Title: Prsentation PowerPoint
1Zendropper
CYTOGENETIC AUTOMATION Automated cell
distributor on standard microscopic glass side
for karyotyping and hybridization in situ
2Why the Zendropper?
- Automatisation of a very sensitive step
- High reproducibility of distribution from one
patient to another and from one slide to another - No inter-technician variations
- Homogeneisation and reliability of distribution
- Samples sparing
- Time saving and rapid execution
-
3Why the Zendropper?
- Optimization of laboratory tasks
- System using a software allowing traceability of
cells dispersion (bar-coded slides session
printing) - Easier laboratory accreditation
- Better valorization of lab technician work
- Environmental factors under control ( T,
humidity, local aeration) with Water Layer
methodology. -
4Zendropper methodologies
Cells preparation
Distribution on slide Hybridization in situ
Distribution on slide Caryotype
Double cells spot on same slide
Water Layer Methodology
51. Karyotype procedure
- High automatic distribution capacity
- Homogeneous cells distribution
- 8 patients with 1 to 5 slides per patient (40
slides on 1 rack) - Timing (dispensing safety 100 )
- /- 25 for 40 slides
- ? Significant time saving
- Protection of technicians not exposed to
methanol and acetic acid emanation from Carroy
solution.
Distribution on slide ZenDropper
6Karyotype Water Layer technology
1
4
Distribution of a water layer on each slide
Ejection of sample on the slide
2
5
Needle containing sample is placed just above
the slide
Aspiration of residual water on slide
6
3
Creation of a central crater in the water layer
due to Carroy solution evaporation
Slide ready for Banding step
7Hybridation in situ
Environment preparation T , Hygrometry
1
Distribution of cell spot I
2
Distribution of cell spot II
3
Slide ready for next step
8Advantages of water layer technology
Karyotype Water Layer technology
- Improved mitosis thanks to contact water acetic
acid/methanol mix on slide surface - ? Chromosomes more spread
- ? Banding (R or G) with better quality
- Creation of a micro-environnement ? The
distribution will always be done with same
hygrometric conditions independently of climatic
conditions
9Zendropper process
10Zendropper process
11Zendropper process validation
Comparative study Manual gtlt Zendropper CERBA
Pasteur - Paris (Genetics departement - More
than 32.000 samples with Zendropper since March
2007)
Contamination Protocol Cells pellets of 240
patients have been distributed on routine
microscopic slide alterning male/female
patient Results No chromosome Y observed on
female samples and confirmed by Quinacrine test
(100 mitosis per patient) Conclusion No
contamination between patients
12Zendropper process validation
- Quality of mitosis
- Less crossing of chromosomes with Zendropper
- Quantity of mitosis
- More reproducible and homogenous dispersion of
cells in mitosis with Zendropper - Quality of Banding
- High level of homogeneity from slide to slide
with Zendropper - Reading
- Easier with Zendropper due to quality of
dispersion - Very efficient particularly for metaphasis
research - Lab technicians reactions
- Easier and fast method than manual
- Less exposition to toxic vapours
- No more difficult work due to work position on
bench or with a glove box.
13Technical Characteristics
- Dimensions 60x60x80 cm
- Weight 62 kg
- Rack with slides 8 patient X max 5 slides
- Tube holder for samples 8
- Extra washing needle solution 2 positions
- 2 needles
- Needle 1 - Mixing, aspiration and distribution
of samples distribution of water - Needle 2 Aspiration of residual water
14Technical characteristics
Zendropper system inclus
- Automate
- PC windows Xp
- Software for operations
- Software for setting up options of user and
instrument calibrations (Ztoolbox) - Printer
- 1 rack of 40 slides
- Option Bar code reader
15Zendropper Software
- Zendropper software Session preparation number
of samples, number of slide/patient (codebar
available) , print of session with information
for tracability - Ztoolbox Technical tool for maintenance and
instrument set up in order to obtain high
homogeneous cells dispersion conditions (Humidity
and T)
16Zendropper maintenance
- Intensive rinsing procedure at start and end of
the day - Regular replacement of needles and tubing due to
intense corrosion of solution
Distribuito in ITALIA da Li StarFish S.r.l. Via
Cavour, 35 - 20063 Cernusco S/N (MI) Tel
02.92150794 - Fax 02.92157285 info_at_listarfish.it
- www.listarfish.it