LED Metrology
  • LED spatial intensity distribution measurement
    research

2
LEDs in present-day industry products
  • LEDs for its optical parameters have a set of
    unique features and cant be compared with
    traditional lamp bulbs
  • Because of large uncertainties are currently
    involved transfer from using lamp bulbs to LEDs
    is not a trivial task

3
LED radiation spatial intensity distribution
  • LED is not a point source, thats why there are
    some difficulties appears during radiant
    intensity measurement
  • Radiant intensity changes with the distance from
    LED to detector
  • Publication CIE 127 recommends to use a new term
    averaged light intensity

4
LED radiant intensity model for approximate
analysis
5
LED radiation spatial intensity distribution
measurement instrument
  • Design of goniometric instrument provide
    measurement in half sphere (solid angle 2p sr)

6
LED radiation spatial intensity distribution
measurement instrument
7
Instrument software
  • Measurement is fully automated and dont need any
    user participation.
  • User have some set of options for measurement
    accuracy adjustment
  • Measurement result presented as 3D diagram
  • Measurement data are saved in xml text file

8
LEDs measurement instrument photos(made by
chapter member)
9
Measurement uncertainties
  • LED radiation properties changes with forward
    current and temperature. Some little current
    fluctuations cause large result uncertainties
  • Detector SNR and ADC error provide additional
    uncertainty for result
  • Errors in LED positioning during measurement give
    an uncertainty that changes with signal
    derivative in diagram region
  • LED angle value is the main factor that doesn't
    allow to receive equally precise results for
    different LEDs