UTMB - PowerPoint PPT Presentation

1 / 17
About This Presentation
Title:

UTMB

Description:

ADAPTIVE PROBE OFF DETECTION TECHNOLOGY (APOD) ... APOD DELIVERS ENHANCED PROTECTION AGAINST ERRONEOUS PULSE RATE AND ARTERIAL ... – PowerPoint PPT presentation

Number of Views:93
Avg rating:3.0/5.0
Slides: 18
Provided by: Mich930
Category:
Tags: utmb | apod

less

Transcript and Presenter's Notes

Title: UTMB


1
UTMB
  • MICHAEL MENSAH
  • PHYSIOLOGIC MONITORING
  • RESPIRATORY CARE

2
MASIMO
TECHNOLOGY
3
MASIMO TECHNOLOGY
  • The Signal Extraction Technology
  • Masimo SET pulse oximetry is a method of
    gathering, processing, and reporting arterial
    oxygen saturation and pulse rate with ability to
    acquire accurate readings in intensive noise/
    motion environments.

4
MASIMO TECHNOLOGY
  • Signal Extraction Technology
  • Masimo SET pulse oximeters technology
    eliminates the problems of motion artifact, low
    peripheral perfusion and most low signal-to-noise
    situations. Its performance has been proven
    clinically and has been valid for several years
    now.

5
MASIMO TECHNOLOGY
  • SIGNAL EXTRACTION TECHNOLOGY
  • 2001 GRAPH

6
MASIMO TECHNOLGY
  • SIGNAL EXTRACTION TECHNOLOGY
  • 2002 GRAPH

7
MASIMO TECHNOLOGY
  • Discrete Saturation Transform Technology (DST)
  • Collect block of red and infrared signals.
  • Calculate the reference signal for a trail
    saturation value ( starting, for example at 0)

8
MASIMO TECHNOLOGY
9
MASIMO TECHNOLOGY
  • Discrete Saturation Transform Technology (DST)
  • Send the infrared signal in the reference signal
    to the adaptive filter.
  • The adaptive removes frequency content in common
    between the IR and reference signal.

10
MASIMO TECHNOLOGY
  • Discrete Saturation Transform Technology (DST)
  • Repeat the process for the full range of
    saturation values ( 0 100)
  • Identify DST power peak which corresponds to
    SPO2.

11
MASIMO TECHNOLOGY
  • ADAPTIVE PROBE OFF DETECTION TECHNOLOGY (APOD)
  • APO IS A SET OF COMPLEX AND POWERFUL SIGNAL
    PROCESSING ALGORITHMS THAT CAREFULLY ANALYZE THE
    INCOMING SIGNAL TO DETERMINE IF THE PULSE
    OXIMETER IS ON OR OFF THE PATIENT.

12
MASIMO TECHNOLOGY
  • ADAPTIVE PROBE OFF DETECTION TECHNOLOGY (APOD)
  • APOD DELIVERS ENHANCED PROTECTION AGAINST
    ERRONEOUS PULSE RATE AND ARTERIAL OXYGEN
    SATURATION READINGS WHEN A SENSOR BECOMES
    DETACHED FROM THE PATIENT.

13
MASIMO TECHNOLOGY
  • NOISE REDUCTION SENSOR TECHNOLOGY
  • MASIMO SET LNOP OXIMETRY SENSORS HAVE SEVERAL
    UNIQUE DESIGN FEATURES.

14
MASIMO TECHNOLOGY
15
MASIMO TECHNOLOGY
  • NOISE REDUCTION SENSOR TECHNOLOGY
  • THE RECESSED PHOTODETECTOR CUP, PRESENT AN ALL
    SIGNAL PATIENT USE LNOP SENSORS, IS DESIGNED TO
    ENHANCE PERFORMANCE BY MINIMIZING DIRECT
    INTERFERENCE FROM AMBIANT LIGHT AND STABILIZING
    THE DETECTOR AREA ON THE SKIN SURFACE.

16
MASIMO TECHNOLOGY
  • NOISE REDUCTION SENSOR TECHNOLOGY
  • Additionally, the single patient-use sensors are
    fabricated of a durable substrate which remains
    clinically viable for substantially longer than
    the competitive sensors.

17
MASIMO TECHNOLOGY
  • References
  • From the Masimo Corporation, 2852 Kelvin Avenue,
    Irvine CA. page 514
  • Department of Anesthesiology, University of
    Arizona School of Medicine, 1501 N. Campbell/ PO
    Box 245114, Tuscan AZ 85724
  • Address Correspondence to Julian M. Goldman, MD,
    Masimo Corporation, 2852 Kelvin Avenue, Irvine CA
    92614, USA
  • Additional Information about Masimo is available
    from www.masimo.com
  • Journal of Clinical Monitoring and Computing
    16475-483, 2000. l Kluwer Academic Publishers.
    Printed in the Netherlands.
Write a Comment