Title: Energy Policy Act 2005: Residential Tax Credits
1RESNETNEW REQUIREMENTSJanuary 1, 2007
RESNET STANDARD 2004 03 BLOWER DOOR TEST
PROCEDURES
2This Presentation Jointly Prepared by
3New Requirements
- Starting Jan 1, 2007, HERS Providers are required
to perform an annual field calibration check on
Blower Door pressure testing instruments and
fans. - Providers must use standards and procedures
recommended by the equipment manufacturer. - Providers must maintain a written log of the
annual calibration check for a period of 3 years. - Similar requirements for all associated pressure
testing instruments.
4Appendix A Building Element Air Leakage
Use the testing protocol described in ASHRAE
Standard 119 Section 5.1. Blower door and
associated pressure testing instruments, which
include but are not limited to hoses, and
Manometers, gauges, and fans shall be FIELD
TESTED annually for calibration by the HERS
provider or rater. The provider shall use a
standard for field testing of calibration
provided by the equipment manufacturer.
Magnehelic Gauges cannot be field tested and
shall be recalibrated by the Blower Door
manufacturer annually. Field check the fan and
flow measuring systems for defects and maintain
them according to manufacturers recommendations.
The HERS provider shall maintain a written log of
the annual calibration check to verify all
equipment accuracy for a period of three (3)
years. These records shall be made available
within 24 hours to a RESNET Quality Assurance
Committee member upon request. It is recommended
all pressure equipment be field checked for
calibration more frequently than is required in
these standards, i.e., monthly, quarterly, etc.
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxx
Determine effective leakage area from a blower
door test. xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
Blower door test.xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
Rated Feature Task
On-Site Inspection Protocol
5Just having a perfect gaugeby no means
ensuresreadings will be accurate
Colin Genge, Retrotec
6Field TestingMay NOTRequire Calibration
7Common Reasons forInaccurate Readings
- Using inappropriate flow ring on fan. (newer
gauges help eliminate this mistake). - Tubing leaking, kinked or plugged up.
- Tubing hooked up improperly.
- Damaged fans (e.g. broken flow sensor, housing).
- Gauges out of calibration.Note If both channels
have an identical error, the effect on the the
result is significantly reduced. - Incorrect test procedures.
8The Energy Conservatory (TEC) Recommended
Calibration Check Procedures
9TEC Recommended Calibration Check Procedures
- Digital Pressure Gauges
- (choose either method)
- Perform the digital gauge field calibration check
procedure found at www.energyconservatory.com
/support/support7.htm - or
- Recalibrate gauge at TEC factory - 50.
10Digital Gauge Field Calibration Check
- Field check procedure consists of comparing
digital gauges against a DG-700 gauge that is
being used as an in-house pressure standard. - This procedure involves comparing pressure
readings from the DG-700 Standard gauge with the
gauge being tested (DG-700 or DG-3) at a number
of different pressures.
11Digital Gauge Field Calibration Check
- Readings are recorded and the difference between
the readings is determined. - If the difference is smaller than the maximum
allowed, gauge passes the field calibration
check. - If difference is larger than the maximum allowed,
gauge should be sent to TEC for calibration.
12Digital Gauge Field Calibration Check
- Ideally, the DG-700 Standard should be
recalibrated once every 6 months and only used as
a pressure standard (not used in field). - Otherwise, using a DG-700 that has just been
returned from TEC calibration is acceptable. - Pressures are created by syringe for field check
(TEC has them available). Syringe is placed on
table and pressures are allowed to stabilize
before readings are made. - Magnehelic gauge is used only to prevent
over-pressurization.
13TEC Factory Digital Gauge Calibration
- Gauge is tested against a NIST traceable pressure
standard at 20 different pressure values. - Gauge is recalibrated and tested again.
- Calibration certificate is provided.
- 50 cost includes ground UPS shipping back to
customer. - Turn around time at TEC is typically 3-4 days
(assuming no repairs needed).
14TEC Factory Digital Gauge Calibration
- Digital gauge calibration certificate looks like
this.
15TEC Recommended Calibration Check Procedures
- Blower Door and Duct Blaster Fans
- (choose either method)
- Perform the fan field check procedure found at
www.energyconservatory.com/support/support7.htm - or
- Recalibrate fan at TEC factory - 200.
-
16Fan Field Check Procedure
- The 2 most important aspects to maintaining
proper fan calibration (Blower Door and Duct
Blaster) - No leaks in flow sensor.
- Sensor in proper position.
- Damaged fan housing (e.g. broken flange) and
broken/ missing blades can also effect
calibration and should be repaired.
17Fan Field Check Procedure
- Simple procedures and forms for checking flow
sensor position and flow sensor integrity are
contained in the fan field check document. - If fan meets specifications listed in the
procedure, the fan passes the field check. - If fan does not meet specifications, contact TEC
for adjustment/repair instructions.
18TECFactory Fan Calibration
- Fan is calibrated using ASTM Standard E1258 with
a test chamber constructed according to ASHRAE
51/AMCA 210. - Calibration certificate is provided including
custom calibration parameters. - 200 cost (shipping is extra).
- Turn around time at TEC is typically 4-5 days
(assuming no repairs needed).
19TECFactory Fan Calibration
- Fan calibration certificate looks like this.
- Custom fan calibration parameters can be input
into TEC software (TECTITE or TECBLAST) for
greater test accuracy.
20TEC Recommended Calibration Check Procedures
- Duct Blaster System Field Check
- (meets requirements for both the Fan and Gauge)
- Perform the Duct Blaster System field check
procedure using an approved orifice plate. - This procedure will be available on the TEC
website in the Spring 07. Approved orifice plates
will also be available at that time. -
21Duct Blaster System Field Check Procedure
- Approved orifice plate is attached to the end of
the square transition piece. - Air is drawn through orifice plate and into the
flex duct. - Fan is adjusted to specified test pressure.
- Measured flow (or hole size) from digital gauge
is compared to standard listed in the procedure. - Pass/fail criteria listed in the procedure.
22TEC Recommended Calibration Check Procedures
- Magnehelic Gauges
- TEC will perform an accuracy check on Magnehelic
gauges for 15 per gauge (at the TEC factory). - The accuracy of the gauges will be tested and
recorded, however, calibration adjustments on
Magnehelic gauges will no longer be performed.
23TEC Recommended Calibration Check Procedures
- Check Tubing for Leaks or Obstructions
- Tubing should be periodically checked for leaks
and obstructions. - Periodically trim off ¼ off the ends of tubing
to remove damaged ends. - If tubing leaks or is obstructed, replace it
immediately.
24RETROTEC Recommended Calibration Check
Procedures
25Retrotec Duc-Tester Calibration Verification
Procedure
This procedure should be used whenever a new
Duc-Tester is received to verify both the system
and the operator. The Verification Plate, which
has a leakage rate of 100 cfm at 25 Pa, is
attached to the Duc-Tester to simulate a duct
system. Readings between 96 and 108 cfm indicate
no significant problems with the test method or
the equipment. If readings fall outside this
range, use the Duc-Tester Problem/Solution
Checklist to fix the problem. Â Repeating this 5
minute test every month or after 20 tests will
detect problems early. No further calibration of
the system is required unless the Duc-Tester
fails this Verification.
26Problems causing inaccurate readings in order of
frequency are
27Retrotec Duc-Tester Calibration Verification
Procedure        Measure the flow rate
at 25 Pa. Â Â Â Flow readings outside of 97 to
106 CFM indicate a problem.
Tape on the Plate (calibrated for 100 cfm at 25
Pa).
28Detailed Retrotec Duc-Tester Calibration
Verification Procedure
29Depressurization is the same as Pressurization
except
- Turn on the DM-2. Actuate the MODE key until
Flow _at_ 25 Pa appears. Actuate the _at_key
until Flow _at_ 25 Pa appears. If units are not
cfm _at_ 25 Pa, adjust using the SETUP key. - Place the Duc-Tester on the floor with flex-duct
attached to the outlet side of the fan such that
the duct is pressurized. - Tape the Retrotec Duc-Tester Calibration
Verification Plate to the flex duct. - Connect the long red tube to the Verification
Plate. Connect the other end to the Blue port on
the Retrotec DM-2 - Connect a yellow tube between the FAN connection
on the Duc-Tester and the yellow port on the
DM-2. - Connect a green tube between one port on the FAN
and the green connector on the DM-2. - Connect speed control and power cords to the FAN.
- Adjust the speed until PrA reads close to 25 Pa.
Allow the flex duct to straighten out to its full
length. - Set Time averaging to 8 seconds.
- Flow should read between 92 and 108 cfm.
- Mark results down so your system can be tracked
with time.
30Â Â Â Â
31Duc-Tester Problem/Solution Checklist Use this
List if Flow is NOT between 97 and 106 cfm for
pressurization or 92 and 108 cfm for
depressurization.Note even though the accuracy
is within 3 of measured flow, this is a
calibration check to see if something is wrong.
32Hole
Water drop
Mid
Clamp
Mid
33(No Transcript)
34Water drop
35 Clamp
36Hole
37Now, move the tubes to the Green and Blue ports
of the DM-2. Compare Channel A B at maximum
pressure and 50 Pa.
38The above Solutions will fix most problems. If
your Duc-Tester still fails the Verification
Test, request an exchange or loaner from your
distributor while yours is sent back to the
factory.
39Retrotec Blower-Door Calibration Verification
Procedure  Blower Doors can be tested in a
similar fashion using larger calibration plates
like the one shown. Â Measure
the background leakage. Measure the increased
leakage with the hole open. The Blower Door
should be able to measure the difference within
15 otherwise something may be wrong. Follow a
procedure similar to the one outlined in the
Duc-Tester Problem/Solution Checklist
Install your Aluminum frame in a small, tight
room with the vents and exhaust fans taped shut.
Install this panel in the upper hole.
Alternatively, cut a piece of cardboard with a
one foot by 3 foot hole in it, centered in above
the fan.
40HERS Provider ShallMaintain a WrittenLog of
Annual Calibration. . . . . . and Shall Be
MadeAvailable Within 24 Hours
41GWS Manometer Calibration Checklist
EXAMPLE
You can make your own
42QUESTIONS ?