Title: Corrections and Adjustments TCA 1800


1
Corrections and AdjustmentsTCA 1800
2
Errors Present in TCA
  • (l,t) Dual Axis Compensator Index Error
  • (i) Vertical Index Error
  • (c) Line of Sight Error
  • (a) Tilting Axis Error
  • (ATR) Target Recognition Axis Error

3
These Errors Can Change Over Time and With
Temperature ...
4
Therefore Should be Redetermined
  • Before the First Use
  • Before Each Precision Survey
  • After Being Transported Over Long Distances
  • After Long Periods of Work
  • Change of Temperature gt 20 degrees C

5
Dual Axis Compensator Errors
  • This Determines the Error in the Longitudinal (l)
    and Transverse (t) Compensators
  • If the instrument is to be Leveled without
    Turning the Alidade, this procedure must be
    conducted periodically

6
Dual Axis Compensator ErrorsPage 2
  • Acclimatize the Instrument
  • Level the Instrument
  • Activate the Calibration Procedure
  • Select ltCALgt from the Main Menu
  • Press ltF1gt (l, t)
  • Follow Prompts

7
Vertical Index Error
  • Zero Point Error of the Vertical Circle in
    relation to the Vertical Axis
  • Setup and Level the Instrument
  • Setup a Target at about 100 m
  • Make sure that the target is within 9 degrees of
    the horizontal plane

8
Vertical Index ErrorPage 2
  • Activate the Calibration Procedure
  • Select ltCALgt from the Main Menu
  • Select ltF2gt (i)
  • The Dual Axis Compensator is automatically turned
    off for the process
  • Follow the Prompts

9
Horizontal Collimation Error
  • This is the Error in the Line of Sight as it
    deviates from the line perpendicular to the
    Tilting Axis
  • Setup and Level the Instrument
  • Setup a Target at about 100 m
  • Make sure that the target is within 9 degrees of
    the horizontal plane

10
Horizontal Collimation ErrorPage 2
  • Activate the Calibration Procedure
  • Select ltCALgt from the Main Menu
  • Select ltF3gt (c)
  • The Dual Axis Compensator is automatically turned
    off for the process
  • Follow the Prompts

11
Tilting Axis Error
  • Deviation of the Tilting Axis from the
    Perpendicular of the Vertical Axis
  • Setup a Target at about 100 m
  • The Target must be positioned at least 27 degrees
    above or below the horizon

12
Tilting Axis ErrorPage 2
  • Activate the Calibration Procedure
  • Select ltCALgt from the Main Menu
  • Select ltF3gt (a)
  • The Dual Axis Compensator is automatically turned
    off for the process
  • Follow the Prompts

13
ATR Collimation Error
To be discussed during the ATR Topic
14