Title: ELECTRON SPECIMEN INTERACTIONS SPECIMEN PREPARATION ELECTRON MICROSCOPES
1ELECTRON -SPECIMEN INTERACTIONSSPECIMEN
PREPARATIONELECTRON MICROSCOPES
2- Electron Target Interactions
- Electron Optical Imaging and chemical analysis
are based upon - Detecting ,
- Collecting, and
- Processing a range of signals that are produced
when an electron beam interacts with either bulk
or foil samples. - Figures 9(a) and (b) show beams of high energy
electrons - interacting with bulk and foil samples
respectively.
3- Figure 9(a). A bulk specimen showing the
spread of the beam and the activated volume
giving rise to (i) backscattered, secondary and
Auger electrons (ii) characteristic X-rays and
(iii) electron current. Typical values given for
the depth of electron penetration and the excited
volume for a 30keV, 100nm diameter electron probe.
4- Figure 9(b). A foil specimen of thickness, t, (t
lt 300nm) where in addition to (a) transmitted
signals give (i) image contrast and electron
diffraction (ii) Kikuchi diffraction and (iii)
plasmon losses.
5- Figure 10 shows the interaction volume as a
function of beam energy (Monte Carlo
calculations). - The penetration of an electron into a solid
depends on the beam energy, which in turn depends
on the accelerating voltage, usually between 20
kV and 1,000 kV.
610 kV
20 kV
730 kV
- Figure 10. Monte Carlo calculations of the
interaction volume in iron as a function of beam
energy (a) 10 kV (b) 20 kV, and 30 kV.
8- SPECIMEN PREPARATION METHODS
- BULK SAMPLES
- Specimens for OPTICAL or SEM examination must be
prepared via a sequence of polishing stages,
except for - (i) Non conducting specimens the specimen
surface must be coated with gold or carbon.
This procedure is not needed if Environmental
Scanning Electron Microscope (ESEM) is used. - (ii) Fracture Evaluation or Failure Analysis -
the specimen must be preserved and the fracture
surface should be covered and padded (if
possible) for protection against further damage.
On no account should the fracture surface be
subjected to any form of polishing.
9- POLISHING STAGES
- (1) Mechanical Grinding This is a form of Rough
Polishing, which involves using different
grades (or grits) of SiC abrasive
papers/belts/cloths, with water as lubricant. A
commonly employed grit sequence uses 120-,
240-, 320-, 400-, and 600- mesh abrasive
papers. The initial grit size depends on the
surface roughness and the depth of damage from
sectioning. Sample orientation must be
continuously changed to avoid comet tailing.
10- (2) Mechanical Polishing After grinding to a
600- grit finish, the sample is polished to
produce a flat, reasonably scratch-free surface
with high reflectivity. (i) Coarse polishing
(30 to 3 ?m finish) - on special cloths smeared
with diamond paste plus some lubricant. (ii)
Fine polishing (? 1 ?m finish) - use diamond
paste plus lubricant or Silica or alumina plus
water lubricant. - (3) Electropolishing or Electrolytic Polishing
This is a process whereby a mechanically
polished sample is made smooth and brighter by
making the sample the anode in an electrolytic
cell. This is only possible when the correct
combination of bath temperature, voltage,
current density and time is employed.
11- (4) Chemical Polishing This procedure can be
used instead of (3). It is a method for
obtaining a polished surface by immersion in or
swabbing with suitable solution without need of
an external electric current. - NOTE
- Specimens for Electron Backscattered Diffraction
(EBSD) analysis or Orientation Imaging Microscopy
(OIM) may require special specimen preparation
methods. This will be discussed latter.
12- THIN FOILS
- Electropolishing window technique disk
method jet thinning - Ion Beam Thinning
- Chemical Polishing
13Figure 11. Typical stages in electropolishing
sheet material using the window technique. The
shaded regions a-d are lacquered.
14Figure 12. (a) Anode current density I vs.
specimen potential E. Curve ABCD shows a
polishing plateau BC whereas curve EF does not.
(b) Schematic diagram illustrating
potentiostatically controlled electropolishing
15Figure 13. Schematic diagram illustrating the
action of an a.c. jet thinning technique due to
Bainbridge and Thorne.
16Figure 14. Schematic diagram of typical ion-beam
thinning equipment
17- Figure 15. Schematic diagrams showing
- (a) the depth of penetration of argon ions as a
function of the angle between the beam and
specimen, and - (b) the variation of thinning rate with angle of
beam incidence
18Other Thin-Foil Preparation Techniques
- Microtomy
- Replicas
- Single-stage replicas 1. Deposit carbon
(100-200A) on surface of interest 2. Remove by
etching the specimen to free the porous film
and floating the film off the specimen 3.
Shadow with a heavy element (gold or silver) - Two-stage replicas
- Extraction replicas - useful for small particle
ID and characterization
19- Artifacts in Thin Foils
- Careless Handling Dislocations, Bend Contours,
Cracks - Careless Washing Surface films/deposits
- Hydrogen Contamination During electropolishing
with acid electrolytes (Ti and Zr alloys) - Interstitial Contamination during
electropolishing or storage (Nb and Ta) - Surface Oxidation even during examination in the
microscope- can effect structure (stacking
faults, extra reflections of Moire patterns if
the oxide is crystalline) or apparent
composition also, surface mottling common
streaks in diffraction pattern
20- Artefacts in Thin Foils Continued
- Electron or ion beam damage in the
microscope--may produce dislocation loops - Ion Beam damage during ion beam thinning
- Heating during ion beam thinning or examination
might cause microstructural changes, e.g.,
crystallization - Uneven polishing--surface etching or mottling
redeposition of precipitates on the surface of
the foil - Dislocations might move and/or annihilate when a
thin foil is produced--especially in materials
with low lattice resistance (Peierls stress)
Likewise, dislocations might dissociate to form
visible stacking faults due to local stresses--
not representative of bulk behavior
21- Artefacts in Thin Foils Continued
- Martensite or other displacive transformations
might form spontaneously in thin regions of the
foil even above Ms (Ti, Zr, and Fe alloys) - Deformation twinning might occur in bent foils
but not in the bulk - Hot stage work might result in precipitates
nucleating on the surface (not representative of
the bulk) - Metastable b.c.c. Ti alloys can transform via
twinning to produce planar features
22SCANNING ELECTRON MICROSCOPY
23- The Instrument-
- SEM is primarily used to study the surface, or
near surface structure of bulk specimens. - The electron gun, condenser lenses and vacuum
system, are similar in SEM and TEM. - A schematic diagram of the electron optical
column for a two lens scanning electron
microscope (SEM), is presented in Figure 16.
24- Figure 16. Schematic diagram of a scanning
electron microscope.
25- The electron gun produces electrons, and
accelerates them at the operating voltage. - The operating voltage ranges from 0 to 30 60 kV
depending upon the type of instrument. - Figure 17 shows a typical configuration of an
electron gun.
26- Figure 17. Gun assembly of a Scanning Electron
Microscope.
27- The microscope is a probe forming system.
- Two or three condenser lenses demagnify the
electron beam until it hits the specimen surface
as a focused spot or probe. - Electron probes of sizes down to 6 nm are
attainable with conventional thermionic emission
sources, although smaller probes 2 nm can be
achieved using field emission sources.
28- The fine beam of electrons is scanned across the
specimen by deflector coils. - The low energy electrons or other radiation
emitted from the surface are detected/collected
and amplified to form a video signal, which
modulates the brightness of a cathode-ray tube
(c.r.t.) display. - The electron beam and the c.r.t. spot are both
scanned in a similar way to a television
receiver, that is in a rectangular set of
straight lines known as raster.
29- Imaging Modes and Information-
- The different signals produced when the electron
beam interacts with the bulk sample in the SEM
are used to create an image. - One of the main features of the SEM is that, in
principle, any radiation emitted from the
specimen or any measurable change in the specimen
can be used to provide the signal to modulate the
c.r.t., and thus provide the contrast. - The various methods used to collect these emitted
electron signals are summarized in Figure 18.
30- Figure 18. Methods of detecting electrons in a
scanning electron microscope (a) secondary
electrons (b) backscattered electrons, solid
state detector (c) backscattered electrons,
scintillation counter, and - (d) absorbed electron current.
31- Table 3. Commonly Used Scanning Electron
Microscopy Imaging Modes Together with Resolution
Attainable
32(a) Topographic Images
- One of the principal uses of SEM is to study the
surface features, or topography of a sample. - Although this form of image may be obtained using
most signals, secondary or backscattered
electrons are usually recommended.
33Secondary Electrons
- Secondary electrons are detected by a
scintillator - photomultiplier system known as
the Everhart-Thorny detector - The number of detected secondary electrons
increase with surface tilt. About 20 to 40
degrees specimen tilt towards the detector may be
necessary. - Topographic images obtained with secondary
electrons look remarkably like images of solid
objects viewed with light. - We find these topographic images easy to
interpret.
34Backscattered Electrons
- Topographical images may be obtained by using
Backscattered electrons, which are detected with
- (i) scintillator - light pipe - photomultiplier
(e.g. Robinson detector) type detectors, known
for their rapid response time, or - (ii) solid state detectors, with a disadvantage
of relatively slow response time.
35(b) Compositional Images
- Backscattered electrons and X-rays from the
specimen is also capable of yielding composition
information. - Backscattered Electrons
- The composition of samples are sensitive to
backscattered coefficient (?), which varies
monotonically with atomic number. - It has been shown that the coefficient can be
expressed as -
13
36- The magnitude of the compositional or atomic
number contrast from two phases of backscattered
coefficients ?1 and ?2 is readily calculated as -
14 - This method is not recommended for phases with
similar atomic number, as the resolution may be
very poor.
37- Quantitative compositional information for phases
of low atomic number can be obtained by comparing
the intensity of the backscattered signal from
the phases with a standard element. Care must be
taken. - Such information is not easily determined using
X-ray method.
38X-rays
- The characteristic X-rays emitted by a specific
element may be identified from either wavelength
or characteristic energy since -
15 - where h is Plancks constant and c is the
velocity of light. - Equation 15 forms the basis of techniques which
use characteristic X-rays for compositional
analysis.
39- The analytical techniques rely on the efficient
detection and discrimination of X-rays. - The detection may essentially be carried out by
any of the following detectors - Wavelength Dispersive (WDX)
- Gas Flow Proportional Counter
- Energy Dispersive Spectrometer (EDX or EDS)
- Electron Energy Loss Spectrometer (EELS), or
40Figure 19(a). Wavelength dispersive crystal
spectrometer Collection of X-rays by
spectrometer generated X-rays have a range of
wavelengths, Dl, but only one is selectively
diffracted to the detector.
41- Figure 19(b). The constant take off angle and
change of crystal position required to retain
focussed condition.
42- Table 4. Crystals Used in Wavelength Dispersive
Spectrometers
43- Figure 20. Gas-flow, argon/methane or xenon,
proportional counter used in conjunction with a
crystal spectrometer.
44- Figure 21. X-rays enter energy dispersive
spectrometer through a thin Be window and
produce electron-hole pairs within the
semiconductor crystal. A typical energy spectrum
obtained from a general area of a ferric
stainless steel is shown.
45(c) Crystallographic Information
- There are three SEM-based techniques for
obtaining crystallographic information from
samples. These are - Kossel technique
- Electron Channeling Diffraction, and
- Electron Backscattered Diffraction technique
46Figure 22. Electron channel pattern The
production of channel patterns
47Figure 23 Electron channel pattern obtained
from a gallium phosphide crystal of (111) of
orientation.