Designing for Radiation Protection - PowerPoint PPT Presentation

1 / 20
About This Presentation
Title:

Designing for Radiation Protection

Description:

Designing for Radiation Protection Design Standards for Radiation Protection Leakage Radiation LEAKAGE RADIATION may not EXCEED TUBE HOUSING 100mR / hr or 1 mGy ... – PowerPoint PPT presentation

Number of Views:118
Avg rating:3.0/5.0
Slides: 21
Provided by: ECC75
Learn more at: https://www.elcamino.edu
Category:

less

Transcript and Presenter's Notes

Title: Designing for Radiation Protection


1
Designing for Radiation Protection
2
Design Standards for Radiation Protection
Leakage Radiation
3
LEAKAGE RADIATIONmay not EXCEED
  • TUBE HOUSING 100mR / hr or 1 mGy/hr
  • _at_ 1 meter

4
Control Panel
  • Must have an indicator when the x-ray tube is
    energized. Sometimes it will be a visible light
    or audible signal or both

5
SID Source to Image Distance
  • Indicator must be accurate to within 2 of the
    desired SID

6
Source-to-Skin Distance
  • SSD is used to describe to distance between the
    tube and patient during fluoroscopy
  • SSD must not be less than 38 cm in fluoro rooms
    and 30 cm for mobile fluoro

7
Collimation or PBL
  • Positive-Beam Limitation PBL
  • Detects the size of the IR used and automatically
    collimates to size of the IR at standard SIDs
  • Used to be required by law between 1974 1994
    but no longer required. PBL still remains
    standard in equipment design

8
Filtration
  • General-purpose tubes
  • (above 70 kVp) and
  • the fluoroscopic
  • beam
  • must be filtered
  • at least 2.5 mm Al

9
Reproducibility Linearity
  • Reproducibility consistent exposures
  • Linearity any mA station X msec will produce
    the same mAs

10
Exposure switch
  • In diagnostic rooms the switch must be fixed to
    prevent the ability to make an exposure in the
    room
  • Mobile switch must be at least 6 ft long and the
    dead-man type
  • Fluoroscopic exposure control should be the
    dead-man type as well

11
Radiography Room Design Factors Affecting
Barrier Thickness
  • Distance between radiation source and barrier
  • Occupancy the use of the area being protected
    (rarely occupied vs nursing station)
  • Control area occupied by radiation workers or
    patients
  • Uncontrolled area occupied by anyone

12
MONITORING WORK AREAS
  • CONTROLLED AREA Used by occupationally exposed
    personnel (monitored)
  • 100mrem or 1mSv / week
  • UNCONTROLLED AREA PUBLIC
  • 2 mrem or 0.02 mSv / week

13
ROOM SHIELDING
  • PRIMARY SHIELD Any wall or portable wall that
    the primary beam could possibly be directed at
  • 1/16 Pb or eq. 7 Feet High
  • 4 inches of Concrete will do!

14
ROOM SHIELDING
Pg. 574
  • SECONDARY NO PRIMARY BEAM
  • 1/32 LEAD
  • CONTROL BOOTH never point the beam at the
    control panel
  • BEAM SCATTERS 2X BEFORE HITTING
  • LEAD WINDOW 1.5 mm LEAD EQ

15
Bucky Slot Cover Protective Curtain
  • What is the
  • minimum
  • Pb eq. required ?

16
Image Intensifier is a primary barrier 2 mm Pb
equivalent
Bucky cover and Curtain Minimum of 0.25 mm/Pb
CUMLATIVE TIMER FOR FLUORO SIGNALS AFTER 5
MINUTES OF FLUORO TO MAKE SURE THE RADIOLOGIST IS
AWARE OF THE BEAM-ON TIME
17
Exposure Calculations
  • Exposure Exposure rate X mAs
  • Exposure rate varies by kVp selection and distance

18
Exposure Calculations
  • At 75 kVp the exposure rate is 2.6 mR/mAs
  • What is the exposure?
  • AP Abdomen 25 mAs
  • LPO L-spine 32 mAs
  • PA Caldwell 200 mA X 0.10 sec
  • AP Pelvis 200 mA X 0.20 sec

19
Exposure Calculations
  • 2.6 mR X 25mAs 65 mR/mAs
  • LPO L-spine 32 mAs
  • PA Caldwell 200 mA X 0.10 sec
  • AP Pelvis 200 mA X 0.20 sec

20
Be Safe! Image gently!
Write a Comment
User Comments (0)
About PowerShow.com