Title: Atomic Spectroscopy
1Atomic Spectroscopy
2Technique Flame Test
3An Introduction to Optical Atomic Spectroscopy
- In optical atomic spectrometry, compounds are
first converted to gaseous molecules followed by
conversion to gaseous atoms. This process is
called atomization and is a prerequisite for
performing atomic spectroscopy. Gaseous atoms
then absorb energy from a beam of radiation or
simply heat. Absorbance can be measured or
emission from excited atoms is measured and is
related to concentration of analyte.
4Atomic Energy Level Diagrams
- As a start, we should be aware that only valence
electrons are responsible for atomic spectra
observed in a process of absorption or emission
of radiation in the UV-Vis region. Valence
electrons in their ground states are assumed to
have an energy equal to zero eV. As an electron
is excited to a higher energy level, it will
absorb energy exactly equal to the energy
difference between the two states. Let us look at
a portion of the sodium energy level diagram
where sodium got one electron in the 3s orbital
5(No Transcript)
6(No Transcript)
7- The dark lines represent most probable
transitions and in an atomic spectrum they would
appear more intense than others. It should also
be indicated that two transitions, of very
comparable energies (589.0 and 589.6 nm), from
the 3s ground state to 3p excited state do take
place. This suggests splitting of the p orbital
into two levels that slightly differ in energy.
Explanation of this splitting may be presented as
a result of electron spin where the electron spin
is either in the direction of the orbital motion
or opposed to it.
8- Both spin and orbital motion create magnetic
fields that may interact in an attractive manner
(if motion is in opposite direction, lower
energy), or in a repulsive manner when both spin
and orbital motion are in the same direction
(higher energy). The same occurs for both d and f
orbitals but the energy difference is so small to
be observed. A Mg ion would show very similar
atomic spectrum as Na since both have one
electron in the 3s orbital.
9- In cases where atoms of large numbers of
electrons are studied, atomic spectra become too
complicated and difficult to interpret. This is
mainly due to presence of a large numbers of
closely spaced energy levels - It should also be indicated that transition from
ground state to excited state is not arbitrary
and unlimited. Transitions follow certain
selection rules that make a specific transition
allowed or forbidden.
10Atomic Emission and Absorption Spectra
- At room temperature, essentially all atoms are in
the ground state. Excitation of electrons in
ground state atoms requires an input of
sufficient energy to transfer the electron to one
of the excited state through an allowed
transition. Excited electrons will only spend a
short time in the excited state (shorter than a
ms) where upon relaxation an excited electron
will emit a photon and return to the ground
state.
11- Each type of atoms would have certain preferred
or most probable transitions (sodium has the
589.0 and the 589.6 nm). Relaxation would result
in very intense lines for these preferred
transitions where these lines are called
resonance lines. - Absorption of energy is most probable for the
resonance lines of each element. Thus intense
absorption lines for sodium will be observed at
589.0 and 589.6 nm.
12Atomic Fluorescence Spectra
- When gaseous atoms at high temperatures are
irradiated with a monochromatic beam of radiation
of enough energy to cause electronic excitation,
emission takes place in all directions. The
emitted radiation from the first excited
electronic level, collected at 90o to the
incident beam, is called resonance fluorescence.
Photons of the same wavelength as the incident
beam are emitted in resonance fluorescence. This
topic will not be further explained in this text
as the merits of the technique are not very clear
compared to instrumental complexity involved
13Atomic Line Width
- It is taken for granted that an atomic line
should have infinitesimally small (or zero) line
width since transition between two quantum states
requires an exact amount of energy. However,
careful examination of atomic lines reveals that
they have finite width. For example, try to look
at the situation where we expand the x-axis
(wavelength axis) of the following line
14(No Transcript)
15- The effective line width in terms of wavelength
units is equal to Dl1/2 and is defined as the
width of the line, in wavelength units, measured
at one half maximum signal (P). The question
which needs a definite answer is what causes the
atomic line to become broad?
16Reasons for Atomic Line Broadening
- There are four reasons for broadening observed in
atomic lines. These include - 1. The Uncertainty Principle
- We have seen earlier that Heisenberg uncertainty
principle suggests that nature places limits on
the precision by which two interrelated physical
quantities can be measured. It is not easy, will
have some uncertainty, to calculate the energy
required for a transition when the lifetime of
the excited state is short.
17- The ground state lifetime is long but the
lifetime of the excited state is very short which
suggests that there is an uncertainty in the
calculation of the transition time. We have seen
earlier that when we are to estimate the energy
of a transition and thus the wavelength (line
width), it is required that the two states where
a transition takes place should have infinite
lifetimes for the uncertainty in energy (or
wavelength) to be zero
18- DEgth/Dt
- Therefore, atomic lines should have some
broadening due to uncertainty in the lifetime of
the excited state. The broadening resulting from
the uncertainty principle is referred to as
natural line width and is unavoidable.
192. Doppler Broadening
- The wavelength of radiation emitted by a fast
moving atom toward a transducer will be different
from that emitted by a fast atom moving away from
a transducer. More wave crests and thus higher
frequency will be measured for atoms moving
towards the transducer. The same occurs for sound
waves
20- Assume your ear is the transducer, when a car
blows its horn toward your ear each successive
wave crest is emitted from a closer distance to
your ear since the car is moving towards you.
Thus a high frequency will be detected. On the
other hand, when the car passes you and blows its
horn, each wave crest is emitted at a distance
successively far away from you and your ear will
definitely sense a lower frequency.
21(No Transcript)
22- The line width (Dl) due to Doppler broadening can
be calculated from the relation - Dl/lo v/c
- Where lo is the wavelength at maximum power and
is equal to (l1 l2)/2, v is the velocity of the
moving atom and c is the speed of light. It is
noteworthy to indicate that an atom moving
perpendicular to the transducer will always have
a lo, i.e. will keep its original frequency and
will not add to line broadening by the Doppler
effect.
23- In the case of absorption lines, you may
visualize the line broadening due to Doppler
effect since fast atoms moving towards the source
will experience more wave crests and thus will
absorb higher frequencies. On the other hand, an
atom moving away from the source will experience
less wave crests and will thus absorb a lower
frequency. The maximum Doppler shifts are
observed for atoms of highest velocities moving
in either direction toward or away from a
transducer (emission) or a source (absorption).
24(No Transcript)
253. Pressure Broadening
- Line broadening caused by collisions of emitting
or absorbing atoms with other atoms, ions, or
other species in the gaseous matrix is called
pressure or collisional broadening. These
collisions result in small changes in ground
state energy levels and thus the energy required
for transition to excited states will be
different and dependent on the ground state
energy level distribution.
26- This will definitely result in important line
broadening. This phenomenon is most astonishing
for xenon where a xenon arc lamp at a high
pressure produces a continuum from 200 to 1100 nm
instead of a line spectrum for atomic xenon. A
high pressure mercury lamp also produces a
continuum output. Both Doppler and pressure
contribution to line broadening in atomic
spectroscopy are far more important than
broadening due to uncertainty principle.
274. Magnetic Effects
- Splitting of the degenerate energy levels does
take place for gaseous atoms in presence of a
magnetic field. The complicated magnetic fields
exerted by electrons in the matrix atoms and
other species will affect the energy levels of
analyte atoms. The simplest situation is one
where an energy level will be split into three
levels, one of the same quantum energy and one of
higher quantum energy, while the third assumes a
lower quantum energy state. A continuum of
magnetic fields exists due to complex matrix
components, and movement of species, thus exist.
Electronic transitions from the thus split levels
will result in line broadening
28The Effect of Temperature on Atomic Spectra
- Atomic spectroscopic methods require the
conversion of atoms to the gaseous state. This
requires the use of high temperatures (in the
range from 2000-6000 oC). Thee high temperature
can be provided through a flame, electrical
heating, an arc or a plasma source. It is
essential that the temperature be of enough value
to convert atoms of the different elements to
gaseous atoms and, in some cases, provide energy
required for excitation. The temperature of a
source should remain constant throughout the
analysis especially in atomic emission
spectroscopy.
29- Quantitative assessment of the effect of
temperature on the number of atoms in the excited
state can be derived from Boltzmann equation - Where Nj is the number of atoms in excited state,
No is the number of atoms in the ground state, Pj
and Po are constants determined by the number of
states having equal energy at each quantum level,
Ej is the energy difference between excited and
ground states, K is the Boltzmann constant, and T
is the absolute temperature.
30Boltzmann distribution
Nj /N0 at 3000 K
Wavelength
Atom
852.1 nm
Cs
7.24 ? 10-3
589.0 nm
Na
5.88 ? 10-4
422.7 nm
Ca
3.69 ? 10-5
213.9 nm
Zn
5.58 ? 10-10
31- To understand the application of this equation
let us consider the situation of sodium atoms in
the 3s state (Po 2) when excited to the 3p
excited state (Pj 6) at two different
temperatures 2500 and 2510K. Now let us apply the
equation to calculate the relative number of
atoms in the ground and excited states - Usually we use the average of the emission lines
from the 3p to 3s where we have two lines at
589.0 and 589.6 nm which is
32(No Transcript)
33Therefore, at higher temperatures, the number of
atoms in the excited state increases. Let us
calculate the percent increase in the number of
atoms in the excited state as a result of this
increase in temperature of only 10 oC
34(No Transcript)
35Effect of Temperature on Atomic Absorption and
Emission
- The question here is which technique would be
affected more as a result of fluctuations in
temperature? The answer to this important
question is rather simple. Atomic emission is the
technique that will be severely affected by
fluctuations in temperature since signal is
dependent on the number of atoms in the excited
state. This number is significantly affected by
fluctuations in temperature as seen from the
example above. However, in the case of atomic
absorption, the signal depends on the number of
atoms in ground state that will absorb energy.
36- very high as related to the number of excited
atoms - Nj/No 1.72x10-4
- or
- 172 excited atoms for each 106 atoms in ground
state - This suggests a very high population of the
ground state even at high temperatures.
Therefore, atomic absorption will not be affected
to any significant extent by fluctuations in
temperature, if compared to atomic emission
spectroscopy.
37- However, there are some indirect effects of
temperature on atomic absorption spectroscopy.
These effects can be summarized as - Better sensitivities are obtained at higher
temperatures since higher temperatures can
increase the number of vaporized atoms at any
time. - Higher temperatures will increase the velocities
of gaseous atoms, thus causing line broadening as
a result of the Doppler and collisional effects. - High temperatures increase the number of ionized
analyte and thus decrease the number of atoms
available for absorption.
38Band and Continuum Spectra Associated with Atomic
Spectra
- When the atomization temperature is insufficient
to cause atomization of all species in the sample
matrix, the existent molecular entities, at the
temperature of the analysis, impose very
important problems on the results of atomic
absorption and emission spectroscopy. The
background band spectrum should be removed for
reasonable determination of analytes. Otherwise,
the sensitivity of the instrument will be
significantly decreased.
39(No Transcript)
40- As the signal for the blank is considered zero
and thus the instrument is made to read zero,
when the analyte is to be determined, it got to
have an absorbance greater than the highest point
on the continuum and the instrument will assume
that the absorbance related to analyte is just
the value exceeding the background blank value.
This will severely limit the sensitivity of the
technique.
41- Putting this conclusion in other words we may say
that if the analyte signal is less than the
background blank, the instrument will read it as
zero. Therefore, it is very important to correct
for the background or simply eliminate it through
use of very high temperatures that will
practically atomize all species in the matrix. We
will come to background correction methods in the
next chapter.
42Atomization Methods
- It is essential, as we have seen from previous
discussion, that all sample components (including
analytes, additives, etc.) should be atomized.
The atoms in the gaseous state absorb or emit
radiation and can thus be determined. Many
ionization methods are available which will be
detailed in the next two chapters. Generally,
atomization methods can be summarized below
43(No Transcript)
44Sample Introduction Methods
- The method of choice for a specific sample will
mainly depend on whether the sample is in
solution or solid form. The method for sample
introduction in atomic spectroscopy affects the
precision, accuracy and detection limit of the
analytical procedure.
45Introduction of Solution Samples
- 1. Pneumatic Nebulizers
- Samples in solution are usually easily introduced
into the atomizer by a simple nebulization,
aspiration, process. Nebulization converts the
solution into an aerosol of very fine droplets
using a jet of compressed gas. The flow of gas
carries the aerosol droplets to the atomization
chamber or region. Several versions of nebulizers
are available and few are shown in the figure
below
46(No Transcript)
47(No Transcript)
48(No Transcript)
49(No Transcript)
50(No Transcript)
512. Ultrasonic Nebulizers
- In this case samples are pumped onto the surface
of a piezoelectric crystal that vibrates in the
kHz to MHz range. Such vibrations convert samples
into homogeneous aerosols that can be driven into
atomizers. Ultrasonic nebulization is preferred
over pneumatic nebulization since finer droplets
and more homogeneous aerosols are usually
achieved. However, most instruments use pneumatic
nebulization.
523. Electrothermal Vaporization
- An accurately measured quantity of sample (few
mL) is introduced into an electrically heated
cylindrical chamber through which an inert gas
flows. Usually, the cylinder is made of pyrolytic
carbon but tungsten cylinders are now available.
The signal produced by instruments which use
electrothermal vaporization (ETV) is a discrete
signal for each sample injection. Electrothemal
vaporizers are called discrete atomizers to
differentiate them from nebulizers which are
called continuous atomizers
534. Hydride Generation Techniques
- Samples that contain arsenic, antimony, tin,
selenium, bismuth, and lead can be vaporized by
converting them to volatile hydrides by addition
of sodium borohydride. Volatile hydrides are then
swept into the atomizer by a stream of an inert
gas.
54Introduction of Solid Samples
- A variety of techniques were used to introduce
solid samples into atomizers. These include - 1. Direct Sample Insertion
- Samples are first powdered and placed in a
boat-like holder (from graphite or tantalum)
which is placed in a flame or an electrothermal
atomizer. - 2. If the sample is conductive and is of a shape
that can be directly used as an electrode (like a
piece of metal or coin), that would be the choice
for sample introduction in arc and spark
techniques. Otherwise, powdered solid samples are
mixed with fine graphite and made into a paste.
Upon drying, this solid composite can be used as
an electrode. The discharge caused by arcs and
sparks interacts with the surface of the solid
sample creating a plume of very fine particulates
and atoms that are swept into the atomizer by a
flow of an inert gas. This process of sample
introduction is called ablation
55- 3. Laser Ablation
- Sufficient energy from a focused intense laser
will interact with the surface of samples (in a
similar manner like arcs and sparks) resulting in
ablation. The formed plume of vapor and fine
particulates are swept into the atomizer by the
flow of an inert gas. Laser ablation is becoming
increasingly used since it is applicable to
conductive and nonconductive samples
56(No Transcript)
57(No Transcript)
584. The Glow Discharge Technique
- A low pressure envelope (1 to 10 torr argon) with
two electrodes with the conductive solid sample
is the cathode, as in the figure below. The
technique is used for sample introduction and
atomization as well. The electrodes are kept at a
250 to 1000 V DC. This high potential is
sufficient to cause ionization of argon which
will be accelerated to the cathode where the
sample is introduced. Collision of the fast
moving energetic argon ions with the sample
(cathode) causes atomization by a process called
sputtering.
59(No Transcript)
60(No Transcript)
61Atomic Absorption Spectroscopy
- We will cover two main techniques of atomic
absorption spectroscopy (AAS), depending of the
type atomizer. Two atomization techniques are
usually used in AAS
621. Flame Atomization
- Flames are regarded as continuous atomizers since
samples are continuously introduced and a
constant or continuous signal is obtained.
Samples in solution form are nebulized by one of
the described nebulization techniques discussed
previously. The most common nebulization
technique is the pneumatic nebulization.
Nebulized solutions are carried into a flame
where atomization takes place.
63- Several processes occur during atomization
including - a. Nebulized samples are sprayed into a flame as
a spray of very fine droplets - b. Droplets will lose their solvent content due
to very high flame temperatures in a process
called desolvation and will thus be converted
into a solid aerosol. - c. The solid aerosol is volatilized to form
gaseous molecules
64- d. Gaseous molecules will then be atomized and
neutral atoms are obtained which can be excited
by absorption of enough energy. If energy is not
enough for atomization, gaseous molecules will
not be atomized and we may see molecular
absorption or emission - e. Atoms in the gaseous state can absorb energy
and are excited. If energy is too much, we may
observe ionization.
65(No Transcript)
66(No Transcript)
67- The different processes occurring in flames are
complicated and are not closely controlled and
predicted. Therefore, it can be fairly stated
that the atomization process in flames may be one
of the important parameters limiting the
precision of the method. It is therefore
justified that we have a closer look at flames
and their characteristics and the different
variables contributing to their performance.
68Types of Flames
- Flames can be classified into several types
depending on fuel/oxidant used. For example, the
following table summarizes the features of most
familiar flames. - Therefore, it can be clearly seen that
significant variations in flame temperatures can
be obtained by changing the composition of fuel
and oxidant.
69- On the other hand, flames are only stable at
certain flow rates and thus the flow rate of the
gas is very important where at low flow rates
(less than the maximum burning velocity) the
flame propagates into the burner body causing
flashback and, in some cases, an explosion. As
the flow rate is increased, the flame starts to
rise above the burner body. Best flames are
obtained when the flow rate of the gas is equal
to the maximum burning velocity. At this equity
ratio the flame is most stable. At higher ratios,
flames will reach a point where they will no
longer form and blow off the burner.
70 Flame Structure
- Three well characterized regions can be
identified in a conventional flame. A lower
region, close to the burner tip, with blue
luminescence. This region is called the primary
combustion zone which is characterized by
existence of some non atomized species and
presence of fuel species (C2 and CH, etc.) that
emit in the blue region of the electromagnetic
spectrum. The second well defined region is
called the interzonal region just above the
primary combustion zone. The interzonal region is
rich in free atoms and is the region of choice
for performing atomic spectroscopy.
71- It also contains the regions of highest
temperatures. The third region in the flame is
the outer region which is called the secondary
combustion region. It is characterized by
reformation of molecules as the temperature at
the edges is much lower than the core. These
regions can be schematically represented by the
following schematic
72(No Transcript)
73Flame Absorbance profiles
- Since the temperature of a flame depends on the
position from its tip, it is necessary to
concentrate our work on one spot in a flame and
preferably adjust the height of the flame to get
best signal. In fact, not all elements require a
specific height above burner tip but rather each
element has its own requirements which largely
reflect some of its properties. For example, one
can use higher distances from the tip so that
higher temperatures are achieved to analyze for
silver. This is possible since silver will not be
easily oxidized.
74Flame Absorption ProfilesWe have seen that there
are different temperature profiles in a flame and
temperature changes as the distance from the
burner tip is change
75- However, best results for the analysis of
chromium occur at lower heights (fuel rich
flames) since at higher heights oxygen from
atmosphere will force chromium to convert to the
oxide which will not be atomized at flame
temperatures. A third situation can be observed
for magnesium where increasing the height above
tip will increase the signal due to increased
atomization at higher temperatures. However, at
higher distances the oxide starts to form leading
to a decrease in signal.
76(No Transcript)
77Flame Atomizers (Continuous Atomizers)
- Flame Atomizers (Continuous Atomizers)
- There are several types of flame atomizers
available. The simplest is a turbulent flow
burner that is very similar to conventional
Bunsen burner. This type of burner suffers from
fluctuations in temperature since there is no
good mechanism for homogeneous mixing of fuel and
oxidant. The drop size of nebulized sample is
also inhomogeneous which adds to fluctuations in
signal. The path length of radiation through the
flame is small which suggests a lower sensitivity
of the technique.
78- Turbulent flow burners are also susceptible to
flashback. These drawbacks were overcome using
the most widely used laminar flow burner where
quite flames and long path length are obtained.
Flashback is avoided and very homogeneous mixing
between fuel, oxidant, and droplets take place.
Larger droplets are excluded and directed to a
waste container. A schematic representation of
the burner is shown below
79(No Transcript)
80(No Transcript)
81(No Transcript)
82(No Transcript)
83Fuel and Oxidant Regulators
- The adjustment of the fuel to oxidant ratio and
flow rate is undoubtedly very crucial. Although
stoichiometric ratios are usually required,
optimization is necessary in order to get highest
signal. However, in the determination of metals
that form stable oxides, a flame with excess fuel
is preferred in order to decrease oxide formation.
84Performance Characteristics of Flame Atomizers
- Reproducibility of flame methods are usually
superior to other atomization techniques.
However, the residence time of an atom in a flame
is in the order of 10-4 s which is very short.
This is reflected in a lower sensitivity of flame
methods as compared to other methods. Also,
conventional flames with reasonable burning
velocities can produce relatively low
temperatures which make them susceptible to
interference from molecular species.
852. Electrothermal Atomization
- These have better sensitivities than flame
methods. The increased sensitivity can be
explained on the basis that a longer atom
residence time is achieved (can be more than 1 s)
as well as atomization of the whole sample in a
very short time. As the name implies, a few mL of
the sample are injected into the atomization
chamber (a cylinder of graphite coated with a
film of pyrolytic carbon) where the following
processes take place
86- a. Evaporation the solvent associated with the
sample is evaporated in a low temperature (120
oC) slow process (seconds) - b. Ashing sample is ashed to burn organics
associated with the sample at moderate
temperatures (600 oC, seconds) - c. Atomization The current is rapidly increased
after ashing so that a temperature in the range
from 2000-3000 oC is obtained in less than1
second.
87Electrothermal Atomizers (Discrete Atomizers)
- The heart of the atomizer, beside efficient
heating elements and electronics, is a
cylindrical graphite tube opened from both ends
and has a central hole for sample introduction.
It was found that porous graphite results in poor
reproducibility since some of the analyzed
materials will diffuse through porous graphite
and will thus lead to a history effect.
88- Therefore, the cylindrical graphite is made from
a special type of nonporous high quality graphite
called pyrolytic graphite. The length of the
cylinder is 2-5 cm and it has less than 1 cm
diameter. When the tube is fixed in place
electrical contacts are achieved which are water
cooled. Two inert gas streams (argon) flow at the
external surface and through the internal space
of the tube to prevent oxidation and clean the
tube after each measurement. Usually, samples are
analyzed in triplicates where three consecutive
reproducible signals are required for each
sample..
89(No Transcript)
90Performance Characteristics of Electrothermal
Atomizers
- Electrothermal atomization is the technique of
choice in case of small sample size. Also, higher
sensitivities than flames are ordinarily
obtained. Unfortunately, the analysis time is in
the few minutes range and the relative precision
is in the range of 5-10 as compared to 1 in
flame methods. In addition, the linear dynamic
range is usually small ( two orders of
magnitude) which requires extra sample
manipulation. It may be also mentioned that
better experienced personnel can achieve the
merits of the technique.
91Atomic Absorption Instrumentation
- Atomic absorption instruments consist of a source
of radiation, a monochromator, a flame or
electrothermal atomizer in which sample is
introduced, and a transducer.
92Radiation Sources
- Although radiation in the UV-Vis region is
required, we can not use broad band sources. This
is because even the best monochromators can not
provide a bandwidth that is narrower than the
atomic absorption line. If the bandwidth of the
incident radiation is wider than the line width,
measurement will fail as absorption will be only
a tiny fraction of a large signal which is
difficult to measure and will result in very low
sensitivities (figure a). Therefore, line sources
with bandwidths narrower than that of the
absorption lines must be used
93(No Transcript)
94- This can be achieved by using a lamp producing
the emission line of the element of interest
where analyte atoms can absorb that line.
Conditions are established to get a narrower
emission line than the absorption line. This can
in fact be achieved by getting an emission line
of interest at the following conditions
95- 1. Low temperatures to decrease Doppler
broadening (which is easily achievable since the
temperature of the source is always much less
than the temperature in flames). - 2. Lower pressures this will decrease pressure
broadening and will thus produce a very narrow
emission line.
96- This may suggest the need for a separate lamp for
each element which is troublesome and
inconvenient. However, recent developments lead
to introduction of multielement lamps. In this
case, the lines from all elements should not
interfere and must be easily resolved by the
monochromator so that, at a specific time, a
single line of one element is leaving the exit
slit
97Hollow Cathode Lamp (HCL)
- This is the most common source in atomic
absorption spectroscopy. It is formed from a
tungsten anode and a cylindrical cathode the
interior surface of which is coated by the metal
of interest. The two electrodes are usually
sealed in a glass tube with a quartz window and
filled with argon at low pressure (1-5 torr).
Ionization of the argon is forced by application
of about 300 V DC where positively charged Ar
heads rapidly towards the negatively charged
cathode causing sputtering. A portion of
sputtered atoms is excited and thus emit photons
as atoms relax to ground state. The cylindrical
shape of the cathode serves to concentrate the
beam in a limited region and enhances
redeposition of sputtered atoms at the hollow
surface.
98(No Transcript)
99(No Transcript)
100- High potentials usually result in high currents
which, in turn, produce more intense radiation.
However, Doppler broadening increases as a
result. In addition, the higher currents will
produce high proportion of unexcited atoms that
will absorb some of the emission beam which is
referred to as self absorption (a lower intensity
at the center of the line is observed in this
case).
101Electrodeless Discharge Lamps (EDL)
- An EDL is a sealed quartz tube containing a few
torr of an inert gas and a small quantity of the
metal of interest. Excitation of the metal is
achieved by a radiofrequency or a microwave
powered coil through ionization of argon, due to
high energetic radiofrequency. Ionized argon will
hit the metal causing excitation of the atoms of
the metal of interest. The output power of the
EDL lamp is higher than the HCL lamp. However,
compared to HCL lamps, EDL lamps are rarely used.
102(No Transcript)
103(No Transcript)
104Emission in Flames
- There can be significant amounts of emission
produced in flames due to presence of flame
constituents (molecular combustible products) and
sometimes impurities in the burner head. This
emitted radiation must be removed for successful
sensitive determinations by AAS, otherwise a
negative error will always be observed. We can
visualize this effect by considering the
schematic below
105(No Transcript)
106- The detector will see the overall signal which is
the power of the transmitted beam (P) in addition
to the power of the emitted radiation from flame
(Pe). Therefore if we are measuring absorbance,
this will result in a negative error as the
detector will measure what it appears as a high
transmittance signal (actually it is P Pe). In
case of emission measurements, there will always
be a positive error since emission from flame is
an additive value to the actual sample emission.
It is therefore obvious that we should get rid of
this interference from emission in flames.
107Source Modulation
- It turned out that excluding the emission signal
from flames can easily be done by an addition of
a chopper to the instrumental design. The chopper
is a motor driven device that has open and solid
(mirrors in some cases) alternating regions as in
the schematic
108(No Transcript)
109- The function of the chopper is to chop the light
leaving the source so that when the incident beam
hits the chopper at the solid surface, the beam
will be blocked and detector will only read the
emitted signal from the flame. As the chopper
rotates and the beam emerges to the detector, the
detector signal will be the sum of the
transmitted signal plus that emitted from the
flame. The signal processor will be able to
subtract the first signal from the second one,
thus excluding the signal from emission in
flames.
110- This can be represented by the following
equations - Signal 1 (Blocked Beam) Pe
- Signal 2 (Transmitted Beam) P Pe
- Overall Difference Signal (P Pe) - Pe P
(Corrected Signal) -
- This correction method for background emission in
flames is called source modulation.
111The schematic of the AAS instrument with source
modulation correction can be represented by the
following schematic
112- It should be recognized that addition of extra
components to an instrument will decrease the
signal to noise ratio and addition of a moving
component is usually regarded as a disadvantage
due to higher need for maintenance. - Another procedure which can overcome the emission
from flames is to use a modulated power supply
that will give fluctuating intensities at some
frequency (say for example pulsed radiation at a
specific frequency).
113- The emission from flames is a continuous signal
but that from the source is modulated. Now if we
use a high pass RC filter, only the fluctuating
signal will be measured as signal while the DC
signal will be considered zero as it can not pass
through the electronic filter. The high pass RC
filter is a device which uses a resistor and a
capacitor the impedance of which is inversely
proportional to the frequency of the modulated
signal. Therefore, only high frequencies will
have low impedance and can pass through the
capacitor while signals of low frequencies will
suffer very high resistance and will not be able
to go through the capacitor.
114AAS Instruments
- Instruments in AAS can be regarded as single or
double beam instruments. -
- Single Beam Atomic Absorption Spectrophotometers
- A single beam instruments is the same as the one
described above (source modulation section) or
generally
115The term spectrophotometer implies that the
instrument uses a dispersive monochromator
(containing a prism or a grating). Also, the
detector is a photomultiplier tube in most cases.
116Double Beam Atomic Absorption Spectrophotometers
- In this type of instruments, the incident beam is
split into two beams of equal intensity by a
chopper with the solid surface being a mirror.
One of the beams will traverse the sample in the
atomizer while the other is considered as a
reference. Detector signals will be consecutive
readings of both the reference and sample beams.
The ratio of the reference to the sample beams is
recorded to give the final signal.
117A schematic representation of a double beam
instrument is shown below
118- It should be emphasized here that in the absence
of sample, Pr is not equal to P since the
reference beam traverses through air while the
other beam traverses through the flame. In
flames, particulates and molecular species
scatter and absorb a portion of incident
radiation, which results in a lower intensity of
the beam. To act as a real double beam, The AA
spectrophotometer reference beam should pass
through a reference flame.
119- But even if we do that, there are no guarantees
that both beams will be of equal intensities
because it is almost impossible to obtain exactly
equivalent flames. It is therefore important to
understand that the excellent features of a
double beam configuration are not achievable in
AAS instrumentation.
120Interferences in Atomic Absorption Spectroscopy
- There are two major classes of interferences
which can be identified in atomic absorption
spectroscopy. The first class is related to
spectral properties of components other than
atomized analyte and is referred to as spectral
interferences. The other class of interferences
is related to the chemical processes occurring in
flames and electrothermal atomizers and their
effects on signal. These are referred to as
chemical interferences and are usually more
important than spectral interferences.
121Spectral Interferences
- 1. Spectral line Interference
- Usually, interferences due to overlapping lines
is rare since atomic lines are very narrow.
However, even in cases of line interference, it
can be simply overcome by choosing to perform the
analysis using another line that has no
interference with other lines. Therefore, line
interference is seldom a problem in atomic
spectroscopy.
1222. Scattering
- Particulates from combustion products and sample
materials scatter radiation that will result in
positive analytical error. The error from
scattering can be corrected for by making a blank
measurement. Scattering phenomenon is most
important when concentrated solutions containing
elements that form refractory oxides (like Ti,
Zr, and W) are present in sample matrix.
123- Metal oxide particles with diameters larger than
the incident wavelength will make scattering a
real problem. In addition, samples containing
organic materials or organic solvents can form
carbonaceous (especially in cases of incomplete
combustion) particles that scatter radiation.
124 3. Broad Band Absorption
- In cases where molecular species from combustion
products or sample matrix are formed in flames or
electrothermal atomizers, a broad band spectrum
will result which will limit the sensitivity of
the technique. It should be indicated here that
spectral interferences by matrix products are not
widely encountered in flame methods. Even if
matrix effects are present in flames, they can be
largely overcome by adjusting various
experimental conditions like fuel/oxidant ratio
or temperature.
125- Another method for overcoming matrix
interferences is to use a much higher
concentration of interferent than that initially
present in sample material, in both sample and
standards (this material is called a radiation
buffer). The contribution from sample matrix will
thus be insignificant. - Spectral interferences due to matrix are severe
in electrothermal methods and must thus be
corrected for.
126Background Correction Methods
- a. The Two Line Correction Method
- In this method, a reference line from the source
(from an impurity in cathode or any emission
line) is selected where this line should have the
following properties - 1. Very close to analyte line
- 2. Not absorbed by analyte
- If such a line exists, since the reference line
is not absorbed by the analyte, its intensity
should remain constant throughout analysis.
127- However, if its intensity decreases, this will be
an indication of absorbance or scattering by
matrix species. The decrease in signal of the
reference line is used to correct for the analyte
line intensity (by subtraction of the absorbance
of the reference from that of the analyte). This
method is very simple but unfortunately it is not
always possible to locate a suitable reference
line.
128b. The Continuum Source Method
- This background correction method is the most
common method although, for reasons to be
discussed shortly, it has major drawbacks and
fails a lot. In this technique, radiation from a
deuterium lamp and a HCL lamp alternately pass
through the graphite tube analyzer. It is
essential to keep the slit width of the
monochromator sufficiently wide in order to pass
a wide bandwidth of the deuterium lamp radiation.
129- In this case, the absorbance by analyte atoms is
negligible and absorbance can be attributed to
molecular species in matrix. The absorbance of
the beam from the deuterium lamp is then
subtracted from the analyte beam (HCL) and thus a
background correction is obtained.
130(No Transcript)
131Problems Associated with Background Correction
Using D2 Lamp
- 1. The very hot medium inside the graphite
tube is inhomogeneous and thus signal is
dependent on the exact path a beam would follow
inside the tube. Therefore, exact alignment of
the D2 and HCL lamps should be made. - 2. The radiant power of the D2 lamp in the
visible is insignificant which precludes the use
of the technique for analysis of analytes in the
visible region. - 3. Addition of an extra lamp and chopper
will decrease the signal to noise ratio.
132c. Background Correction Based on Zeeman
Effect
- Zeeman has observed that when gaseous atoms (but
not molecules) are placed in a strong magnetic
field ( 1 tesla), splitting of electronic energy
levels takes place. The simplest splitting of one
energy level results in three energy levels, one
at a higher energy, another at a lower energy
(two s satellite lines) and the third remains at
the same energy as the level in absence of the
magnetic field (central p line). Furthermore, the
p line has twice the absorbance of a s line and
absorbs polarized light parallel to direction of
the magnetic field while the two s lines absorb
light perpendicular to magnetic field.
133- Light from a HCL lamp will pass through a
rotating polarizer that passes polarized light
parallel to external magnetic field at one cycle
and passes light perpendicular to field in the
other cycle. The idea of background correction
using this method is to allow light to traverse
the sample in the graphite furnace atomizer and
record the signal for both polarizer cycles using
the wavelength at the p line.
134- a. First cycle light parallel to field
the p line of the analyte absorbs in addition to
absorbance by matrix (molecular matrix absorb
both polarized light parallel or perpendicular to
field) - Signal a Ap AMatrix
- b. Second cycle light perpendicular to
field the p line of analyte will not absorb
light perpendicular to field and s lines will
also not affect absorbance at the p line
wavelength. Only matrix will absorb. - Signal b AMatrix
135- The overall signal is the difference of the two
signals Ap - Therefore, excellent background correction is
achieved using the Zeeman effect. This background
correction method results in good correction and
is usually one of the best methods available.
136(No Transcript)
137Chemical Interferences
- These are interferences resulting from chemical
processes occurring in flames and electrothermal
atomizers and affect the absorption signal. To
quantitatively assess the effects of the
different chemical processes occurring in flames,
one should regard the burnt gases as behaving
like a solvent. This is necessary since our
knowledge of gaseous state reaction equilibria is
rather limited. Chemical interferences include
three major processes
138Formation of Compounds of Low Volatility
- Anionic species forming compounds of low
volatility are the most important. The formation
of low volatility species will result in a
negative error or at least will decrease the
sensitivity. For example, the absorption signal
of calcium will be decreased as higher
concentrations of sulfate or phosphate are
introduced. Cations forming combined products
with the analyte will also decrease the signal
obtained for the analyte. For example aluminum
forms a heat stable compound with magnesium.
139Elimination of Low Volatility Compounds
- Addition of a releasing agent cations that can
replace the analyte (preferentially react with
the anion) are called releasing agents. In this
case the analyte is released from the compound of
low volatility and replaced by the releaseing
agent. Lanthanum or strontium are good releasing
agents in the determination of calcium in
presence of phosphate or sulfate. Also, lanthanum
or strontium are good releasing agents in the
determination of magnesium in presence of
aluminum since both can replace magnesium.
140- Addition of a protective agent organic ligands
that form stable volatile species with analytes
are called protective agents. An example is EDTA
and 8-hydroxyquinoline which will form complexes
with calcium even in presence of sulfate and
phosphate or aluminum. - Use of higher temperature is the simplest
procedure to try if it is possible
141Dissociation Equilibria
- Dissociation reactions occur in flames where the
outcome of the process is desired to produce the
atoms of analyte. For example, metal oxides and
hydroxides will dissociate in flames to produce
the atoms as in the equations - MO M O
- M(OH)2 M 2 OH
142- Remember that we are not working in solution to
dissociate the compounds into ionic species. In
fact, not much is known about equilibrium
reactions in flames. It should also be remembered
that alkaline earth oxides and hydroxides are
relatively stable and will definitely show
characteristic broad band spectra (more intense
than line spectra), except at very high
temperatures. The opposite behavior is observed
fro alkali metals oxides and hydroxides which are
instable even at lower flame temperatures and
thus produce line spectra.
143- An equilibrium can be established for the
dissociation of compounds containing atoms other
than oxygen, like NaCl where - NaCl Na Cl
- Now, if the signal from a solution of NaCl was
studied in presence of variable amounts of Cl
(from HCl, as an example), the signal will be
observed to decrease as the concentration of Cl
is increased a behavior predicted by the Le
Chatelier principle in solutions.
144- The same phenomenon is observed when a metal
oxide is analyzed using a fuel rich flame or a
lean flame. Signal will be increased in fuel rich
flames since the dissociation of metal oxides is
easier due to less oxygen while the opposite
takes place in lean flames (oxygen rich).
145- A good example on dissociation equilibria can be
presented for the analysis of vanadium in
presence of aluminum and titanium, fuel rich
flames result in higher absorbance signal for
vanadium since the little oxygen present in
flames will be mainly captured by Al and Ti, thus
more V atoms are available. However, in lean
flames, excess oxygen is present and thus
vanadium will form the oxide and addition of
extra Ti and Al will not affect the signal.
146Ionization Equilibria
- Ionization in fuel/air flames is very limited
due to relatively low temperatures. However, in
fuel/nitrous oxide or fuel/oxygen mixtures,
ionization is significant. Therefore, at higher
temperatures an important portion of atoms can be
converted to ions - M M e
- K Me/M
147- Ionization in flames may explain the decrease in
absorption signal for alkali metals at very high
temperatures where as the temperature is
increased signal will increase till an extent at
some temperature where it starts to decrease as
temperature is further increased a consequence
of ionization. Therefore, usually lower flame
temperatures are used for determination of alkali
metals. A material that is added to samples in
order to produce large number of electrons is
referred to as an ionization suppressor, the
addition of which results in higher
sensitivities.
148Practical Details in AAS
- 1. Sample Preparation
- The most unfortunate requirement of AAS may be
the need for introduction of samples in the
solution form. This necessitates the dissolution
of the sample where in many cases the procedure
is lengthy and requires very good experience.
Care should be particularly taken in order not to
lose any portion of the analyte and to make sure
that the reagents, acids, etc. used in the
dissolution and pretreatment of the sample are
free from analyte impurities.
149- I suggest that you follow exact procedures for
preparation of specific samples for analysis by
AAS. In some cases where the sample can be
introduced directly to an electrothermal atomizer
without pretreatment (like serum samples),
definitely, electrothermal atomizers will have an
obvious advantage over flame methods which
require nebulization.
150Organic Solvents
- Increased nebulization rate due to lower surface
tension of organic solvents which produces
smaller droplets as well as faster evaporation of
solvents in flames will result in better
sensitivities. - Immiscible organic solvents containing organic
ligands are used to extract metal ions of
interest and thus concentrate them in a small
volume (thus increasing sensitivity) and
excluding possible interferences due to matrix
components.
151 Calibration Curves
- The absorbance of a solution is directly
proportional to its concentration but due to the
large number of variables in AAS, usually this
direct relationship may slightly deviate from
linearity. The standard procedure to do is to
construct a relation between the absorbance and
concentration for a series of solutions of
different concentrations. The thus constructed
graph is called a calibration curve.
152- The unknown analyte absorbance is found and the
concentration is calculated or located on the
curve. Neither interpolation nor extrapolation is
permitted to the calibration curve. A sample can
be diluted or the calibration curve may be
extended but always the analyte absorbance should
be within the standard absorbance range recorded.
Usually, the concentration axis has the ppm or
ppb units.
153Standard Addition method
- Chemical and spectral interferences can be
partially or wholly overcome by the use of a
special technique of calibration called the
method of standard addition. In addition, the use
of this method provides better correlations
between standards and sample results due to
constant nebulization rates. The method involves
addition of the same sample volume to a set of
tubes or containers.
154- Variable volumes of a standard are added to the
tube set followed by completion to a specific
volume. Now, all tubes contain the same amount of
sample but different concentrations of analyte. A
plot is then made for the volume of standard and
absorbance. This plot will have an intercept (b)
with the y axis and a slope equals m.
155- The concentration of the analyte can be
determined by the relation - Cx bCs/mVx
- Where, Cx and Vx are concentration and volume of
analyte and Cs is the concentration of standard. - One can only use two points to get the analyte
concentration using the relation - Cx AxCsVs/(At Ax)Vx
156Detection Limits
- Usually, atomic absorption based on
electrothermal atomization has better
sensitivities and detection limits than methods
based on flames. In general, flame methods have
detection limits in the range from 1-20 ppm while
electrothermal methods have detection limits in
the range from 1-20 ppb.
157- This range can significantly change for specific
elements where not all elements have the same
detection limits. For example, detection limits
fro mercury and magnesium using electrothermal
atomization are 100 and 0.02 ppb while the
detection limits for the same elements using
flame methods are 500 and 0.1 ppm, respectively.
158Accuracy
- Flame methods are superior to electrothermal
methods in terms of accuracy. The relative error
in flame method can be less than 1 while that
for electrothermal method occurs in the range
from 5-10. Also, electrothermal methods are more
susceptible to molecular interferences from the
matrix components. Therefore, unless a good
background correction method is used, large
errors can be encountered in electrothermal
methods depending on the nature of sample
analyzed.
159Flame Photometry
- The technique referred to as flame photometry is
a flame emission technique. We introduce it here
because we will not be back to flame methods in
later chapters. The basics of the technique are
extremely simple where a sample is nebulized into
a flame. Atomization occurs due to high flame
temperatures and also excitation of easily
excitable atoms can occur.
160- Emission of excited atoms is proportional to
concentration of analyte. Flame emission is good
for such atoms that do not require high
temperatures for atomization and excitation, like
Na, K, Li, Ca, and Mg. The instrument is very
simple and excludes the need for a source lamp.
The filter is exchangeable in order to determine
the analyte of interest and, in most cases, a
photomultiplier tube is used as the detector.
161(No Transcript)
162(No Transcript)
163b. Charge Transfer and Charge Coupled Transducers
- The photosensitive elements are, in contrary to
PDAs, arranged in two dimensions in both charge
injection devices (CID) and charge-coupled
devices (CCD). Therefore, these are very similar
to photographic films. For example, a
commercially available transducer is formed from
244 rows with each row containing 388 detector
elements. This will add up to a two-dimensional
array holding 16672 detector elements (pixels) on
silicon chip that is 6.5 mm by 8.7 mm.
164(No Transcript)
165(No Transcript)
166- The full description of the system and its
mechanism will not be covered here as this is
behind the scope of this course. However, we
should qualitatively know that these important
transducers function by first collecting the
photogenerated charges in different pixels and
then measuring the quantity of the charge
accumulated in a brief period. Measurement is
accomplished by transferring the charge from a
collection area to a detection area.
167Optical Atomic Spectra
- We have briefly described in an introductory
chapter that atomic spectra are usually line,
rather than band, spectra due to absence of
vibrational and rotational levels. The existence
of quantized electronic energy levels explains
the origin of the observed line spectra and exact
locations of possible lines