Title: Tekran


1
Tekran Automated Mercury Speciation
  • F.H. Schaedlich1 , Robert K. Stevens2,
  • D.R. Schneeberger2 , Eric Prestbo3,
  • Steve Lindberg4 , Gerald Keeler5

1 Tekran Inc., 1-132 Railside Rd. Toronto, Canada
M3A 1A3, 2 Florida Dept. Environmental
Protection at USEPA, Res Triangle Park, NC
27711, 3 Frontier Geosciences Inc., 414 Pontius
Ave. N., Suite B, Seattle, WA 98109 4 Oak Ridge
National Lab. PO Box 2008, Oak Ridge, TN
37831-6038 5 University of Michigan, School of
Public Health, 109 Observatory Street, MI 48109.
2
Objectives
  • Development of automated methods to measure and
    differentiate between
  • Reactive Gaseous Mercury (RGM) in ambient air
  • Total Particulate Mercury (TPM) in ambient air
  • Elemental mercury (Hg0) in ambient air

3
Ambient Air Mercury Speciation
  • Different forms of gaseous Hg have very different
    behaviors.
  • Forms can interconvert in the atmosphere and in
    various reservoirs.
  • Classes of mercury in ambient air
  • Elemental mercury Hg0.
  • Reactive mercury HgII, RGM, Hg2.
  • Particulate mercury HgP, TPM.

4
Model 2537A Mercury Analyzer
  • Provides continuous total gaseous (TGM) readings
    with update rate as low as 2.5 minutes.
  • Detection limit lt 0.1 ng/m3 (5 min. samples).
  • Automatic recalibration with internal Hg0
    permeation source.
  • Capable of unattended operation for extended
    periods.

5
Model 2537A Analyzer
6
Flow Diagram of Model 2537A Mercury Analyzer
7
Elemental Mercury Hg0
  • Typically 90 of atmospheric Hg loadings.
  • Relatively inert. Long residence time.
  • Hg0 sources impact large areas.
  • Sources chlor-alkali plants, gold and Hg mining,
    thermal power plants.

8
Part 1Reactive Gaseous Mercury Monitoring
9
Why Bother Measuring RGM ?
  • RGM (as HgCl2) is emitted from power plants and
    incinerators.
  • RGM is removed rapidly via wet and dry
    deposition.
  • Often responsible for contamination of sensitive
    nearby ecosystems. (Florida Everglades, Arctic).

10
Reactive Mercury Hg(II) or RGM
  • Consists of reactive, water soluble forms,
    primarily HgCl2.
  • Usually only a few percent of total gaseous
    mercury present in the atmosphere.
  • Short range deposits relatively close to the
    source of emission.
  • Primary sources Coal burning power plants, waste
    incinerators. (Chlor-alkali plants?)

11
Difficulties Measuring RGM
  • Method must be 1-2 orders of magnitude more
    sensitive than total mercury methods.
  • Reject much larger elemental component.
  • Exclude particulate bound mercury, however,
    particulate filters have problems passing RGM.
  • Apparatus must pass RGM to the collector
    quantitatively.

12
Model 1130 - Principles of Operation
  • Quartz, specially coated annular denuder is
    thermally desorbed and regenerated.
  • Sampling Phase.
  • Absorbs RGM while passing all elemental Hg.
  • Model 2537A reads Hg0 during this phase.
  • Desorption/Analysis Phase.
  • Zero air used as carrier.
  • RGM released as elemental mercury.

13
RGM Sampling Phase
  • Denuder is slightly warm.
  • Inlet slightly warm.
  • Denuder captures all RGM while passing Hg0.
  • Total sample flow is 10 lpm.
  • Analyzer measures Hg0 during this time.

14
Analysis Phase
  • Denuder is heated to 500C.
  • Inlet heated to clean upstream glassware.
  • Denuder releases captured RGM as Hg0.
  • Excess zero air flow cleans upstream components.
  • Instrument measures accumulated RGM as Hg0.

15
Lab RGM Apparatus
16
Quartz Denuder Assembly
17
Model 1130Denuder Pump Module
18
Preliminary Results - Toronto
19
Preliminary Results - Toronto
20
Lab Testing Frontier Geosciences
21
Seattle RGM Data
Ambient Air - Downtown Seattle
Courtesy Frontier Geosciences
22
Mercury Levels - Rural Michigan
Courtesy UMAQL
23
Summary - RGM
  • A quartz annular denuder coated with KCl
    quantitatively collects HgCl2 but does not retain
    Hg0.
  • HgCl2 collected on a denuder is thermally
    desorbed as Hg0.
  • The automated instrument that uses this denuder
    technology measures RGM concentrations down to 1
    pg/m3.
  • Denuder lasts 1 - 2 weeks between recoatings.

24
Part 2 Particulate Mercury Monitoring
25
Why Bother Measuring TPM ?
  • Particulate bound mercury (TPM / HgP) is created
    directly by some industrial processes and also
    formed downwind of sources by the combination of
    mercury with existing particulates.
  • TPM is removed rapidly via wet and dry deposition.

26
Particulate Mercury TPM
  • Consists of various compounds of mercury bound
    onto particles.
  • Size range Generally lt 3 um.
  • Usually only a few percent of total mercury
    present in the atmosphere.
  • Short range deposits relatively close to the
    source of emission.

27
Difficulties Measuring TPM
  • Method must be 1-2 orders of magnitude more
    sensitive than total mercury methods.
  • Must reject much larger elemental component.
  • Method must exclude reactive gaseous mercury,
    however, normal particulate filters have problems
    in that they retain RGM.

28
Model 1135 - Principles of Operation
  • Sampling Phase
  • Quartz, regenerable particulate filter (RPF)
    captures fine fraction (lt 2.5 um) particulates
    after air has passed through impactor and denuder
  • Desorption/Analysis Phase
  • Zero air used as carrier
  • Downstream pyrolyser is activated first
  • RPF is heated to desorb captured particulates

29
Model 1135 Particulate Mercury Monitor
  • Unit stacks on top of Model 1130
  • Allows simultaneous deter-mination of Hg0, RGM
    and HgP
  • Detection limit 1 pg/m3

30
Model 1130 Model 1135 Flow Path
31
Ambient Air Speciation
32
Conclusions - HgP
  • A quartz RPF quantitatively collects particulate
    mercury but does not retain Hg0
  • HgP collected on a denuder is thermally desorbed
    and converted to Hg0 in the pyrolyzer
  • Carbon based particulate matter on the filter is
    oxidized to CO2 during desorption cycle
  • Filter is usable for 1 2 weeks before requiring
    restuffing