Title: 1. dia
1Chromatographs
column
eluent tank
injector
pump
detector
PC
CHROMATOGRAM
- Qualitative
-
- Quantitative information
2GasChromatography (GC) 1952 A.T. James A.J.P.
Martin
- High performancy
- Qualitative Quantitative information
- Complicated samples
- Separation
- 1956 van Deemter kinetic theory
- M. Golay capillary columns
MOBILE PHASE GAS STATIONARY PHASE solid or
liquid on solid support (GSC, GLC)
COLUMN ELUTION TECHNIQUE
- Base of separation
- Boiling point (vaporization)
- Structure
GASCHROMATOGRAPHY analysis in vapor phase
12 billion organic compounds 50 000
evaporative without destruction
- Evaporization depends on
- Molar mass
- polarity
Thermal stability
3GASCHROMATOGRAPHY (GC)
Sample introduction to the mobile
phase gas/vapor
- Sample can be
- gas
- liquid vaporization
- solid dissolution in liquid
Pressure and flow regulators
Gas tank
Gas cleaner
Gaschromatograph (GC)
4GASCHROMATOGRAPH (GC)
Gas tank
5(No Transcript)
6Eluent gas
- Depending on the type of detector
- H2
- Ar
- N2
- He
Name sign ppm
pure 2.5 99,5 5000
pure 3.0 99.9 1000
Very pure 3.5 99,95 500
Very pure 4.0 99,99 100
Very pure 4.5 99,995 50
Very pure 5.0 99,999 10
Very pure 6.0 99,9999 1
Ultra pure 7.0 99,99999 0,1
Reductor valve Type depends on the quality and
pressure of the gas
Flow-rates
Inside apparatus Pressure and flow controllers
7Sample introduction
- Injection in a very short time
- Vapor/gas phase
- Mixible with eluent gas
volume 0,1 ?l-1 ml
Liquid vaporization 100-10000 X volume increase
Syringe For gas liquid sample
six-port valve
8FLASH INJECTOR
- Samle introduction
- Vaporization
- Inlet to column
Packed columns greater diameter greater sample
volume Capillary columns small sample volume
9Flash injector
vaporization
Injection
- Sample vaporization
- Liquids 100 1000 X volume increase
- Mixing with eluent
- Stick needle into the septum
- Push the syringe piston
- Remove syringe
10Quick injection
Slow injection
Eluent gas moves the sample to the column.
11- SPLIT
- SPLITLESS
- ON-COLUMN
- PTV
Type of injectors
Split-injector
Carrier gas
Septum wash
split-gas
Split/splitless ratio determines amount of
sample moving to the column
122001
51
13Splitless injector
Purge Off
Purge On
14On-column
PTV (Programmed Temperature Vaporizer)
Injection directly to the column
15Columns
Capillary
polyimid, 350 oC
quarz
d
Stationary phase
microbore d lt 150 ?m standard capillary 150 ?m
lt d lt 500 ?m widebore d gt 500 ?m
Adsorption mechanism PLOT (Porous Layer Open
Tubular)
Distribution mechanism WCOT Wall Coated
OT SCOT Support Coated OT
16Interaction between stationary phase and sample
Active side silanol groups
- tailing
- Non-symmetric peaks
desactivation sylil reagents
Quartz surface
17Stationary phases I.
- Thermal stability
- No bleeding
- Known chemical structure
- Chemical inertnees
- Low price
Adsorbents (GSC) porous, with large special suface
modified adszorbents Based on carbon or silicagel
- inorganic adsorbents
- silicagel
- aluminium-oxide
- zeolits (molekulasziták)
- Organic adsorbents
- active carbon
- polymers
Analytes Hydrocarbons with small molar mass, He,
Ne, Ar, Kr, Xe
(PLOT)
18Stationary phases II. (GLC)
(absorption dissolution of gas and liquids in
liquids)
Polymers WCOT polymers on the surface of
capillary)
Relative small number 12-15
substituted polysiloxans (silicons) long lifetime
R substituents on polysiloxans
Thermal stability up 250-300 ?C
Methyl -CH3
Phenyl
- Substituents
- Methyl
- phenyl
- Cianopropyl
- Trifluoropropyl
Cianopropyl -CH2CH2CH2CN
Trifluoropropyl -CH2CH2CF3
19- methyl-phenyl
- cianopropyl-phenyl
- etc.
substitution how much of Si atoms
100 metil 5 fenil 95 metil
20Polyethyleneglycols (PEG)
Special separation
- Disadvantage
- Lower thermal stability
- oxygen-sensitivity
Carbowax
21- Polarity of stationary phase
- Structure of stationary phase
- Quality of functional groups
- Number of functional groups
- Apolar stationary phases
- 100 methyl
- 5 phenyl
- Polar phases
- cyanopropyl
- PEG
- Midium polar phases
- 35 phenyl
- 50 phenyl
Selectivity depends on the interaction between
stationary phase and analyte
- Interactions depend on
- Quality of analytes
- Structure of stationary phase
22Thermostate
- Type of working
- Izotherm
- Programmable heating
T (oC)
thermostate
- Large temperature range -50 400 ?C
- Programmable heating 0- 40 oC/min
- cooling
t (min)
Decrease of analysis time Good peak shape
23Detectors
Quantitative analysis signal of detector is
proportional with concentration of analytes in
detector
universal signal for every compounds selective
signal for a groups of compounds specific
signal for special compounds
destruktiv non destruktiv
Dinamic range change of concentration results a
change in signal linearity T mc (deviation lt 5
) sensitivity m (ratio of signal/concentration)
Limit of detection (LOD) signal to noise
ratio 3 Limit of quantitation (LOQ) signal to
noise ratio 10
24Detectors
Thermal Conductivity Detector (katharometer)
Change of impedance
Wheatstone-bridge
W-filaments 100-200 mA heating current
Carrier gas H2, He N2
dinamic range 105 LOD 5-50 ng
non destructív universal
25Flame-ionization detector (FID)
hydrogen/air microburner with a pair of electrodes
Carrier gas non ionizable gas N2, Ar, He, H2
Organic compounds leaving the column are burning
in burner jet, ions are forming
Ions result a small current
Carbon-detector it is good for organics, except
formic acid
destructív
Dinamic range 105-106 LOD 0,05-0,5 ng
26 High Performance Liquid Chromatography (HPLC )
Mobile phase liquid Stationary phase adsorbent
(LSC) or liquid on a support(LLC)
Column Elution technique
Sample liquid
pump
Gas removal
eluent tank
injector
PC
detector
27HPLC
Gas removal
pump
automated injector
detector
(thermostate)
28Eluent
- Should (have) be
- Low viscosity
- inert no reaction with analytes
- Chemical stability
- No corrosion
- No toxycity
- Higher boiling point
- Low price
- Good quality and purity
- Compatible with detector
- UV-absortion low
purity HPLC grade Water and buffers too !!!
29Eluent
Analytes distributed between stationary and
mobile phase interaction of analytes with both
phases
Polarity of molecule mobile phase stationary
phase
- Change of polarity
- Change of quality of mobile phase
- Mixing of solvents
hexane chloroform tetrahidrofuran acetonitrile iso
propanol ethanol methanol water
Mixed solvents should be mixcible
Eluent strength determined on silicagel on the
bease of heat of adsorption of solvents
izoeluotrope mixture eluent strength is the
same k, Rs may change !!!
Izocratic elution fixed mobile phase
composition Gradient elution eluent strength is
increasing in time
Use of buffers adjusting of pH in the case of
analysis of ionisable components
30Pumps To carry of eluent
- Should be
- pressure (400 bar)
- Stable flow-rate
- Compatible with different solventsno corrosion
- Small hold-up volume
- No pulsation
Flow-rates in classical analytical HPLC 0,1-1,5
ml/min (0-5 ml/min)
Syringe-type pump
31 Reciprocating pump
Pulsation double pistons (phase-deviation)
Volume 10-100 ?l Change of flow rate easy
32Gas removals Liquids contain dissolved gases
- Effect of gas bubbles
- In pump
- Pressure pulsation
- Different flow-rates
- Mechanical instability
- In detector
- Increased noise (retention time changes)
Remove of gas from the solvent
- Ultrasound
- cheap
- Non effective
- Vacuum
- Higher price
- effective
- He-purge
- Higher price
- effective
33Sample loading
- Quick
- Sample should be mixable with eluent
Sample volume 10-50 ?l
Micro syringe
Six-port valve
34Columns
Function separation
Liquid chromatography NP LC Normal Phase RP
LC Reversed Phase
NPLC polarity of stationary phase gt polarity of
mobile phase RPLC polarity of stationary phase
lt polarity of mobile phase
- Material of column
- Stainless steel
- Glass
- PEEK (poly(ether-ether-ketone)
Packing regular spherical
- Size of column
- diameter 2-5 mm
- length 5-25 cm
35Modified silica gel
Modifying groups C18 octadecyl C18H37 C8
octyl C8H17 C4 buthyl C4H9 Amino
CH2CH2CH2NH2 Ciano CH2CH2CH2CN Phenyl C6H5
RPLC C18 stationary phase methanol/water
mobile phase NPLC silicagel stationary phase
hexane/alcohol mobile phase
Guard column avoid contamination of analytical
column
36Detectors
Quantitative analysis signal of detector is
proportional with concentration of analytes in
detector
universal signal for every compounds selective
signal for a groups of compounds specific
signal for special compounds
destruktiv non destruktiv
Dinamic range change of concentration results a
change in signal linearity T mc (deviation lt 5
) sensitivity m (ratio of signal/concentration)
Limit of detection (LOD) signal to noise
ratio 3 Limit of quantitation (LOQ) signal to
noise ratio 10
37UV-Vis spectrophotometer Application UV-Vis range
Lambeert-Beer A? e? c l
A lg I0/I
Cuvette quartz l5-10 mm
Light source UV deuterium lamp Vis wolfram lamp
Detector fotodiode
Most usable HPLC detector 190 nm lt ? lt 800 nm
38Dioda Array Detector (DAD)
polychromator
lence
Light source
cuvette
Diode array
Advantage Spectra and chromatogram at the same
time
39Paper and thin-layer chromatography
Planar arrangement
Stationary phase paper silica gel or
aluminium-oxide on a glass plate
- Evaluation of chromatogram
- Dropping liquid sample on the one edge of the
plate with capillary - Evaporation (drying) the solvent
- Place the plate to the closed container saturated
with vapors of developing solvent - Running of analytes based on capillary activity
- After development of chromatogram, remove plate
from container and dry it - Locating analytes on the plate spraying with
chemical reagents, like iodine, - sulfuric acid or UV-light
Selection of mobile phaselike in Normal Phase
HPLC
Qualitative data retardation factor
(Rf) Quantitative data intensity of spots
40(No Transcript)