - Inspect for film breaks, and repeat this portion
of the cleaning cycle if necessary. - Finally, fill the pipet with distilled water to
perhaps one third of its capacity and carefully
rotate it so that the entire interior surface is
wetted. - Repeat this rinsing step at least twice.
51- Measuring an Aliquot
- Draw a small volume of the sample liquid into the
pipet and thoroughly wet the entire interior
surface. Repeat twice. - Carefully fill the pipet to a level above the
graduation mark. Be sure that there are no
bubbles. - Touch the pipet tip to the wall of a glass vessel
and slowly allow the liquid level to drop. - When the bottom of the meniscus coincides exactly
with the graduation mark, stop the flow. - Remove the pipet from the volumetric flask, tilt
it until liquid is drawn slightly up into the
pipet, and wipe the tip with a lintless tissue. - Place the pipet tip well within the receiving
vessel, and allow the liquid to drain. - Finally, withdraw the pipet with a rotating
motion to remove any liquid adhering to the tip. - Rinse the pipet thoroughly after use.
52Figure 2-22 Dispensing an aliquot.
53- Directions for Using a Buret
- Cleaning
- Clean the tube with detergent and a long brush.
Rinse thoroughly with tap water and then with
distilled water. Inspect for water breaks. Repeat
if necessary. - Lubricating a Glass Stopcock
- Carefully remove all old grease from a glass
stopcock and its barrel with a paper towel and
dry both parts completely. Lightly grease the
stopcock. Insert the stopcock into the barrel and
rotate it vigorously with slight inward pressure.
- A proper amount of lubricant has been used when
- (1) the area of contact between stopcock and
barrel appears nearly transparent, - (2) the seal is liquid-tight, and (3) no grease
has worked its way into the tip. - Buret readings should be estimated to the nearest
0.01 mL.
54- Filling
- Ensure that the stopcock is closed.
- Add 5 to 10 mL of the titrant, and carefully
rotate the buret to wet the interior completely. - Allow the liquid to drain through the tip. Repeat
twice. - Then fill the buret well above the zero mark.
- Free the tip of air bubbles by rapidly rotating
the stopcock and permitting small quantities of
the titrant to pass. - Lower the level of the liquid just to or somewhat
below the zero mark. - Allow for drainage (lt1 min), and then record the
initial volume reading, estimating to the nearest
0.01 mL.
55- Titration
- Figure 2-23 Recommended method
- for manipulating a buret stopcock.
- Be sure the tip of the buret is well within the
- titration flask, and introduce the titrant in
- increments of about 1 mL.
- Swirl (or stir) constantly.
- Decrease the volume of the
- increments progresseively toward the end point.
- When only a few more drops are needed to reach
the end point, rinse - the walls of the container.
- Allow the titrant to drain from the inner wall of
the buret (at least 30 seconds) at the completion
of the titration. Then, record the final volume,
again to the nearest 0.01 mL.
56- Directions for Using a Volumetric Flask
- Volumetric flasks should be washed with detergent
and thoroughly rinsed. - Only rarely do they need to be dried.
- Drying is best accomplished by clamping the flask
in an inverted position. - Direct Weighing into a Volumetric Flask
- The direct preparation of a standard solution
requires the introduction of a known mass of
solute to a volumetric flask. - Use of a powder funnel minimizes the possibility
of losing solid during the transfer. - Rinse the funnel thoroughly, and collect the
washings in the flask.
57- Quantitative Transfer of Liquid to a Volumetric
Flask - The solute should be completely dissolved before
diluting to the mark. - Insert a funnel into the neck of the volumetric
flask, and use a stirring rod to direct the flow
of liquid from the beaker into the funnel. - With the stirring rod, tip off the last drop of
liquid on the spout of the beaker. - Rinse both the stirring rod and the interior of
the beaker with distilled water and transfer the
washings to the volumetric flask as before.
Repeat the rinsing process.
58- Diluting to the Mark
- After the solute has been transferred, fill the
flask about half full and swirl the contents to
hasten solution. - Add more solvent and again mix well.
- Bring the liquid level almost to the mark, and
allow time for drainage (1 min). Then, use a
medicine dropper to make any necessary final
additions of solvent. - Firmly stopper the flask, and invert it
repeatedly to ensure thorough mixing. - Transfer the contents to a storage bottle that
either is dry or has been thoroughly rinsed with
several small portions of the solution from the
flask.
59- 2H Calibrating volumetric glassware
- Measure the mass of a liquid of known density and
temperature that is contained in (or delivered
by) the volumetric ware. -
- A buoyancy correction must be made since the
density of water is quite different from that of
the masses. - The volume of the apparatus at the temperature of
calibration (T) is obtained by dividing the
density of the liquid at that temperature into
the corrected mass. - Finally, this volume is corrected to the standard
temperature of 20C.
60- General Directions for Calibration
- All volumetric ware should be freed of water
breaks before being calibrated. - Burets and pipets need not be
- dry, but volumetric flasks should
- be thoroughly drained and dried
- at room temperature.
61- Calibrating a Volumetric Pipet
- Determine the empty mass of the stoppered
receiver to the nearest milligram. - Transfer a portion of temperature-equilibrated
water to the receiver with the pipet, weigh the
receiver and its contents (to the nearest
milligram), and calculate the mass of water
delivered from the difference in these masses. - Calculate the volume delivered.
- Repeat the calibration several times, and
calculate the mean volume delivered and its
standard deviation.
62- Calibrating a Buret
- Fill the buret with temperature-equilibrated
water. - Lower the liquid level to bring the bottom of the
meniscus to the 0.00-mL mark. - Touch the tip to the wall of a beaker to remove
any adhering drop. Wait 10 minutes and recheck
the volume. - Weigh (to the nearest milligram) a 125-mL
conical flask fitted with a rubber stopper. - Slowly transfer (at about 10 mL/min)
approximately 10 mL of water to the flask. Touch
the tip to the wall of the flask. - Wait 1 minute, record the volume that was
apparently delivered, and refill the buret.
63- Weigh the flask and its contents to the nearest
milligram. The difference between this mass and
the initial value is the mass of water delivered.
Convert this mass to the true volume. Subtract
the apparent volume from the true volume. - This difference is the correction that should be
applied to the apparent volume to give the true
volume. - Repeat the calibration until agreement within
60.02 mL is achieved. - Calibrating a Volumetric Flask
- Weigh the clean, dry flask to the nearest
milligram. - Then fill to the mark with equilibrated water and
reweigh. - With the aid of the table, calculate the volume
contained.
64- 2I The laboratory notebook
- It is needed to record measurements and
observations concerning an analysis. It should be
permanently bound with consecutively numbered
pages. - Maintaining a Laboratory Notebook
- Record all data and observations directly into
the notebook in ink. - Supply each entry or series of entries with a
heading or label. - Date each page of the notebook as it is used.
- Never attempt to erase or obliterate an incorrect
entry. - Never remove a page from the notebook. Draw
diagonal lines across any page that is to be
disregarded. Provide a brief rationale for
disregarding the page.
65- Notebook Format
- In one convention, data and observations are
recorded on consecutive pages as they occur. - The completed analysis is then summarized on the
next available page spread. - The first of these two facing pages should
contain the following entries - 1. The title of the experiment.
- 2. A brief statement of the principles on which
the analysis is based. - 3. A complete summary of the weighing,
volumetric, and/or instrument response data
needed to calculate the results. - 4. A report of the best value for the set and a
statement of its precision. - The second page should contain the following
items - 1. Equations for the principal reactions in the
analysis. - 2. An equation showing how the results were
calculated. - 3. A summary of observations that appear to bear
on the validity.
66Figure 2-24 Laboratory notebook data page.
67- 2J Safety in the laboratory
- 1. Learn the location of the nearest eye
fountain, fire blanket, shower, and fire
extinguisher. Do not hesitate to use this
equipment if the need arises. - 2. Wear eye protection at all times.
- 3. Most of the chemicals in a laboratory are
toxic. Avoid contact with these liquids. In the
event of such contact, immediately flood the
affected area with large quantities of water. - 4. NEVER perform an unauthorized experiment.
Unauthorized experiments are - grounds for disqualification at many
institutions. - 5. Never work alone in the laboratory. Always be
certain that someone is within earshot. - 6. Never bring food or beverages into the
laboratory. NEVER drink from laboratory
glassware. NEVER smoke in the laboratory.
68-
- Always use a bulb or other device to draw liquids
into a pipet. NEVER pipet by mouth. - 8. Wear adequate foot covering (no sandals).
Confine long hair with a net. A laboratory coat
or apron will provide some protection and may be
required. - 9. Be extremely tentative in touching objects
that have been heated because hot glass looks
exactly like cold glass. - Always fire-polish the ends of freshly cut glass
tubing. NEVER attempt to force glass tubing
through the hole of a stopper. - ?11. Use fume hoods whenever toxic or noxious
gases are likely to be evolved. Be cautious in
testing for odors. - ?12. Notify your instructor immediately in the
event of an injury. - ?13. Dispose off solutions and chemicals as
instructed. It is illegal to flush solutions - containing heavy metal ions or organic liquids
down the drain in most localities.