Title: Ultrasound Physics
1Ultrasound Physics
Our Hospital Physics Group
George David, M.S. Associate Professor of
Radiology
2Nothing to do with Anything
3Artifacts
- Assumptions can cause artifacts when assumed
conditions are not true - sound travels at 1540 m/s
- sound travels in a straight line
- All sound attenuation exactly0.5 dB/cm/MHz
4Distance from Transducer
- Calculation of Distance
- scanner accurately measures time delay between
sound generation echo reception
Distance Assumed Speed X Measured Delay / 2
Actual Distance to interface
58 usec
1380 m/s X 58usec / 2 4 cm
4 cm
Calculated Distance to interface
V 1380 m/s
1540 m/s X 58usec / 2 4.47 cm
(Average speed of sound in soft tissue)
5Distance from Transducer
- Echo positioning on image
- distance from transducer calculated from assumed
speed of sound - can place reflector too close to or too far from
transducer - can alter size or shape of reflector
V 1380 m/s
X
Actual Object Position
Position of Object on Image
X
V 1540 m/s
6Attenuation
- For all scanning your scanner assumes
- soft tissue attenuation
- .5 dB/cm per MHz
- Your scanners action
- compensate for assumedattenuation
- allow operator fine tuning
- TGC
7Shadowing
- Clinical Manifestation
- reduction in imaged reflector amplitude
- Cause
- object between this reflector transducer
attenuates ultrasound more than assumed - assumed compensation not enough to provide proper
signal amplitude - intensity under-compensated
- Opposite of Enhancement
Attenuates more than .5 dB/cm/MHz
Shadowed Reflector
8Shadowing
Attenuates more than .5 dB/cm/MHz
Shadowed Reflector
http//raddi.uah.ualberta.ca/hennig/teach/cases/a
rtifact/noframe/imag2-f2.htm
9Enhancement
- Clinical Manifestation
- increase in imaged reflector amplitude
- Cause
- object between reflector transducer attenuates
ultrasound less than assumed - assumed compensation more than needed to provide
proper signal amplitude - intensity over-compensated
- Opposite of Shadowing
Attenuates less .5 dB/cm/MHz
Enhanced reflector
10Enhancement
Attenuates less .5 dB/cm/MHz
Enhanced reflector
http//raddi.uah.ualberta.ca/hennig/teach/cases/a
rtifact/noframe/imag6-f1.htm
11Shadowing / Enhancing
- these artifacts not necessarily bad
- can help in determining nature of masses
upstream of artifact which caused shadowing /
enhancing
12Scanner Assumptions
- Echo positioning on image
- direction of all sound travel assumed to be
direction that sound was transmitted
X
Refraction
13Refraction Artifact
- refraction alters beam direction
- direction of sound travel assumed to be direction
sound transmitted
X
Refraction
14Refraction Artifact
- refraction alters beam direction
- scanner places dot in wrong location along line
of assumed beam direction - can alter reflector shape
15Lobe Artifacts
- Side Lobes
- beams propagating from a single element
transducer in directions different from primary
beam - reflections from objects here will be placed on
main sound transmission line - Grating Lobes
- same as above except for transducer arrays
X
16Range Ambiguity
- Reflection from 1st pulse reaches transducer
after 2nd pulse emitted - scanner assumes this is reflection from 2nd pulse
- places echo too close in wrong direction
1
2
17Range Ambiguity
- To improve any 1 of 3, at least 1 of other 2 must
be reduced. - many scanners automatically reduce frame rate as
depth increases
Depth
RangeAmbiguityTrade-off
Lines / Frame
Frames / sec(dynamics)
18Scanner Assumptions
Multipath Artifact
Actual Object Position
Position of Object on Image
X
X
19Multiple Reflection Scenario
- reflection from reflector B splits at A
- some intensity re-reflected toward B
- Result
- later false echoes heard
- scanner places dots behind reflector B
2
3
1
A
B
real
1
2
false
3
20Artifacts
- Reverberation (multiple echo) artifact
- comet tail effect is 1 example
- can have dozens of multiple reflections between
- transducer reflector
- 2 reflectors
- Mirror Image
- common around diaphragm pleura
21Artifacts
http//raddi.uah.ualberta.ca/hennig/teach/cases/a
rtifact/noframe/imag1-f1.htm
Caused by Shotgun Pellets
22Multiple Reflection Scenario
http//raddi.uah.ualberta.ca/hennig/teach/cases/a
rtifact/noframe/imag5-f2.htm
23Resolution Artifacts
- Axial and Lateral Resolution Limitations
- results in failure to resolve 2 adjacent
structures as separate - minimum image size equal to resolution in each
direction
24Section Thickness Artifact
- anatomy may not be uniform over its thickness
- universal problem of imaging 3D anatomy
- in CT MRI this is known as partial volume effect
Thickness
25Constructive Interference
- 2 echoes received at same time
- in phase
- Result
- higher intensity
26Destructive Interference
- 2 echoes received at same time
- Exactly 180o out of phase
- Result
- flat (zero) wave
-
27Acoustic Speckle
- texture seen on image may not correspond to
tissue texture - Results from interference effects between
multiple reflectors received simultaneously which
can - add together
- constructive interference
- subtract from one another
- destructive interference
28Mirror Image Doppler
- Analogous to mirror image artifact discussed
previously - mirrored structures can include mirrored vessel
- duplicate image visible on opposite side of
strong reflector - example bone
- Doppler data also duplicated
- flow spectrum copied from original vessel
29Spectral Duplication
- mirror image of Doppler spectrum appears on
opposite side of baseline - causes
- electronic duplication caused by receiver gain
set too high - overloads receiver
- True sensing caused by too large Doppler angle
- beam covers flow in both directions
Blood flows toward transducer
Blood flows away from transducer
30Doppler Artifacts
- Doppler spectrum speckle
- Cause
- same as acoustic speckle
- random constructive destructive interference
from sound scattered in blood
31Aliasing
- Results in detection of improper flow direction
- occurs because sampling rate too slow
- Similar to wagon wheels rotating backwards in
movies
32Aliasing
Sufficient Sampling
Insufficient Sampling
33Aliasing
- Which way is this shape turning?
OR
1
2
3
34Aliasing
- Did the shape turn 1/4 turn right or 3/4 turn
left? - 1 1/4 turn right?
1
2
3
35Aliasing
- Does it help to sample more often?
2
1
1A
2A
3A
3
36Aliasing
- Maximum detectable Doppler shift equals half the
pulse repetition frequency - Sampling rate
- Same as pulse repetition frequency
- Must be at least twice highest frequency to be
sensed - Aliasing occurs when Doppler shift exceeds 0.5
PRF
37Aliasing
- Maximum detectable Doppler Shift not limited for
continuous wave Doppler - Maximum detectable Doppler Shift is limited for
pulsed instruments
Maximum Detectable Doppler Shift half pulse
repetition frequency
38Coping with Aliasing
- decrease transducer frequency
- reduces Doppler shift
- shift proportional to operating frequency
- increase pulse repetition frequency
- decreases maximum imaging depth
- increases likelihood of range ambiguity for
pulsed instruments
77 X fD (kHz) v (cm/s)
-------------------------- fo
(MHz) X cosq
39Coping with Aliasing
- increase Doppler angle
- Reduces relative flow rate between blood
transducer - Reduces Doppler shift sensed by scanner
77 X fD (kHz) v (cm/s)
-------------------------- fo
(MHz) X cosq
40Coping with AliasingBaseline Shifting
- operator instructs scanner to assume that
aliasing is occurring - scanner does calculations based on operators
assumption - scanner has no way of determining where in image
aliasing occurs
Yes
No
41Any Questions?