Title: Designing for Radiation Protection
1Designing for Radiation Protection
2Design Standards for Radiation Protection
Leakage Radiation
3LEAKAGE RADIATIONmay not EXCEED
- TUBE HOUSING 100mR / hr or 1 mGy/hr
- _at_ 1 meter
4Control Panel
- Must have an indicator when the x-ray tube is
energized. Sometimes it will be a visible light
or audible signal or both
5SID Source to Image Distance
- Indicator must be accurate to within 2 of the
desired SID
6Source-to-Skin Distance
- SSD is used to describe to distance between the
tube and patient during fluoroscopy - SSD must not be less than 38 cm in fluoro rooms
and 30 cm for mobile fluoro
7Collimation or PBL
- Positive-Beam Limitation PBL
- Detects the size of the IR used and automatically
collimates to size of the IR at standard SIDs - Used to be required by law between 1974 1994
but no longer required. PBL still remains
standard in equipment design
8Filtration
- General-purpose tubes
- (above 70 kVp) and
- the fluoroscopic
- beam
- must be filtered
- at least 2.5 mm Al
9Reproducibility Linearity
- Reproducibility consistent exposures
- Linearity any mA station X msec will produce
the same mAs
10Exposure switch
- In diagnostic rooms the switch must be fixed to
prevent the ability to make an exposure in the
room - Mobile switch must be at least 6 ft long and the
dead-man type - Fluoroscopic exposure control should be the
dead-man type as well
11Radiography Room Design Factors Affecting
Barrier Thickness
- Distance between radiation source and barrier
- Occupancy the use of the area being protected
(rarely occupied vs nursing station) - Control area occupied by radiation workers or
patients - Uncontrolled area occupied by anyone
12MONITORING WORK AREAS
- CONTROLLED AREA Used by occupationally exposed
personnel (monitored) - 100mrem or 1mSv / week
- UNCONTROLLED AREA PUBLIC
- 2 mrem or 0.02 mSv / week
13ROOM SHIELDING
- PRIMARY SHIELD Any wall or portable wall that
the primary beam could possibly be directed at - 1/16 Pb or eq. 7 Feet High
- 4 inches of Concrete will do!
14ROOM SHIELDING
Pg. 574
- SECONDARY NO PRIMARY BEAM
- 1/32 LEAD
- CONTROL BOOTH never point the beam at the
control panel - BEAM SCATTERS 2X BEFORE HITTING
- LEAD WINDOW 1.5 mm LEAD EQ
15Bucky Slot Cover Protective Curtain
- What is the
- minimum
- Pb eq. required ?
16Image Intensifier is a primary barrier 2 mm Pb
equivalent
Bucky cover and Curtain Minimum of 0.25 mm/Pb
CUMLATIVE TIMER FOR FLUORO SIGNALS AFTER 5
MINUTES OF FLUORO TO MAKE SURE THE RADIOLOGIST IS
AWARE OF THE BEAM-ON TIME
17Exposure Calculations
- Exposure Exposure rate X mAs
- Exposure rate varies by kVp selection and distance
18Exposure Calculations
- At 75 kVp the exposure rate is 2.6 mR/mAs
- What is the exposure?
- AP Abdomen 25 mAs
- LPO L-spine 32 mAs
- PA Caldwell 200 mA X 0.10 sec
- AP Pelvis 200 mA X 0.20 sec
19Exposure Calculations
- 2.6 mR X 25mAs 65 mR/mAs
- LPO L-spine 32 mAs
- PA Caldwell 200 mA X 0.10 sec
- AP Pelvis 200 mA X 0.20 sec
20 Be Safe! Image gently!