Title: Gas Chromatography
1Gas Chromatography Acetates
- Gas Chromatography, Refractive Index
Distillation - The next two (2) experiments introduce Gas
Chromatography and Simple Fractional
Distillation. - They are then tied together along with the
Refractive Index technique in a third experiment.
- This Week
- Gas Chromatography Acetates
- Pavia p. 817 836
- Slayden p. 39 - 31
- 2nd Week
- Distillation of a Mixture
- Slayden p. 43 - 46
- 3rd Week
- Gas Chromatography and Refractive Index of
Distillates from Distillation of Mixture
Experiment - Slayden p. 47
2Gas Chromatography Acetates
- Gas Chromatography
- Uses
- Separation and analysis of organic compounds
- Testing purity of compounds
- Determine relative amounts of components in
mixture - Compound identification
- Isolation of pure compounds (microscale work)
- Similar to column chromatography, but differs in
3 ways - Partitioning process carried out between Moving
Gas Phase and Stationary Liquid Phase - Temperature of gas can be controlled
- Concentration of compound in gas phase is a
function of the vapor pressure only. - GC also known as Vapor-Phase Chromatography (VPC)
and Gas-Liquid Partition Chromatography (GLPC)
3Gas Chromatography Acetates
- Gas Chromatograph
- Microliter Syringe
- Heated injection port with rubber septum for
inserting sample - Heating chamber with carrier gas injection port
- Oven containing copper, stainless steel, or glass
column - Column packed with the Stationary Liquid Phase, a
non-volatile liquid, wax, or low melting
solid-high boiling hydrocarbons, silicone oils,
waxes or polymeric esters, ethers, and amides. We
use DC200 from Dow Chemical - Liquid phase is coated onto a support material,
generally crushed firebrick
4Gas Chromatography Acetates
- Principals of Separation
- Column is selected, packed with Liquid Phase, and
installed - Sample injected with microliter syringe into the
injection port where it is vaporized and mixed
into the Carrier Gas stream (helium, nitrogen,
argon) - Sample vapor becomes partitioned between Moving
Gas Phase and Stationary Liquid Phase - The time the different compounds in the sample
spend in the Vapor Phase is a function of their
Vapor Pressure - The more volatile (Low Boiling Point / Higher
Vapor Pressure) compounds arrive at the end of
the column first and pass into the detector
5Gas Chromatography Acetates
- Principals of Detection
- Two Detector Types
- Thermal Conductivity Detector (TCD) (we use this)
- Flame Ionization
- TCD is electrically heated Hot Wire placed in
carrier gas stream - Thermal conductivity of carrier gas (helium in
our case) is higher than most organic substances - Presence of sample compounds in gas stream
reduces thermal conductivity of stream - Wire heats up and resistance decreases
- Two detectors used one exposed to sample gas and
the other exposed to reference flow of carrier
gas - Detectors form arms of Wheatstone Bridge, which
becomes unbalanced by sample gas - Unbalanced bridge generates electrical signal,
which is amplified and sent to recorder
6Gas Chromatography Acetates
- Factors Affecting Separation
- Boiling Points of Components in Sample
- Low boiling point compounds have higher vapor
pressures - High boiling point compounds have lower vapor
pressures requiring more energy to reach
equilibrium vapor pressure, i.e., atmospheric
pressure - Boiling point increases as molecular weight
increases - Flow Rate of Carrier Gas
- Choice of Liquid Phase
- Molecular weights, functional groups, and
polarities of component molecules are factors in
selecting liquid phase - Length of Column
- Similar compounds require longer columns than
dissimilar compounds. Isomeric mixtures often
require quite long columns
7Gas Chromatography Acetates
- The Experiment
- Purpose Introduce the theory and technique of
gas chromatography - Identify a compound by it retention time
- From the relationship between peak area and
mole content calculate the mole fraction and
mole percent of a compound in a mixture - Approach
- Obtain chromatograph of a known equimolar mixture
of four (4) esters - Ethyl, Propyl, Butyl,
Hexyl Acetate - Obtain chromatograph of unknown mixture (one or
more compounds in the known mixture) - Determine Retention Times
- Calculate Peak Areas
- Adjust Peak Areas for Thermal Response
- Calculate Total Area from Adjusted Areas
- Calculate Mole Fraction
- Calculate Mole Percentage
8Gas Chromatography Acetates
- The Experiment (Cont)
- Groups Work in groups of three (2)
- Each group will obtain 2 copies of the
chromatogram for the standard (equimolar) mixture - Each Student will run their own unknown
- Samples
- The Standard (Equimolar) Mixture has 4 esters
- Ethyl Acetate, Propyl Acetate, Butyl Acetate,
Hexyl Acetate - The Unknowns have from 2 to 4 of the compounds in
the standard mixture
9Gas Chromatography Acetates
- The Report
- The Gas Chromatograph instrument settings and the
processing of the samples to get the
chromatograms are considered one (1) procedure - When multiple samples or sub-samples are
processed with the same procedure, it is not
necessary to set up a separate procedure for each
sample - Setup a suitable template in Results section to
report all of the results obtained - Thus, the process to obtain Gas Chromatograms of
the Known mixture of 4 acetates and the
Unknown mixture utilize the same procedure - The computation of the Peak Area, Adjusted Peak
Areas, Total Peak area, Mole Fraction, and Mole
are considered separate procedures
10Gas Chromatography Acetates
- Data Summary Procedure Using complete
sentences summarize, in paragraph form, all of
the results obtained in the experiment - Analysis Conclusion Section
- Develop a set of arguments to prove the identity
of the unknown compounds in the unknown mixture - Comment on the equivalency of the peak areas and
equimolar content of the known mixture - Why was it necessary to apply the Thermal
Response Correction Factor to the measured peak
areas? - Chromatograms
- Copied chromatogram sets for each team member
must be copied at the same scale, otherwise
retention time computations will be wrong - Tape the trimmed chromatograms to a blank sheet
of paper and attach to end of report
11Gas Chromatography Acetates
- Record Instrument readings (Place in GC procedure
Results) - Injection Port Temp
- Column Temp
- Detector Temp
- Gas Flow Rate 60 mL / min
- Chart Speed Generally 5 cm /min
- Moving Liquid Phase (DC-200)
- Injecting the Sample
- Sample is injected into the B port with the
Microsyringe - The Microsyringe is fragile and expensive BE
CAREFUL - Mark Starting Point on chart short vertical
line - Insert needle fully into B Port through the
rubber septum - Coordinating with chart recorder operator, inject
the sample into the heated chamber, while
simultaneously starting the chart recorder
For the instrument in the 407 lab, all three
temperatures are read from the single dial on the
front of cabinet
12Gas Chromatography Acetates
- Determine the Retention Time
- The period following injection that is required
for a compound to pass through the column to the
point where the detector current is maximum, i.e.
maximum pen deflection or maximum peak height - For a given set of constant conditions (carrier
gas, flow rate of carrier gas, column
temperature, column length, liquid phase,
injection port temperature), the retention time
of any compound is always constant - Retention Time is similar to the Retardation
Factor, Rf in Thin Layer Chromatography - Compute Retention Time from the Chart Speed (5 or
10 cm/min) and the distance on the chart from the
time of injection to the point on the chart where
the perpendicular line drawn from the peak height
intersects the base line
13Gas Chromatography Acetates
- Determination of Retention Time
- Since Velocity (v) Distance / Time d / t
- Ret Time (t) Distance(cm) / Velocity(cm/min)
d / v
Note Disregard Air Peaks in all calculations
Retention Time Distances Mark Starting Point On
Chart (t 0) Draw vertical Line from Peak Top to
Base Line Measure Distance from Starting Point to
Base of Peak
14Gas Chromatography Acetates
- Quantitative Analysis
- The area under a gas chromatograph peak is
proportional to the amount (moles) of the
compounds eluted - The molar percentage composition of a mixture can
be approximated by comparing the relative areas
of the peaks in the chromatogram - This approach assumes that the detector is
equally sensitive to all compounds and its
response is linear - This assumption is usually not valid and will be
addressed by adjusting the peak areas using the
Thermal Response algorithm described on slides
17-24
15Gas Chromatography Acetates
- Triangulation Method of Determining Area Under
Peak - Multiply the height of peak (in mm) above the
baseline by the width of the peak at half the
height. - Baseline is a straight line connecting side arms
of the peak. Best if peaks are symmetrical. - Add the individual areas to get the total area
- Divide each area by total area to get mole
fraction - Multiply mole fraction by 100 to get adjusted
mole - See algorithm development on next slide
- Adjust the peak areas for non-linear thermal
response using the algorithm described in slides
17-28
16Gas Chromatography Acetates
- Draw Baseline connecting peak bottoms
- Peak Area by the Triangulation Method
- Peak Area h w½
- Where h Peak Height from
baseline - w½ width of peak at
½ the peak height - Adjust Peak Area for thermal response
- See discussion on following slides
- Total Adjusted Peak Area (TA) A B
- Mole Fraction (MF)
- A/TA B/TA
- Mole Percent MF x 100
17Gas Chromatography Acetates
- Thermal Response Factor
- The areas of gas chromatogram peaks are
proportional to the molar content of the mixture - Compounds with different functional groups or
widely varying molecular weights do not all have
the same thermal conductivity. This can cause
the instrument to produce response variations,
which cause deviations (non-linearity) in the
relationship between peak area and molar content - A correction factor called The Thermal Response
Factor for a given compound can be established
from the relative peak areas of an equimolar
solution - Equimolar mixtures contain compounds with the
same molar content, i.e., the same number of
moles - Thus, equimolar mixtures should produce peaks of
equal area, if the instrument response is linear
18Gas Chromatography Acetates
- Thermal Response Ratios
- GC Peak Area Correction Factor (approach 1)
- The ratio of one peak area to another in a GC
chromatogram should be proportional to the molar
ratio of the components in the mixture - The expression for modifying the Peak Areas for a
non-linear area instrument response is
constructed as follows - Determine the area of each peak in an equimolar
mixture - Compute the ratio of one of the peaks selected as
the basis for computation relative to each peak
area -
19Gas Chromatography Acetates
- Thermal Response Correction Factor (cont)
- Multiply the area of each peak by the respective
Thermal Response Factor (TRx) - Compute the Total Adjusted Area
- Compute the Adjusted Mole Fraction
- Compute the Adjusted Mole Percent
20Gas Chromatography Acetates
- Thermal Response Ratios
- Example Ethyl Acetate (S2) is used as basis
for calculations
EtAc (2) ProAc (3) BuAc (4) HexAc (6)
Standard Equimolar Mixture Measured Peak Area 1.44 1.09 1.16 0.98
Standard Equimolar Mixture TRs/TRi As/Ai (s2) 1.44 1.00 1.44 1.44 1.32 1.09 1.44 1.24 1.16 1.44 1.47 0.98
Unknown Mixture Measured Peak Area 2.14 2.18 2.12 1.54
Unknown Mixture Adjusted Areas 2.14 1.00 2.14 2.18 1.32 2.88 2.12 1.24 2.63 1.54 1.47 2.26
Total Adjusted Area ? 2.14 2.88 2.63
2.26 9.91 Mole Fraction EtAc 2.14 /
9.91 0.216 Mole Fraction ProAc 2.88 / 9.91
0.291 Mole Fraction Bu Ac 2.63 / 9.91
0.265 Mole Fraction HexAc 2.26 / 9.91
0.228
21Gas Chromatography Acetates
- Thermal Response Ratios
- GC Peak Area Correction Factor (alternate
approach) - The ratio of one peak area to another in a GC
chromatogram should be proportional to the molar
ratio of the components in the mixture - If the peaks of an equimolar mixture do not have
the same area, the relationship between the area
of a peak and the mole fraction of the compound
in the mixture is incorrect and would have to be
adjusted by some factor - The Thermal Response Factor (TR) is determined
from an Equimolar Mixture
22Gas Chromatography Acetates
- The derivation that follows utilizes ratios
between any two compounds in a mixture, one of
which will be designated as the basis for
computation - Assuming an equimolar mixture of 4 acetates
- Ethyl Acetate, Propyl Acetate, Butyl Acetate,
Hexyl Acetate - In the equation development below, the subscript
i will be used to designate the compounds in a
mixture - i(1,2,3,4) Ethyl(1), Propyl(2), Butyl(3),
Hexyl(4) - In the derivation and examples that follow, Ethyl
Acetate will be used as the basis for the
calculations (designated by subscript (s), but
any of the other compounds could also be used,
such as in the case where the unknown mixture
does not contain any Ethyl Acetate
23Gas Chromatography Acetates
- Thermal Response Ratios (Cont)
- The following expression equates corrected area
ratios to an adjustment of the molar ratios - The area ratio (mole ratio) of each component (i)
is shown relative to the selected base of
computation compound (s) in the mixture - If the equation is rearranged to indicate an
adjustment to the measured areas - Note the subscripts relative to the TR factor
24Gas Chromatography Acetates
- Compute the TRs/TRi ratios from the measured peak
areas from the standard equimolar mixture - For an equimolar mixture molei/moles 1
- Thus, substitution in equation 2 gives
- Again note the relative position of the
subscripts - From equation (3), each individual TRs/TRi ratio
is calculated from the peak areas of the standard
equimolar mixture
25Gas Chromatography Acetates
- Thermal Response Ratios (Cont)
- Adjusting the Peak Areas of the Unknown Mixture
- Using each TRs/TRi ratio, the mole ratio of each
component in the unknown mixture, relative to the
base compound, is calculated from equation (2) - The Molei/Moles values from equation 2 now
represent adjusted peak areas, and thus are
proportional to the molar content of the unknown
mixture - The adjusted Molei/Moles values are summed
- The new Mole Fractions are computed by dividing
each Molei/Moles value by the total - The new Mole is computed by multiplying the
mole fraction by 100
.
26Gas Chromatography Acetates
- Thermal Response Ratios (Cont)
- Example Ethyl Acetate (S2) is used as basis
for calculations
EtAc (2) ProAc (3) BuAc (4) HexAc (6)
Standard Equimolar Mixture Measured Peak Area 1.44 1.09 1.16 0.98
Standard Equimolar Mixture TRs/TRi As/Ai (s2) 1.44 1.00 1.44 1.44 1.32 1.09 1.44 1.24 1.16 1.44 1.47 0.98
Unknown Mixture Measured Peak Area 2.14 2.18 2.12 1.54
Unknown Mixture areai/areas (s2) 2.14 1.00 2.14 2.18 1.02 2.14 2.12 0.99 2.14 1.54 0.72 2.14
Apply TRs/Tri correction factor to measured area
ratios using equation 2
EtAc / EtAc mol2 / mol2 area2 /
area2 ? TR2 / TR2 2.14 / 2.14 ? 1.00
1.00 ProAc / EtAc mol3 / mol2 area3 /
area2 ? TR2 / TR3 2.18 / 2.14 ? 1.32
1.34 BuAc / EtAc mol4 / mol2 area4 /
area2 ? TR2 / TR4 2.12 / 2.14 ? 1.24 1.23
HexAc / EtAc mol6 / mol2 area6 / area2 ?
TR2 / TR6 1.54 / 2.14 ? 1.47 1.06 ?
moli/mol2 1.00 1.34 1.23 1.06
4.63 ? mole EtAc 1.00 / 4.63 100
21.6 ? mole ProAc 1.34 / 4.63 100
28.9 ? mole BuAc 1.23 / 4.63 100
26.6 ? mole HexAc 1.06 / 4.63 100
22.9
27Gas Chromatography Acetates
Thermal Response Ratios (Cont) Example 2
Ethyl Acetate (S2) is used as basis for
calculations
EtAc (2) ProAc (3) BuAc (4) HexAc (6)
Standard Equimolar Mixture Measured Peak Area 128 186 208 210
Standard Equimolar Mixture TRs/TRi As/Ai (s2) 128 1.00 128 128 0.69 186 128 0.62 208 128 0.61 210
Unknown Mixture Measured Peak Area 2.14 2.18 2.12 1.54
Unknown Mixture areai/areas (s2) 2.14 1.00 2.14 2.18 1.01 2.14 2.12 0.99 2.14 1.54 0.72 2.14
Apply TRs/Tri correction factor to measured area
ratios using equation 2
EtAc / EtAc mol2 / mol2 area2 / area2 ? TR2
/ TR2 2.14 / 2.14 ? 1.00 1.00 ProAc /
EtAc mol3 / mol2 area3 / area2 ? TR2 /
TR3 2.18 / 2.14 ? 0.69 0.70 BuAc /
EtAc mol4 / mol2 area4 / area2 ? TR2 /
TR4 2.12 / 2.14 ? 0.62 0.61 HexAc /
EtAc mol6 / mol2 area6 / area2 ? TR2 /
TR6 1.54 / 2.14 ? 0.61 0.44 ?
moli/mol2 1.00 0.70 0.61 0.44
2.75 ? mole EtAc 1.00 / 2.75 100
36.4 ? mole ProAc 0.70 / 2.75 100
25.4 ? mole BuAc 0.61 / 2.75 100
22.2 ? mole HexAc 0.44 / 2.75 100 16.0
28Gas Chromatography Acetates
Thermal Response Ratios (Cont) Ex. 3 - Assumes
the unknown is missing Ethyl Acetate and
Propyl Acetate (S3) is used as basis for
calculations
ProAc (3) BuAc (4) HexAc (6)
Standard Equimolar Mixture Measured Peak Area 186 208 210
Standard Equimolar Mixture TRs/TRi As/Ai (s3) 186 1.0 186 186 0.89 208 186 0.89 210
Unknown Mixture Measured Peak Area 2.18 2.12 1.54
Unknown Mixture areai/areas (s3) 2.18 1.0 2.18 2.12 0.97 2.18 1.54 0.71 2.18
Apply TRs/Tri correction factor to measured area
ratios using equation 2
- ProAc / EtAc mol3 / mol3 area3 / area3 ?
TR3 / TR3 2.18 / 2.18 ? 1.00 1.00 - BuAc / EtAc mol4 / mol3 area4 / area3 ?
TR3 / TR4 2.12 / 2.18 ? 0.89 0.87 - HexAc / EtAc mol6 / mol3 area6 / area3 ?
TR3 / TR6 1.54 / 2.18 ? 0.89 0.63 -
- moli/mol3 1.00 0.87 0.63 2.50
- ? mole ProAc 1.00 / 2.50 100 40.0
- ? mole BuAc 0.87 / 2.50 100 34.8
- ? mole HexAc 0.63 / 2.50 100 25.2