Scatter Radiation
  • Compton interaction
  • Higher kVp energies in diagnostic range
  • As scatter increases, radiographs lose contrast
  • Radiographs will look dull, blurred, and fogged

3
Factors Affecting Scatter Radiation
  • kVp
  • Field size
  • Patient or part thickness

4
kVp
  • Compton interaction increases with increasing kVp
  • Scatter radiation increases
  • Scatter radiation is reduced by using collimators
    and grids
  • Decreasing kVp is not always desirable
  • Increases patient dose

5
Beam Field Size
  • As field size is increased, scatter radiation
    increases
  • Restriction of field size improves image quality
  • Even more important during fluoroscopy

6
Patient Or Part Thickness
  • Scatter increases with patient or part thickness
  • A chest x-ray will result in more scatter than an
    arm x-ray
  • Compression devices reduce patient thickness and
    improve quality

7
Beam-Restricting Devices
  • Two types of devices to reduce scatter
  • Beam restricting devices
  • Grids
  • Restricted for two reasons
  • Reduce patient dose
  • Improve image contrast
  • Three types of beam-restricting devices
  • Aperture diaphragm
  • Extension cones and cylinders
  • Variable aperture collimator
  • Positive beam limitation

8
Aperture Diaphragm
  • Simplest of all beam-restricting devices
  • Lead or lead-lined plate with an opening
  • Used with older head units
  • Cannot see a light

9
Extension Cones And Cylinders
  • Modifications of aperture diaphragms
  • Primarily used for headwork, L5-S1
  • Fixed openings
  • Generally not light localizing

10
Variable-Aperture Collimator
  • Light localizing
  • Most commonly used
  • A series of shutters and mirrors used
  • Light field must coincide with x-ray beam

11
Positive Beam Limitation Devices
  • Automatically collimates to the size of film used
  • Does not relieve of technologist of having to
    collimate to anatomy

12
Collimator Filtration
  • Some collimator housing units come with
    interchangeable filtration
  • Generally, different millimeter thicknesses of
    aluminum is used