Title: Introduction to Microscopy
1Introduction to Microscopy
2What is a microscope? How does it work?
- A microscope is an instrument designed to make
fine details visible. - The microscope must accomplish three tasks
- produce a magnified image of the specimen
(magnification), - separate the details in the image (resolution),
and - render the details visible to the eye, camera, or
other imaging device (contrast).
3The Human Eye
- Our eyes are capable of distinguishing color in
the visible portion of the spectrum from violet
to blue to green to yellow to orange to red the
eye cannot perceive ultra-violet or infra-red
rays. - The eye also is able to sense differences in
brightness or intensity ranging from black to
white and all the grays in-between.
4Light
- What is light and how does it interact with
matter? - Light has both particle and wave natures.
- For microscopists it is the wave theory that is
applied in most cases. - Wavelength of light is perceived as colour and
the amplitude is seen as brightness.
5Light
- Light travels in straight lines
- Its path can be deflected or reflected by means
of mirrors or right angle prisms. - Light can be bent or refracted by means of
glass lenses that are thicker or thinner at their
center or their periphery. - Light travels at different speeds in air and in
glass (faster in air which is usually taken as
the standard of 1). - Light is slowed and bent or refracted when it
passes through air and enters a lens.
6Lenses
- The two most common types of lenses are concave
and convex. A common bi-convex lens is considered
a positive lens because it causes light rays to
converge or concentrate forming a real image.
7Focal point and image formation
- Every convex lens has two focal points one on
each side of the lens. If an object is placed on
the far side of the front focal point (F), an
inverted image is projected on a screen placed on
the opposite side of the lens. The inverted image
is known as the real image. - If the object is brought to a point between the
focal point and the lens, it appears enlarged and
is transposed to the same side of lens as the
object. The enlarged image is known as the
virtual image.
8Focal point and image formation
9SIMPLE MICROSCOPE
- More than five hundred years ago, simple glass
magnifiers were developed. - These were convex lenses (thicker in the center
than the periphery). The specimen or object could
be focused by use of the magnifier placed between
the object and the eye. - These simple microscopes, along with the cornea
and eye lens, could spread the image on the
retina by magnification through increasing the
visual angle on the retina.
10SIMPLE MICROSCOPE
11Compound Microscope
- In the compound microscope we find the object
that is magnified by the object (forming the real
image) is further magnified by the ocular lens to
form the virtual image. - When you look into a microscope, you are not
looking at the specimen, you are looking at an
IMAGE of the specimen. - The image is floating in space about 10 mm
below the top of the observation tube (at the
level of the fixed diaphragm of the eyepiece)
where the eyepiece is inserted.
12COMPOUND MICROSCOPE
- The compound microscope achieves a two-stage
magnification. The objective projects a magnified
image into the body tube of the microscope and
the eyepiece further magnifies the image
projected by the objective (more of how this is
done later). - For example, the total visual magnification using
a 10X objective and a 15X eyepiece is 150X.
13Types of Microscopes
14Optical Components
- The most important optical component of the
microscope is the OBJECTIVE. - 1. Its basic function is to gather the light
passing through the specimen and then to project
an accurate, real, inverted IMAGE of the specimen
up into the body of the microscope. - 2. Other related functions of the objective are
to house special devices such as an iris for
darkfield microscopy, a correction collar for
counteracting spherical aberration or a phase
plate for phase contrast microscopy. - 3. The higher power objectives should have a
retractable front lens housing to protect the
front lens where the objective requires focusing
very close to the specimen. - 4. To the extent possible, corrections for lens
errors (aberrations) should be made within the
objective itself.
15Examples of Objective Lens
16Optical Components
- A second important optical component is the
EYEPIECE or Ocular Lens. - 1. Its basic function is to look at the
focused, magnified real image projected by the
objective and magnify that image a second time as
a virtual image seen as if 10 inches from the
eye. - 2. The eyepiece can be fitted with scales or
markers or pointers or crosshairs that will be in
simultaneous focus with the focused image.
17Optical Components
- The third important optical component is the
SUBSTAGE CONDENSER. - 1. Its basic function is to gather the light
coming from the light source and to concentrate
that light in a collection of parallel beams onto
the specimen. - 2. The light gathered by the condenser comes to a
focus at the back focal plane of the objective. - 3. In appropriately set up illumination the image
of the light source comes to focus at the level
of the front focal plane of the condenser. - 4. Correction for lens errors are incorporated in
the finest condensers, an important feature for
research and photography. - 5. Where desired, the condenser can be designed
to house special accessories for phase contrast
or differential interference or darkfield
microscopy
18Examples of Condensers
19Mechanical/Electrical Components
- The STAND of the microscope houses the
mechanical/electrical parts of the microscope. It
provides a sturdy, vibration-resistant base for
the various attachments. - The BASE of the microscope is Y-shaped or
U-shaped for greater stability. It houses the
electrical components for operating and
controlling the intensity of the lamp. The lamp
may be placed, depending on the instrument, at
the lower rear of the stand or directly under the
condenser fitting. The base also houses the
variable field diaphragm.
20Mechanical/Electrical Components
- Built into the stand is a fitting to receive the
microscope STAGE. The stage has an opening for
passing the light. The specimen is placed on top
of the stage and held in place by a specimen
holder. Attached to the stage are concentric X-Y
control knobs which move the specimen forward
/back or left/right. - On the lower right and left sides of the stand
are the concentric COARSE and FINE FOCUSING
KNOBS. These raise or lower the stage in larger /
smaller increments to bring the specimen into
focus.
21Mechanical/Electrical Components
- 4. Under the stage there is a built-in ring or a
U-shaped CONDENSER HOLDER. Adjacent to the
condenser holder there are either one or two
knobs for raising or lowering the condenser. - 5. Above the stage, the stand has a NOSEPIECE
(may be fixed or removable) for holding the
objectives of various magnifications. The
rotation of the nosepiece can bring any one of
the attached objectives into the light path
(optical axis).
22Using the microscope
23A method for setting up the microscope correctly
- Köhler illumination was first introduced in 1893
by August Köhler of the Carl Zeiss corporation as
a method of providing the optimum specimen
illumination.
24Setting Up Köhler illumination
- After switching on the lamp of the microscope,
open up fully both the field diaphragm (in the
light port of the microscope) and the
iris/aperture diaphragm (usually built into the
substage condenser). Check you have light by
insert a piece of paper over the light.
25Setting Up Köhler illumination
- Then set the interpupillary distance via the
folding bridge of the binocular tube. The correct
setting is reached when you see one light circle
instead of two. - Rotate the nosepiece to bring the 10X objective
into the light path. - Place the specimen on the microscope stage and
focus the specimen using the coarse and fine
focusing knobs.
26Setting Up Köhler illumination
- Close down the field diaphragm most of the way.
Now raise the substage condenser (using the
condenser focusing knob) and focus the image of
the field diaphragm sharply onto the
already-focused specimen. - This image of the field diaphragm should appear
as a focused polygon.
27Setting Up Köhler illumination
- If the image of the field diaphragm is not
centered in the field of view, use the condenser
centering screws (or knobs) to center the image
of the field diaphragm. - Then open up the field diaphragm until it just
disappears from view.
28Setting Up Köhler illumination
- Centering the image of the field diaphragm
29Setting Up Köhler illumination
- As a rule of thumb adjust the iris/apeture
diaphragm so that it is 2/3 to 3/4 open. This
setting usually represents the best compromise
between resolution and contrast.
30Setting Up Köhler illumination
- You have now set up Köhler illumination with the
10X objective. If you wish to switch to a higher
power objective, you must again adjust BOTH the
field and the iris/aperture diaphragms. - For example, if you switch to the 40X objective,
you will have to close the field diaphragm
somewhat and may have to recenter it (looking at
a smaller area of the specimen). - You will also have to open up the iris diaphragm
somewhat (the 40X objective has a higher
numerical aperture - light-grasping ability -
than does the 10X objective). - Thus, every time you change objectives, you must
adjust both diaphragms in accordance with the
steps given above.
31Parfocal
- The parfocal distance, that is, the distance from
the objective shoulder to the specimen, is set at
45mm for all objectives. - This means that focusing can be performed quickly
and easily, with a minimum of fine adjustment,
even when switching from a low to a high
magnification objective.
32Resolution
- Related to the wavelength of light used and to
the numerical aperture of the objective lens. - R wavelength of light (?)
- 2NA
33Resolution
- Numerical aperture is determined by the
refractive index of the medium between the
specimen and the objective and the sine of the
angle a. - In the case of dry lenses, n 1 (air), and for
oil immersion lenses it has a value depending on
the refractive index of the oil used, usually
1.515.
34Resolution
- Resolution increases with increasing numerical
aperture, or decreasing wavelength of light. - Limit to resolution for an optical microscope is
0.2µm or 0.25µm.
35Care and Cleaning of your Microscope
- Parts that require cleaning include
- Oculars and condenser
- Objective lenses
- Light source lens and filters located at the base
of the microscope - Microscope body including the stage and stage
clips
36Care and Cleaning of your Microscope
- Cleaning Supplies
- Canned air or duster
- Lint-free tissue
- Lens tissue and/or cotton-tipped sticks stored in
a dust-free container - Microscope Cleaning Solution (MCS)
37Suggested cleaning fluids (MCS)
- Distilled water
- Commercial photographic cleaning solutions
- Windex or sparkle
- Ethanol
- Methlylated spirits
- Petroleum ether 85 , Isopropanol 15
- Histoclear/Histolene
38Cleaning the Oculars and Condenser
- Remove slide from the stage blow off dust with
canned air or air blower. - Dampen a new cotton-tipped stick or lens tissue
with MCS. - Start in the center of the ocular or condenser,
and spiral to the outside edge while rotating the
cotton tip. - Repeat with a new, dry cotton-tipped stick.
- Repeat the above until the view is clear.
39Cleaning the Oculars and Condenser
- Use only new cotton-tipped sticks in a rotating
fashion when touching the lens surface to avoid
scratching the lens with any debris that is being
removed. - The oculars and the condenser should be cleaned
by starting at the center and spiraling outward
while rotating the cotton tip. Be sure to only
use new clean cotton-tipped sticks to touch the
lens surface to avoid scratching the lens.
40Objective Care
- Remove/install objectives using both hands.
Loosely cup with one hand and twist the barrel
with the other, being very careful not to touch
the front lens with your fingers. Take extreme
caution not to drop the objective! - Never apply strong physical force to an
objective. To move another objective into
position, move the rotating turret do not grab
the objective and pull on it. - To remove a stuck objective, never use vice-grips
or a pipe wrench, which can severely damage the
optics. Attempt to loosen the threads by applying
a few drops of water to dissolve salts, or an
oil-dissolving agent if immersion oil is the
culprit.
41Cleaning the Objectives
- Remove slide from the stage.
- To clean, remove the objective from the turret.
- Dampen a new cotton-tipped stick or lens tissue
(folded into fourths) with MCS. - Hold the cotton-tipped stick at a 45 degree angle
on the objective lens and twirl. - Repeat with a clean dry cotton-tipped stick.
- Repeat the above until the view is clear.
42Cleaning the Objectives
- The 100x objective may need to be cleaned during
a microscope session as well as after the session
if blurry optics are noticed when looking at an
object on oil. - Oil objectives should be cleaned after each
microscope session. Dampen a cotton-tipped stick
with MCS. Cleaning should be done at a 45 degree
angle to the objective. Be sure to only use new
cotton-tipped sticks to touch the objective to
avoid scratching.
43Cleaning Objectives
- Examine the lens carefully by removing the
microscopes eyepiece, looking through it
backwards, and holding it up to the edge of the
objective, to see a magnified image of the lens. - Angle the objective so that the room light is
brightly illuminating the lens surface it should
appear spotlessly clean. If not, repeat the above
procedure. - This is also a good way to examine a lens closely
for scratches or other imperfections.
44Cleaning Stage, Body, Light Source Lens and
Filters
- Remove slide from the stage.
- Use canned air or blower to remove dust from the
light source lens, microscope stage and body. - Use lens paper dampened with MCS to wipe off the
surface of the light source lens. - Use lint-free tissue dampened with MCS, to wipe
off the stage and stage clip remove the stage
clip if necessary. - Repeat as needed, making sure to use only new
lens paper to touch the lens surface to avoid
scratching. - Wipe down all surfaces of the microscope body
with lint-free tissue dampened with MCS.
45Cleaning Stage, Body, Light Source Lens and
Filters
- Use dampened paper towels to clean off the work
bench area around the microscope after completing
each day of work. Be sure to use your microscope
dust cover or clean plastic bag when the
equipment is not in use. - The light source lens, microscope stage and body
should be cleaned every day of use. Use canned
air to blow dust off the light source lens and
microscope body. - Use lens paper dampened with MCS to wipe off the
surface of the light source lens. - Use only new lens paper to touch the lens surface
to avoid scratching. - Wipe off the stage and stage clip.
- Cover the microscope when the equipment is not in
use.