Atomic Spectroscopy - PowerPoint PPT Presentation

1 / 167
About This Presentation
Title:

Atomic Spectroscopy

Description:

Atomic Spectroscopy Introduction An Introduction to Optical Atomic Spectroscopy In optical atomic spectrometry, compounds are first converted to gaseous molecules ... – PowerPoint PPT presentation

Number of Views:410
Avg rating:3.0/5.0
Slides: 168
Provided by: monzirpal
Category:

less

Transcript and Presenter's Notes

Title: Atomic Spectroscopy


1
Atomic Spectroscopy
  • Introduction

2
Technique Flame Test
3
An Introduction to Optical Atomic Spectroscopy
  • In optical atomic spectrometry, compounds are
    first converted to gaseous molecules followed by
    conversion to gaseous atoms. This process is
    called atomization and is a prerequisite for
    performing atomic spectroscopy. Gaseous atoms
    then absorb energy from a beam of radiation or
    simply heat. Absorbance can be measured or
    emission from excited atoms is measured and is
    related to concentration of analyte.

4
Atomic Energy Level Diagrams
  • As a start, we should be aware that only valence
    electrons are responsible for atomic spectra
    observed in a process of absorption or emission
    of radiation in the UV-Vis region. Valence
    electrons in their ground states are assumed to
    have an energy equal to zero eV. As an electron
    is excited to a higher energy level, it will
    absorb energy exactly equal to the energy
    difference between the two states. Let us look at
    a portion of the sodium energy level diagram
    where sodium got one electron in the 3s orbital

5
(No Transcript)
6
(No Transcript)
7
  • The dark lines represent most probable
    transitions and in an atomic spectrum they would
    appear more intense than others. It should also
    be indicated that two transitions, of very
    comparable energies (589.0 and 589.6 nm), from
    the 3s ground state to 3p excited state do take
    place. This suggests splitting of the p orbital
    into two levels that slightly differ in energy.
    Explanation of this splitting may be presented as
    a result of electron spin where the electron spin
    is either in the direction of the orbital motion
    or opposed to it.

8
  • Both spin and orbital motion create magnetic
    fields that may interact in an attractive manner
    (if motion is in opposite direction, lower
    energy), or in a repulsive manner when both spin
    and orbital motion are in the same direction
    (higher energy). The same occurs for both d and f
    orbitals but the energy difference is so small to
    be observed. A Mg ion would show very similar
    atomic spectrum as Na since both have one
    electron in the 3s orbital.

9
  • In cases where atoms of large numbers of
    electrons are studied, atomic spectra become too
    complicated and difficult to interpret. This is
    mainly due to presence of a large numbers of
    closely spaced energy levels
  • It should also be indicated that transition from
    ground state to excited state is not arbitrary
    and unlimited. Transitions follow certain
    selection rules that make a specific transition
    allowed or forbidden.

10
Atomic Emission and Absorption Spectra
  • At room temperature, essentially all atoms are in
    the ground state. Excitation of electrons in
    ground state atoms requires an input of
    sufficient energy to transfer the electron to one
    of the excited state through an allowed
    transition. Excited electrons will only spend a
    short time in the excited state (shorter than a
    ms) where upon relaxation an excited electron
    will emit a photon and return to the ground
    state.

11
  • Each type of atoms would have certain preferred
    or most probable transitions (sodium has the
    589.0 and the 589.6 nm). Relaxation would result
    in very intense lines for these preferred
    transitions where these lines are called
    resonance lines.
  • Absorption of energy is most probable for the
    resonance lines of each element. Thus intense
    absorption lines for sodium will be observed at
    589.0 and 589.6 nm.

12
Atomic Fluorescence Spectra
  • When gaseous atoms at high temperatures are
    irradiated with a monochromatic beam of radiation
    of enough energy to cause electronic excitation,
    emission takes place in all directions. The
    emitted radiation from the first excited
    electronic level, collected at 90o to the
    incident beam, is called resonance fluorescence.
    Photons of the same wavelength as the incident
    beam are emitted in resonance fluorescence. This
    topic will not be further explained in this text
    as the merits of the technique are not very clear
    compared to instrumental complexity involved

13
Atomic Line Width
  • It is taken for granted that an atomic line
    should have infinitesimally small (or zero) line
    width since transition between two quantum states
    requires an exact amount of energy. However,
    careful examination of atomic lines reveals that
    they have finite width. For example, try to look
    at the situation where we expand the x-axis
    (wavelength axis) of the following line

14
(No Transcript)
15
  • The effective line width in terms of wavelength
    units is equal to Dl1/2 and is defined as the
    width of the line, in wavelength units, measured
    at one half maximum signal (P). The question
    which needs a definite answer is what causes the
    atomic line to become broad?

16
Reasons for Atomic Line Broadening
  • There are four reasons for broadening observed in
    atomic lines. These include
  • 1. The Uncertainty Principle
  • We have seen earlier that Heisenberg uncertainty
    principle suggests that nature places limits on
    the precision by which two interrelated physical
    quantities can be measured. It is not easy, will
    have some uncertainty, to calculate the energy
    required for a transition when the lifetime of
    the excited state is short.

17
  • The ground state lifetime is long but the
    lifetime of the excited state is very short which
    suggests that there is an uncertainty in the
    calculation of the transition time. We have seen
    earlier that when we are to estimate the energy
    of a transition and thus the wavelength (line
    width), it is required that the two states where
    a transition takes place should have infinite
    lifetimes for the uncertainty in energy (or
    wavelength) to be zero

18
  • DEgth/Dt
  • Therefore, atomic lines should have some
    broadening due to uncertainty in the lifetime of
    the excited state. The broadening resulting from
    the uncertainty principle is referred to as
    natural line width and is unavoidable.

19
2. Doppler Broadening
  • The wavelength of radiation emitted by a fast
    moving atom toward a transducer will be different
    from that emitted by a fast atom moving away from
    a transducer. More wave crests and thus higher
    frequency will be measured for atoms moving
    towards the transducer. The same occurs for sound
    waves

20
  • Assume your ear is the transducer, when a car
    blows its horn toward your ear each successive
    wave crest is emitted from a closer distance to
    your ear since the car is moving towards you.
    Thus a high frequency will be detected. On the
    other hand, when the car passes you and blows its
    horn, each wave crest is emitted at a distance
    successively far away from you and your ear will
    definitely sense a lower frequency.

21
(No Transcript)
22
  • The line width (Dl) due to Doppler broadening can
    be calculated from the relation
  • Dl/lo v/c
  • Where lo is the wavelength at maximum power and
    is equal to (l1 l2)/2, v is the velocity of the
    moving atom and c is the speed of light. It is
    noteworthy to indicate that an atom moving
    perpendicular to the transducer will always have
    a lo, i.e. will keep its original frequency and
    will not add to line broadening by the Doppler
    effect.

23
  • In the case of absorption lines, you may
    visualize the line broadening due to Doppler
    effect since fast atoms moving towards the source
    will experience more wave crests and thus will
    absorb higher frequencies. On the other hand, an
    atom moving away from the source will experience
    less wave crests and will thus absorb a lower
    frequency. The maximum Doppler shifts are
    observed for atoms of highest velocities moving
    in either direction toward or away from a
    transducer (emission) or a source (absorption).

24
(No Transcript)
25
3. Pressure Broadening
  • Line broadening caused by collisions of emitting
    or absorbing atoms with other atoms, ions, or
    other species in the gaseous matrix is called
    pressure or collisional broadening. These
    collisions result in small changes in ground
    state energy levels and thus the energy required
    for transition to excited states will be
    different and dependent on the ground state
    energy level distribution.

26
  • This will definitely result in important line
    broadening. This phenomenon is most astonishing
    for xenon where a xenon arc lamp at a high
    pressure produces a continuum from 200 to 1100 nm
    instead of a line spectrum for atomic xenon. A
    high pressure mercury lamp also produces a
    continuum output. Both Doppler and pressure
    contribution to line broadening in atomic
    spectroscopy are far more important than
    broadening due to uncertainty principle.

27
4. Magnetic Effects
  • Splitting of the degenerate energy levels does
    take place for gaseous atoms in presence of a
    magnetic field. The complicated magnetic fields
    exerted by electrons in the matrix atoms and
    other species will affect the energy levels of
    analyte atoms. The simplest situation is one
    where an energy level will be split into three
    levels, one of the same quantum energy and one of
    higher quantum energy, while the third assumes a
    lower quantum energy state. A continuum of
    magnetic fields exists due to complex matrix
    components, and movement of species, thus exist.
    Electronic transitions from the thus split levels
    will result in line broadening

28
The Effect of Temperature on Atomic Spectra
  • Atomic spectroscopic methods require the
    conversion of atoms to the gaseous state. This
    requires the use of high temperatures (in the
    range from 2000-6000 oC). Thee high temperature
    can be provided through a flame, electrical
    heating, an arc or a plasma source. It is
    essential that the temperature be of enough value
    to convert atoms of the different elements to
    gaseous atoms and, in some cases, provide energy
    required for excitation. The temperature of a
    source should remain constant throughout the
    analysis especially in atomic emission
    spectroscopy.

29
  • Quantitative assessment of the effect of
    temperature on the number of atoms in the excited
    state can be derived from Boltzmann equation
  • Where Nj is the number of atoms in excited state,
    No is the number of atoms in the ground state, Pj
    and Po are constants determined by the number of
    states having equal energy at each quantum level,
    Ej is the energy difference between excited and
    ground states, K is the Boltzmann constant, and T
    is the absolute temperature.

30
Boltzmann distribution
Nj /N0 at 3000 K
Wavelength
Atom
852.1 nm
Cs
7.24 ? 10-3
589.0 nm
Na
5.88 ? 10-4
422.7 nm
Ca
3.69 ? 10-5
213.9 nm
Zn
5.58 ? 10-10
31
  • To understand the application of this equation
    let us consider the situation of sodium atoms in
    the 3s state (Po 2) when excited to the 3p
    excited state (Pj 6) at two different
    temperatures 2500 and 2510K. Now let us apply the
    equation to calculate the relative number of
    atoms in the ground and excited states
  • Usually we use the average of the emission lines
    from the 3p to 3s where we have two lines at
    589.0 and 589.6 nm which is

32
(No Transcript)
33
Therefore, at higher temperatures, the number of
atoms in the excited state increases. Let us
calculate the percent increase in the number of
atoms in the excited state as a result of this
increase in temperature of only 10 oC
34
(No Transcript)
35
Effect of Temperature on Atomic Absorption and
Emission
  • The question here is which technique would be
    affected more as a result of fluctuations in
    temperature? The answer to this important
    question is rather simple. Atomic emission is the
    technique that will be severely affected by
    fluctuations in temperature since signal is
    dependent on the number of atoms in the excited
    state. This number is significantly affected by
    fluctuations in temperature as seen from the
    example above. However, in the case of atomic
    absorption, the signal depends on the number of
    atoms in ground state that will absorb energy.

36
  • very high as related to the number of excited
    atoms
  • Nj/No 1.72x10-4
  • or
  • 172 excited atoms for each 106 atoms in ground
    state
  • This suggests a very high population of the
    ground state even at high temperatures.
    Therefore, atomic absorption will not be affected
    to any significant extent by fluctuations in
    temperature, if compared to atomic emission
    spectroscopy.

37
  • However, there are some indirect effects of
    temperature on atomic absorption spectroscopy.
    These effects can be summarized as
  • Better sensitivities are obtained at higher
    temperatures since higher temperatures can
    increase the number of vaporized atoms at any
    time.
  • Higher temperatures will increase the velocities
    of gaseous atoms, thus causing line broadening as
    a result of the Doppler and collisional effects.
  • High temperatures increase the number of ionized
    analyte and thus decrease the number of atoms
    available for absorption.

38
Band and Continuum Spectra Associated with Atomic
Spectra
  • When the atomization temperature is insufficient
    to cause atomization of all species in the sample
    matrix, the existent molecular entities, at the
    temperature of the analysis, impose very
    important problems on the results of atomic
    absorption and emission spectroscopy. The
    background band spectrum should be removed for
    reasonable determination of analytes. Otherwise,
    the sensitivity of the instrument will be
    significantly decreased.

39
(No Transcript)
40
  • As the signal for the blank is considered zero
    and thus the instrument is made to read zero,
    when the analyte is to be determined, it got to
    have an absorbance greater than the highest point
    on the continuum and the instrument will assume
    that the absorbance related to analyte is just
    the value exceeding the background blank value.
    This will severely limit the sensitivity of the
    technique.

41
  • Putting this conclusion in other words we may say
    that if the analyte signal is less than the
    background blank, the instrument will read it as
    zero. Therefore, it is very important to correct
    for the background or simply eliminate it through
    use of very high temperatures that will
    practically atomize all species in the matrix. We
    will come to background correction methods in the
    next chapter.

42
Atomization Methods
  • It is essential, as we have seen from previous
    discussion, that all sample components (including
    analytes, additives, etc.) should be atomized.
    The atoms in the gaseous state absorb or emit
    radiation and can thus be determined. Many
    ionization methods are available which will be
    detailed in the next two chapters. Generally,
    atomization methods can be summarized below

43
(No Transcript)
44
Sample Introduction Methods
  • The method of choice for a specific sample will
    mainly depend on whether the sample is in
    solution or solid form. The method for sample
    introduction in atomic spectroscopy affects the
    precision, accuracy and detection limit of the
    analytical procedure.

45
Introduction of Solution Samples
  • 1. Pneumatic Nebulizers
  • Samples in solution are usually easily introduced
    into the atomizer by a simple nebulization,
    aspiration, process. Nebulization converts the
    solution into an aerosol of very fine droplets
    using a jet of compressed gas. The flow of gas
    carries the aerosol droplets to the atomization
    chamber or region. Several versions of nebulizers
    are available and few are shown in the figure
    below

46
(No Transcript)
47
(No Transcript)
48
(No Transcript)
49
(No Transcript)
50
(No Transcript)
51
2. Ultrasonic Nebulizers
  • In this case samples are pumped onto the surface
    of a piezoelectric crystal that vibrates in the
    kHz to MHz range. Such vibrations convert samples
    into homogeneous aerosols that can be driven into
    atomizers. Ultrasonic nebulization is preferred
    over pneumatic nebulization since finer droplets
    and more homogeneous aerosols are usually
    achieved. However, most instruments use pneumatic
    nebulization.

52
3. Electrothermal Vaporization
  • An accurately measured quantity of sample (few
    mL) is introduced into an electrically heated
    cylindrical chamber through which an inert gas
    flows. Usually, the cylinder is made of pyrolytic
    carbon but tungsten cylinders are now available.
    The signal produced by instruments which use
    electrothermal vaporization (ETV) is a discrete
    signal for each sample injection. Electrothemal
    vaporizers are called discrete atomizers to
    differentiate them from nebulizers which are
    called continuous atomizers

53
4. Hydride Generation Techniques
  • Samples that contain arsenic, antimony, tin,
    selenium, bismuth, and lead can be vaporized by
    converting them to volatile hydrides by addition
    of sodium borohydride. Volatile hydrides are then
    swept into the atomizer by a stream of an inert
    gas.

54
Introduction of Solid Samples
  • A variety of techniques were used to introduce
    solid samples into atomizers. These include
  • 1. Direct Sample Insertion
  • Samples are first powdered and placed in a
    boat-like holder (from graphite or tantalum)
    which is placed in a flame or an electrothermal
    atomizer.
  • 2. If the sample is conductive and is of a shape
    that can be directly used as an electrode (like a
    piece of metal or coin), that would be the choice
    for sample introduction in arc and spark
    techniques. Otherwise, powdered solid samples are
    mixed with fine graphite and made into a paste.
    Upon drying, this solid composite can be used as
    an electrode. The discharge caused by arcs and
    sparks interacts with the surface of the solid
    sample creating a plume of very fine particulates
    and atoms that are swept into the atomizer by a
    flow of an inert gas. This process of sample
    introduction is called ablation

55
  • 3. Laser Ablation
  • Sufficient energy from a focused intense laser
    will interact with the surface of samples (in a
    similar manner like arcs and sparks) resulting in
    ablation. The formed plume of vapor and fine
    particulates are swept into the atomizer by the
    flow of an inert gas. Laser ablation is becoming
    increasingly used since it is applicable to
    conductive and nonconductive samples

56
(No Transcript)
57
(No Transcript)
58
4. The Glow Discharge Technique
  • A low pressure envelope (1 to 10 torr argon) with
    two electrodes with the conductive solid sample
    is the cathode, as in the figure below. The
    technique is used for sample introduction and
    atomization as well. The electrodes are kept at a
    250 to 1000 V DC. This high potential is
    sufficient to cause ionization of argon which
    will be accelerated to the cathode where the
    sample is introduced. Collision of the fast
    moving energetic argon ions with the sample
    (cathode) causes atomization by a process called
    sputtering.

59
(No Transcript)
60
(No Transcript)
61
Atomic Absorption Spectroscopy
  • We will cover two main techniques of atomic
    absorption spectroscopy (AAS), depending of the
    type atomizer. Two atomization techniques are
    usually used in AAS

62
1. Flame Atomization
  • Flames are regarded as continuous atomizers since
    samples are continuously introduced and a
    constant or continuous signal is obtained.
    Samples in solution form are nebulized by one of
    the described nebulization techniques discussed
    previously. The most common nebulization
    technique is the pneumatic nebulization.
    Nebulized solutions are carried into a flame
    where atomization takes place.

63
  • Several processes occur during atomization
    including
  • a. Nebulized samples are sprayed into a flame as
    a spray of very fine droplets
  • b. Droplets will lose their solvent content due
    to very high flame temperatures in a process
    called desolvation and will thus be converted
    into a solid aerosol.
  • c. The solid aerosol is volatilized to form
    gaseous molecules

64
  • d. Gaseous molecules will then be atomized and
    neutral atoms are obtained which can be excited
    by absorption of enough energy. If energy is not
    enough for atomization, gaseous molecules will
    not be atomized and we may see molecular
    absorption or emission
  • e. Atoms in the gaseous state can absorb energy
    and are excited. If energy is too much, we may
    observe ionization.

65
(No Transcript)
66
(No Transcript)
67
  • The different processes occurring in flames are
    complicated and are not closely controlled and
    predicted. Therefore, it can be fairly stated
    that the atomization process in flames may be one
    of the important parameters limiting the
    precision of the method. It is therefore
    justified that we have a closer look at flames
    and their characteristics and the different
    variables contributing to their performance.

68
Types of Flames
  • Flames can be classified into several types
    depending on fuel/oxidant used. For example, the
    following table summarizes the features of most
    familiar flames.
  • Therefore, it can be clearly seen that
    significant variations in flame temperatures can
    be obtained by changing the composition of fuel
    and oxidant.

69
  • On the other hand, flames are only stable at
    certain flow rates and thus the flow rate of the
    gas is very important where at low flow rates
    (less than the maximum burning velocity) the
    flame propagates into the burner body causing
    flashback and, in some cases, an explosion. As
    the flow rate is increased, the flame starts to
    rise above the burner body. Best flames are
    obtained when the flow rate of the gas is equal
    to the maximum burning velocity. At this equity
    ratio the flame is most stable. At higher ratios,
    flames will reach a point where they will no
    longer form and blow off the burner.

70
 Flame Structure
  • Three well characterized regions can be
    identified in a conventional flame. A lower
    region, close to the burner tip, with blue
    luminescence. This region is called the primary
    combustion zone which is characterized by
    existence of some non atomized species and
    presence of fuel species (C2 and CH, etc.) that
    emit in the blue region of the electromagnetic
    spectrum. The second well defined region is
    called the interzonal region just above the
    primary combustion zone. The interzonal region is
    rich in free atoms and is the region of choice
    for performing atomic spectroscopy.

71
  • It also contains the regions of highest
    temperatures. The third region in the flame is
    the outer region which is called the secondary
    combustion region. It is characterized by
    reformation of molecules as the temperature at
    the edges is much lower than the core. These
    regions can be schematically represented by the
    following schematic

72
(No Transcript)
73
Flame Absorbance profiles
  • Since the temperature of a flame depends on the
    position from its tip, it is necessary to
    concentrate our work on one spot in a flame and
    preferably adjust the height of the flame to get
    best signal. In fact, not all elements require a
    specific height above burner tip but rather each
    element has its own requirements which largely
    reflect some of its properties. For example, one
    can use higher distances from the tip so that
    higher temperatures are achieved to analyze for
    silver. This is possible since silver will not be
    easily oxidized.

74
Flame Absorption ProfilesWe have seen that there
are different temperature profiles in a flame and
temperature changes as the distance from the
burner tip is change
75
  • However, best results for the analysis of
    chromium occur at lower heights (fuel rich
    flames) since at higher heights oxygen from
    atmosphere will force chromium to convert to the
    oxide which will not be atomized at flame
    temperatures. A third situation can be observed
    for magnesium where increasing the height above
    tip will increase the signal due to increased
    atomization at higher temperatures. However, at
    higher distances the oxide starts to form leading
    to a decrease in signal.

76
(No Transcript)
77
Flame Atomizers (Continuous Atomizers)
  • Flame Atomizers (Continuous Atomizers)
  • There are several types of flame atomizers
    available. The simplest is a turbulent flow
    burner that is very similar to conventional
    Bunsen burner. This type of burner suffers from
    fluctuations in temperature since there is no
    good mechanism for homogeneous mixing of fuel and
    oxidant. The drop size of nebulized sample is
    also inhomogeneous which adds to fluctuations in
    signal. The path length of radiation through the
    flame is small which suggests a lower sensitivity
    of the technique.

78
  • Turbulent flow burners are also susceptible to
    flashback. These drawbacks were overcome using
    the most widely used laminar flow burner where
    quite flames and long path length are obtained.
    Flashback is avoided and very homogeneous mixing
    between fuel, oxidant, and droplets take place.
    Larger droplets are excluded and directed to a
    waste container. A schematic representation of
    the burner is shown below

79
(No Transcript)
80
(No Transcript)
81
(No Transcript)
82
(No Transcript)
83
Fuel and Oxidant Regulators
  • The adjustment of the fuel to oxidant ratio and
    flow rate is undoubtedly very crucial. Although
    stoichiometric ratios are usually required,
    optimization is necessary in order to get highest
    signal. However, in the determination of metals
    that form stable oxides, a flame with excess fuel
    is preferred in order to decrease oxide formation.

84
Performance Characteristics of Flame Atomizers
  • Reproducibility of flame methods are usually
    superior to other atomization techniques.
    However, the residence time of an atom in a flame
    is in the order of 10-4 s which is very short.
    This is reflected in a lower sensitivity of flame
    methods as compared to other methods. Also,
    conventional flames with reasonable burning
    velocities can produce relatively low
    temperatures which make them susceptible to
    interference from molecular species.

85
2. Electrothermal Atomization
  • These have better sensitivities than flame
    methods. The increased sensitivity can be
    explained on the basis that a longer atom
    residence time is achieved (can be more than 1 s)
    as well as atomization of the whole sample in a
    very short time. As the name implies, a few mL of
    the sample are injected into the atomization
    chamber (a cylinder of graphite coated with a
    film of pyrolytic carbon) where the following
    processes take place

86
  • a. Evaporation the solvent associated with the
    sample is evaporated in a low temperature (120
    oC) slow process (seconds)
  • b. Ashing sample is ashed to burn organics
    associated with the sample at moderate
    temperatures (600 oC, seconds)
  • c. Atomization The current is rapidly increased
    after ashing so that a temperature in the range
    from 2000-3000 oC is obtained in less than1
    second.

87
Electrothermal Atomizers (Discrete Atomizers)
  • The heart of the atomizer, beside efficient
    heating elements and electronics, is a
    cylindrical graphite tube opened from both ends
    and has a central hole for sample introduction.
    It was found that porous graphite results in poor
    reproducibility since some of the analyzed
    materials will diffuse through porous graphite
    and will thus lead to a history effect.

88
  • Therefore, the cylindrical graphite is made from
    a special type of nonporous high quality graphite
    called pyrolytic graphite. The length of the
    cylinder is 2-5 cm and it has less than 1 cm
    diameter. When the tube is fixed in place
    electrical contacts are achieved which are water
    cooled. Two inert gas streams (argon) flow at the
    external surface and through the internal space
    of the tube to prevent oxidation and clean the
    tube after each measurement. Usually, samples are
    analyzed in triplicates where three consecutive
    reproducible signals are required for each
    sample..

89
(No Transcript)
90
Performance Characteristics of Electrothermal
Atomizers
  • Electrothermal atomization is the technique of
    choice in case of small sample size. Also, higher
    sensitivities than flames are ordinarily
    obtained. Unfortunately, the analysis time is in
    the few minutes range and the relative precision
    is in the range of 5-10 as compared to 1 in
    flame methods. In addition, the linear dynamic
    range is usually small ( two orders of
    magnitude) which requires extra sample
    manipulation. It may be also mentioned that
    better experienced personnel can achieve the
    merits of the technique.

91
Atomic Absorption Instrumentation
  • Atomic absorption instruments consist of a source
    of radiation, a monochromator, a flame or
    electrothermal atomizer in which sample is
    introduced, and a transducer.

92
Radiation Sources
  • Although radiation in the UV-Vis region is
    required, we can not use broad band sources. This
    is because even the best monochromators can not
    provide a bandwidth that is narrower than the
    atomic absorption line. If the bandwidth of the
    incident radiation is wider than the line width,
    measurement will fail as absorption will be only
    a tiny fraction of a large signal which is
    difficult to measure and will result in very low
    sensitivities (figure a). Therefore, line sources
    with bandwidths narrower than that of the
    absorption lines must be used

93
(No Transcript)
94
  • This can be achieved by using a lamp producing
    the emission line of the element of interest
    where analyte atoms can absorb that line.
    Conditions are established to get a narrower
    emission line than the absorption line. This can
    in fact be achieved by getting an emission line
    of interest at the following conditions

95
  • 1. Low temperatures to decrease Doppler
    broadening (which is easily achievable since the
    temperature of the source is always much less
    than the temperature in flames).
  • 2. Lower pressures this will decrease pressure
    broadening and will thus produce a very narrow
    emission line.

96
  • This may suggest the need for a separate lamp for
    each element which is troublesome and
    inconvenient. However, recent developments lead
    to introduction of multielement lamps. In this
    case, the lines from all elements should not
    interfere and must be easily resolved by the
    monochromator so that, at a specific time, a
    single line of one element is leaving the exit
    slit

97
Hollow Cathode Lamp (HCL)
  • This is the most common source in atomic
    absorption spectroscopy. It is formed from a
    tungsten anode and a cylindrical cathode the
    interior surface of which is coated by the metal
    of interest. The two electrodes are usually
    sealed in a glass tube with a quartz window and
    filled with argon at low pressure (1-5 torr).
    Ionization of the argon is forced by application
    of about 300 V DC where positively charged Ar
    heads rapidly towards the negatively charged
    cathode causing sputtering. A portion of
    sputtered atoms is excited and thus emit photons
    as atoms relax to ground state. The cylindrical
    shape of the cathode serves to concentrate the
    beam in a limited region and enhances
    redeposition of sputtered atoms at the hollow
    surface.

98
(No Transcript)
99
(No Transcript)
100
  • High potentials usually result in high currents
    which, in turn, produce more intense radiation.
    However, Doppler broadening increases as a
    result. In addition, the higher currents will
    produce high proportion of unexcited atoms that
    will absorb some of the emission beam which is
    referred to as self absorption (a lower intensity
    at the center of the line is observed in this
    case).

101
Electrodeless Discharge Lamps (EDL)
  • An EDL is a sealed quartz tube containing a few
    torr of an inert gas and a small quantity of the
    metal of interest. Excitation of the metal is
    achieved by a radiofrequency or a microwave
    powered coil through ionization of argon, due to
    high energetic radiofrequency. Ionized argon will
    hit the metal causing excitation of the atoms of
    the metal of interest. The output power of the
    EDL lamp is higher than the HCL lamp. However,
    compared to HCL lamps, EDL lamps are rarely used.

102
(No Transcript)
103
(No Transcript)
104
Emission in Flames
  • There can be significant amounts of emission
    produced in flames due to presence of flame
    constituents (molecular combustible products) and
    sometimes impurities in the burner head. This
    emitted radiation must be removed for successful
    sensitive determinations by AAS, otherwise a
    negative error will always be observed. We can
    visualize this effect by considering the
    schematic below

105
(No Transcript)
106
  • The detector will see the overall signal which is
    the power of the transmitted beam (P) in addition
    to the power of the emitted radiation from flame
    (Pe). Therefore if we are measuring absorbance,
    this will result in a negative error as the
    detector will measure what it appears as a high
    transmittance signal (actually it is P Pe). In
    case of emission measurements, there will always
    be a positive error since emission from flame is
    an additive value to the actual sample emission.
    It is therefore obvious that we should get rid of
    this interference from emission in flames.

107
Source Modulation
  • It turned out that excluding the emission signal
    from flames can easily be done by an addition of
    a chopper to the instrumental design. The chopper
    is a motor driven device that has open and solid
    (mirrors in some cases) alternating regions as in
    the schematic

108
(No Transcript)
109
  • The function of the chopper is to chop the light
    leaving the source so that when the incident beam
    hits the chopper at the solid surface, the beam
    will be blocked and detector will only read the
    emitted signal from the flame. As the chopper
    rotates and the beam emerges to the detector, the
    detector signal will be the sum of the
    transmitted signal plus that emitted from the
    flame. The signal processor will be able to
    subtract the first signal from the second one,
    thus excluding the signal from emission in
    flames.

110
  • This can be represented by the following
    equations
  • Signal 1 (Blocked Beam)  Pe
  • Signal 2 (Transmitted Beam) P Pe
  • Overall Difference Signal (P Pe) - Pe P
    (Corrected Signal)
  •  
  • This correction method for background emission in
    flames is called source modulation.

111
The schematic of the AAS instrument with source
modulation correction can be represented by the
following schematic
112
  • It should be recognized that addition of extra
    components to an instrument will decrease the
    signal to noise ratio and addition of a moving
    component is usually regarded as a disadvantage
    due to higher need for maintenance.
  • Another procedure which can overcome the emission
    from flames is to use a modulated power supply
    that will give fluctuating intensities at some
    frequency (say for example pulsed radiation at a
    specific frequency).

113
  • The emission from flames is a continuous signal
    but that from the source is modulated. Now if we
    use a high pass RC filter, only the fluctuating
    signal will be measured as signal while the DC
    signal will be considered zero as it can not pass
    through the electronic filter. The high pass RC
    filter is a device which uses a resistor and a
    capacitor the impedance of which is inversely
    proportional to the frequency of the modulated
    signal. Therefore, only high frequencies will
    have low impedance and can pass through the
    capacitor while signals of low frequencies will
    suffer very high resistance and will not be able
    to go through the capacitor.

114
AAS Instruments
  •  Instruments in AAS can be regarded as single or
    double beam instruments.
  •  
  • Single Beam Atomic Absorption Spectrophotometers
  • A single beam instruments is the same as the one
    described above (source modulation section) or
    generally

115
The term spectrophotometer implies that the
instrument uses a dispersive monochromator
(containing a prism or a grating). Also, the
detector is a photomultiplier tube in most cases.
116
Double Beam Atomic Absorption Spectrophotometers
  • In this type of instruments, the incident beam is
    split into two beams of equal intensity by a
    chopper with the solid surface being a mirror.
    One of the beams will traverse the sample in the
    atomizer while the other is considered as a
    reference. Detector signals will be consecutive
    readings of both the reference and sample beams.
    The ratio of the reference to the sample beams is
    recorded to give the final signal.

117
A schematic representation of a double beam
instrument is shown below
118
  • It should be emphasized here that in the absence
    of sample, Pr is not equal to P since the
    reference beam traverses through air while the
    other beam traverses through the flame. In
    flames, particulates and molecular species
    scatter and absorb a portion of incident
    radiation, which results in a lower intensity of
    the beam. To act as a real double beam, The AA
    spectrophotometer reference beam should pass
    through a reference flame.

119
  • But even if we do that, there are no guarantees
    that both beams will be of equal intensities
    because it is almost impossible to obtain exactly
    equivalent flames. It is therefore important to
    understand that the excellent features of a
    double beam configuration are not achievable in
    AAS instrumentation.

120
Interferences in Atomic Absorption Spectroscopy
  • There are two major classes of interferences
    which can be identified in atomic absorption
    spectroscopy. The first class is related to
    spectral properties of components other than
    atomized analyte and is referred to as spectral
    interferences. The other class of interferences
    is related to the chemical processes occurring in
    flames and electrothermal atomizers and their
    effects on signal. These are referred to as
    chemical interferences and are usually more
    important than spectral interferences.

121
Spectral Interferences
  • 1. Spectral line Interference
  • Usually, interferences due to overlapping lines
    is rare since atomic lines are very narrow.
    However, even in cases of line interference, it
    can be simply overcome by choosing to perform the
    analysis using another line that has no
    interference with other lines. Therefore, line
    interference is seldom a problem in atomic
    spectroscopy.

122
2. Scattering
  • Particulates from combustion products and sample
    materials scatter radiation that will result in
    positive analytical error. The error from
    scattering can be corrected for by making a blank
    measurement. Scattering phenomenon is most
    important when concentrated solutions containing
    elements that form refractory oxides (like Ti,
    Zr, and W) are present in sample matrix.

123
  • Metal oxide particles with diameters larger than
    the incident wavelength will make scattering a
    real problem. In addition, samples containing
    organic materials or organic solvents can form
    carbonaceous (especially in cases of incomplete
    combustion) particles that scatter radiation.

124
 3. Broad Band Absorption
  • In cases where molecular species from combustion
    products or sample matrix are formed in flames or
    electrothermal atomizers, a broad band spectrum
    will result which will limit the sensitivity of
    the technique. It should be indicated here that
    spectral interferences by matrix products are not
    widely encountered in flame methods. Even if
    matrix effects are present in flames, they can be
    largely overcome by adjusting various
    experimental conditions like fuel/oxidant ratio
    or temperature.

125
  • Another method for overcoming matrix
    interferences is to use a much higher
    concentration of interferent than that initially
    present in sample material, in both sample and
    standards (this material is called a radiation
    buffer). The contribution from sample matrix will
    thus be insignificant.
  • Spectral interferences due to matrix are severe
    in electrothermal methods and must thus be
    corrected for.

126
Background Correction Methods
  • a.      The Two Line Correction Method
  • In this method, a reference line from the source
    (from an impurity in cathode or any emission
    line) is selected where this line should have the
    following properties
  • 1.      Very close to analyte line
  • 2.      Not absorbed by analyte
  • If such a line exists, since the reference line
    is not absorbed by the analyte, its intensity
    should remain constant throughout analysis.

127
  • However, if its intensity decreases, this will be
    an indication of absorbance or scattering by
    matrix species. The decrease in signal of the
    reference line is used to correct for the analyte
    line intensity (by subtraction of the absorbance
    of the reference from that of the analyte). This
    method is very simple but unfortunately it is not
    always possible to locate a suitable reference
    line.

128
b.      The Continuum Source Method
  • This background correction method is the most
    common method although, for reasons to be
    discussed shortly, it has major drawbacks and
    fails a lot. In this technique, radiation from a
    deuterium lamp and a HCL lamp alternately pass
    through the graphite tube analyzer. It is
    essential to keep the slit width of the
    monochromator sufficiently wide in order to pass
    a wide bandwidth of the deuterium lamp radiation.

129
  • In this case, the absorbance by analyte atoms is
    negligible and absorbance can be attributed to
    molecular species in matrix. The absorbance of
    the beam from the deuterium lamp is then
    subtracted from the analyte beam (HCL) and thus a
    background correction is obtained.

130
(No Transcript)
131
Problems Associated with Background Correction
Using D2 Lamp
  • 1.      The very hot medium inside the graphite
    tube is inhomogeneous and thus signal is
    dependent on the exact path a beam would follow
    inside the tube. Therefore, exact alignment of
    the D2 and HCL lamps should be made.
  • 2.      The radiant power of the D2 lamp in the
    visible is insignificant which precludes the use
    of the technique for analysis of analytes in the
    visible region.
  • 3.      Addition of an extra lamp and chopper
    will decrease the signal to noise ratio.

132
c.       Background Correction Based on Zeeman
Effect
  • Zeeman has observed that when gaseous atoms (but
    not molecules) are placed in a strong magnetic
    field ( 1 tesla), splitting of electronic energy
    levels takes place. The simplest splitting of one
    energy level results in three energy levels, one
    at a higher energy, another at a lower energy
    (two s satellite lines) and the third remains at
    the same energy as the level in absence of the
    magnetic field (central p line). Furthermore, the
    p line has twice the absorbance of a s line and
    absorbs polarized light parallel to direction of
    the magnetic field while the two s lines absorb
    light perpendicular to magnetic field.

133
  • Light from a HCL lamp will pass through a
    rotating polarizer that passes polarized light
    parallel to external magnetic field at one cycle
    and passes light perpendicular to field in the
    other cycle. The idea of background correction
    using this method is to allow light to traverse
    the sample in the graphite furnace atomizer and
    record the signal for both polarizer cycles using
    the wavelength at the p line.

134
  • a.       First cycle light parallel to field
    the p line of the analyte absorbs in addition to
    absorbance by matrix (molecular matrix absorb
    both polarized light parallel or perpendicular to
    field)
  • Signal a Ap AMatrix
  • b.      Second cycle light perpendicular to
    field the p line of analyte will not absorb
    light perpendicular to field and s lines will
    also not affect absorbance at the p line
    wavelength. Only matrix will absorb.
  • Signal b AMatrix

135
  • The overall signal is the difference of the two
    signals Ap
  • Therefore, excellent background correction is
    achieved using the Zeeman effect. This background
    correction method results in good correction and
    is usually one of the best methods available.

136
(No Transcript)
137
Chemical Interferences
  • These are interferences resulting from chemical
    processes occurring in flames and electrothermal
    atomizers and affect the absorption signal. To
    quantitatively assess the effects of the
    different chemical processes occurring in flames,
    one should regard the burnt gases as behaving
    like a solvent. This is necessary since our
    knowledge of gaseous state reaction equilibria is
    rather limited. Chemical interferences include
    three major processes

138
Formation of Compounds of Low Volatility
  • Anionic species forming compounds of low
    volatility are the most important. The formation
    of low volatility species will result in a
    negative error or at least will decrease the
    sensitivity. For example, the absorption signal
    of calcium will be decreased as higher
    concentrations of sulfate or phosphate are
    introduced. Cations forming combined products
    with the analyte will also decrease the signal
    obtained for the analyte. For example aluminum
    forms a heat stable compound with magnesium.

139
Elimination of Low Volatility Compounds
  • Addition of a releasing agent cations that can
    replace the analyte (preferentially react with
    the anion) are called releasing agents. In this
    case the analyte is released from the compound of
    low volatility and replaced by the releaseing
    agent. Lanthanum or strontium are good releasing
    agents in the determination of calcium in
    presence of phosphate or sulfate. Also, lanthanum
    or strontium are good releasing agents in the
    determination of magnesium in presence of
    aluminum since both can replace magnesium.

140
  • Addition of a protective agent organic ligands
    that form stable volatile species with analytes
    are called protective agents. An example is EDTA
    and 8-hydroxyquinoline which will form complexes
    with calcium even in presence of sulfate and
    phosphate or aluminum.
  • Use  of higher temperature is the simplest
    procedure to try if it is possible

141
Dissociation Equilibria
  • Dissociation reactions occur in flames where the
    outcome of the process is desired to produce the
    atoms of analyte. For example, metal oxides and
    hydroxides will dissociate in flames to produce
    the atoms as in the equations
  • MO M O
  • M(OH)2 M 2 OH

142
  • Remember that we are not working in solution to
    dissociate the compounds into ionic species. In
    fact, not much is known about equilibrium
    reactions in flames. It should also be remembered
    that alkaline earth oxides and hydroxides are
    relatively stable and will definitely show
    characteristic broad band spectra (more intense
    than line spectra), except at very high
    temperatures. The opposite behavior is observed
    fro alkali metals oxides and hydroxides which are
    instable even at lower flame temperatures and
    thus produce line spectra.

143
  • An equilibrium can be established for the
    dissociation of compounds containing atoms other
    than oxygen, like NaCl where
  • NaCl Na Cl
  • Now, if the signal from a solution of NaCl was
    studied in presence of variable amounts of Cl
    (from HCl, as an example), the signal will be
    observed to decrease as the concentration of Cl
    is increased a behavior predicted by the Le
    Chatelier principle in solutions.

144
  • The same phenomenon is observed when a metal
    oxide is analyzed using a fuel rich flame or a
    lean flame. Signal will be increased in fuel rich
    flames since the dissociation of metal oxides is
    easier due to less oxygen while the opposite
    takes place in lean flames (oxygen rich).

145
  • A good example on dissociation equilibria can be
    presented for the analysis of vanadium in
    presence of aluminum and titanium, fuel rich
    flames result in higher absorbance signal for
    vanadium since the little oxygen present in
    flames will be mainly captured by Al and Ti, thus
    more V atoms are available. However, in lean
    flames, excess oxygen is present and thus
    vanadium will form the oxide and addition of
    extra Ti and Al will not affect the signal. 

146
Ionization Equilibria
  •  Ionization in fuel/air flames is very limited
    due to relatively low temperatures. However, in
    fuel/nitrous oxide or fuel/oxygen mixtures,
    ionization is significant. Therefore, at higher
    temperatures an important portion of atoms can be
    converted to ions
  • M M e
  • K Me/M

147
  • Ionization in flames may explain the decrease in
    absorption signal for alkali metals at very high
    temperatures where as the temperature is
    increased signal will increase till an extent at
    some temperature where it starts to decrease as
    temperature is further increased a consequence
    of ionization. Therefore, usually lower flame
    temperatures are used for determination of alkali
    metals. A material that is added to samples in
    order to produce large number of electrons is
    referred to as an ionization suppressor, the
    addition of which results in higher
    sensitivities.

148
Practical Details in AAS
  • 1.        Sample Preparation
  • The most unfortunate requirement of AAS may be
    the need for introduction of samples in the
    solution form. This necessitates the dissolution
    of the sample where in many cases the procedure
    is lengthy and requires very good experience.
    Care should be particularly taken in order not to
    lose any portion of the analyte and to make sure
    that the reagents, acids, etc. used in the
    dissolution and pretreatment of the sample are
    free from analyte impurities.

149
  • I suggest that you follow exact procedures for
    preparation of specific samples for analysis by
    AAS. In some cases where the sample can be
    introduced directly to an electrothermal atomizer
    without pretreatment (like serum samples),
    definitely, electrothermal atomizers will have an
    obvious advantage over flame methods which
    require nebulization.

150
Organic Solvents
  • Increased nebulization rate due to lower surface
    tension of organic solvents which produces
    smaller droplets as well as faster evaporation of
    solvents in flames will result in better
    sensitivities.
  • Immiscible organic solvents containing organic
    ligands are used to extract metal ions of
    interest and thus concentrate them in a small
    volume (thus increasing sensitivity) and
    excluding possible interferences due to matrix
    components.

151
  Calibration Curves
  • The absorbance of a solution is directly
    proportional to its concentration but due to the
    large number of variables in AAS, usually this
    direct relationship may slightly deviate from
    linearity. The standard procedure to do is to
    construct a relation between the absorbance and
    concentration for a series of solutions of
    different concentrations. The thus constructed
    graph is called a calibration curve.

152
  • The unknown analyte absorbance is found and the
    concentration is calculated or located on the
    curve. Neither interpolation nor extrapolation is
    permitted to the calibration curve. A sample can
    be diluted or the calibration curve may be
    extended but always the analyte absorbance should
    be within the standard absorbance range recorded.
    Usually, the concentration axis has the ppm or
    ppb units.

153
Standard Addition method
  • Chemical and spectral interferences can be
    partially or wholly overcome by the use of a
    special technique of calibration called the
    method of standard addition. In addition, the use
    of this method provides better correlations
    between standards and sample results due to
    constant nebulization rates. The method involves
    addition of the same sample volume to a set of
    tubes or containers.

154
  • Variable volumes of a standard are added to the
    tube set followed by completion to a specific
    volume. Now, all tubes contain the same amount of
    sample but different concentrations of analyte. A
    plot is then made for the volume of standard and
    absorbance. This plot will have an intercept (b)
    with the y axis and a slope equals m.

155
  • The concentration of the analyte can be
    determined by the relation
  • Cx bCs/mVx
  • Where, Cx and Vx are concentration and volume of
    analyte and Cs is the concentration of standard.
  • One can only use two points to get the analyte
    concentration using the relation
  • Cx AxCsVs/(At Ax)Vx

156
Detection Limits
  •  Usually, atomic absorption based on
    electrothermal atomization has better
    sensitivities and detection limits than methods
    based on flames. In general, flame methods have
    detection limits in the range from 1-20 ppm while
    electrothermal methods have detection limits in
    the range from 1-20 ppb.

157
  • This range can significantly change for specific
    elements where not all elements have the same
    detection limits. For example, detection limits
    fro mercury and magnesium using electrothermal
    atomization are 100 and 0.02 ppb while the
    detection limits for the same elements using
    flame methods are 500 and 0.1 ppm, respectively.

158
Accuracy
  • Flame methods are superior to electrothermal
    methods in terms of accuracy. The relative error
    in flame method can be less than 1 while that
    for electrothermal method occurs in the range
    from 5-10. Also, electrothermal methods are more
    susceptible to molecular interferences from the
    matrix components. Therefore, unless a good
    background correction method is used, large
    errors can be encountered in electrothermal
    methods depending on the nature of sample
    analyzed.

159
Flame Photometry
  • The technique referred to as flame photometry is
    a flame emission technique. We introduce it here
    because we will not be back to flame methods in
    later chapters. The basics of the technique are
    extremely simple where a sample is nebulized into
    a flame. Atomization occurs due to high flame
    temperatures and also excitation of easily
    excitable atoms can occur.

160
  • Emission of excited atoms is proportional to
    concentration of analyte. Flame emission is good
    for such atoms that do not require high
    temperatures for atomization and excitation, like
    Na, K, Li, Ca, and Mg. The instrument is very
    simple and excludes the need for a source lamp.
    The filter is exchangeable in order to determine
    the analyte of interest and, in most cases, a
    photomultiplier tube is used as the detector.

161
(No Transcript)
162
(No Transcript)
163
b. Charge Transfer and Charge Coupled Transducers
  • The photosensitive elements are, in contrary to
    PDAs, arranged in two dimensions in both charge
    injection devices (CID) and charge-coupled
    devices (CCD). Therefore, these are very similar
    to photographic films. For example, a
    commercially available transducer is formed from
    244 rows with each row containing 388 detector
    elements. This will add up to a two-dimensional
    array holding 16672 detector elements (pixels) on
    silicon chip that is 6.5 mm by 8.7 mm.

164
(No Transcript)
165
(No Transcript)
166
  • The full description of the system and its
    mechanism will not be covered here as this is
    behind the scope of this course. However, we
    should qualitatively know that these important
    transducers function by first collecting the
    photogenerated charges in different pixels and
    then measuring the quantity of the charge
    accumulated in a brief period. Measurement is
    accomplished by transferring the charge from a
    collection area to a detection area.

167
Optical Atomic Spectra
  • We have briefly described in an introductory
    chapter that atomic spectra are usually line,
    rather than band, spectra due to absence of
    vibrational and rotational levels. The existence
    of quantized electronic energy levels explains
    the origin of the observed line spectra and exact
    locations of possible lines
Write a Comment
User Comments (0)
About PowerShow.com