Radiation Detection - PowerPoint PPT Presentation

1 / 38
About This Presentation
Title:

Radiation Detection

Description:

Many radiation detectors produce electrical pulses whose amplitudes are ... Barium x-ray photopeak caused by absorption of barium K-shell x-rays (31 to 37 keV) ... – PowerPoint PPT presentation

Number of Views:445
Avg rating:3.0/5.0
Slides: 39
Provided by: michae354
Category:

less

Transcript and Presenter's Notes

Title: Radiation Detection


1
Radiation Detection Measurement II
  • Pulse height spectroscopy

2
Pulse height analyzers
  • Many radiation detectors produce electrical
    pulses whose amplitudes are proportional to the
    energies deposited in the detector by individual
    interactions
  • PHAs are electronic systems that may be used with
    these detectors to perform pulse height
    spectroscopy and energy-selective counting
  • In energy-selective counting, only interactions
    that deposit energies within a certain energy
    range are counted

3
PHAs (cont.)
  • Energy-selective counting can be used to
  • Reduce the effects of background radiation
  • Reduce the effects of scatter
  • Separate events caused by different radionuclides
    in a mixed radionuclide sample
  • Two types of PHAs single-channel analyzers
    (SCAs) and multichannel analyzers (MCAs)
  • Pulse height discrimination circuits incorporated
    in scintillation cameras and other nuclear
    medicine imaging devices to reduce effects of
    scatter

4
Single-channel analyzer systems
  • High-voltage power supply typically provides 800
    to 1,200 volts to the PMT
  • Raising voltage increases magnitude of voltage
    pulses from PMT
  • Preamp connected to PMT using very short cable
  • Amplifies voltage pulses to minimize distortion
    and attenuation of signal during transmission to
    remainder of system

5
(No Transcript)
6
SCA systems (cont.)
  • Amplifier further amplifies the pulses and
    modifies their shapes gain typically adjustable
  • SCA allows user to set two voltage levels, a
    lower level and an upper level
  • If input pulse has voltage within this range,
    output from SCA is a single logic pulse (fixed
    amplitude and duration)
  • Counter counts the logic pulses from the SCA for
    a time interval set by the timer

7
(No Transcript)
8
SCA energy modes
  • LL/UL mode one knob directly sets the lower
    level and the other sets the upper level
  • Window mode one knob (often labeled E) sets the
    midpoint of the range of acceptable pulse heights
    and the other knob (often labeled ?E or window)
    sets a range of voltages around this value.
  • Lower-level voltage is E - ?E/2 and upper-level
    voltage is E ?E/2

9
(No Transcript)
10
Plotting a spectrum using a SCA
  • The SCA is placed in window mode, the E setting
    is set to zero, and a small window (?E) is
    selected
  • A series of counts is taken for a fixed length of
    time per count, with the E setting increased
    before each count but without changing the window
    setting
  • Each count is plotted on graph paper as a
    function of baseline (E) setting

11
Energy calibration of SCA
  • On most SCAs, each of the two knobs permits
    values from 0 to 1,000 to be selected
  • By adjusting the amplification of the pulses
    reaching the SCA either by changing the voltage
    applied to the PMT or by changing the amplifier
    gain the system can be calibrated so that these
    knob settings directly indicate keV
  • A Cs-137 source, which emits 662-keV gamma rays,
    is often used for calibration

12
Multichannel analyzer systems
  • An MCA system permits an energy spectrum to be
    automatically acquired much more quickly and
    easily than does a SCA system
  • The detector, HV power supply, preamp, and
    amplifier are the same as for SCA systems
  • The MCA consists of an analog-to-digital
    converter, a memory containing many storage
    locations called channels, control circuitry, a
    timer, and a display

13
(No Transcript)
14
(No Transcript)
15
(No Transcript)
16
(No Transcript)
17
Interactions of photons with a spectrometer
  • An incident photon can deposit its full energy
    by
  • A photoelectric interaction (A)
  • One or more Compton scatters followed by a
    photoelectric interaction (B)
  • A photon will deposit only a fraction of its
    energy if it interacts by Compton scattering and
    the scattered photon escapes the detector (C)
  • Energy deposited depends on scattering angle,
    with larger angle scatters depositing larger
    energies

18
(No Transcript)
19
Interactions (cont.)
  • Even if the incident photon interacts by the
    photoelectric effect, less than its total energy
    will be deposited if the inner-shell electron
    vacancy created by the interaction results in
    emission of a characteristic x-ray that escapes
    the detector (D)

20
(No Transcript)
21
Interactions (cont.)
  • Detectors normally shielded to reduce effects of
    natural background radiation and nearby radiation
    sources
  • An x-ray or gamma-ray may interact in the shield
    of the detector and deposit energy in the
    detector
  • Compton scatter in the shield, with the scattered
    photon striking the detector (E)
  • A characteristic x-ray from the shield may
    interact with the detector (F)

22
Spectrum of Cesium-137
  • Cs-137 decays by beta particle emission to
    Ba-137m, leaving the Ba-137m nucleus in an
    excited state
  • The Ba-137m nucleus attains its ground state by
    the emission of a 662-keV gamma ray 90 of the
    time
  • In 10 of decays, a conversion electron is
    emitted instead, followed by a 32-keV K-shell
    characteristic x-ray

23
(No Transcript)
24
Reasons for differences in spectra
  • First, there are a number of mechanisms by which
    an x-ray or gamma-ray can deposit energy in the
    detector, several of which deposit only a
    fraction of the incident photon energy
  • Second, there are random variations in the
    processes by which the energy deposited in the
    detector is converted into an electrical signal

25
NaI(Tl) crystal/PMT
  • Random variations in
  • The fraction of deposited energy converted into
    light
  • The fraction of the light that reaches the
    photocathode of the PMT
  • The number of electrons ejected from the back of
    the photocathode per unit energy deposited by the
    light
  • Cause random variations in the size of the
    voltage pulses produced by the detector, even
    when the incident x-rays or gamma rays deposit
    exactly the same energy

26
(No Transcript)
27
Pulse height spectrum of Cs-137
  • Photopeak corresponding to interactions in which
    the energy of an incident 662-keV photon is
    entirely absorbed in the crystal
  • Compton continuum caused by 662-keV photons that
    scatter in the crystal, with the scattered photon
    escaping the crystal
  • The Compton edge is the upper limit of the
    Compton continuum

28
(No Transcript)
29
Pulse height spectrum (cont.)
  • Backscatter peak caused by 662-keV photons that
    scatter from the shielding around the detector
    into the detector
  • Barium x-ray photopeak caused by absorption of
    barium K-shell x-rays (31 to 37 keV)
  • Photopeak caused by lead K-shell x-rays (72 to 88
    keV) from the shield

30
Spectrum of Technetium-99m
  • Tc-99m is an isomer of Tc-99 that decays by
    isomeric transition to its ground state, with the
    emission of a 140.5-keV gamma ray
  • In 11 of the transitions, a conversion electron
    is emitted instead of a gamma ray

31
(No Transcript)
32
Tc-99m (cont.)
  • Photopeak caused by total absorption of the
    140-keV gamma rays
  • Escape peak caused by 140-keV gamma rays that
    interact with the crystal by photoelectric effect
    but with resultant iodine K-shell x-rays (28 to
    33 keV) escaping the crystal
  • Photopeak caused by absorption of lead K-shell
    x-rays from the shield
  • Compton continuum is quite small because the
    photoelectric effect predominates in iodine at
    140 keV

33
(No Transcript)
34
Energy resolution
  • Energy resolution of a spectrometer is a measure
    of its ability to differentiate between particles
    or photons of different energies
  • Determined by irradiating detector with
    monoenergetic particles or photons and measuring
    width of resulting peak in the pulse height
    spectrum
  • Statistical effects in the detection process
    cause the amplitudes of pulses from detector to
    randomly vary about the mean pulse height, giving
    the peak a Gaussian shape

35
Energy resolution (cont.)
  • Width is usually measured at half the maximal
    height of the peak called the full width at
    half-maximum (FWHM)

36
(No Transcript)
37
Count-rate effects
  • If two interactions occur in a detector,
    separated by a very short time interval, the
    detector produces a single pulse
  • Sum of the individual signals from the two
    interactions
  • Higher amplitude than the signal from either
    individual interaction
  • Operating a pulse height spectrometer at a high
    count rate causes loss of counts and misplacement
    of counts in the spectrum

38
(No Transcript)
Write a Comment
User Comments (0)
About PowerShow.com