Title: Gas Tightness Test and Water Content Measurements
1Gas Tightness Test and Water Content Measurements
- Peter Vankov
- LHCb week, CERN, 01.12.2004
2Outline
- Consistency Check of Flow-Meter Measurements
- Leak rate calculation
- Results and analysis
- Water content measurements
3Consistency Check of Flow-Meter Measurements
Standard Method (the one we use to measure leak
rate)
Gas
Module
closed
F1
?
Flow meter
Bubbler, 7mbar
F1nominal (F2F1)-(F3F4)in-out
Consistency check with Flow InOut method
F2
Gas
F4
Module
F1
F3
Bubbler, 7mbar
Flow meter
4M023 F1nominal 0.60 0.00 (F2F1)-(F3F4)i
n-out 0.70 0.18 M020 F1nominal 55.27
0.00 (F2F1)-(F3F4)in-out 54.72 0.16
5Gas tightness modules summary
Module dp/3min(mbar) _at_7mbar Flow(l/s) _at_7mbar
0. 1.90
0.1280E-02 1.
2.
0.50 0.3960E-03
3. 0.00
0.1440E-04 4. 0.10
0.2160E-04 5.
1.20 0.9790E-03 6.
0.60
0.1610E-03 7. 1.70
0.1550E-02 8.
0.00 0.2290E-04 9.
0.00
0.1510E-04 10. 1.35
0.7150E-03 11.
0.00 0.5140E-04 12.
0.80 0.4480E-03
13. 1.60
0.1450E-02 14. 0.70
0.3280E-03 15.
2.30 0.1230E-02 16.
0.20 0.7360E-04
17. 0.30
0.1600E-03 18. 0.75
0.3510E-03 19.
1.90 0.1200E-02 20.
1.70 0.1050E-02
21. 0.00
0.2080E-04 22. 0.60
0.3460E-03 23.
0.00 0.1030E-04 24.
0.00 0.1040E-04
25. 0.00
0.9970E-05 26. 0.45
0.1790E-03 27.
0.45 0.1940E-03 28.
0.00 0.1130E-04
29. 0.00
0.1070E-04 30. 0.00
0.1470E-04 31.
1.30 0.5840E-03
6Leak rate calculation
At equilibrium
Keeping only lowest order terms
CERN
This term accounts for the module bulging under
overpressure
7Leak rate calculation
Assume the module bulging under overpressure
Volume loss and leak-rate (initial overpressure
H0)
For constant Temperature and Pressure
8Leak rate calculation
Assuming no change in temperature and atmospheric
pressure and no module deformation (?0)
(ii),
9Overpressure Decay Measurement Results and
analysis
Module 14
?7.400.13
By fitting eq. (ii) to data
10Module Bulging Model
A, d the nominal surface and thickness
e deflection of each panel per unit
overpressure
In rough agreement with direct measurements !
if ?7.4
e39 µm/mbar
11Average Leak Rate
For small measuring intervals and small leak
rates, a good linear approximation is given by
Length 5 m
12Gas Volume Loss
An overpressure decay ?H -1mbar in ?t3min from
an initial overpressure H07mbar corresponds to a
leak rate
In the experiment, the refresh rate of the gas
volume will be
_at_ 2 mbar overpressure
Scales to
10 loss of the gas volume during a refresh
cycle
_at_ 7 mbar overpressure
13Water Content measurements
Presented by S.Bachmann during PRR
14Water content of the nitrogen from the bottle
Water content ppm
Humidity meter measures the Dew point
1.5 ppm
15M008 gas tight one. M007 not very gas tight
one. Both modules have been flushed with
Nitrogen 2.5 mounts before the test _at_ 8.8 l/h
16000 l in total
16The effect of disconnected tubes
17Summary and Outlook
- The consistency check of the flow rate
measurement is successful. - 10 loss of the gas volume during a refresh cycle
if (dp/dt)1mbar/3min. - Analyse the results from M007 and repeat the
water content measurements with just build, fresh
modules.